Abstract

Purpose

The discrepancy of continuously decreasing opportunities for clinical training and assessment and the increasing complexity of interventions in surgery has led to the development of different training and assessment options like anatomical models, computer-based simulators or cadaver trainings. However, trainees, following training, assessment and ultimately performing patient treatment, still face a steep learning curve.

Methods

To address this problem for C-arm-based surgery, we introduce a realistic radiation-free simulation system that combines patient-based 3D printed anatomy and simulated X-ray imaging using a physical C-arm. To explore the fidelity and usefulness of the proposed mixed-reality system for training and assessment, we conducted a user study with six surgical experts performing a facet joint injection on the simulator.

Results

In a technical evaluation, we show that our system simulates X-ray images accurately with an RMSE of 1.85 mm compared to real X-ray imaging. The participants expressed agreement with the overall realism of the simulation, the usefulness of the system for assessment and strong agreement with the usefulness of such a mixed-reality system for training of novices and experts. In a quantitative analysis, we furthermore evaluated the suitability of the system for the assessment of surgical skills and gather preliminary evidence for validity.

Conclusion

The proposed mixed-reality simulation system facilitates a transition to C-arm-based surgery and has the potential to complement or even replace large parts of cadaver training, to provide a safe assessment environment and to reduce the risk for errors when proceeding to patient treatment. We propose an assessment concept and outline the steps necessary to expand the system into a test instrument that provides reliable and justified assessments scores indicative of surgical proficiency with sufficient evidence for validity.

Sie haben noch keine Lizenz? Dann Informieren Sie sich jetzt über unsere Produkte:

Springer Professional "Wirtschaft+Technik"

Online-Abonnement

Mit Springer Professional "Wirtschaft+Technik" erhalten Sie Zugriff auf:

  • über 102.000 Bücher
  • über 537 Zeitschriften

aus folgenden Fachgebieten:

  • Automobil + Motoren
  • Bauwesen + Immobilien
  • Business IT + Informatik
  • Elektrotechnik + Elektronik
  • Energie + Nachhaltigkeit
  • Finance + Banking
  • Management + Führung
  • Marketing + Vertrieb
  • Maschinenbau + Werkstoffe
  • Versicherung + Risiko

Jetzt Wissensvorsprung sichern!

Springer Professional "Wirtschaft"

Online-Abonnement

Mit Springer Professional "Wirtschaft" erhalten Sie Zugriff auf:

  • über 67.000 Bücher
  • über 340 Zeitschriften

aus folgenden Fachgebieten:

  • Bauwesen + Immobilien
  • Business IT + Informatik
  • Finance + Banking
  • Management + Führung
  • Marketing + Vertrieb
  • Versicherung + Risiko




Jetzt Wissensvorsprung sichern!

Springer Professional "Technik"

Online-Abonnement

Mit Springer Professional "Technik" erhalten Sie Zugriff auf:

  • über 67.000 Bücher
  • über 390 Zeitschriften

aus folgenden Fachgebieten:

  • Automobil + Motoren
  • Bauwesen + Immobilien
  • Business IT + Informatik
  • Elektrotechnik + Elektronik
  • Energie + Nachhaltigkeit
  • Maschinenbau + Werkstoffe




 

Jetzt Wissensvorsprung sichern!

Anhänge
Nur mit Berechtigung zugänglich
Literatur
1.
Zurück zum Zitat Peters TM, Linte CA (2016) Image-guided interventions and computer-integrated therapy: Quo vadis? Med Image Anal 33:56–63CrossRefPubMed Peters TM, Linte CA (2016) Image-guided interventions and computer-integrated therapy: Quo vadis? Med Image Anal 33:56–63CrossRefPubMed
2.
Zurück zum Zitat Härtl R, Lam KS, Wang J, Korge A, Kandziora F, Audige L (2013) Worldwide survey on the use of navigation in spine surgery. World Neurosurg 79(1):162–172CrossRefPubMed Härtl R, Lam KS, Wang J, Korge A, Kandziora F, Audige L (2013) Worldwide survey on the use of navigation in spine surgery. World Neurosurg 79(1):162–172CrossRefPubMed
3.
Zurück zum Zitat Bott OJ, Teistler M, Duwenkamp C, Wagner M, Marschollek M, Plischke M, Raab BW, Stürmer KM, Pretschner DP, Dresing K (2008) virtX—evaluation of a computer-based training system for mobile C-arm systems in trauma and orthopedic surgery. Methods Inf Med 47(3):270–278PubMed Bott OJ, Teistler M, Duwenkamp C, Wagner M, Marschollek M, Plischke M, Raab BW, Stürmer KM, Pretschner DP, Dresing K (2008) virtX—evaluation of a computer-based training system for mobile C-arm systems in trauma and orthopedic surgery. Methods Inf Med 47(3):270–278PubMed
4.
Zurück zum Zitat Ghaderi I, Manji F, Park YS, Juul D, Ott M, Harris I, Farrell TM (2015) Technical skills assessment toolbox: a review using the unitary framework of validity. Ann Surg 261(2):251–262CrossRefPubMed Ghaderi I, Manji F, Park YS, Juul D, Ott M, Harris I, Farrell TM (2015) Technical skills assessment toolbox: a review using the unitary framework of validity. Ann Surg 261(2):251–262CrossRefPubMed
5.
Zurück zum Zitat Ahmed N, Devitt KS, Keshet I, Spicer J, Imrie K, Feldman L, Cools-Lartigue J, Kayssi A, Lipsman N, Elmi M (2014) A systematic review of the effects of resident duty hour restrictions in surgery: impact on resident wellness, training, and patient outcomes. Ann Surg 259(6):1041–1053CrossRefPubMedPubMedCentral Ahmed N, Devitt KS, Keshet I, Spicer J, Imrie K, Feldman L, Cools-Lartigue J, Kayssi A, Lipsman N, Elmi M (2014) A systematic review of the effects of resident duty hour restrictions in surgery: impact on resident wellness, training, and patient outcomes. Ann Surg 259(6):1041–1053CrossRefPubMedPubMedCentral
6.
Zurück zum Zitat Gong RH, Jenkins B, Sze RW, Yaniv Z (2014) A cost effective and high fidelity fluoroscopy simulator using the image-guided surgery toolkit (IGSTK). In: SPIE medical imaging, p 903618 Gong RH, Jenkins B, Sze RW, Yaniv Z (2014) A cost effective and high fidelity fluoroscopy simulator using the image-guided surgery toolkit (IGSTK). In: SPIE medical imaging, p 903618
7.
Zurück zum Zitat Bott OJ, Dresing K, Wagner M, Raab BW, Teistler M (2011) Informatics in radiology: use of a C-arm fluoroscopy simulator to support training in intraoperative radiography. Radiographics 31(3):E65–E75CrossRefPubMed Bott OJ, Dresing K, Wagner M, Raab BW, Teistler M (2011) Informatics in radiology: use of a C-arm fluoroscopy simulator to support training in intraoperative radiography. Radiographics 31(3):E65–E75CrossRefPubMed
8.
Zurück zum Zitat Wucherer P, Stefan P, Abhari K, Fallavollita P, Weigl M, Lazarovici M, Winkler A, Weidert S, Peters T, de Ribaupierre S, Eagleson R, Navab N (2015) Vertebroplasty performance on simulator for 19 surgeons using hierarchical task analysis. IEEE Trans Med Imaging 34(8):1730–1737CrossRefPubMed Wucherer P, Stefan P, Abhari K, Fallavollita P, Weigl M, Lazarovici M, Winkler A, Weidert S, Peters T, de Ribaupierre S, Eagleson R, Navab N (2015) Vertebroplasty performance on simulator for 19 surgeons using hierarchical task analysis. IEEE Trans Med Imaging 34(8):1730–1737CrossRefPubMed
9.
Zurück zum Zitat Harrop J, Rezai AR, Hoh DJ, Ghobrial GM, Sharan A (2013) Neurosurgical training with a novel cervical spine simulator: posterior foraminotomy and laminectomy. Neurosurgery 73:S94–S99CrossRef Harrop J, Rezai AR, Hoh DJ, Ghobrial GM, Sharan A (2013) Neurosurgical training with a novel cervical spine simulator: posterior foraminotomy and laminectomy. Neurosurgery 73:S94–S99CrossRef
10.
Zurück zum Zitat Rudarakanchana N, Herzeele IV, Bicknell CD, Riga CV, Rolls A, Cheshire NJW, Hamady MS (2013) Endovascular repair of ruptured abdominal aortic aneurysm: technical and team training in an immersive virtual reality environment. Cardiovasc Interv Radiol 37(4):920–927 Rudarakanchana N, Herzeele IV, Bicknell CD, Riga CV, Rolls A, Cheshire NJW, Hamady MS (2013) Endovascular repair of ruptured abdominal aortic aneurysm: technical and team training in an immersive virtual reality environment. Cardiovasc Interv Radiol 37(4):920–927
11.
Zurück zum Zitat Waran V, Narayanan V, Karuppiah R, Owen SL, Aziz T (2014) Utility of multimaterial 3d printers in creating models with pathological entities to enhance the training experience of neurosurgeons. J Neurosurg 120(2):489–492CrossRefPubMed Waran V, Narayanan V, Karuppiah R, Owen SL, Aziz T (2014) Utility of multimaterial 3d printers in creating models with pathological entities to enhance the training experience of neurosurgeons. J Neurosurg 120(2):489–492CrossRefPubMed
12.
Zurück zum Zitat Pustka D, Huber M, Bauer M, Klinker G (2006) Spatial relationship patterns: elements of reusable tracking and calibration systems. In: International symposium on mixed and augmented reality, pp 88–97 Pustka D, Huber M, Bauer M, Klinker G (2006) Spatial relationship patterns: elements of reusable tracking and calibration systems. In: International symposium on mixed and augmented reality, pp 88–97
13.
Zurück zum Zitat Tsai RY, Lenz RK (1988) Real time versatile robotics hand/eye calibration using 3d machine vision. In: IEEE international conference on robotics and automation, pp 554–561 Tsai RY, Lenz RK (1988) Real time versatile robotics hand/eye calibration using 3d machine vision. In: IEEE international conference on robotics and automation, pp 554–561
14.
Zurück zum Zitat Tuceryan M, Greer DS, Whitaker RT, Breen DE, Crampton C, Rose E, Ahlers KH (1995) Calibration requirements and procedures for a monitor-based augmented reality system. IEEE Trans Vis Comput Graph 1(3):255–273CrossRef Tuceryan M, Greer DS, Whitaker RT, Breen DE, Crampton C, Rose E, Ahlers KH (1995) Calibration requirements and procedures for a monitor-based augmented reality system. IEEE Trans Vis Comput Graph 1(3):255–273CrossRef
15.
Zurück zum Zitat Besl P, McKay N (1992) A method for registration of 3-d shapes. Trans Pattern Anal Mach Intell 14(2):239–256CrossRef Besl P, McKay N (1992) A method for registration of 3-d shapes. Trans Pattern Anal Mach Intell 14(2):239–256CrossRef
17.
Zurück zum Zitat Wang L (2012) Novel techniques for integrating video augmented X-ray imaging into orthopedic and trauma surgery. Ph.D. thesis, Technische Universität München Wang L (2012) Novel techniques for integrating video augmented X-ray imaging into orthopedic and trauma surgery. Ph.D. thesis, Technische Universität München
18.
Zurück zum Zitat Navab N, Heining SM, Traub J (2010) Camera augmented mobile C-arm (camc): calibration, accuracy study, and clinical applications. IEEE Trans Med Imaging 29(7):1412–1423CrossRefPubMed Navab N, Heining SM, Traub J (2010) Camera augmented mobile C-arm (camc): calibration, accuracy study, and clinical applications. IEEE Trans Med Imaging 29(7):1412–1423CrossRefPubMed
20.
Zurück zum Zitat Koo TK, Li MY (2016) A guideline of selecting and reporting intraclass correlation coefficients for reliability research. J Chiropr Med 15(2):155–163CrossRefPubMedPubMedCentral Koo TK, Li MY (2016) A guideline of selecting and reporting intraclass correlation coefficients for reliability research. J Chiropr Med 15(2):155–163CrossRefPubMedPubMedCentral
21.
Zurück zum Zitat Torres I, De Luccia N (2017) A simulator for training in endovascular aneurysm repair: the use of three dimensional printers. Eur J Vasc Endovasc Surg 54(2):247–253CrossRefPubMed Torres I, De Luccia N (2017) A simulator for training in endovascular aneurysm repair: the use of three dimensional printers. Eur J Vasc Endovasc Surg 54(2):247–253CrossRefPubMed
22.
Zurück zum Zitat Tjardes T, Shafizadeh S, Rixen D, Paffrath T, Bouillon B, Steinhausen ES, Baethis H (2009) Image-guided spine surgery: state of the art and future directions. Eur Spine J 19(1):25–45CrossRefPubMedPubMedCentral Tjardes T, Shafizadeh S, Rixen D, Paffrath T, Bouillon B, Steinhausen ES, Baethis H (2009) Image-guided spine surgery: state of the art and future directions. Eur Spine J 19(1):25–45CrossRefPubMedPubMedCentral
23.
Zurück zum Zitat Downing SM (2003) Validity: on meaningful interpretation of assessment data. Med Educ 37(9):830–837CrossRefPubMed Downing SM (2003) Validity: on meaningful interpretation of assessment data. Med Educ 37(9):830–837CrossRefPubMed
24.
Zurück zum Zitat Korndorffer JR Jr, Kasten SJ, Downing SM (2010) A call for the utilization of consensus standards in the surgical education literature. Am J Surg 199(1):99–104CrossRefPubMed Korndorffer JR Jr, Kasten SJ, Downing SM (2010) A call for the utilization of consensus standards in the surgical education literature. Am J Surg 199(1):99–104CrossRefPubMed
25.
Zurück zum Zitat Rodas NL, Padoy N (2014) 3d global estimation and augmented reality visualization of intra-operative x-ray dose. In: International conference on medical image computing and computer-assisted intervention. Springer, pp 415–422 Rodas NL, Padoy N (2014) 3d global estimation and augmented reality visualization of intra-operative x-ray dose. In: International conference on medical image computing and computer-assisted intervention. Springer, pp 415–422
26.
Zurück zum Zitat Martin JA, Regehr G, Reznick R, MacRae H, Murnaghan J, Hutchison C, Brown M (1997) Objective structured assessment of technical skill (OSATS) for surgical residents. Br J Surg 84(2):273–278CrossRefPubMed Martin JA, Regehr G, Reznick R, MacRae H, Murnaghan J, Hutchison C, Brown M (1997) Objective structured assessment of technical skill (OSATS) for surgical residents. Br J Surg 84(2):273–278CrossRefPubMed
27.
Zurück zum Zitat Thinggaard E, Bjerrum F, Strandbygaard J, Ggenur I, Konge L (2015) Validity of a cross-specialty test in basic laparoscopic techniques (TABLT). Br J Surg 102(9):1106–1113CrossRefPubMedPubMedCentral Thinggaard E, Bjerrum F, Strandbygaard J, Ggenur I, Konge L (2015) Validity of a cross-specialty test in basic laparoscopic techniques (TABLT). Br J Surg 102(9):1106–1113CrossRefPubMedPubMedCentral
Metadaten
Titel
A radiation-free mixed-reality training environment and assessment concept for C-arm-based surgery
verfasst von
Philipp Stefan
Séverine Habert
Alexander Winkler
Marc Lazarovici
Julian Fürmetz
Ulrich Eck
Nassir Navab
Publikationsdatum
25.06.2018
Verlag
Springer International Publishing
Erschienen in
International Journal of Computer Assisted Radiology and Surgery / Ausgabe 9/2018
Print ISSN: 1861-6410
Elektronische ISSN: 1861-6429
DOI
https://doi.org/10.1007/s11548-018-1807-6

Weitere Artikel der Ausgabe 9/2018

International Journal of Computer Assisted Radiology and Surgery 9/2018 Zur Ausgabe