5. Design and Analysis of a Whole-Body Noncontact Electromagnetic Subthreshold Stimulation Device with Field Modulation Targeting Nonspecific Neuropathic Pain

verfasst von : Sergey Makarov, Gene Bogdanov, Gregory Noetscher, William Appleyard, Reinhold Ludwig, Juho Joutsa, Zhi-De Deng

Erschienen in: Brain and Human Body Modeling

Verlag: Springer International Publishing

Aktivieren Sie unsere intelligente Suche, um passende Fachinhalte oder Patente zu finden.

search-config
loading …

Abstract

Chronic pain represents a major health problem. Approximately 100 million US adults suffer from common chronic pain conditions, more than the number affected by heart disease, diabetes, and cancer combined. The economic cost of chronic pain in adults is $560–630 billion annually. Chronic low back pain accounts for 22% of all cases of chronic pain and for 35% of most persistent pain sites. The most common diffuse low back neuropathic pain is classified as nonspecific low back pain.
Evidence suggests that electrical stimulation may modify both cause and perception of chronic pain. The objective of this study is to describe the design of a novel, noncontact, large-scale electrostimulation device, which utilizes the concept of a spacious, high-quality electromagnetic resonator operating in the kHz range and at mild subthreshold stimulation levels. Our new resonant neurostimulation approach could probably combine the best of transcutaneous electric nerve stimulation, or TENS, and transcranial magnetic stimulation, or TMS. In the first case, it is a continuous and flexible operation, while in the second case, it is painless, noncontact, and provides deep field penetration.
This study is driven by the limitations of TENS. We introduce a conceptually different electromagnetic stimulation device. Instead of local high-intensity and suprathreshold TENS, we suggest to stimulate the PNS (peripheral nervous system) and muscular system of the entire lower body in a noncontact, patient-friendly way. At the same time, we suggest to use low or subthreshold power levels. In other words, we propose mild yet more broad electromagnetic treatment potentially beneficial for nonspecific chronic pain. The proposed device would primarily affect peripheral nerves, spinal cord, muscles, joints, and bone. Simultaneously, it could influence the somatosensory cortex via many affected pathways, in line with the modern concept of central control of pain.

5.1 Introduction

Pain is distinguished by duration as acute (less than 6 weeks), subacute (6–12 weeks), and chronic (12 weeks or more) pain. Approximately 100 million US adults suffer from common chronic pain conditions, more than the number affected by heart disease, diabetes, and cancer combined [1]. The economic cost of chronic pain in adults, including health care expenses and lost productivity, is $560–630 billion annually [1]. Seven in ten Americans feel that pain research and management should be one of the medical community’s top few priorities (16%) or a high priority (55%) [2].
One form of chronic pain is nociceptive pain, which is the normal response to injury of tissues such as muscles, visceral organs, joints, or bones. Another form is neuropathic pain, which involves dysfunction of (i) the peripheral nervous system (PNS) or (ii) the central nervous system (CNS). The latter case is amplification and generation of pain within the CNS itself due to distorted sensory processing, malfunctioning of pain-inhibitory mechanisms, and enhancing pain-facilitatory mechanisms [36]. An example is psychogenic pain, which does not usually have a physical origin [7, 8]. Highly prevalent symptoms in chronic pain are depression and anxiety [711], which are reported by more than 50% of patients with chronic pain [9]. Pain and depression may create a vicious cycle in which pain worsens depression and vice versa [10].
Low back pain, either acute or chronic, dominates other pain types [7] and affects about 80–84% of the population at least once at some point in life [1214]. In the US Armed Services alone, low back pain was the primary diagnosis for more than seven million ambulatory care visits between 2000 and 2009 [15]. Current estimates are that approximately 25% of people with acute low back pain experience recurrent episodes, while 7–10% progress to a chronic state [15] and can experience significant physical, psychological, and social sequelae that affect their long-term functioning and quality of life [16]. According to [11], 70% of subjects with chronic low back pain report fatigue and 18% report depression. According to [17], 59% of the patients with chronic low back pain report poor sleep.
Chronic low back pain accounts for 22% of all chronic pain cases and for 35% of most persistent pain sites [7]. The classification of low back pain is complicated by the varying presentation and complex nature of pain [14]. The most common diffuse neuropathic pain without radiating beyond the buttocks is classified as nonspecific low back pain [14], which makes up 60% of individuals suffering from chronic low back pain [18].
The initial treatment for acute nonspecific low back pain is conservative, including nonopioid analgesics (acetaminophen, aspirin), nonsteroidal anti-inflammatory drugs (ibuprofen, ketoprofen), physiotherapy, dynamic strengthening exercises, thermotherapy, and, if necessary, a short course of muscle relaxants [13, 14]. Further, conservative methods include traction treatment, manual therapy, and transcutaneous electrical nerve stimulation (TENS) [14]. A commonly prescribed treatment for chronic pain is opioids (codeine, oxycodone, hydrocodone, and morphine) [13, 19]. The use of opioids is controversial due to severe addiction and misuse [13, 15, 16, 20, 21]. It is argued that chronic use of opioids is detrimental to people with back pain because they can aggravate depression, leading to a worsening of the pain [13]. Whenever possible, opioid medications in chronic noncancer pain should be avoided [20].
TENS is a common drug-free alternative treatment technique that stimulates selected sensory nerves and muscles via electrodes placed on the skin over the painful area [2228]. The electrodes inject electric currents and, most importantly, electric fields into body. The theory is that the local electric field stimulation can modify both cause and perception of chronic pain. A number of systematic reviews of the effect of TENS on various painful conditions, such as labor pain, rheumatoid arthritis, phantom limb pain, and chronic lower back pain, are available [22, 2428]. However, these reviews indicate that most controlled randomized clinical trials failed to show significant effects of the existing small-scale TENS systems with a strongly localized electric field distribution.
This design-based study is driven by the limitations of TENS. We introduce a conceptually different electromagnetic stimulation device. Instead of local high-intensity and suprathreshold TENS, we suggest to stimulate the PNS and muscular system of the entire lower body in a noncontact, patient-friendly way. At the same time, we suggest to use low or subthreshold power levels. In other words, we propose mild yet more broad electromagnetic treatment, potentially beneficial for nonspecific chronic pain. The proposed device [29, 30] would primarily affect peripheral nerves, the spinal cord, muscles, joints, and bone. Simultaneously, it could influence the somatosensory cortex via many affected pathways, in line with the modern concept of central control of pain [8]. Based on the numbers cited above [7, 14, 18], we can estimate that some 14% of all chronic pain cases might be subject to the proposed alternative treatment.
The text is organized as follows. Section 5.2 describes a theoretical device model, specifies the field distribution within the resonator, and describes hardware design, test, and functionality, including semiautomatic operation/tuning and representative continuous run times. Section 5.3 provides computational results for the electric field distribution within the body obtained via two independent numerical methods. Section 5.4 discusses possible device modifications, as well as potential application scenarios. Section 5.5 concludes the chapter.

5.2 Materials and Methods

5.2.1 Suprathreshold Versus Subthreshold Stimulation

The vast majority of transcutaneous electrical nerve stimulation (TENS) devices [14, 2228] today are suprathreshold. They excite local currents/electric fields that are strong enough to produce an action potential in peripheral axons assembled in bundles of sensory nerve fibers. A case in point is a small pulsed-current InTENSity™ Twin Stim® III device [31] similar to that shown in Fig. 5.1a [32], intended for home use. It creates a prominent zapping sensation when the electrodes are connected to the lower back. The majority of PEMF (pulsed electromagnetic field) devices operate in a similar fashion.
Alternatively, a continuous current injection with a more powerful stationary device may be employed by chiropractors as shown in Fig. 5.1b [33]. When connected to the lower back, it creates a rather strong burning sensation. The electric field may also be injected via a noncontact induction coil [34, 35] as in Fig. 5.1c. Although a broader coverage with many electrodes is now possible, local suprathreshold effects still dominate.
With respect to treatment-persistent depression, only three nonpharmacological therapies have been approved by the FDA to date; all of them are electromagnetic and suprathreshold: transcranial magnetic stimulation [36], vagus nerve stimulation [37], and electroconvulsive therapy [38].
Subthreshold refers to a low-power electromagnetic stimulation that is too small to elicit action potentials. However, it still alters the axonal membrane potential [39]. This effect accumulates and maximizes toward axon terminals, i.e., synapses [39]. It is synaptic efficacy (or natural neurotransmission efficacy) that is altered and enhanced by the subthreshold stimulation [39]. In a chain of neurons, this stimulation could cause an incremental relay effect, which may further enhance neuronal network activity [39]. The theory of subthreshold stimulation [3942] has been developed in application to transcranial current stimulation [3947]. Low-field magnetic stimulation of depression is an active area of research [4854]. The subthreshold technique is also a modern research direction in spinal cord stimulation [5559], as it eliminates noxious and off-target paresthesia, while being efficient, as well as in vagus [60], occipital [61], proximal [62], and parasympathetic [63] nerve stimulation, and in neuromuscular stimulation [64]. Electrocutaneous subthreshold stimulation has been found to improve sleep and reduce reactive anxiety/depression [65].

5.2.2 Concept of the Magnetic Stimulator. Two-Dimensional Analytical Solution for Solenoidal E-Field

Figure 5.2 shows the anticipated device concept. An external, uniform, rotating (or circularly-polarized) magnetic flux density B with amplitude B0 is created in the transverse plane around a tissue volume, depicted in Fig. 5.2a. By Faraday’s law of induction, this field excites an axial, rotating solenoidal electric field E in free space or in a tissue volume, which is expressed in terms of the magnetic vector potential A,
$$ E=-\raisebox{1ex}{$\partial A$}\!\left/ \!\raisebox{-1ex}{$\partial t$}\right. $$
(5.1)
as shown in Fig. 5.2b; we set B =  ∇  × A. Thus, when a biological body is placed into this volume, a significant noninvasively excited electric field in the axial direction will appear parallel to the major peripheral nerves, spinal cord, long bones, major arteries, veins, and other structures. This is in contrast to a solenoidal coil wound around the body creating electric fields and currents in the less desirable transverse plane.
The rotational character of the field also assures that not only one body cross section (e.g., coronal or sagittal) will be subject to the electric field excitation, but the entire body volume.
In the ideal, two-dimensional case and for any conducting target with a strict cylindrical symmetry placed into the device, either homogeneous or not, the corresponding two-dimensional problem, shown in Fig. 5.2a, b, will have an exact analytical solution in the quasi-static (or eddy current) approximation. The electric field within the target is given by [66].
$$ {E}_x=0,\kern1em {E}_y=0,\kern1em {E}_z=\omega {B}_0r\cos \left(\omega t-\phi \right) $$
(5.2)
$$ J=\sigma E $$
(5.3)
where r is the radial distance from the coil axis in cylindrical coordinates, ω is the angular frequency, ϕ is an arbitrary phase, J is induced current density, and σ is the (local) medium conductivity which is either constant or obeys cylindrical symmetry. Although Eqs. (5.2) and (5.3) might be used to roughly estimate the electric field in the body based on a cylindrically symmetric assumption, its actual value will deviate as shown below.

5.2.3 Three-Dimensional Coil Resonator Design. Solenoidal E-Field

The external rotating magnetic flux density B is created using a volumetric resonator in the form of a low-pass birdcage coil. Resonators of this type (called “birdcage coils”) are routinely used as MRI radio frequency (RF) coils [67, 68], but for a completely different purpose, namely atomic spin excitation and RF signal acquisition. The resonant frequency in this application is very high, typically 64 MHz (for 1.5 T magnets) or higher. For our purpose, we decided to reconstruct that design for a much lower frequency band of 100 kHz or less. In particular, the band of 10 kHz has recently demonstrated great promise for spinal cord stimulation for back and leg chronic pain management [5559, 6977] with and without previous back surgeries [70, 72], and is utilized by TENS [78]. In addition to superior pain relief, the 10 kHz band may provide long-term improvements in quality of life and functionality for subjects with chronic low back and leg pain [77]. On the other hand, a wider band of 4–30 kHz has been used for polyneuropathy (a general degeneration of peripheral nerves that spreads toward the center of the body) electrostimulation treatment [7982].
We chose the birdcage coil because it can produce a very homogeneous B-field in the transverse plane, and because it can produce a circularly polarized B-field. These features relax the requirements for accurate patient positioning relative to the coil. The patient has significant freedom of movement transversely within the coil, including freedom of rotation (thanks to circular polarization). This should enhance patient comfort and permit long treatment sessions.
The resonator concept will allow for an arbitrary “tonic” modulation [30] of the carrier frequency, which was found to beneficially address the variable nature of chronic pain across different patients [76]. This modulation can be either open- or closed-loop (e.g., a single-channel EEG signal fed back to the modulator).
When the resonant frequency becomes low as in the present case, the standard RF birdcage coil will possess very low inductance L. Tuning such a coil toward resonance at low frequencies would require large capacitance C. This, however, means a low Q-factor (quality factor \( Q=\sqrt{L/C}/R \) is the “gain” of the series resonator) and higher costs, as well as higher fabrication complexity [83, 84], and will restrict the use of the conventional birdcage coil to frequencies above at least a few megahertz. Different methods to overcome this difficulty have been suggested [8388], but they are all limited to small-size coils.
Our design described in the subsequent patent application [29] utilizes a unique large-scale low-pass birdcage coil topology with an intentionally very large number (144) of long rungs (boosting inductance) and bridging capacitors seen in Figs. 5.6 and 5.7. After this point, the improvement in adding more rungs is small. Several other reasons for the number of rungs include:
  • To distribute the required capacitors uniformly around the coil circumference
  • To improve the mechanical stiffness of this self-supporting coil structure
  • To reduce rung tube diameter, simplifying assembly
  • To facilitate direct drive of the coil at a single rung (for each mode). The equivalent parallel resistance at resonance was significantly above 50 ohms for the 144 rung configuration, permitting the use of capacitive-only matching networks. Later in the development process, we decided not to use direct drive, but the option remains
However, we are not claiming that the particular number of rungs we used is optimal. All the coil geometrical parameters (number of rungs, rung and end ring tube diameters, coil length and diameter) are subject to optimization during design and construction of the next prototype. We may even consider switching to a fundamentally different coil winding design, such a saddle coil. That said, any optimization is not expected to dramatically improve the coil’s B-field. Improvements up to 20% may be possible (slightly more if the coil is made smaller).
Along with this, we employ carefully designed inductive power coupling. The inductive coupling blocks DC and acts as a balun (balanced-unbalanced transformer). This is useful in terms of both circuit design (we avoid having to install a transformer) and safety concerns (no direct path from the AC to the coil). It also allows adjusting the load resistance at resonance. Inductive coupling perturbs the current distribution in the coil significantly less than directly driving an individual rung. As a result, the resonator possesses a superior quality factor of approximately 300 [30]. Therefore, we may achieve any desired electric field levels of up to 50–100 V/m within the lower body due to the resonance effect and still use standard power electronics equipment.
A computational model of a particular resonator constructed in this study is shown in Fig. 5.3a with the electric current distribution to scale. The coil has a diameter of 0.94 m and a length of 1.10 m; the coil resonates at 100 kHz or at 145 kHz depending on the values of the bridging capacitors. The coil consists of two rings (top and bottom) joined via multiple straight rungs, each bridged with a lumped capacitor at its center. The capacitors control the coil’s resonant frequency. The resonating coil is fed via two lumped ports in quadrature, or using inductive coupling with two loops in quadrature as explained below.
From the modeling point of view, the resonant electric current in both rings at any fixed time instant behaves like a full period of a sine function of polar angle φ. This ring current distribution is shown in Fig. 5.3a. As time progresses, the ring current distribution shown in Fig. 5.3a rotates with angular frequency ω. As a result, the time-domain ring current i(t, φ) in the top and bottom rings can be expressed in the form
$$ i\left(t,\varphi \right)=\pm {I}_0\cos \left(\omega t-\varphi +\pi /2\right),\kern0.5em \varphi \in 0,2\pi $$
(5.4)
where I0 is the current amplitude determined by the excitation power and by the quality factor (or the “gain”) of the resonator.
The AC current in each rung shown in Fig. 5.3a does not change along its length. Simultaneously, at any fixed time instant, it also varies from rung to rung as a harmonic function of the polar angle φ with the full period corresponding to the ring circumference. This rung current distribution is shown in Fig. 5.3a. As time progresses, the rung current distribution shown in Fig. 5.3a also rotates with angular frequency ω. Each individual time-domain rung current density j(t, φ) can be expressed in the form
$$ j\left(t,\varphi \right)=\frac{I_0}{F}\cos \left(\varphi +\omega t\right),\kern1em F=\frac{1}{2}\sum \limits_{m=0}^{\frac{N}{2}-1}\sin \left(\frac{2\pi m}{N}\right),\kern1em \varphi \in 0,2\pi $$
(5.5)
where N is the total number of rungs. This form obeys the current conservation law, or Kirchhoff’s Current Law (KCL), at every ring-rung junction.
The useful current, which creates a nearly constant horizontal rotating magnetic field Br with amplitude B0 and axial rotating electric field Ez according to Eq. (5.2), is the rung current density j(t, φ). Contributions of each rung add up in a constructive manner. The ring current, on the other hand, does not contribute to the axial (or vertical) electric field, Ez. However, it may create a strong transverse electric field very close to the rings.
It should be pointed out that Eqs. (5.4) and (5.5) describe the rotating current behavior, which is a combination of two elementary resonant modes. Each elementary mode does not rotate and appears as depicted in Fig. 5.3a. However, when excited in quadrature (with a 90 degree phase shift and a 90 degree excitation offset along the coil circumference), both modes combine to create the current distribution given by Eqs. (5.4) and (5.5) and the associated rotating electric field. The rotation phenomenon enables us to treat the entire body and not merely a singular component or region.

5.2.4 Solenoidal Electric Field Distribution with and without a Simple Conducting Object

Figure 5.3b–d shows the resulting electric field distribution in the coil (coronal plane) when the current amplitude I0 = 1 ampere in either ring given by Eq. (5.4). The results are given for one resonant mode, as shown in Fig. 5.3a. Due to linearity, this result can simply be scaled for other excitation levels. Accurate field computations have been performed with the fast multipole method described in [89]. The magnitude of the axial component Ez in V/m for an empty coil is shown in Fig. 5.3b. The electric field is indeed zero at the coil’s center.
When a conducting object representing a load is inserted into the coil, the field distribution changes. Figure 5.3c shows the distribution when a conducting cylinder with a diameter of 0.4 m and a length of 1 m is inserted into the coil along its axis. The particular conductivity value σ does not matter since only the conductivity contrast, (σ − σair)/(σ + σair), is present in the solution [66]. This value is always unity since σair = 0.
An interesting and useful effect is observed in Fig. 5.3c: we see a “pulling” of the electric field into the cylinder close to the coil center. This is due to surface charges that appear at and near the cylinder tips. As a result, the electric field close to the cylinder surface at the center plane of the coil increases by nearly 36%.
Another remarkable observation (this effect is common in MRI RF coils) follows from Fig. 5.3d where the conducting cylinder has been shifted from the coil axis to the right by 0.2 m. While the electric field outside the conducting cylinder clearly changes, the field within the cylinder remains nearly the same, as observed in Fig. 5.3c. This may be explained as a result of the electric field being induced by the magnetic field, similar to eddy currents. Since the magnetic field is relatively homogeneous in the transverse plane of the coil, the induced electric fields in a load should not strongly depend on the transverse position of the load within the coil.
These rudimentary simulations allow us to establish two basic facts relevant for the analysis of realistic electric field distributions in a human body within the coil. First, we expect that the average transcutaneous electric field will be slightly higher than predicted by the air-filled coil model in dorsal, abdominal, and lumbar body regions. Second, we expect that the field within the body will not change significantly when the body is moved within the coil in the transverse plane; this circumstance seems to be useful from a practical point of view.

5.2.5 Contribution of Unpaired Electric Charges

Generally, the total electric field within the coil is expressed in terms of two auxiliary potentials. Instead of Eq. (5.1), one has
$$ E=-\partial A/\partial t-\nabla \varphi $$
(5.6)
where φ is the scalar electric potential and A is the magnetic vector potential. In the quasistatic approximation to Maxwell’s equations, the time derivative in the Lorentz gauge \( \frac{1}{c^2}\partial \varphi /\partial t+\nabla \cdot A=0 \) is neglected (which gives us the Coulomb gauge, ∇ · A = 0), while it is still kept in Eq. (5.6). As a result, the − ∇ φ term in Eq. (5.6) becomes a conservative electric field contribution due to charge density alone, while the −A/∂t term is a true solenoidal electric field contribution due to current density alone.
In accordance with the (quasi)electrostatic theory [90], the conservative electric field is blocked by charges induced on a surface of a conducting object and does not penetrate into the object. Therefore, its contribution is ignored in the present study, similar to the theoretical models of transcranial magnetic stimulation or TMS. Note that this charge contribution may be quite large in the present problem, close to the bridging capacitors.
We have also performed full-wave ANSYS ED simulations of this coil and found that the capacitor voltage drop E-field does not significantly affect the E-field within the patient. The externally applied conservative E-field is expelled from the patient by the high conductivity of tissues. Only the E-field induced from the B-field is important.

5.2.6 Power Amplifier/Driver

In order to create the rotating magnetic and electric fields, as seen in Figs. 5.2a, b, two resonant modes are excited in the coil resonator. These modes display the same current distribution as shown in Fig. 5.3a, but rotated by 90 degrees about the coil axis with respect to each other as well as having an additional temporal phase shift of 90 degrees.
To accomplish this, a custom designed class-D, high-efficiency, single-frequency power amplifier (PA), whose circuit schematic is presented in Fig. 5.4a, was constructed and prototyped, as shown in Fig. 5.4b, c. The upper block in Fig. 5.4a is a class-S modulator, which is followed by two class-D output stages in quadrature, exciting the two resonant modes. The PA has two outputs, one for each resonant mode, and generates a harmonic power RF signal at a fixed carrier frequency at each output. At present, this frequency is typically around 100 kHz. At the same time, the same PA may be tuned to operate at any carrier frequency from 30 kHz to 300 kHz in the LF band. The PA operation, including variable power levels, an optional variable modulation or tonic frequency, and a semiautomatic patient-specific RF frequency tuning procedure, which is automated via a microcomputer board, can be seen in Figs. 5.4c and 5.5d.
The PA output stage is powered by a 3 kW Sorensen DCS 150–20 Variable Regulated DC (direct current) power supply seen at the bottom of Fig. 5.4c. When connected to a standard three-phase 208 VAC outlet, the max RF output power is about 2.9 kW, based on 3 kW DC power. Alternatively, when connected into a single-phase 240 VAC outlet, the max RF power reduces to about 2.3 kW based on 2.4 kW DC power.
Arbitrary modulation (pulse or CW) of the carrier signal with a maximum modulation frequency component of 1 kHz is available via the modulator. The modulation bandwidth is limited mainly by the coil envelope time constant of about 1 ms. Typical modulation is sinusoidal in the 0.5–100 Hz range, generated by the PA firmware.
The PA also monitors its output power and load impedance. It uses this information to automatically adjust the carrier frequency in a narrow band such that the output power remains on target. The amplifier cost, including the DC power supply, is under $10,000. The prototype 100 kHz PA was assembled in a rackmount case shown in Figs. 5.4b, c.
The reason for designing a custom, fixed-frequency PA is the lack of an affordable and appropriate commercial model. Industrial low-frequency RF power supplies, e.g., Comdel’s CLB3000, are costly and require a matched 50 Ω load. Because our load impedance varies widely with frequency, keeping the load matched is a challenge. It would require load impedance monitoring and fine frequency control (potentially difficult with a commercial unit), and/or a software-controlled matching network (costly). Additionally, generating two outputs in quadrature would require either a 90° hybrid (another costly component), or phase-locking two commercial PAs at a 90° phase difference, which can be difficult. Finally, the majority of commercial PAs require water cooling, whereas our PA relies on air cooling. One disadvantage of our custom design is its unknown reliability, a factor that will be proven over time.

5.2.7 Coupling and Matching the Power Amplifier to the Resonating Coil

The amplifier is coupled to the resonating coil inductively via two proximate loops. One such loop is shown in Fig. 5.5c. Apart from certain technical advantages of the inductive coupling, this methodology assures that there is no direct current path from the AC power outlet to the coil. This design enhances overall device safety at any power level, including high-power operation.
The matching network for a single coil port is shown in Fig. 5.5a. Two ports with identical matching networks are located 90° apart around the coil structure, as shown in Fig. 5.5b. The port matching network consists of a series capacitance C1, series inductance L1, and the fixed inductance L2 of the inductive loop seen in Fig. 5.5c.
The matching network is tuned such that the load looks mostly resistive over a small frequency band around the coil’s resonance. For example, the load reactance stays quite low from 99.85 kHz to 100.15 kHz, while the resistance varies from 1.3 Ω to 6 Ω. Because the coil resonance shifts as the coil heats up, the operating frequency must be actively adjusted to compensate for this change, or the output power will vary.
We used two Cornell Dubilier Electronics 940C20S47K-F per rung (C = 0.094 μF per rung). These are 0.047 μF, 2 kV DC, 500 V AC-rated polypropylene film capacitors. They have a typical ESR of 12 mΩ at 100 kHz (Q = 2800), and a max RMS current rating of 5.2 A at 70 °C. We exceed this current rating by about 50% at full power. However, we have measured capacitor temperatures using an IR camera. They are below 70 °C, well within the operating range.
Another important safety feature of the matching network is its benign power envelope step response. The matching network avoids large spikes in PA output current while energy is building up in the resonating coil.
Finally, the matching network presents a sufficiently inductive impedance to higher harmonics of the PA output voltage. This protects the output stage, and ensures that voltage transitions occur when the output current is low, thereby improving efficiency. The efficiency of the PA with the expected load is estimated to be greater than 90% over a wide output power range.

5.2.8 Tuning Procedure

The primary adjustable components are the series capacitance C1 in Fig. 5.5a and the coil rung capacitors at the numbered locations in Fig. 5.5b. First, the coil needs to be manually tuned by installing smaller capacitors in parallel with the primary coil capacitors at strategic locations. Coil adjustments include mode decoupling, tuning of each mode to the same frequency, and impedance matching for each mode (at the coupling loop). Since capacitance is normally only added, the coil only tunes down in frequency.
The semiautomatic tuning procedure implies adjusting PA frequency in a very narrow frequency range. It ensures that the reflection coefficient of both modes stays below −25 dB when matched to the maximum-power coil impedance of Z0 = 2.5 Ω and that both resonances are within a 20 Hz band. The tuning procedure is controlled and guided by the Smith chart/reflection coefficient display of the PA controller seen in Fig. 5.5d. It includes a number of well-defined steps, and is applied to the coil at its designated operating location in an effort to account for the presence of large nearby metal objects. The tuning procedure is simple to perform.

5.2.9 Coil Assembly, Device Setup, and Operation

A resonator coil prototype made of thin-walled, light copper tubing was constructed; it is shown in Figs. 5.6a, b and 5.7. Tubing thickness was kept at ¾ mm or greater to avoid excessive eddy current losses in the copper. The capacitor size in Fig. 5.6a is relatively large since those components must operate at significant currents levels, up to 18 A RMS per rung for a maximum power of 3 kW, and at large voltages, up to 300 V RMS across the capacitor. The total coil weight without the frame is approximately 120 lbs. (54 kg).
This durable coil prototype was then framed, augmented with a horizontal bed, and placed horizontally to enable a subject to rest in the coil, as shown in Fig. 5.6a, which simultaneously shows the complete device setup. The entire coil frame is portable. The distance between the PA, which is connected to the inductive coupling loops of the coil via two isolated cables, can vary from 1 to 3 m, although larger distances may be possible. As mentioned above, there is no direct ohmic current path from the AC power outlet to the coil, which is an important safety feature.
An operator sets the power level, the modulation frequency, and performs RF tuning at the beginning of the resonator operation and for a particular coil load. At the maximum power level, the ring conductors of the coil heat up to approximately 70–75 °C at continuous operation, as illustrated in Fig. 5.8. Continuous coil operation at the maximum input power of 3 kW was tested multiple times with an uninterrupted operation time of up to 2 h and with a cumulative operating time in excess of 100 h.

5.2.10 Quality Factor of the Resonator and the Magnetic Field Strength

The achievable field strength in the coil is determined by three factors: the strength of the PA, the quality factor Q or the “gain” of the resonator, and the coil volume. When the quality factor is high, large field values within the coil can be achieved at a modest input power.
When measured across one of its rung capacitors, the birdcage coil behaves like a parallel resonator in a narrow frequency range around the resonant mode. Using a setup with a signal generator and oscilloscope, the resonator’s quality factor has been estimated in the form:
$$ Q=\frac{f_0}{\left({f}_{\mathrm{U}}-{f}_{\mathrm{L}}\right)\left(1-\frac{V_1}{V_0}\right)} $$
(5.7)
where f0 is the resonance frequency and voltage V0 is the open-circuited generator voltage. Derivation of Eq. (5.7) is given in Appendix A. Voltage V1 is measured at resonance (where it is maximized). fL and fU are the lower and upper frequencies, respectively, where voltage V1 drops by 3 dB from its peak at resonance. This method is accurate in the high Q limit. The experimental data for 100 kHz and 145 kHz are given in Table 5.1. Table 5.1 reports a Q-factor value of about 300 and a minimum difference between loaded (with a human body) and unloaded coil, which is to be expected at this low frequency. These values agree with the theoretical/simulation predictions to within 10%. With decreasing frequency, the Q-factor will decrease approximately proportional to the square root of the resonant frequency.
Table 5.1
Measured Q-factors for the coil resonator at 100 kHz and 145 kHz, respectively
Coil
f0, kHz
fL, kHz
fU, kHz
Q
Unloaded
145.30
144.567
145.875
295.8
Loaded
145.28
144.566
145.879
292.2
Unloaded
101.42
101.008
101.827
277.3
Loaded
101.43
101.009
101.835
275.0
The load is a 200 lb. subject
The established quality factor values are superior to the values reported in the literature for known low-frequency resonator coils (used for low-field MRI) in Table 5.2. Note that all listed competitors have a much smaller coil size/volume and typically a lower quality factor.
Table 5.2
Characteristics of existing low-frequency RF coils given for comparison with the present resonator prototype
Ref.#
Type of the coil
Frequency, kHz
Q (unloaded)
[83]
Wound birdcage coil (84 mm long and has a diameter of 73 mm)
386
180
[84]
Wound birdcage coils and a solenoid. The diameter and the length of the coils are 70 mm
238/425
100–280
[85]
27 tTurn saddle coil made of Litz wire with 8 cm diameter
373
105
[86]
4-Coil Whiting-Lee configuration, 33 cm long
83.6
100
[87]
Solenoidal coils; 6–46 cm in length and 4–52 cm in diameter
210/275
60–30, reduced Q
[88]
Cylindrical saddle-shaped loops (5 saddle pairs of 10 turns each), coil diameter is 26 mm
87
NA
It is important to point out again that the quality factor in Table 5.1 is weakly affected by body loading, in contrast to conventional high-frequency MRI RF coils. This observation, also mentioned in Ref. [84] and other sources, is a limitation of the present electromagnetic stimulator. The RF power losses are mostly in the coil itself, and not in the human body.
B-field measurements have been performed via a calibrated single-axis coil probe located at the coil axis. The B-field magnitude was 1.01 mT at the coil center and at full power (3 kW DC) at 100 kHz. The measured and theoretical results differ by no more than 10%. At the full input power level of 3 kW, the amplitude of the resonant ring current I0 at 100 kHz in two coil rings reaches 603 A, while the amplitude of the rung current reaches 26 A.

5.3 Device Safety Estimates

5.3.1 Peripheral Nervous System (PNS) Stimulation Threshold

The present low-frequency subthreshold electrostimulation device must not exceed the PNS simulation threshold to operate safely and without unpleasant sensation. Guidelines from the International Commission on Non-Ionizing Radiation Protection (Table 5.2 of [91]) require the occupational exposure to an electric field to be limited to a value of approximately 27 V/m RMS at 100 kHz and by a value of 39 V/m RMS at 145 kHz (the so-called basic restrictions [91]). These restrictions are mainly due to limits on peripheral nerve stimulation [91] and should therefore be respected. Other relevant results on the PNS stimulation thresholds at lower frequencies are presented in Refs. [9294].

5.3.2 Specific Absorption Rate (SAR)

Safety estimates also rely upon the levels of the specific absorption rate (SAR) within the body. The SAR is the energy absorption rate that causes body temperature to rise due to an imposed electromagnetic field. The maximum value of SAR1g in the body must be below 10 W/kg required by the FDA-accepted safety standard [95, 96]. The global-body SAR must be below the 2 W/kg limit [95, 96].

5.3.3 Method of Analysis

SAR and electric field measurements cannot be performed easily for human subjects in vivo. SAR and device performance estimates are typically derived and accepted today from computational electromagnetics (CEM) simulations performed with detailed virtual humans [97]. In this study, we use the multi-tissue CEM phantom VHP-Female v. 5.0 (female/60 year/162 cm/88 kg, obese) [97103] derived from the cryosection dataset archived within the Visible Human Project® of the US National Library of Medicine [104]. The phantom includes about 250 individual tissues and is augmented with material property values from the IT’IS database [105]. The average-body conductivity is assigned as 0.25 S/m, which reflects a mixture of muscle and fat.
The primary CEM software used in this study is the accurate commercial FEM solver ANSYS® Electromagnetic Suite 18.2.0 with rigorous adaptive mesh refinement. In addition, and for verification/validation purposes, we employ an in-house boundary element fast multipole method (BEM-FMM) described in Ref. [89]. In the latter case, a higher near-surface resolution can be achieved and the original surface phantom model can be refined and smoothed from approximately 0.5 M triangles to 3.5 M triangles. In this particular study, the human model is placed in the coil at the shoulder landmark, as shown in Fig. 5.9, so that the top of the shoulder coincides with the ring plane. Other configurations have also been considered.
Results obtained with both software packages differ by no more than 2% in the unloaded coil (field at the coil center) and by no more than 25–50% in the coil loaded with the multi-tissue human body. The latter deviation may be explained by somewhat different surface meshes.
Below we report simulations at two power levels: 1.5 kW input power and 3 kW input power. The first power level is the half power level of the amplifier driver; the second power level corresponds to full power. At full power level, the amplitude of the resonant ring current I0 in Eqs. (5.4 and 5.5) reaches 603 A, while the amplitude of the rung current reaches 26 A.

5.3.4 Electric Field Levels

Figure 5.10 shows the simulated RMS levels of the electric field in the body at 100 kHz and at the input power level of 1.5 kW obtained via the BEM-FMM simulations. We observe that, at half power level of the amplifier driver, the fields everywhere in the body do not generally exceed 30 V/m RMS and are thus within the ICNIRP guidelines. As expected, higher field levels are observed closer to the surface; the field gradually decreases toward the center of the body.
Quantitative estimates of the average electric field for every particular tissue obtained via ANSYS Electromagnetic Suite 18.2.0 are given in Table. 5.3. Note the lower electric fields in the intracranial volume. Additionally, we observe higher electric fields in the individual body muscles. It is also interesting to observe that the fields in bone may be quite high, in particular in the femur and pelvic bones.
Table 5.3
Computed electric field levels (V/m RMS) in every individual tissue at 1.5 kW input power (ANSYS® Electromagnetic Suite 18.2.0)
Mesh
Tissue
Avg. E field (V/m RMS)
Mesh
Tissue
Avg. E field (V/m RMS)
1
Air Internal Maxillary Sinus Left
7.7
39
Cuneiform Medial right
0.6
2
Air Internal Maxillary Sinus Right
6.9
40
discC02C03
10.3
3
Arteries
10.5
41
discC03C04
11.6
4
Bladder
28.0
42
discC04C05
14.3
5
C01
14.5
43
discC05C06
16.9
6
C02
13.6
44
discC06C07
19.0
7
C03
14.8
45
discC07T01
20.6
8
C04
18.5
46
discL01L02
13.9
9
C05
21.7
47
discL02L03
11.5
10
C06
26.1
48
discL03L04
8.4
11
C07
29.6
49
discL04L05
4.7
12
Calcaneous left
0.6
50
discL05L06
7.1
13
Calcaneous right
1.1
51
discL06S00
13.8
14
Cartilage1 Left
18.5
52
discT01T02
20.2
15
Cartilage1 Right
19.9
53
discT02T03
17.6
16
Cartilage2 Left
19.7
54
discT03T04
17.9
17
Cartilage2 Right
20.3
55
discT04T05
17.4
18
Cartilage3 Left
21.2
56
discT05T06
15.8
19
Cartilage3 Right
20.9
57
discT06T07
15.2
20
Cartilage4 Left
22.5
58
discT07T08
14.4
21
Cartilage4 Right
21.8
59
discT08T09
13.6
22
Cartilage5 Left
24.3
60
discT09T10
13.2
23
Cartilage5 Right
22.6
61
discT10T11
12.5
24
Cartilage6 Left
36.3
62
discT11T12
13.2
25
Cartilage6 Right
35.9
63
discT12L01
13.4
26
Cerebellum
1.4
64
Eye Left
5.0
27
Clavicle left
55.6
65
Eye Right
5.3
28
Clavicle right
34.9
66
Feet1Phalange left
0.5
29
Coccyx
42.6
67
Feet1Phalange right
0.4
30
CSF OuterShell
3.5
68
Feet2Phalange left
0.4
31
CSF Ventricles
0.4
69
Feet2Phalange right
0.4
32
Cuboid Left
0.9
70
Feet3Phalange left
0.3
33
Cuboid Right
0.6
71
Feet3Phalange right
0.4
34
Cuneiform Intermediate left
1.3
72
Feet4Phalange left
0.4
35
Cuneiform Intermediate right
0.5
73
Feet4Phalange right
0.6
36
Cuneiform Lateral left
1.1
74
Feet5Phalange left
0.4
37
Cuneiform Lateral right
0.4
75
Feet5Phalange right
0.7
38
Cuneiform Medial left
1.3
76
Femur Bone Marrow Left
7.1
77
Femur Bone Marrow Right
8.6
117
Humerus right
23.1
78
Femur left
69.3
118
Intestine
20.6
79
Femur right
83.7
119
Jaw lower
10.0
80
Fibula left
5.9
120
Kidney left
29.3
81
Fibula right
5.4
121
Kidney right
27.2
82
Gray Matter Spinal Cord
1.5
122
L01
27.9
83
Hands1 1Phalange left
10.2
123
L02
25.3
84
Hands1 1Phalange right
9.5
124
L03
22.1
85
Hands1 2Phalange left
9.5
125
L04
19.0
86
Hands1 2Phalange right
12.1
126
L05
14.4
87
Hands1 3Phalange left
12.2
127
L06
17.9
88
Hands1 3Phalange right
13.4
128
Liver
29.8
89
Hands2 1Phalange left
7.6
129
Lungs
19.4
90
Hands2 1Phalange right
7.1
130
Median Nerve left
11.6
91
Hands2 2Phalange left
8.1
131
Median Nerve right
13.0
92
Hands2 2Phalange right
8.5
132
Muscle Bicep left
11.9
93
Hands2 3Phalange left
7.1
133
Muscle Bicep right
12.9
94
Hands2 3Phalange right
9.5
134
Muscle Calf left
5.0
95
Hands3 1Phalange left
6.4
135
Muscle Calf right
5.2
96
Hands3 1Phalange right
6.2
136
Muscle Deltoid left
18.7
97
Hands3 2Phalange left
8.1
137
Muscle Deltoid right
19.3
98
Hands3 2Phalange right
8.6
138
Muscle Erector spinae left
26.8
99
Hands3 3Phalange left
9.3
139
Muscle Erector spinae right
26.9
100
Hands3 3Phalange right
11.0
140
Muscle Forearm Extensors left
6.9
101
Hands4 1Phalange left
7.2
141
Muscle Forearm Extensors right
8.6
102
Hands4 1Phalange right
6.9
142
Muscle Forearm Flexors left
6.9
103
Hands4 2Phalange left
10.4
143
Muscle Forearm Flexors right
7.2
104
Hands4 2Phalange right
10.0
144
Muscle Gluteus left
27.9
105
Hands4 3Phalange left
11.1
145
Muscle Gluteus right
27.2
106
Hands4 3Phalange right
10.7
146
Muscle Hamstring left
18.9
107
Hands5 1Phalange left
9.0
147
Muscle Hamstring right
19.1
108
Hands5 1Phalange right
10.3
148
Muscle Latissimus Dorsi left
36.6
109
Hands5 2Phalange left
11.0
149
Muscle Latissimus Dorsi right
38.5
110
Hands5 2Phalange right
12.6
150
Muscle Neck Combined left
13.6
111
Hands5 3Phalange left
10.3
151
Muscle Neck Combined right
13.4
112
Hands5 3Phalange right
12.1
152
Muscle Obliques left
39.5
113
Heart Muscle
14.2
153
Muscle Obliques right
40.1
114
Hip left
60.0
154
Muscle Pectoralis major left
21.7
115
Hip right
61.4
155
Muscle Pectoralis major right
20.9
116
Humerus left
20.7
156
Muscle Pectoralis minor left
19.2
157
Muscle Pectoralis minor right
18.6
194
Ribs left8
47.3
158
Muscle Pelvic Combined left
25.7
195
Ribs left9
46.1
159
Muscle Pelvic Combined right
25.0
196
Ribs left10
48.5
160
Muscle Psoas left
13.9
197
Ribs left11
51.7
161
Muscle Psoas right
13.9
198
Ribs left12
39.0
162
Muscle Quadriceps left
20.2
199
Ribs right1
29.6
163
Muscle Quadriceps right
19.9
200
Ribs right2
26.2
164
Muscle Rectus Abdominis left bottom
32.3
201
Ribs right3
25.2
165
Muscle Rectus Abdominis left middle
34.9
202
Ribs right4
26.1
166
Muscle Rectus Abdominis left top
39.1
203
Ribs right5
27.2
167
Muscle Rectus Abdominis right bottom
32.5
204
Ribs right6
29.9
168
Muscle Rectus Abdominis right middle
35.4
205
Ribs right7
35.6
169
Muscle Rectus Abdominis right top
38.1
206
Ribs right8
43.2
170
Muscle Sartorius left
18.6
207
Ribs right9
53.9
171
Muscle Sartorius right
17.5
208
Ribs right10
58.9
172
Muscle Tibialis Anterior left
6.2
209
Ribs right11
56.9
173
Muscle Tibialis Anterior right
5.8
210
Ribs right12
40.9
174
Muscle Trapezius left
23.6
211
Sacrum
45.7
175
Muscle Trapezius right
24.0
212
Scapula left
38.2
176
Muscle Tricep left
12.0
213
Scapula right
38.8
177
Muscle Tricep right
14.0
214
Skin Shell
27.8
178
Navicular left
1.8
215
Skull
22.8
179
Navicular right
0.7
216
Sphenoid
8.9
180
Patella left
24.3
217
Spleen
33.9
181
Patella right
22.6
218
Sternum
25.2
182
Peripheral Nerve left
17.1
219
Stomach
22.6
183
Peripheral Nerve Right
14.1
220
T01
28.4
184
Pubic Symphysis
32.1
221
T02
27.2
185
Radial Nerve left
14.6
222
T03
27.2
186
Radial Nerve right
12.4
223
T04
26.9
187
Ribs left1
26.4
224
T05
25.7
188
Ribs left2
30.1
225
T06
25.5
189
Ribs left3
26.4
226
T07
26.2
190
Ribs left4
26.7
227
T08
26.4
191
Ribs left5
28.3
228
T09
26.9
192
Ribs left6
31.5
229
T10
26.1
193
Ribs left7
37.3
230
T11
26.4
231
T12
27.5
240
Trabecular upper right
0.9
232
Talus left
1.3
241
Trachea Sinus
12.4
233
Talus right
0.6
242
Ulna Radius left
8.1
234
Tibia left
8.3
243
Ulna Radius right
7.8
235
Tibia right
7.9
244
Uterus
17.3
236
Tongue
5.2
245
Veins lower
12.5
237
Trabecular lower left
0.5
246
Veins upper
12.4
238
Trabecular lower right
0.8
247
White Matter
1.0
239
Trabecular upper left
0.7
However, the computed local electric fields may considerably exceed the values reported in Table 5.3, in particular by 1.5–6 times. These peak values are less accurate. One potential source of the numerical error is insufficient resolution of lengthy and time-consuming full-body computations very close to the interfaces where higher fields are usually observed.

5.3.5 SAR Levels

The body-averaged or whole-body (global-body) SARbody is given by averaging the local SAR over the entire body volume. In terms of the complex field phasor E(r), one has
$$ SA{R}_{\mathrm{body}}=\frac{1}{V_{\mathrm{body}}}{\int}_{V_{\mathrm{body}}}\frac{\sigma (r)}{2\rho (r)}\kern0.5em E(r)\cdot E{(r)}^{\ast } dV $$
(5.8)
Here, σ(r) is the local tissue conductivity and ρ(r) is the local mass density. At full power of 3 kW and positioned at the shoulder landmark, the global-body SAR computed via ANSYS Electromagnetic Suite 18.2.0 is 0.25 W/kg. Thus, the total power dissipation in the body does not exceed 30 W, i.e., 1% of the total power. The same percentage ratio is valid at half input power.
The second critical estimate is SAR1g, which is given by averaging over a contiguous volume with the weight of 1 g,
$$ SA{R}_{1g}(r)=\frac{1}{V_{1g}}{\int}_{V_{1g}}\frac{\sigma (r)}{2\rho (r)}\kern0.5em E(r)\cdot E{(r)}^{\ast } dV $$
(5.9)
The maximum value of SAR1g in the body computed via ANSYS Electromagnetic Suite 18.2.0 at the full power of 3 kW and located at the shoulder landmark is 4.55 W/kg.
Although this last value might appear to be relatively high, it is still within the corresponding SAR limits in MRI machines [95, 96]. In particular, the major applicable MRI safety standard, issued by the International Electrotechnical Commission (IEC) and also accepted by the U.S. Food and Drug Administration, in the normal mode (mode of operation that causes no physiological stress to patients) limits global-body SAR to 2 W/kg, global-head SAR to 3.2 W/kg, local head and torso SAR to 10 W/kg, and local extremity SAR to 20 W/kg [96]. The global SAR limits are intended to ensure a body core temperature of 39 °C or less [95, 96].

5.4 Discussion

5.4.1 Efficacy of Stimulation

The present study establishes safety and potential feasibility of the resonant neurostimulation device. However, its efficacy for treatment of chronic back pain remains largely unknown. Only clinical trials, which would ideally thoroughly investigate both short-term and cumulative effects of the suggested lower-body electromagnetic treatment, could probably answer this question. Our aim is to provide a doctor with the possibility to vary power, resonant frequency, tonic frequency, and electromagnetic pulse envelope to enable the best possible outcome during the anticipated clinical trial.

5.4.2 Integrated Effect of Stimulation

The present subthreshold stimulation device will not only affect the PNS of the lower back but also muscles, bones, tendons, and cartilage. Evidence suggests that subthreshold pulsed electromagnetic fields may stimulate osteogenesis in vitro and in vivo [106, 107], improve bone quality in osteoporotic and nonosteoporotic cell-based studies [108, 109], human studies [110114], animal studies [115121], and augment bone fracture healing [107, 122124]. Further evidence suggests that TENS therapy stimulates a change in the biochemical and physiological muscle conditions that may lead to muscle relaxation [125, 126]. Some evidence also suggests that the kHz stimulation of the lower body will increase the vascular endothelial growth factor receptor on circulating hematopoietic stem cells [81], whose local niche (the bone marrow of the pelvis, femur, and sternum [127]) might be well affected by the present stimulation device.
Another extremely interesting effect of the kHz peripheral nerve stimulation observed previously in [65, 80] and implicitly in the present device is a potential for sleep improvement. It is not clear how to describe and account these integrated effects of the stimulation. We will attempt to carefully document and report prior relevant literature findings and the corresponding stimulation conditions for nerve/muscle/bone/marrow, and link them to the present stimulation conditions.

5.4.3 Operation as an EMAT

The present electrostimulation device may also operate as an electromagnetic acoustic transducer (EMAT) when a DC current is injected into the tissue via surface electrodes at a specified location. The Lorentz force will excite an ultrasonic field whose frequency is the resonant frequency.

5.4.4 Variation of Resonant Frequency

While fine tuning with a low-loss ferromagnetic load is straightforward, it is quite challenging, however, to vary the resonant frequency of a power resonator, which is usually cast in stone, allowing only a narrow tuning range. In order to do this, we have studied three different methods: a bank of electronically controlled switched power capacitors, a bank of fixed capacitors with low-resistance power relay switches, and a mechanically replaceable bolted joints-based fixed-capacitor bank. Such banks need be constructed for each of the 144 rungs of the coil in Fig. 5.6 or Fig. 5.7 in order to vary the resonant or carrier frequency over the band of, say, 10–100 kHz.
Although the first two approaches are fast and elegant, they are unfeasible. The key is the equivalent series resistance (ESR) of switched capacitors and power relay switches. Existing series switches increase ESR by about 10 times or even more. This dramatically lowers the resonator quality factor Q, resulting in about three times lower field values for a given input power. The switched capacitor solution has other serious drawbacks. Each switched capacitor block is much larger than a fixed capacitor, which will result in issues related to physically accommodating all components. On the other hand, for a relay with an exceptionally small contact resistance of 5 mΩ, Q will change by a factor of 0.6 at 100 kHz and 0.3 at 10 kHz. As the relays cycle, their contact resistance may go up significantly, especially if we do not follow the guidelines for minimum switched current (to create an arc that cleans the contacts). Thus, Q could continue dropping with cycling. Therefore, we plan to implement low-ESR mechanically replaceable bolted-joints based fixed-capacitor banks. The frequency-switching operation will take approximately 3 h.

5.5 Conclusion

In this technical study, we described a whole-body noncontact subthreshold electromagnetic stimulation device based on the concept of a familiar MRI RF resonating coil, but at a much lower resonant frequency (100–150 kHz and potentially down to 10 kHz), with a field modulation option (0.5–100 Hz), and with an input power level of up to 3 kW. Its unique features include a relatively high electric field level within the subject’s biological tissue due to the resonant effect but at low power dissipation, or SAR level, in the body itself.
We emphasize that in the low-frequency limit and at moderate field levels, SAR rather weakly correlates with the deposited electric field. One reason for this is that SAR is proportional to the field squared, and is thus quite small at moderate and low field levels. A second reason is that the tissue conductivity itself is lower (at 100 kHz, it is twice as low as at 100 MHz for muscle and five times lower for fat [105]).
Due to the large resonator volume and its noncontact nature, the subject may be conveniently located anywhere within the resonating coil over a prolonged period of time at moderate and safe electric field levels. The electric field effect does not depend on a particular body position within the resonator. The field penetration is deep everywhere in the body, including the extremities; muscles, bones, and peripheral tissues are mostly affected. Over a shorter period of time, the electric field levels could be increased to relatively large values with an amplitude of about 1 V/cm.
We envision treatment of chronic pain, and particularly neuropathic pain, as the primary potential clinical application for the device. The device enables whole-body coverage, which could be useful in the treatment of widespread pain conditions, such as painful polyneuropathy or fibromyalgia. In addition, a deeper tissue penetration can be achieved without side-effects caused by high current density in the skin associated with the traditional contact electrodes of TENS. It should be noted that these potential clinical applications are speculative and warrant empirical testing in the future.
Considerable attention has been paid to device safety including both the AC power safety and human exposure to electromagnetic fields. In the former case, we have used inductive coupling, which assures that there is no direct current path from the AC power outlet to the coil. This design enhances overall device safety at any power level, including high-power operation. As with more traditional MRI devices, no large metal objects should be located in the immediate vicinity of the coil.
Human exposure to the electromagnetic field within the coil has been evaluated by performing extensive modeling with two independent numerical methods and with an anatomically realistic multi-tissue human phantom. We have shown that the SAR levels within the body correspond to the safety standards of the International Electrotechnical Commission when the input power level of the amplifier driver does not exceed 3 kW. We have also shown that the electric field levels generally comply with the safety standards of the International Commission on Non-Ionizing Radiation Protection when the input power level of the amplifier driver does not exceed 1.5 kW.

Acknowledgments

The authors are thankful to Dr. James O’Rourke, Dr. John McNeill, Ms. Leah Morales, Mr. Brandon Weyant (all from Worcester Polytechnic Institute), MD Irina V. Zhdanova (ClockCoach), and Dr. Aapo Nummenmaa (Massachusetts General Hospital) for useful discussions. Dr. Deng is supported by the Intramural Research Program of the National Institute of Mental Health, NIH.
Open Access This chapter is licensed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license and indicate if changes were made.
The images or other third party material in this chapter are included in the chapter's Creative Commons license, unless indicated otherwise in a credit line to the material. If material is not included in the chapter's Creative Commons license and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder.
Anhänge

Appendix A: Derivation of Eq. (5.7) and Coil Q

The corresponding measurement circuit is given in Fig. 5.11.
The derivation of Eq. (5.7) is as follows. The coil, when measured at a rung capacitor, looks like a parallel resonator (assuming the two degenerate modes are decoupled). In a narrow band around the high-Q resonance, the parallel resonator impedance can be approximated as:
$$ {Z}_{\mathrm{res}}=\frac{R_1}{1+j2Q\frac{\varDelta f}{f_0}} $$
(5.A1)
where f0 is the resonance frequency and Δf is the deviation from it. Then,
$$ \frac{V_1}{V_0}=\frac{Z_{\mathrm{res}}}{R_0+{Z}_{\mathrm{res}}} $$
(5.A2)
Taking the absolute value and simplifying, one has
$$ \mid \frac{V_1}{V_0}\mid =\frac{R_1}{\sqrt{{\left({R}_0+{R}_1\right)}^2+4{R}_0^2{Q}^2\frac{\varDelta {\mathrm{f}}^2}{f_0^2}}} $$
(5.A3)
At resonance,
$$ \frac{V_1}{V_0}=\frac{R_1}{R_0+{R}_1} $$
(5.A4)
At the V1 3 dB frequencies,
$$ \frac{R_1}{\sqrt{{\left({R}_0+{R}_1\right)}^2+4{R}_0^2{Q}^2\frac{\varDelta {f}_{3\mathrm{dB}}^2}{f_0^2}}}=\frac{R_1}{\sqrt{2}\left({R}_0+{R}_1\right)} $$
(5.A5)
Solving for Q yields
$$ Q=\frac{f_0}{2\varDelta {f}_{3\mathrm{dB}}}\left(\frac{R_0+{R}_1}{R_0}\right) $$
(5.A6)
From V1/V0 at resonance, we can express
$$ 1-\frac{V_1}{V_0}=1-\frac{R_1}{R_0+{R}_1}=\frac{R_0}{R_0+{R}_1} $$
(5.A7)
Therefore,
$$ Q=\frac{f_0}{2\varDelta {f}_{3\mathrm{dB}}\left(1-\frac{V_1}{V_0}\right)} $$
(5.A8)
where V1 is measured at resonance, and 2Δf3dB = fU − fL, with fU and fL being the upper and lower 3 dB frequencies for V1.
Regarding the envelope time constant
$$ \tau =\frac{2Q}{\omega } $$
(5.A9)
we mention that for Q = 300 and f = 100 kHz, τ = 0.955 ms. This can be rounded to 1 ms if high precision is not needed. We do not measure Q through this time constant. It is only mentioned as a limiter of modulation bandwidth.
The quality factor is difficult to predict accurately. Estimated Q from ANSYS HFSS-circuit co-simulation typically comes out about 10–30% higher than in the real circuit. How much higher depends on the level of refinement of the FEM model. The important parameter is the actual achieved Q. ANSYS can still be used for rough coil Q optimization, as long as the level of refinement is kept about the same.
Literatur
1.
Zurück zum Zitat The Americian Academy of Pain Medicine. (2011). Relieving pain in America: A blueprint for transforming prevention, care, education, and research. Institute of Medicine. Committee on Advancing Pain Research, Care, and Education. Washington (DC): National Academies Press (US). ISBN 978–0–309-21484-1. The Americian Academy of Pain Medicine. (2011). Relieving pain in America: A blueprint for transforming prevention, care, education, and research. Institute of Medicine. Committee on Advancing Pain Research, Care, and Education. Washington (DC): National Academies Press (US). ISBN 978–0–309-21484-1.
5.
Zurück zum Zitat Nijs, J., Kosek, E., Van Oosterwijck, J., & Meeus, M. (2012). Dysfunctional endogenous analgesia during exercise in patients with chronic pain: to exercise or not to exercise? Pain Physician, 1, 205–214. Nijs, J., Kosek, E., Van Oosterwijck, J., & Meeus, M. (2012). Dysfunctional endogenous analgesia during exercise in patients with chronic pain: to exercise or not to exercise? Pain Physician, 1, 205–214.
6.
Zurück zum Zitat Kuppens, K., Hans, G., Roussel, N., Struyf, F., Fransen, E., Cras, P., Van Wilgen, C. P., & Nijs, J. (2018). Sensory processing and central pain modulation in patients with chronic shoulder pain: A case-control study. Scandinavian Journal of Medicine & Science in Sports, 28(3), 1183–1192. https://doi.org/10.1111/sms.12982.CrossRef Kuppens, K., Hans, G., Roussel, N., Struyf, F., Fransen, E., Cras, P., Van Wilgen, C. P., & Nijs, J. (2018). Sensory processing and central pain modulation in patients with chronic shoulder pain: A case-control study. Scandinavian Journal of Medicine & Science in Sports, 28(3), 1183–1192. https://​doi.​org/​10.​1111/​sms.​12982.CrossRef
7.
Zurück zum Zitat Nakamura, M., Nishiwaki, Y., Sumitani, M., Ushida, T., Yamashita, T., Konno, S., Taguchi, T., & Toyama, Y. (2014). Investigation of chronic musculoskeletal pain (third report): with special reference to the importance of neuropathic pain and psychogenic pain. Journal of Orthopaedic Science, 19(4), 667–675. https://doi.org/10.1007/s00776-014-0567-6.CrossRef Nakamura, M., Nishiwaki, Y., Sumitani, M., Ushida, T., Yamashita, T., Konno, S., Taguchi, T., & Toyama, Y. (2014). Investigation of chronic musculoskeletal pain (third report): with special reference to the importance of neuropathic pain and psychogenic pain. Journal of Orthopaedic Science, 19(4), 667–675. https://​doi.​org/​10.​1007/​s00776-014-0567-6.CrossRef
13.
Zurück zum Zitat National Inst. of Neurological Disorders and Stroke. Low back pain fact sheet. 2018-08-07. NIH Publication No. 15–5161. National Inst. of Neurological Disorders and Stroke. Low back pain fact sheet. 2018-08-07. NIH Publication No. 15–5161.
14.
Zurück zum Zitat Krath, A., Klüter, T., Stukenberg, M., Zielhardt, P., Gollwitzer, H., Harrasser, N., Hausdorf, J., Ringeisen, M., & Gerdesmeyer, L. (2017). Electromagnetic transduction therapy in non-specific low back pain: A prospective randomised controlled trial. Journal of Orthopaedics, 14(3), 410–415. https://doi.org/10.1016/j.jor.2017.06.016.CrossRef Krath, A., Klüter, T., Stukenberg, M., Zielhardt, P., Gollwitzer, H., Harrasser, N., Hausdorf, J., Ringeisen, M., & Gerdesmeyer, L. (2017). Electromagnetic transduction therapy in non-specific low back pain: A prospective randomised controlled trial. Journal of Orthopaedics, 14(3), 410–415. https://​doi.​org/​10.​1016/​j.​jor.​2017.​06.​016.CrossRef
16.
Zurück zum Zitat Outcalt, S. D., Kroenke, K., Krebs, E. E., Chumbler, N. R., Wu, J., Yu, Z., & Bair, M. J. (2015). Chronic pain and comorbid mental health conditions: Independent associations of posttraumatic stress disorder and depression with pain, disability, and quality of life. Journal of Behavioral Medicine, 38(3), 535–543. https://doi.org/10.1007/s10865-015-9628-3.CrossRef Outcalt, S. D., Kroenke, K., Krebs, E. E., Chumbler, N. R., Wu, J., Yu, Z., & Bair, M. J. (2015). Chronic pain and comorbid mental health conditions: Independent associations of posttraumatic stress disorder and depression with pain, disability, and quality of life. Journal of Behavioral Medicine, 38(3), 535–543. https://​doi.​org/​10.​1007/​s10865-015-9628-3.CrossRef
18.
Zurück zum Zitat Iizuka, Y., Iizuka, H., Mieda, T., Tsunoda, D., Sasaki, T., Tajika, T., Yamamoto, A., & Takagishi, K. (2017). Prevalence of chronic nonspecific low back pain and its associated factors among middle-aged and elderly people: an analysis based on data from a musculoskeletal examination in Japan. Asian Spine Journal, 11(6), 989–997. https://doi.org/10.4184/asj.2017.11.6.989.CrossRef Iizuka, Y., Iizuka, H., Mieda, T., Tsunoda, D., Sasaki, T., Tajika, T., Yamamoto, A., & Takagishi, K. (2017). Prevalence of chronic nonspecific low back pain and its associated factors among middle-aged and elderly people: an analysis based on data from a musculoskeletal examination in Japan. Asian Spine Journal, 11(6), 989–997. https://​doi.​org/​10.​4184/​asj.​2017.​11.​6.​989.CrossRef
19.
Zurück zum Zitat Dworkin, R. H., O’Connor, A. B., Backonja, M., Farrar, J. T., Finnerup, N. B., Jensen, T. S., Kalso, E. A., Loeser, J. D., Miaskowski, C., Nurmikko, T. J., Portenoy, R. K., Rice, A. S., Stacey, B. R., Treede, R. D., Turk, D. C., & Wallace, M. S. (2007). Pharmacologic management of neuropathic pain: Evidence-based recommendations. Pain, 132(3), 237–251. https://doi.org/10.1016/j.pain.2007.08.033.CrossRef Dworkin, R. H., O’Connor, A. B., Backonja, M., Farrar, J. T., Finnerup, N. B., Jensen, T. S., Kalso, E. A., Loeser, J. D., Miaskowski, C., Nurmikko, T. J., Portenoy, R. K., Rice, A. S., Stacey, B. R., Treede, R. D., Turk, D. C., & Wallace, M. S. (2007). Pharmacologic management of neuropathic pain: Evidence-based recommendations. Pain, 132(3), 237–251. https://​doi.​org/​10.​1016/​j.​pain.​2007.​08.​033.CrossRef
24.
Zurück zum Zitat Koca, I., Boyaci, A., Tutoglu, A., Ucar, M., & Kocaturk, O. (2014). Assessment of the effectiveness of interferential current therapy and TENS in the management of carpal tunnel syndrome: a randomized controlled study. Rheumatology International, 34, 1639–1645. https://doi.org/10.1007/s00296-014-3005-3.CrossRef Koca, I., Boyaci, A., Tutoglu, A., Ucar, M., & Kocaturk, O. (2014). Assessment of the effectiveness of interferential current therapy and TENS in the management of carpal tunnel syndrome: a randomized controlled study. Rheumatology International, 34, 1639–1645. https://​doi.​org/​10.​1007/​s00296-014-3005-3.CrossRef
25.
29.
Zurück zum Zitat Makarov, S. N., Bogdanov, G., Makarov, V. S., Noetscher, G. M., & Deng, Z-D. (2018). A whole body non-contact electrical stimulation device with variable parameters. US Patent Application #15868038 of Jan. 11th 2018. Pub # 20180196113; Pub Date July 12th 2018. Revision Oct 2018. Makarov, S. N., Bogdanov, G., Makarov, V. S., Noetscher, G. M., & Deng, Z-D. (2018). A whole body non-contact electrical stimulation device with variable parameters. US Patent Application #15868038 of Jan. 11th 2018. Pub # 20180196113; Pub Date July 12th 2018. Revision Oct 2018.
30.
Zurück zum Zitat Makarov, S. N., Bogdanov, G., Noetscher, G. M., Allpleyard, W., Ludwig, R., Joutsa, U. T., & Deng, Z.-D. (2018). Design and analysis of a whole body non-contact electromagnetic stimulation device with field modulation at low and medium power levels. bioRxiv, 416065. https://doi.org/10.1101/416065. Makarov, S. N., Bogdanov, G., Noetscher, G. M., Allpleyard, W., Ludwig, R., Joutsa, U. T., & Deng, Z.-D. (2018). Design and analysis of a whole body non-contact electromagnetic stimulation device with field modulation at low and medium power levels. bioRxiv, 416065. https://​doi.​org/​10.​1101/​416065.
36.
Zurück zum Zitat O’Reardon, J. P., Solvason, H. B., Janicak, P. G., Sampson, S., Isenberg, K. E., Nahas, Z., McDonald, W. M., Avery, D., Fitzgerald, P. B., Loo, C., Demitrack, M. A., George, M. S., & Sackeim, H. A. (2007). Efficacy and safety of transcranial magnetic stimulation in the acute treatment of major depression: a multisite randomized controlled trial. Biological Psychiatry, 62(11), 1208–1216. https://doi.org/10.1016/j.biopsych.2007.01.018.CrossRef O’Reardon, J. P., Solvason, H. B., Janicak, P. G., Sampson, S., Isenberg, K. E., Nahas, Z., McDonald, W. M., Avery, D., Fitzgerald, P. B., Loo, C., Demitrack, M. A., George, M. S., & Sackeim, H. A. (2007). Efficacy and safety of transcranial magnetic stimulation in the acute treatment of major depression: a multisite randomized controlled trial. Biological Psychiatry, 62(11), 1208–1216. https://​doi.​org/​10.​1016/​j.​biopsych.​2007.​01.​018.CrossRef
37.
Zurück zum Zitat Bajbouj, M., Merkl, A., Schlaepfer, T. E., Frick, C., Zobel, A., Maier, W., O’Keane, V., Corcoran, C., Adolfsson, R., Trimble, M., Rau, H., Hoff, H. J., Padberg, F., Müller-Siecheneder, F., Audenaert, K., van den Abbeele, D., Matthews, K., Christmas, D., Eljamel, S., & Heuser, I. (2010). Two-year outcome of vagus nerve stimulation in treatment-resistant depression. Journal of Clinical Psychopharmacology, 30(3), 273–281. https://doi.org/10.1097/JCP.0b013e3181db8831.CrossRef Bajbouj, M., Merkl, A., Schlaepfer, T. E., Frick, C., Zobel, A., Maier, W., O’Keane, V., Corcoran, C., Adolfsson, R., Trimble, M., Rau, H., Hoff, H. J., Padberg, F., Müller-Siecheneder, F., Audenaert, K., van den Abbeele, D., Matthews, K., Christmas, D., Eljamel, S., & Heuser, I. (2010). Two-year outcome of vagus nerve stimulation in treatment-resistant depression. Journal of Clinical Psychopharmacology, 30(3), 273–281. https://​doi.​org/​10.​1097/​JCP.​0b013e3181db8831​.CrossRef
40.
Zurück zum Zitat Jefferys, J. G. (1981). Influence of electric fields on the excitability of granule cells in guinea-pig hippocampal slices. The Journal of Physiology, 319, 143–152.CrossRef Jefferys, J. G. (1981). Influence of electric fields on the excitability of granule cells in guinea-pig hippocampal slices. The Journal of Physiology, 319, 143–152.CrossRef
43.
Zurück zum Zitat Kerezoudis, P., Grewal, S. S., Stead, M., Lundstrom, B. N., Britton, J. W., Shin, C., Cascino, G. D., Brinkmann, B. H., Worrell, G. A., & Van Gompel, J. J. (2018). Chronic subthreshold cortical stimulation for adult drug-resistant focal epilepsy: safety, feasibility, and technique. Journal of Neurosurgery, 129(2), 533–543. https://doi.org/10.3171/2017.5.JNS163134.CrossRef Kerezoudis, P., Grewal, S. S., Stead, M., Lundstrom, B. N., Britton, J. W., Shin, C., Cascino, G. D., Brinkmann, B. H., Worrell, G. A., & Van Gompel, J. J. (2018). Chronic subthreshold cortical stimulation for adult drug-resistant focal epilepsy: safety, feasibility, and technique. Journal of Neurosurgery, 129(2), 533–543. https://​doi.​org/​10.​3171/​2017.​5.​JNS163134.CrossRef
48.
Zurück zum Zitat Fava, M., Freeman, M. P., Flynn, M., Hoeppner, B. B., Shelton, R., Iosifescu, D. V., Murrough, J. W., Mischoulon, D., Cusin, C., Rapaport, M., Dunlop, B. W., Trivedi, M. H., Jha, M., Sanacora, G., Hermes, G., & Papakostas, G. I. (2018). Double-blind, proof-of-concept (POC) trial of low-field magnetic stimulation (LFMS) augmentation of antidepressant therapy in treatment-resistant depression (TRD). Brain Stimulation, 11(1), 75–84. https://doi.org/10.1016/j.brs.2017.09.010.CrossRef Fava, M., Freeman, M. P., Flynn, M., Hoeppner, B. B., Shelton, R., Iosifescu, D. V., Murrough, J. W., Mischoulon, D., Cusin, C., Rapaport, M., Dunlop, B. W., Trivedi, M. H., Jha, M., Sanacora, G., Hermes, G., & Papakostas, G. I. (2018). Double-blind, proof-of-concept (POC) trial of low-field magnetic stimulation (LFMS) augmentation of antidepressant therapy in treatment-resistant depression (TRD). Brain Stimulation, 11(1), 75–84. https://​doi.​org/​10.​1016/​j.​brs.​2017.​09.​010.CrossRef
51.
57.
Zurück zum Zitat Provenzano, D. A., Rebman, J., Kuhel, C., Trenz, H., & Kilgore, J. (2017). The efficacy of high-density spinal cord stimulation among trial, implant, and conversion patients: A retrospective case series. Neuromodulation, 20(7), 654–660. https://doi.org/10.1111/ner.12612.CrossRef Provenzano, D. A., Rebman, J., Kuhel, C., Trenz, H., & Kilgore, J. (2017). The efficacy of high-density spinal cord stimulation among trial, implant, and conversion patients: A retrospective case series. Neuromodulation, 20(7), 654–660. https://​doi.​org/​10.​1111/​ner.​12612.CrossRef
58.
Zurück zum Zitat De Jaeger, M., van Hooff, R. J., Goudman, L., Valenzuela Espinoza, A., Brouns, R., Puylaert, M., Duyvendak, W., & Moens, M. (2017). High-density in spinal cord stimulation: virtual expert registry (DISCOVER): Study protocol for a prospective observational trial. Anesthesiology and Pain Medicine, 7(3), e13640. https://doi.org/10.5812/aapm.13640.CrossRef De Jaeger, M., van Hooff, R. J., Goudman, L., Valenzuela Espinoza, A., Brouns, R., Puylaert, M., Duyvendak, W., & Moens, M. (2017). High-density in spinal cord stimulation: virtual expert registry (DISCOVER): Study protocol for a prospective observational trial. Anesthesiology and Pain Medicine, 7(3), e13640. https://​doi.​org/​10.​5812/​aapm.​13640.CrossRef
59.
Zurück zum Zitat Al-Kaisy, A., Palmisani, S., Pang, D., Sanderson, K., Wesley, S., Tan, Y., McCammon, S., & Trescott, A. (2018). Prospective, randomized, sham-control, double blind, crossover trial of subthreshold spinal cord stimulation at various kilohertz frequencies in subjects suffering from failed back surgery syndrome (SCS frequency study). Neuromodulation, 21(5), 457–465. https://doi.org/10.1111/ner.12771.CrossRef Al-Kaisy, A., Palmisani, S., Pang, D., Sanderson, K., Wesley, S., Tan, Y., McCammon, S., & Trescott, A. (2018). Prospective, randomized, sham-control, double blind, crossover trial of subthreshold spinal cord stimulation at various kilohertz frequencies in subjects suffering from failed back surgery syndrome (SCS frequency study). Neuromodulation, 21(5), 457–465. https://​doi.​org/​10.​1111/​ner.​12771.CrossRef
60.
Zurück zum Zitat Zhang, L., Lu, Y., Sun, J., Zhou, X., & Tang, B. (2016). Subthreshold vagal stimulation suppresses ventricular arrhythmia and inflammatory response in a canine model of acute cardiac ischaemia and reperfusion. Experimental Physiology, 101(1), 41–49. https://doi.org/10.1113/EP085518.CrossRef Zhang, L., Lu, Y., Sun, J., Zhou, X., & Tang, B. (2016). Subthreshold vagal stimulation suppresses ventricular arrhythmia and inflammatory response in a canine model of acute cardiac ischaemia and reperfusion. Experimental Physiology, 101(1), 41–49. https://​doi.​org/​10.​1113/​EP085518.CrossRef
63.
Zurück zum Zitat Hotta, H., Onda, A., Suzuki, H., Milliken, P., & Sridhar, A. (2017). Modulation of calcitonin, parathyroid hormone, and thyroid hormone secretion by electrical stimulation of sympathetic and parasympathetic nerves in anesthetized rats. Frontiers in Neuroscience, 11, 375. https://doi.org/10.3389/fnins.2017.00375.CrossRef Hotta, H., Onda, A., Suzuki, H., Milliken, P., & Sridhar, A. (2017). Modulation of calcitonin, parathyroid hormone, and thyroid hormone secretion by electrical stimulation of sympathetic and parasympathetic nerves in anesthetized rats. Frontiers in Neuroscience, 11, 375. https://​doi.​org/​10.​3389/​fnins.​2017.​00375.CrossRef
65.
Zurück zum Zitat Gulyaev, Y. V., Bugaev, A. S., Indursky, P. A., Shakhnarovich, V. M., & Dementienko, V. V. (2017). Improvement of the night sleep quality by electrocutaneous subthreshold stimulation synchronized with the slow wave sleep. Doklady Biological Sciences, 474(1), 132–134. https://doi.org/10.1134/S0012496617030139.CrossRef Gulyaev, Y. V., Bugaev, A. S., Indursky, P. A., Shakhnarovich, V. M., & Dementienko, V. V. (2017). Improvement of the night sleep quality by electrocutaneous subthreshold stimulation synchronized with the slow wave sleep. Doklady Biological Sciences, 474(1), 132–134. https://​doi.​org/​10.​1134/​S001249661703013​9.CrossRef
66.
Zurück zum Zitat Makarov, S. N., Noetscher, G. M., & Nazarian, A. (2015). A low-frequency electromagnetic modeling for electrical and biological systems using MATLAB. New York: Wiley. Makarov, S. N., Noetscher, G. M., & Nazarian, A. (2015). A low-frequency electromagnetic modeling for electrical and biological systems using MATLAB. New York: Wiley.
70.
Zurück zum Zitat Reddy, C. G., Dalm, B. D., Flouty, O. E., Gillies, G. T., Howard, M. A., & Brennan, T. J. (2016). Comparison of conventional and kilohertz frequency epidural stimulation in patients undergoing trialing for spinal cord stimulation: clinical considerations. World Neurosurgery, 88, 586–591. https://doi.org/10.1016/j.wneu.2015.10.088.CrossRef Reddy, C. G., Dalm, B. D., Flouty, O. E., Gillies, G. T., Howard, M. A., & Brennan, T. J. (2016). Comparison of conventional and kilohertz frequency epidural stimulation in patients undergoing trialing for spinal cord stimulation: clinical considerations. World Neurosurgery, 88, 586–591. https://​doi.​org/​10.​1016/​j.​wneu.​2015.​10.​088.CrossRef
71.
Zurück zum Zitat Shechter, R., Yang, F., Xu, Q., Cheong, Y. K., He, S. Q., Sdrulla, A., Carteret, A. F., Wacnik, P. W., Dong, X., Meyer, R. A., Raja, S. N., & Guan, Y. (2013). Conventional and kilohertz-frequency spinal cord stimulation produces intensity- and frequency-dependent inhibition of mechanical hypersensitivity in a rat model of neuropathic pain. Anesthesiology, 119(2), 422–432. https://doi.org/10.1097/ALN.0b013e31829bd9e2.CrossRef Shechter, R., Yang, F., Xu, Q., Cheong, Y. K., He, S. Q., Sdrulla, A., Carteret, A. F., Wacnik, P. W., Dong, X., Meyer, R. A., Raja, S. N., & Guan, Y. (2013). Conventional and kilohertz-frequency spinal cord stimulation produces intensity- and frequency-dependent inhibition of mechanical hypersensitivity in a rat model of neuropathic pain. Anesthesiology, 119(2), 422–432. https://​doi.​org/​10.​1097/​ALN.​0b013e31829bd9e2​.CrossRef
72.
Zurück zum Zitat Thomson, S. J., Tavakkolizadeh, M., Love-Jones, S., Patel, N. K., Gu, J. W., Bains, A., Doan, Q., & Moffitt, M. (2018). Effects of rate on analgesia in kilohertz frequency spinal cord stimulation: Results of the proco randomized controlled trial. Neuromodulation, 21(1), 67–76. https://doi.org/10.1111/ner.12746.CrossRef Thomson, S. J., Tavakkolizadeh, M., Love-Jones, S., Patel, N. K., Gu, J. W., Bains, A., Doan, Q., & Moffitt, M. (2018). Effects of rate on analgesia in kilohertz frequency spinal cord stimulation: Results of the proco randomized controlled trial. Neuromodulation, 21(1), 67–76. https://​doi.​org/​10.​1111/​ner.​12746.CrossRef
73.
76.
Zurück zum Zitat Haider, N., Ligham, D., Quave, B., Harum, K. E., Garcia, E. A., Gilmore, C. A., Miller, N., Moore, G. A., Bains, A., Lechleiter, K., & Jain, R. (2018). Spinal cord stimulation (SCS) trial outcomes after conversion to a multiple waveform SCS system. Neuromodulation, 21(5), 504–507. https://doi.org/10.1111/ner.12783.CrossRef Haider, N., Ligham, D., Quave, B., Harum, K. E., Garcia, E. A., Gilmore, C. A., Miller, N., Moore, G. A., Bains, A., Lechleiter, K., & Jain, R. (2018). Spinal cord stimulation (SCS) trial outcomes after conversion to a multiple waveform SCS system. Neuromodulation, 21(5), 504–507. https://​doi.​org/​10.​1111/​ner.​12783.CrossRef
77.
Zurück zum Zitat Amirdelfan, K., Yu, C., Doust, M. W., Gliner, B. E., Morgan, D. M., Kapural, L., Vallejo, R., Sitzman, B. T., Yearwood, T. L., Bundschu, R., Yang, T., Benyamin, R., Burgher, A. H., Brooks, E. S., Powell, A. A., & Subbaroyan, J. (2018). Long-term quality of life improvement for chronic intractable back and leg pain patients using spinal cord stimulation: 12-month results from the SENZA-RCT. Quality of Life Research, 27(8), 2035–2044. https://doi.org/10.1007/s11136-018-1890-8.CrossRef Amirdelfan, K., Yu, C., Doust, M. W., Gliner, B. E., Morgan, D. M., Kapural, L., Vallejo, R., Sitzman, B. T., Yearwood, T. L., Bundschu, R., Yang, T., Benyamin, R., Burgher, A. H., Brooks, E. S., Powell, A. A., & Subbaroyan, J. (2018). Long-term quality of life improvement for chronic intractable back and leg pain patients using spinal cord stimulation: 12-month results from the SENZA-RCT. Quality of Life Research, 27(8), 2035–2044. https://​doi.​org/​10.​1007/​s11136-018-1890-8.CrossRef
80.
81.
83.
Zurück zum Zitat Borsboom, H. M., Claasen-Vujcic, T., Gaykema, H. J. G., & Mehlkopf, T. (1997). Low-frequency quadrature mode birdcage resonator. Magma, 5(1), 33–37.CrossRef Borsboom, H. M., Claasen-Vujcic, T., Gaykema, H. J. G., & Mehlkopf, T. (1997). Low-frequency quadrature mode birdcage resonator. Magma, 5(1), 33–37.CrossRef
84.
Zurück zum Zitat Claasen-Vujcic, T., Borsboom, H. M., Gaykema, H. J. G., & Mehlkopf, T. (1996). Transverse low-field RF coils in MRI transverse low-field RF coils in MRI. Magnetic Resonance in Medicine, 36(1), 111–116.CrossRef Claasen-Vujcic, T., Borsboom, H. M., Gaykema, H. J. G., & Mehlkopf, T. (1996). Transverse low-field RF coils in MRI transverse low-field RF coils in MRI. Magnetic Resonance in Medicine, 36(1), 111–116.CrossRef
85.
Zurück zum Zitat Galante, A., Sinibaldi, R., Conti, A., De Luca, C., Catallo, N., Sebastiani, P., Pizzella, V., Romani, G. L., Sotgiu, A., & Della Penna, S. (2015). Fast room temperature very low field-magnetic resonance imaging system compatible with magnetoencephalography environment. PLoS One, 10(12), e0142701. https://doi.org/10.1371/journal.pone.0142701.CrossRef Galante, A., Sinibaldi, R., Conti, A., De Luca, C., Catallo, N., Sebastiani, P., Pizzella, V., Romani, G. L., Sotgiu, A., & Della Penna, S. (2015). Fast room temperature very low field-magnetic resonance imaging system compatible with magnetoencephalography environment. PLoS One, 10(12), e0142701. https://​doi.​org/​10.​1371/​journal.​pone.​0142701.CrossRef
90.
Zurück zum Zitat Smythe, W. R. (1950). Static and dynamic electricity (2nd ed.). New York: McGraw-Hill.MATH Smythe, W. R. (1950). Static and dynamic electricity (2nd ed.). New York: McGraw-Hill.MATH
92.
94.
Zurück zum Zitat Davids, M., Guérin, B., Vom Endt, A., Schad, L. R., & Wald, L. L. (2018). Prediction of peripheral nerve stimulation thresholds of MRI gradient coils using coupled electromagnetic and neurodynamic simulations. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.27382.CrossRef Davids, M., Guérin, B., Vom Endt, A., Schad, L. R., & Wald, L. L. (2018). Prediction of peripheral nerve stimulation thresholds of MRI gradient coils using coupled electromagnetic and neurodynamic simulations. Magnetic Resonance in Medicine. https://​doi.​org/​10.​1002/​mrm.​27382.CrossRef
95.
Zurück zum Zitat International Electrotechnical Commission. (2010). International standard, Medical equipment – IEC 60601-2-33: Particular requirements for the safety of magnetic resonance equipment, 3rd Ed., Amendment 1 (2013), Amendment 2 (2015). Online: https://webstore.iec.ch/publication/22705 International Electrotechnical Commission. (2010). International standard, Medical equipment – IEC 60601-2-33: Particular requirements for the safety of magnetic resonance equipment, 3rd Ed., Amendment 1 (2013), Amendment 2 (2015). Online: https://​webstore.​iec.​ch/​publication/​22705
96.
Zurück zum Zitat Homann, H. (2012). SAR prediction and SAR management for parallel transmit MRI. Karlsruhe Translations on Biomedical Engineering, 16, 1–124. Homann, H. (2012). SAR prediction and SAR management for parallel transmit MRI. Karlsruhe Translations on Biomedical Engineering, 16, 1–124.
98.
Zurück zum Zitat Makarov, S. N., Yanamadala, J., Piazza, M. W., Helderman, A. M., Thang, N. S., Burnham, E. H., & Pascual-Leone, A. (2016). Preliminary upper estimate of peak currents in transcranial magnetic stimulation at distant locations from a TMS coil. IEEE Transactions on Biomedical Engineering, 63(9), 1944–1955. https://doi.org/10.1109/TBME.2015.2507572.CrossRef Makarov, S. N., Yanamadala, J., Piazza, M. W., Helderman, A. M., Thang, N. S., Burnham, E. H., & Pascual-Leone, A. (2016). Preliminary upper estimate of peak currents in transcranial magnetic stimulation at distant locations from a TMS coil. IEEE Transactions on Biomedical Engineering, 63(9), 1944–1955. https://​doi.​org/​10.​1109/​TBME.​2015.​2507572.CrossRef
99.
Zurück zum Zitat Kozlov, M., Tankaria, H., Noetscher, G. M., & Makarov, S. N. (2017). Comparative analysis of different versions of a human model located inside a 1.5T MRI whole body RF coil. Conference Proceedings: Annual International Conference of the IEEE Engineering in Medicine and Biology Society, 2017, 1477–1480. https://doi.org/10.1109/EMBC.2017.8037114.CrossRef Kozlov, M., Tankaria, H., Noetscher, G. M., & Makarov, S. N. (2017). Comparative analysis of different versions of a human model located inside a 1.5T MRI whole body RF coil. Conference Proceedings: Annual International Conference of the IEEE Engineering in Medicine and Biology Society, 2017, 1477–1480. https://​doi.​org/​10.​1109/​EMBC.​2017.​8037114.CrossRef
100.
Zurück zum Zitat Tran, A. L., & Makarov, S. N. (2017). Degree of RF MRI coil detuning for an anatomically realistic respiratory cycle modeled with the finite element method. Conference Proceedings: Annual International Conference of the IEEE Engineering in Medicine and Biology Society, 2017, 1405–1408. https://doi.org/10.1109/EMBC.2017.8037096.CrossRef Tran, A. L., & Makarov, S. N. (2017). Degree of RF MRI coil detuning for an anatomically realistic respiratory cycle modeled with the finite element method. Conference Proceedings: Annual International Conference of the IEEE Engineering in Medicine and Biology Society, 2017, 1405–1408. https://​doi.​org/​10.​1109/​EMBC.​2017.​8037096.CrossRef
101.
Zurück zum Zitat Noetscher, G. M., Yanamadala, J., Tankaria, H., Louie, S., Prokop, A., Nazarian, A., & Makarov, S. N. (2016). Computational human model VHP-FEMALE derived from datasets of the National Library of Medicine. Conference Proceedings: Annual International Conference of the IEEE Engineering in Medicine and Biology Society, 2016, 3350–3353. https://doi.org/10.1109/EMBC.2016.7591445.CrossRef Noetscher, G. M., Yanamadala, J., Tankaria, H., Louie, S., Prokop, A., Nazarian, A., & Makarov, S. N. (2016). Computational human model VHP-FEMALE derived from datasets of the National Library of Medicine. Conference Proceedings: Annual International Conference of the IEEE Engineering in Medicine and Biology Society, 2016, 3350–3353. https://​doi.​org/​10.​1109/​EMBC.​2016.​7591445.CrossRef
102.
Zurück zum Zitat Tankaria, H., Jackson, X. J., Borwankar, R., Srichandhru, G. N., Le Tran, A., Yanamadala, J., Noetscher, G. M., Nazarian, A., Louie, S., & Makarov, S. N. (2016). VHP-Female full-body human CAD model for cross-platform FEM simulations: recent development and validations. Conference Proceedings: Annual International Conference of the IEEE Engineering in Medicine and Biology Society, 2016, 2232–2235. https://doi.org/10.1109/EMBC.2016.7591173.CrossRef Tankaria, H., Jackson, X. J., Borwankar, R., Srichandhru, G. N., Le Tran, A., Yanamadala, J., Noetscher, G. M., Nazarian, A., Louie, S., & Makarov, S. N. (2016). VHP-Female full-body human CAD model for cross-platform FEM simulations: recent development and validations. Conference Proceedings: Annual International Conference of the IEEE Engineering in Medicine and Biology Society, 2016, 2232–2235. https://​doi.​org/​10.​1109/​EMBC.​2016.​7591173.CrossRef
103.
Zurück zum Zitat Yanamadala, J., Noetscher, G. M., Rathi, V. K., Maliye, S., Win, H. A., Tran, A. L., Jackson, X. J., Htet, A. T., Kozlov, M., Nazarian, A., Louie, S., & Makarov, S. N. (2015). New VHP-Female v. 2.0 full-body computational phantom and its performance metrics using FEM simulator ANSYS HFSS. Conference Proceedings: Annual International Conference of the IEEE Engineering in Medicine and Biology Society, 2015, 3237–3241. https://doi.org/10.1109/EMBC.2015.7319082.CrossRef Yanamadala, J., Noetscher, G. M., Rathi, V. K., Maliye, S., Win, H. A., Tran, A. L., Jackson, X. J., Htet, A. T., Kozlov, M., Nazarian, A., Louie, S., & Makarov, S. N. (2015). New VHP-Female v. 2.0 full-body computational phantom and its performance metrics using FEM simulator ANSYS HFSS. Conference Proceedings: Annual International Conference of the IEEE Engineering in Medicine and Biology Society, 2015, 3237–3241. https://​doi.​org/​10.​1109/​EMBC.​2015.​7319082.CrossRef
105.
Zurück zum Zitat Hasgall PA, Neufeld E, Gosselin MC, Klingenböck A, Kuster N, Klingenbock A, et al. 2015 IT’IS Database for thermal and electromagnetic parameters of biological tissues. Online: www.itis.ethz.ch/database. Hasgall PA, Neufeld E, Gosselin MC, Klingenböck A, Kuster N, Klingenbock A, et al. 2015 IT’IS Database for thermal and electromagnetic parameters of biological tissues. Online: www.​itis.​ethz.​ch/​database.
106.
107.
Zurück zum Zitat Assiotis, A., Chalidis, B. E., & Sachinis, N. P. (2012 Jun 8). Pulsed electromagnetic fields for the treatment of tibial delayed unions and nonunions. A prospective clinical study and review of the literature. Journal of Orthopaedic Surgery and Research, 7, 24. https://doi.org/10.1186/1749-799X-7-24.CrossRef Assiotis, A., Chalidis, B. E., & Sachinis, N. P. (2012 Jun 8). Pulsed electromagnetic fields for the treatment of tibial delayed unions and nonunions. A prospective clinical study and review of the literature. Journal of Orthopaedic Surgery and Research, 7, 24. https://​doi.​org/​10.​1186/​1749-799X-7-24.CrossRef
108.
Zurück zum Zitat Wu, S., Yu, Q., Sun, Y., & Tian, J. (2018). Synergistic effect of a LPEMF and SPIONs on BMMSC proliferation, directional migration, and osteoblastogenesis. American Journal of Translational Research, 10(5), 1431–1443. Wu, S., Yu, Q., Sun, Y., & Tian, J. (2018). Synergistic effect of a LPEMF and SPIONs on BMMSC proliferation, directional migration, and osteoblastogenesis. American Journal of Translational Research, 10(5), 1431–1443.
109.
Zurück zum Zitat Ehnert, S., Fentz, A. K., Schreiner, A., Birk, J., Wilbrand, B., Ziegler, P., Reumann, M. K., Wang, H., Falldorf, K., & Nussler, A. K. (2017). Extremely low frequency pulsed electromagnetic fields cause antioxidative defense mechanisms in human osteoblasts via induction of O2- and H2O2. Scientific Reports, 7(1), 14544. https://doi.org/10.1038/s41598-017-14983-9.CrossRef Ehnert, S., Fentz, A. K., Schreiner, A., Birk, J., Wilbrand, B., Ziegler, P., Reumann, M. K., Wang, H., Falldorf, K., & Nussler, A. K. (2017). Extremely low frequency pulsed electromagnetic fields cause antioxidative defense mechanisms in human osteoblasts via induction of O2- and H2O2. Scientific Reports, 7(1), 14544. https://​doi.​org/​10.​1038/​s41598-017-14983-9.CrossRef
111.
Zurück zum Zitat Ashraf, A. M., Abdelaal, P. D., Mona, M. T., Doaa, I. A., & Amira, H. D. (2017). Effect of pulsed electromagnetic therapy versus low-level laser therapy on bone mineral density in the elderly with primary osteoporosis: a randomized, controlled trial. Bulletin of Faculty of Physical Therapy, 22(1), 34–39. Ashraf, A. M., Abdelaal, P. D., Mona, M. T., Doaa, I. A., & Amira, H. D. (2017). Effect of pulsed electromagnetic therapy versus low-level laser therapy on bone mineral density in the elderly with primary osteoporosis: a randomized, controlled trial. Bulletin of Faculty of Physical Therapy, 22(1), 34–39.
113.
Zurück zum Zitat Woods, B., Manca, A., Weatherly, H., Saramago, P., Sideris, E., Giannopoulou, C., Rice, S., Corbett, M., Vickers, A., Bowes, M., MacPherson, H., & Sculpher, M. (2017). Cost-effectiveness of adjunct non-pharmacological interventions for osteoarthritis of the knee. PLoS One, 12(3), e0172749. https://doi.org/10.1371/journal.pone.0172749.CrossRef Woods, B., Manca, A., Weatherly, H., Saramago, P., Sideris, E., Giannopoulou, C., Rice, S., Corbett, M., Vickers, A., Bowes, M., MacPherson, H., & Sculpher, M. (2017). Cost-effectiveness of adjunct non-pharmacological interventions for osteoarthritis of the knee. PLoS One, 12(3), e0172749. https://​doi.​org/​10.​1371/​journal.​pone.​0172749.CrossRef
114.
Zurück zum Zitat Bilgin, H. M., Çelik, F., Gem, M., Akpolat, V., Yıldız, İ., Ekinci, A., Özerdem, M. S., & Tunik, S. (2017). Effects of local vibration and pulsed electromagnetic field on bone fracture: A comparative study. Bioelectromagnetics, 38(5), 339–348. https://doi.org/10.1002/bem.22043.CrossRef Bilgin, H. M., Çelik, F., Gem, M., Akpolat, V., Yıldız, İ., Ekinci, A., Özerdem, M. S., & Tunik, S. (2017). Effects of local vibration and pulsed electromagnetic field on bone fracture: A comparative study. Bioelectromagnetics, 38(5), 339–348. https://​doi.​org/​10.​1002/​bem.​22043.CrossRef
118.
Zurück zum Zitat Jing, D., Cai, J., Wu, Y., Shen, G., Li, F., Xu, Q., Xie, K., Tang, C., Liu, J., Guo, W., Wu, X., Jiang, M., & Luo, E. (2014). Pulsed electromagnetic fields partially preserve bone mass, microarchitecture, and strength by promoting bone formation in hindlimb-suspended rats. Journal of Bone and Mineral Research, 29, 2250–2261. https://doi.org/10.1002/jbmr.2260.CrossRef Jing, D., Cai, J., Wu, Y., Shen, G., Li, F., Xu, Q., Xie, K., Tang, C., Liu, J., Guo, W., Wu, X., Jiang, M., & Luo, E. (2014). Pulsed electromagnetic fields partially preserve bone mass, microarchitecture, and strength by promoting bone formation in hindlimb-suspended rats. Journal of Bone and Mineral Research, 29, 2250–2261. https://​doi.​org/​10.​1002/​jbmr.​2260.CrossRef
119.
Zurück zum Zitat Lei, T., Li, F., Liang, Z., Tang, C., Xie, K., Wang, P., Dong, X., Shan, S., Liu, J., Xu, Q., Luo, E., & Shen, G. Effects of four kinds of electromagnetic fields (EMF) with different frequency spectrum bands on ovariectomized osteoporosis in mice. Scientific Reports, 7(1), 553. https://doi.org/10.1038/s41598-017-00668-w. Lei, T., Li, F., Liang, Z., Tang, C., Xie, K., Wang, P., Dong, X., Shan, S., Liu, J., Xu, Q., Luo, E., & Shen, G. Effects of four kinds of electromagnetic fields (EMF) with different frequency spectrum bands on ovariectomized osteoporosis in mice. Scientific Reports, 7(1), 553. https://​doi.​org/​10.​1038/​s41598-017-00668-w.
120.
Zurück zum Zitat Li, J., Zeng, Z., Zhao, Y., Jing, D., Tang, C., Ding, Y., & Feng, X. Effects of low-intensity pulsed electromagnetic fields on bone microarchitecture, mechanical strength and bone turnover in type 2 diabetic db/db mice. Scientific Reports, 7, 10834. https://doi.org/10.1038/s41598-017-11090-7. Li, J., Zeng, Z., Zhao, Y., Jing, D., Tang, C., Ding, Y., & Feng, X. Effects of low-intensity pulsed electromagnetic fields on bone microarchitecture, mechanical strength and bone turnover in type 2 diabetic db/db mice. Scientific Reports, 7, 10834. https://​doi.​org/​10.​1038/​s41598-017-11090-7.
121.
Zurück zum Zitat Aksoy, M. Ç., Topal, O., Özkavak, H. V., Doğuç, D. K., Ilhan, A. A., & Çömlekçi, S. (2017). Effects of pulsed electromagnetic field on mineral density, biomechanical properties, and metabolism of bone tissue in heparin-induced osteoporosis in male rats. Biomedical Research, 28(6), 2724–2729. Aksoy, M. Ç., Topal, O., Özkavak, H. V., Doğuç, D. K., Ilhan, A. A., & Çömlekçi, S. (2017). Effects of pulsed electromagnetic field on mineral density, biomechanical properties, and metabolism of bone tissue in heparin-induced osteoporosis in male rats. Biomedical Research, 28(6), 2724–2729.
122.
Zurück zum Zitat Andrew, C., Bassett, C. A., Pawluk, R. J., & Pilla, A. A. (1974). Augmentation of bone repair by inductively coupled electromagnetic fields. Science, 184, 575–577.CrossRef Andrew, C., Bassett, C. A., Pawluk, R. J., & Pilla, A. A. (1974). Augmentation of bone repair by inductively coupled electromagnetic fields. Science, 184, 575–577.CrossRef
123.
Zurück zum Zitat Bassett, C. A., Mitchell, S. N., & Gasto, S. R. (1982). Pulsing electromagnetic field treatment in ununited fractures and failed arthrodeses. Journal of the American Medical Association (JAMA), 247, 623–628.CrossRef Bassett, C. A., Mitchell, S. N., & Gasto, S. R. (1982). Pulsing electromagnetic field treatment in ununited fractures and failed arthrodeses. Journal of the American Medical Association (JAMA), 247, 623–628.CrossRef
125.
Zurück zum Zitat Ramos, L. A. V., Callegari, B., França, F. J. R., Magalhães, M. O., Burke, T. N., Carvalho E Silva, A. P. M. C., Almeida, G. P. L., Comachio, J., & Marques, A. P. (2018). Comparison between transcutaneous electrical nerve stimulation and stabilization exercises in fatigue and transversus abdominis activation in patients with lumbar disk herniation: A randomized study. Journal of Manipulative and Physiological Therapeutics, 41(4), 323–331. https://doi.org/10.1016/j.jmpt.2017.10.010.CrossRef Ramos, L. A. V., Callegari, B., França, F. J. R., Magalhães, M. O., Burke, T. N., Carvalho E Silva, A. P. M. C., Almeida, G. P. L., Comachio, J., & Marques, A. P. (2018). Comparison between transcutaneous electrical nerve stimulation and stabilization exercises in fatigue and transversus abdominis activation in patients with lumbar disk herniation: A randomized study. Journal of Manipulative and Physiological Therapeutics, 41(4), 323–331. https://​doi.​org/​10.​1016/​j.​jmpt.​2017.​10.​010.CrossRef
Metadaten
Titel
Design and Analysis of a Whole-Body Noncontact Electromagnetic Subthreshold Stimulation Device with Field Modulation Targeting Nonspecific Neuropathic Pain
verfasst von
Sergey Makarov
Gene Bogdanov
Gregory Noetscher
William Appleyard
Reinhold Ludwig
Juho Joutsa
Zhi-De Deng
Copyright-Jahr
2019
DOI
https://doi.org/10.1007/978-3-030-21293-3_5

Neuer Inhalt