Abstract
Presents the results obtained in a study of LiF TLD-100 performance for mailed dosimetry in radiotherapy, using glow curve analysis for TL evaluation and reusable chips. An excellent reusability is assured by a simple thermal treatment of 1 h at 400 degrees C followed by a controlled and reproducible cooling to room temperature that accurately reproduces the sensitivity of the detectors. Particular attention was devoted to studying sensitivity changes during room temperature storage of TLD-100 detectors. These changes can be one of the major sources of uncertainty in mailed dosimetry, where long intervals can occur between detector preparation, irradiation and readout. Based on the features of the individual peak evolution during storage a rather simple method to correct for these sensitivity modifications is proposed. The analysis of the different influential factors permitted the authors to estimate a combined uncertainty for the measurement of absorbed doses within the 0.54 Gy range of 1.0%.