Algorithm to improve accuracy of energy expended in a room calorimeter

verfasst von: Haiying Quan, Wenrui Hao, Lu Li, Ming Sun, Kuan Zhang

Erschienen in: Medical & Biological Engineering & Computing | Ausgabe 8/2017

Einloggen

Aktivieren Sie unsere intelligente Suche, um passende Fachinhalte oder Patente zu finden.

search-config
loading …

Abstract

The whole-room indirect calorimeter is considered as important equipment for human energy expenditure measurement, but noise reduction in the system remains a challenge. A selective filtering method (SFM) was designed to improve the accuracy of the computation of O2 consumption rate (\( \dot{V}_{{{\text{O}}_{ 2} }} \)) and CO2 production rate (\( \dot{V}_{{{\text{CO}}_{ 2} }} \)), based on two facts: (1) the rapid changes of \( \dot{V}_{{{\text{O}}_{ 2} }} \), \( \dot{V}_{{{\text{CO}}_{ 2} }} \) and respiratory quotient (RQ) in human should be accompanied by physical activity; (2) the oxygen consumption and the carbon dioxide production should not be negative because living humans do not generate oxygen, nor consume carbon dioxide. The performance of SFM was compared with the moving average method, the central difference method and the wavelet de-noising method. The range of \( \dot{V}_{{{\text{O}}_{ 2} }} \) and \( \dot{V}_{{{\text{CO}}_{ 2} }} \) in the empty room (the background noise) is reduced from −130.00–146.00 ml/min to −26.00–24.00 ml/min, and from −20.50–12.50 ml/min to −3.99–4.19 ml/min, by SFM. The background noise was added to simulated rectangular and sinusoidal signals that were used to evaluate the four methods over different time periods (64, 32, 16 and 8 min). The highest signal-to-noise ratio and the lowest deviation were achieved by SFM. Abnormal metabolic rates and RQs were corrected and compensated with measurement accuracy of 98.51 ± 0.3 % for 24-h alcohol burning tests. The results of the study showed that SFM can significantly improve \( \dot{V}_{{\text{O}_{2} }} \) and \( \dot{V}_{{{\text{CO}}_{ 2} }} \) measurements.

Sie haben noch keine Lizenz? Dann Informieren Sie sich jetzt über unsere Produkte:

Springer Professional "Wirtschaft+Technik"

Online-Abonnement

Mit Springer Professional "Wirtschaft+Technik" erhalten Sie Zugriff auf:

  • über 102.000 Bücher
  • über 537 Zeitschriften

aus folgenden Fachgebieten:

  • Automobil + Motoren
  • Bauwesen + Immobilien
  • Business IT + Informatik
  • Elektrotechnik + Elektronik
  • Energie + Nachhaltigkeit
  • Finance + Banking
  • Management + Führung
  • Marketing + Vertrieb
  • Maschinenbau + Werkstoffe
  • Versicherung + Risiko

Jetzt Wissensvorsprung sichern!

Springer Professional "Technik"

Online-Abonnement

Mit Springer Professional "Technik" erhalten Sie Zugriff auf:

  • über 67.000 Bücher
  • über 390 Zeitschriften

aus folgenden Fachgebieten:

  • Automobil + Motoren
  • Bauwesen + Immobilien
  • Business IT + Informatik
  • Elektrotechnik + Elektronik
  • Energie + Nachhaltigkeit
  • Maschinenbau + Werkstoffe




 

Jetzt Wissensvorsprung sichern!

Springer Professional "Wirtschaft"

Online-Abonnement

Mit Springer Professional "Wirtschaft" erhalten Sie Zugriff auf:

  • über 67.000 Bücher
  • über 340 Zeitschriften

aus folgenden Fachgebieten:

  • Bauwesen + Immobilien
  • Business IT + Informatik
  • Finance + Banking
  • Management + Führung
  • Marketing + Vertrieb
  • Versicherung + Risiko




Jetzt Wissensvorsprung sichern!

Anhänge
Nur mit Berechtigung zugänglich
Glossar
RQ
Respiratory quotient
MR
Metabolic rate
SD
Standard deviation
\( \dot{V}_{{{\text{O}}_{ 2} }} \)
O2 consumption rate
\( \dot{V}_{{{\text{CO}}_{ 2} }} \)
CO2 production rate
\( \dot{V}_{G} \)
Input data set
\( \dot{V}_{F} \)
Filtering data set
SNR
Signal-to-noise ratio
SFM
Selective filtering method
CDM
Central difference method
MAM
Moving average method
WDM
Wavelet de-noising method
Literatur
1.
Zurück zum Zitat Atwater WO, Rosa EB (1899) Description of a new respiration calorimeter and experiments on the conservation of energy in the human body. Exp Stn Bull (United States Office of Experiment Stations); no 63 Atwater WO, Rosa EB (1899) Description of a new respiration calorimeter and experiments on the conservation of energy in the human body. Exp Stn Bull (United States Office of Experiment Stations); no 63
2.
Zurück zum Zitat Beaver WL, Wasserman K, Whipp BJ (1973) On-line computer analysis and breath-by-breath graphical display of exercise function tests. J Appl Physiol 34:128–132PubMed Beaver WL, Wasserman K, Whipp BJ (1973) On-line computer analysis and breath-by-breath graphical display of exercise function tests. J Appl Physiol 34:128–132PubMed
3.
Zurück zum Zitat Brown D, Cole T, Dauncey M, Marrs R, Murgatroyd P (1984) Analysis of gaseous exchange in open-circuit indirect calorimetry. Med Biol Eng Comput 22:333–338CrossRefPubMed Brown D, Cole T, Dauncey M, Marrs R, Murgatroyd P (1984) Analysis of gaseous exchange in open-circuit indirect calorimetry. Med Biol Eng Comput 22:333–338CrossRefPubMed
4.
Zurück zum Zitat Brychta RJ, Rothney MP, Skarulis MC, Chen KY (2009) Optimizing energy expenditure detection in human metabolic chambers. In: Conference proceedings: annual international conference of the IEEE Engineering in Medicine and Biology Society IEEE Engineering in Medicine and Biology Society Annual Conference 2009:6864–6868. doi:10.1109/IEMBS.2009.5333121 Brychta RJ, Rothney MP, Skarulis MC, Chen KY (2009) Optimizing energy expenditure detection in human metabolic chambers. In: Conference proceedings: annual international conference of the IEEE Engineering in Medicine and Biology Society IEEE Engineering in Medicine and Biology Society Annual Conference 2009:6864–6868. doi:10.​1109/​IEMBS.​2009.​5333121
5.
Zurück zum Zitat Chavan MS, Chavan MMN, Gaikwad M (2010) Studies on implementation of wavelet for denoising speech signal. Int J Comput Appl 3 Chavan MS, Chavan MMN, Gaikwad M (2010) Studies on implementation of wavelet for denoising speech signal. Int J Comput Appl 3
6.
Zurück zum Zitat De Waele E, Spapen H, Honoré PM, Mattens S, Van Gorp V, Diltoer M, Huyghens L (2013) Introducing a new generation indirect calorimeter for estimating energy requirements in adult intensive care unit patients: feasibility, practical considerations, and comparison with a mathematical equation. J Crit Care 28:884 (e881-884. e886) CrossRefPubMed De Waele E, Spapen H, Honoré PM, Mattens S, Van Gorp V, Diltoer M, Huyghens L (2013) Introducing a new generation indirect calorimeter for estimating energy requirements in adult intensive care unit patients: feasibility, practical considerations, and comparison with a mathematical equation. J Crit Care 28:884 (e881-884. e886) CrossRefPubMed
8.
Zurück zum Zitat Ergen B (2012) Signal and image denoising using wavelet transform. INTECH Open Access Publisher Ergen B (2012) Signal and image denoising using wavelet transform. INTECH Open Access Publisher
9.
Zurück zum Zitat Graf S, Karsegard VL, Viatte V, Maisonneuve N, Pichard C, Genton L (2013) Comparison of three indirect calorimetry devices and three methods of gas collection: a prospective observational study. Clin Nutr 32:1067–1072CrossRefPubMed Graf S, Karsegard VL, Viatte V, Maisonneuve N, Pichard C, Genton L (2013) Comparison of three indirect calorimetry devices and three methods of gas collection: a prospective observational study. Clin Nutr 32:1067–1072CrossRefPubMed
10.
Zurück zum Zitat Granato L, Brandes A, Bruni C, Greco AV, Mingrone G (2004) VO2, VCO2, and RQ in a respiratory chamber: accurate estimation based on a new mathematical model using the Kalman–Bucy method. J Appl Physiol 96:1045–1054CrossRefPubMed Granato L, Brandes A, Bruni C, Greco AV, Mingrone G (2004) VO2, VCO2, and RQ in a respiratory chamber: accurate estimation based on a new mathematical model using the Kalman–Bucy method. J Appl Physiol 96:1045–1054CrossRefPubMed
11.
Zurück zum Zitat Henning B, Löfgren R, Sjöström L (1996) Chamber for indirect calorimetry with improved transient response. Med Biol Eng Comput 34:207–212CrossRefPubMed Henning B, Löfgren R, Sjöström L (1996) Chamber for indirect calorimetry with improved transient response. Med Biol Eng Comput 34:207–212CrossRefPubMed
12.
Zurück zum Zitat Lighton JR (2008) Measuring metabolic rates: a manual for scientists. Oxford University Press, OxfordCrossRef Lighton JR (2008) Measuring metabolic rates: a manual for scientists. Oxford University Press, OxfordCrossRef
13.
Zurück zum Zitat Melzer K, Kayser B, Schutz Y (2014) Respiratory quotient evolution during normal pregnancy: what nutritional or clinical information can we get out of it? Eur J Obstet Gynecol Reprod Biol 176:5–9CrossRefPubMed Melzer K, Kayser B, Schutz Y (2014) Respiratory quotient evolution during normal pregnancy: what nutritional or clinical information can we get out of it? Eur J Obstet Gynecol Reprod Biol 176:5–9CrossRefPubMed
14.
Zurück zum Zitat Moon JK, Vohra FA, Valerio Jimenez OS, Puyau MR, Butte NF (1995) Closed-loop control of carbon dioxide concentration and pressure improves response of room respiration calorimeters. J Nutr 125:220PubMed Moon JK, Vohra FA, Valerio Jimenez OS, Puyau MR, Butte NF (1995) Closed-loop control of carbon dioxide concentration and pressure improves response of room respiration calorimeters. J Nutr 125:220PubMed
15.
Zurück zum Zitat Murgatroyd P, Shetty P, Prentice A (1993) Techniques for the measurement of human energy expenditure: a practical guide. Int J Obes Relat Metab Disord J Int Assoc Study Obes 17:549–568 Murgatroyd P, Shetty P, Prentice A (1993) Techniques for the measurement of human energy expenditure: a practical guide. Int J Obes Relat Metab Disord J Int Assoc Study Obes 17:549–568
16.
Zurück zum Zitat Nguyen T, De Jonge L, Smith S, Bray G (2003) Chamber for indirect calorimetry with accurate measurement and time discrimination of metabolic plateaus of over 20 min. Med Biol Eng Comput 41:572–578CrossRefPubMed Nguyen T, De Jonge L, Smith S, Bray G (2003) Chamber for indirect calorimetry with accurate measurement and time discrimination of metabolic plateaus of over 20 min. Med Biol Eng Comput 41:572–578CrossRefPubMed
17.
Zurück zum Zitat Pettenkofer M (1862) Ueber die respiration. AnnlChem 123:1–52 Pettenkofer M (1862) Ueber die respiration. AnnlChem 123:1–52
18.
Zurück zum Zitat Rumpler WV, Seale JL, Conway JM, Moe PW (1990) Repeatability of 24-h energy expenditure measurements in humans by indirect calorimetry. Am J Clin Nutr 51:147–152PubMed Rumpler WV, Seale JL, Conway JM, Moe PW (1990) Repeatability of 24-h energy expenditure measurements in humans by indirect calorimetry. Am J Clin Nutr 51:147–152PubMed
19.
Zurück zum Zitat Seale JL, Rumpler WV, Moe PW (1991) Description of a direct-indirect room-sized calorimeter. Am J Physiol Endocrinol Metab 260:E306–E320 Seale JL, Rumpler WV, Moe PW (1991) Description of a direct-indirect room-sized calorimeter. Am J Physiol Endocrinol Metab 260:E306–E320
20.
Zurück zum Zitat Siervo M, Bertoli S, Battezzati A, Wells J, Lara J, Ferraris C, Tagliabue A (2014) Accuracy of predictive equations for the measurement of resting energy expenditure in older subjects. Clin Nutr 33:613–619CrossRefPubMed Siervo M, Bertoli S, Battezzati A, Wells J, Lara J, Ferraris C, Tagliabue A (2014) Accuracy of predictive equations for the measurement of resting energy expenditure in older subjects. Clin Nutr 33:613–619CrossRefPubMed
21.
Zurück zum Zitat Sion-Sarid R, Cohen J, Houri Z, Singer P (2013) Indirect calorimetry: a guide for optimizing nutritional support in the critically ill child. Nutrition 29:1094–1099CrossRefPubMed Sion-Sarid R, Cohen J, Houri Z, Singer P (2013) Indirect calorimetry: a guide for optimizing nutritional support in the critically ill child. Nutrition 29:1094–1099CrossRefPubMed
22.
Zurück zum Zitat Smith SW (1997) The scientist and engineer’s guide to digital signal processing Smith SW (1997) The scientist and engineer’s guide to digital signal processing
23.
Zurück zum Zitat Snell B, Fullmer S, Eggett DL (2014) Reading and listening to music increase resting energy expenditure during an indirect calorimetry test. J Acad Nutr Diet 114:1939–1942CrossRefPubMed Snell B, Fullmer S, Eggett DL (2014) Reading and listening to music increase resting energy expenditure during an indirect calorimetry test. J Acad Nutr Diet 114:1939–1942CrossRefPubMed
24.
Zurück zum Zitat Sun M, Reed G, Hill J (1994) Modification of a whole room indirect calorimeter for measurement of rapid changes in energy expenditure. J Appl Physiol 76:2686–2691PubMed Sun M, Reed G, Hill J (1994) Modification of a whole room indirect calorimeter for measurement of rapid changes in energy expenditure. J Appl Physiol 76:2686–2691PubMed
25.
Zurück zum Zitat Tokuyama K, Ogata H, Katayose Y, Satoh M (2009) Algorithm for transient response of whole body indirect calorimeter: deconvolution with a regularization parameter. J Appl Physiol 106:640–650CrossRefPubMed Tokuyama K, Ogata H, Katayose Y, Satoh M (2009) Algorithm for transient response of whole body indirect calorimeter: deconvolution with a regularization parameter. J Appl Physiol 106:640–650CrossRefPubMed
27.
Zurück zum Zitat White M, Bouchard G, Buemann B, Almeras N, Despres J, Bouchard C, Tremblay A (1996) Reproducibility of 24-h energy expenditure and macronutrient oxidation rates in an indirect calorimeter. J Appl Physiol 80:133–139PubMed White M, Bouchard G, Buemann B, Almeras N, Despres J, Bouchard C, Tremblay A (1996) Reproducibility of 24-h energy expenditure and macronutrient oxidation rates in an indirect calorimeter. J Appl Physiol 80:133–139PubMed
Metadaten
Titel
Algorithm to improve accuracy of energy expended in a room calorimeter
verfasst von
Haiying Quan
Wenrui Hao
Lu Li
Ming Sun
Kuan Zhang
Publikationsdatum
20.10.2016
Verlag
Springer Berlin Heidelberg
Erschienen in
Medical & Biological Engineering & Computing / Ausgabe 8/2017
Print ISSN: 0140-0118
Elektronische ISSN: 1741-0444
DOI
https://doi.org/10.1007/s11517-016-1583-9

Weitere Artikel der Ausgabe 8/2017

Medical & Biological Engineering & Computing 8/2017 Zur Ausgabe