Abstract

Purpose

This paper describes the design, principles, performances, and applications of a novel image-guided master–slave robotic system for vascular intervention (VI), including the performance evaluation and in vivo trials.

Methods

Based on the peer-to-peer (P2P) remote communication system, the kinetics analysis, the sliding-mode neural network self-adaptive control model and the feedback system, this new robotic system can accomplish in real time a number of VI operations, including guidewire translation and rotation, balloon catheter translation, and contrast agent injection. The master–slave design prevents surgeons from being exposed to X-ray radiation, which means that they are not required to wear a heavy lead suit. We also conducted a performance evaluation of the new system, which assessed the speed, position tracking, and accuracy, as well as in vivo swine trials.

Results

The speed and position tracking effects are really good, which contribute to the high level of performance in terms of the translational (error ≤ 0.45%) and rotational (error ≤ 2.6°) accuracy. In addition, the accuracy of the contrast agent injection is less than 0.2 ml. The robotic system successfully performed both the stent revascularization of an arteria carotis and four in vivo trials. The haptic feedback data correspond with the robotic-assisted procedure, and peaks and troughs of data occur regularly.

Conclusions

By means of the performance evaluation and four successful in vivo trials, the feasibility and efficiency of the new robotic system are validated, which should prove helpful for further research.

Sie haben noch keine Lizenz? Dann Informieren Sie sich jetzt über unsere Produkte:

Springer Professional "Wirtschaft+Technik"

Online-Abonnement

Mit Springer Professional "Wirtschaft+Technik" erhalten Sie Zugriff auf:

  • über 102.000 Bücher
  • über 537 Zeitschriften

aus folgenden Fachgebieten:

  • Automobil + Motoren
  • Bauwesen + Immobilien
  • Business IT + Informatik
  • Elektrotechnik + Elektronik
  • Energie + Nachhaltigkeit
  • Finance + Banking
  • Management + Führung
  • Marketing + Vertrieb
  • Maschinenbau + Werkstoffe
  • Versicherung + Risiko

Jetzt Wissensvorsprung sichern!

Springer Professional "Wirtschaft"

Online-Abonnement

Mit Springer Professional "Wirtschaft" erhalten Sie Zugriff auf:

  • über 67.000 Bücher
  • über 340 Zeitschriften

aus folgenden Fachgebieten:

  • Bauwesen + Immobilien
  • Business IT + Informatik
  • Finance + Banking
  • Management + Führung
  • Marketing + Vertrieb
  • Versicherung + Risiko




Jetzt Wissensvorsprung sichern!

Springer Professional "Technik"

Online-Abonnement

Mit Springer Professional "Technik" erhalten Sie Zugriff auf:

  • über 67.000 Bücher
  • über 390 Zeitschriften

aus folgenden Fachgebieten:

  • Automobil + Motoren
  • Bauwesen + Immobilien
  • Business IT + Informatik
  • Elektrotechnik + Elektronik
  • Energie + Nachhaltigkeit
  • Maschinenbau + Werkstoffe




 

Jetzt Wissensvorsprung sichern!

Literatur
1.
Zurück zum Zitat Venketasubramanian N, Yoon BW, Pandian J, Navarro JC (2017) Stroke epidemiology in south, east, and South-East Asia: a review. J Stroke 19(3):286–294CrossRefPubMedPubMedCentral Venketasubramanian N, Yoon BW, Pandian J, Navarro JC (2017) Stroke epidemiology in south, east, and South-East Asia: a review. J Stroke 19(3):286–294CrossRefPubMedPubMedCentral
2.
Zurück zum Zitat World Health Organization (2017) World health statistics 2017: monitoring health for the SDGs. World Health Organization, Geneva World Health Organization (2017) World health statistics 2017: monitoring health for the SDGs. World Health Organization, Geneva
3.
Zurück zum Zitat Rafii-Tari H, Payne CJ, Yang GZ (2014) Current and emerging robot-assisted endovascular catheterization technologies: a review. Ann Biomed Eng 42(4):697–715CrossRefPubMed Rafii-Tari H, Payne CJ, Yang GZ (2014) Current and emerging robot-assisted endovascular catheterization technologies: a review. Ann Biomed Eng 42(4):697–715CrossRefPubMed
4.
Zurück zum Zitat Lin JC (2013) The role of robotic surgical system in the management of vascular disease. Ann Vasc Surg 27(7):976–983CrossRefPubMed Lin JC (2013) The role of robotic surgical system in the management of vascular disease. Ann Vasc Surg 27(7):976–983CrossRefPubMed
5.
Zurück zum Zitat Da L, Zhang D, Wang T (2008) Overview of the vascular interventional robot. Int J Med Robot Comp 4(4):289CrossRef Da L, Zhang D, Wang T (2008) Overview of the vascular interventional robot. Int J Med Robot Comp 4(4):289CrossRef
6.
Zurück zum Zitat Daneshmand M, Bilici O, Bolotnikova A, Anbarjafari G (2017) Medical robots with potential applications in participatory and opportunistic remote sensing: a review. Robot Auton Syst 95:160–180CrossRef Daneshmand M, Bilici O, Bolotnikova A, Anbarjafari G (2017) Medical robots with potential applications in participatory and opportunistic remote sensing: a review. Robot Auton Syst 95:160–180CrossRef
7.
Zurück zum Zitat Arai F, Fujimura R, Fukuda T, Negoro M (2002) New catheter driving method using linear stepping mechanism for intravascular neurosurgery. In: IEEE ICRA, 2002. Proceedings, Vol 3, pp 2944–2949 Arai F, Fujimura R, Fukuda T, Negoro M (2002) New catheter driving method using linear stepping mechanism for intravascular neurosurgery. In: IEEE ICRA, 2002. Proceedings, Vol 3, pp 2944–2949
8.
Zurück zum Zitat Xiao N, Guo J, Guo S, Tamiya T (2012) A robotic catheter system with real-time force feedback and monitor. Australas Phys Eng Sci Med 35(3):283–289CrossRefPubMed Xiao N, Guo J, Guo S, Tamiya T (2012) A robotic catheter system with real-time force feedback and monitor. Australas Phys Eng Sci Med 35(3):283–289CrossRefPubMed
9.
Zurück zum Zitat Guo S, Kondo H, Wang J, Guo J, Tamiya T (2007) A new catheter operating system for medical applications. In: IEEE/ICME, pp 82–86 Guo S, Kondo H, Wang J, Guo J, Tamiya T (2007) A new catheter operating system for medical applications. In: IEEE/ICME, pp 82–86
10.
Zurück zum Zitat Srimathveeravalli G, Kesavadas T, Li X (2010) Design and fabrication of a robotic mechanism for remote steering and positioning of interventional devices. Int J Med Robot Comp 6(2):160–170 Srimathveeravalli G, Kesavadas T, Li X (2010) Design and fabrication of a robotic mechanism for remote steering and positioning of interventional devices. Int J Med Robot Comp 6(2):160–170
12.
Zurück zum Zitat Romo E, Hart SJ, Covington T (2016) Low friction instrument driver interface for robotic systems [P]. US 20160338783 A1 Romo E, Hart SJ, Covington T (2016) Low friction instrument driver interface for robotic systems [P]. US 20160338783 A1
13.
Zurück zum Zitat Pankratov MM (2011) Robotically assisted PCI with CorPath 200 system: early evidence of potential benefits for patient and physician. Cardiovasc Revasc Med 12(3):e40–e40CrossRef Pankratov MM (2011) Robotically assisted PCI with CorPath 200 system: early evidence of potential benefits for patient and physician. Cardiovasc Revasc Med 12(3):e40–e40CrossRef
14.
Zurück zum Zitat Pedersen A (2011) Trial to evaluate CorPath robotic system for PCI. Med Device Daily 15(18):1 Pedersen A (2011) Trial to evaluate CorPath robotic system for PCI. Med Device Daily 15(18):1
15.
Zurück zum Zitat Granada JF, Delgado JA, Uribe MP, Fernandez A, Blanco G, Leon MB, Weisz Giora (2011) First-in-human evaluation of a novel robotic-assisted coronary angioplasty system. JACC-Cardiovasc Interv 4(4):460–465CrossRefPubMed Granada JF, Delgado JA, Uribe MP, Fernandez A, Blanco G, Leon MB, Weisz Giora (2011) First-in-human evaluation of a novel robotic-assisted coronary angioplasty system. JACC-Cardiovasc Interv 4(4):460–465CrossRefPubMed
16.
Zurück zum Zitat Khan EM, Ng GA, Neelagaru S, Abi-Samra FM, Lee J, Giudici M, Gohn D, Winkle RA, Sussman J, Knight BP, Berman A, Calkins H (2013) First experience with a novel robotic remote catheter system: amigo™ mapping trial. J Interv Card Electr 37(2):121–129CrossRef Khan EM, Ng GA, Neelagaru S, Abi-Samra FM, Lee J, Giudici M, Gohn D, Winkle RA, Sussman J, Knight BP, Berman A, Calkins H (2013) First experience with a novel robotic remote catheter system: amigo™ mapping trial. J Interv Card Electr 37(2):121–129CrossRef
17.
Zurück zum Zitat Won JY, Yi BJ, Cha HJ (2016) Review of new master/slave catheter driving vascular intervention robot system: intervention radiologist’s perspective. HMR 36(4):225 Won JY, Yi BJ, Cha HJ (2016) Review of new master/slave catheter driving vascular intervention robot system: intervention radiologist’s perspective. HMR 36(4):225
18.
Zurück zum Zitat Shen H, Wang C, Xie L, Zhou S, Gu L, Xie H (2018) A novel remote-controlled robotic system for cerebrovascular intervention. Int J Med Robot Comput Assist Surg Mrcas, 14: e1943 Shen H, Wang C, Xie L, Zhou S, Gu L, Xie H (2018) A novel remote-controlled robotic system for cerebrovascular intervention. Int J Med Robot Comput Assist Surg Mrcas, 14: e1943
19.
Zurück zum Zitat Zeng Q, Zhou SJ, Shen H, Wang C (2018). A network communication protocols for robotic-assisted vascular intervention systems. BST Zeng Q, Zhou SJ, Shen H, Wang C (2018). A network communication protocols for robotic-assisted vascular intervention systems. BST
20.
Zurück zum Zitat Li Zi-Yan, Jing-Li Fu, Chen Li-Qun (2009) Euler–lagrange equation from nonlocal-in-time kinetic energy of nonconservative system. Phys Lett A 374(2):106–109CrossRef Li Zi-Yan, Jing-Li Fu, Chen Li-Qun (2009) Euler–lagrange equation from nonlocal-in-time kinetic energy of nonconservative system. Phys Lett A 374(2):106–109CrossRef
21.
Zurück zum Zitat Bai Z (2004) Study of a 3-dof parallel manipulator dynamics based on lagrange’s equation. J B Univ Aeronaut Astronaut 30(1):51–54 Bai Z (2004) Study of a 3-dof parallel manipulator dynamics based on lagrange’s equation. J B Univ Aeronaut Astronaut 30(1):51–54
22.
Zurück zum Zitat Xia HY, Zhan WB (2003) Application and development expression of lagrange equation. J Luoyang Teach Coll 2:27–29 Xia HY, Zhan WB (2003) Application and development expression of lagrange equation. J Luoyang Teach Coll 2:27–29
23.
Zurück zum Zitat Edwards C, Spurgeon S (1998) Sliding mode control: theory and applications. CRC Press, Boca RatonCrossRef Edwards C, Spurgeon S (1998) Sliding mode control: theory and applications. CRC Press, Boca RatonCrossRef
24.
Zurück zum Zitat Utkin VI, Chang HC (2007) Sliding mode control on electro-mechanical systems. Math Probl Eng 8(4–5):451–473 Utkin VI, Chang HC (2007) Sliding mode control on electro-mechanical systems. Math Probl Eng 8(4–5):451–473
25.
Zurück zum Zitat Liu J, Wang X (2011) Advanced sliding mode control for mechanical systems. Tsinghua University Press, Springer, BerlinCrossRef Liu J, Wang X (2011) Advanced sliding mode control for mechanical systems. Tsinghua University Press, Springer, BerlinCrossRef
26.
Zurück zum Zitat Liu J (2013) Radial basis function (rbf) neural network control for mechanical systems. Springer, Berlin, pp 19–53CrossRef Liu J (2013) Radial basis function (rbf) neural network control for mechanical systems. Springer, Berlin, pp 19–53CrossRef
27.
Zurück zum Zitat Seshagiri S, Khalil HK (2000) Output feedback control of nonlinear systems using RBF neural networks. In: Proceedings of the ACC, 1999, Vol 11, pp 2808–2812 Seshagiri S, Khalil HK (2000) Output feedback control of nonlinear systems using RBF neural networks. In: Proceedings of the ACC, 1999, Vol 11, pp 2808–2812
28.
Zurück zum Zitat Welch G, Bishop G (1995). An introduction to the kalman filter. Course Notes 8 of ACM SIGGRAPH 2001, 8(7): 127–132 Welch G, Bishop G (1995). An introduction to the kalman filter. Course Notes 8 of ACM SIGGRAPH 2001, 8(7): 127–132
29.
Zurück zum Zitat Evensen G (2003) The ensemble kalman filter: theoretical formulation and practical implementation. Ocean Dyn 53(4):343–367CrossRef Evensen G (2003) The ensemble kalman filter: theoretical formulation and practical implementation. Ocean Dyn 53(4):343–367CrossRef
30.
Zurück zum Zitat Brown RG, Hwang PYC (1997) Introduction to random signals and applied Kalman filtering: with MATLAB exercises and solutions. Wiley, New York Brown RG, Hwang PYC (1997) Introduction to random signals and applied Kalman filtering: with MATLAB exercises and solutions. Wiley, New York
31.
Zurück zum Zitat Ovando-Shelley E, Santoyo E, de Oliveir M (2013) Intervention techniques. Geotechnical engineering for the preservation of monuments and historic sites, pp 75–91 Ovando-Shelley E, Santoyo E, de Oliveir M (2013) Intervention techniques. Geotechnical engineering for the preservation of monuments and historic sites, pp 75–91
Metadaten
Titel
A novel robotic system for vascular intervention: principles, performances, and applications
verfasst von
Hao Shen
Cheng Wang
Le Xie
Shoujun Zhou
Lixu Gu
Hongzhi Xie
Publikationsdatum
09.02.2019
Verlag
Springer International Publishing
Erschienen in
International Journal of Computer Assisted Radiology and Surgery / Ausgabe 4/2019
Print ISSN: 1861-6410
Elektronische ISSN: 1861-6429
DOI
https://doi.org/10.1007/s11548-018-01906-w

Weitere Artikel der Ausgabe 4/2019

International Journal of Computer Assisted Radiology and Surgery 4/2019 Zur Ausgabe