Abstract

Purpose

Epidural and spinal anesthesia are common procedures that require a needle to be inserted into the patient’s spine to deliver an anesthetic. Traditionally, these procedures were performed without image guidance, using only palpation to identify the correct vertebral interspace. More recently, ultrasound has seen widespread use in guiding spinal needle interventions. Dural pulsation is a valuable cue for finding a path through the vertebral interspace and for determining needle insertion depth. However, dural pulsation is challenging to detect and not perceptible in many cases. Here, a method for automatically detecting very subtle dural pulsation from live ultrasound video is presented.

Methods

A periodic model is fit to the B-mode intenstity values through extended Kalman filtering. The fitted frequencies and amplitudes are used to detect and visualize dural pulsation. The method is validated retrospectively on synthetic and human video and used in real time on an interventional spinal phantom.

Results

This method was capable of quickly identifying subtle dural pulsation and was robust to background noise and motion. The pulsation visualization reduced both the normalized path length and number of attempts required in a mock epidural procedure.

Conclusion

This technique is able to localize the dura and help find a clear needle trajectory to the epidural space. It can be run in real time on commercial ultrasound systems and has the potential to improve ultrasound guidance of spine needle interventions.

Sie haben noch keine Lizenz? Dann Informieren Sie sich jetzt über unsere Produkte:

Springer Professional "Wirtschaft+Technik"

Online-Abonnement

Mit Springer Professional "Wirtschaft+Technik" erhalten Sie Zugriff auf:

  • über 102.000 Bücher
  • über 537 Zeitschriften

aus folgenden Fachgebieten:

  • Automobil + Motoren
  • Bauwesen + Immobilien
  • Business IT + Informatik
  • Elektrotechnik + Elektronik
  • Energie + Nachhaltigkeit
  • Finance + Banking
  • Management + Führung
  • Marketing + Vertrieb
  • Maschinenbau + Werkstoffe
  • Versicherung + Risiko

Jetzt Wissensvorsprung sichern!

Springer Professional "Wirtschaft"

Online-Abonnement

Mit Springer Professional "Wirtschaft" erhalten Sie Zugriff auf:

  • über 67.000 Bücher
  • über 340 Zeitschriften

aus folgenden Fachgebieten:

  • Bauwesen + Immobilien
  • Business IT + Informatik
  • Finance + Banking
  • Management + Führung
  • Marketing + Vertrieb
  • Versicherung + Risiko




Jetzt Wissensvorsprung sichern!

Springer Professional "Technik"

Online-Abonnement

Mit Springer Professional "Technik" erhalten Sie Zugriff auf:

  • über 67.000 Bücher
  • über 390 Zeitschriften

aus folgenden Fachgebieten:

  • Automobil + Motoren
  • Bauwesen + Immobilien
  • Business IT + Informatik
  • Elektrotechnik + Elektronik
  • Energie + Nachhaltigkeit
  • Maschinenbau + Werkstoffe




 

Jetzt Wissensvorsprung sichern!

Literatur
1.
Zurück zum Zitat Amir-Khalili A, Peyrat JM, Abinahed J, Al-Alao O, Al-Ansari A, Hamarneh G, Abugharbieh R (2014) Auto localization and segmentation of occluded vessels in robot-assisted partial nephrectomy. In: Golland P, Hata N, Barillot C, Hornegger J, Howe R (eds) Medical image computing and computer-assisted intervention-MICCAI 2014, Springer, pp 407–414 Amir-Khalili A, Peyrat JM, Abinahed J, Al-Alao O, Al-Ansari A, Hamarneh G, Abugharbieh R (2014) Auto localization and segmentation of occluded vessels in robot-assisted partial nephrectomy. In: Golland P, Hata N, Barillot C, Hornegger J, Howe R (eds) Medical image computing and computer-assisted intervention-MICCAI 2014, Springer, pp 407–414
2.
Zurück zum Zitat Bittanti S, Savaresi SM (2000) On the parametrization and design of an extended Kalman filter frequency tracker. Autom Control IEEE Trans 45(9):1718–1724CrossRef Bittanti S, Savaresi SM (2000) On the parametrization and design of an extended Kalman filter frequency tracker. Autom Control IEEE Trans 45(9):1718–1724CrossRef
3.
Zurück zum Zitat Briassouli A, Ahuja N (2007) Extraction and analysis of multiple periodic motions in video sequences. Pattern Anal Mach Intell IEEE Trans 29(7):1244–1261CrossRef Briassouli A, Ahuja N (2007) Extraction and analysis of multiple periodic motions in video sequences. Pattern Anal Mach Intell IEEE Trans 29(7):1244–1261CrossRef
4.
Zurück zum Zitat Broadbent C, Maxwell W, Ferrie R, Wilson D, Gawne-Cain M, Russell R (2000) Ability of anaesthetists to identify a marked lumbar interspace. Anaesthesia 55(11):1122–1126CrossRefPubMed Broadbent C, Maxwell W, Ferrie R, Wilson D, Gawne-Cain M, Russell R (2000) Ability of anaesthetists to identify a marked lumbar interspace. Anaesthesia 55(11):1122–1126CrossRefPubMed
5.
Zurück zum Zitat Chen ECS, Ameri G, Li H, Sondekoppam RV, Ganapathy S, Peters TM (2014) Navigated simulator for spinal needle interventions. Studies in health technology and informatics 196, pp 56–60. doi:10.3233/978-1-61499-375-9-56 Chen ECS, Ameri G, Li H, Sondekoppam RV, Ganapathy S, Peters TM (2014) Navigated simulator for spinal needle interventions. Studies in health technology and informatics 196, pp 56–60. doi:10.​3233/​978-1-61499-375-9-56
6.
Zurück zum Zitat Conroy P, Luyet C, McCartney C, McHardy P (2013) Real-time ultrasound-guided spinal anaesthesia: a prospective observational study of a new approach. Anesthesiol Res Pract 2013. doi:10.1155/2013/525818 Conroy P, Luyet C, McCartney C, McHardy P (2013) Real-time ultrasound-guided spinal anaesthesia: a prospective observational study of a new approach. Anesthesiol Res Pract 2013. doi:10.​1155/​2013/​525818
7.
Zurück zum Zitat Cook T, Counsell D, Wildsmith J (2009) Major complications of central neuraxial block: report on the third national audit project of the royal college of anaesthetists. Br J Anaesth 102(2):179–190CrossRefPubMed Cook T, Counsell D, Wildsmith J (2009) Major complications of central neuraxial block: report on the third national audit project of the royal college of anaesthetists. Br J Anaesth 102(2):179–190CrossRefPubMed
8.
Zurück zum Zitat Cutler R, Davis LS (2000) Robust real-time periodic motion detection, analysis, and applications. Pattern Anal Mach Intell IEEE Trans 22(8):781–796CrossRef Cutler R, Davis LS (2000) Robust real-time periodic motion detection, analysis, and applications. Pattern Anal Mach Intell IEEE Trans 22(8):781–796CrossRef
9.
Zurück zum Zitat Fisher RA (1929) Tests of significance in harmonic analysis. Proc R Soc London Ser A 125(796):54–59CrossRef Fisher RA (1929) Tests of significance in harmonic analysis. Proc R Soc London Ser A 125(796):54–59CrossRef
10.
Zurück zum Zitat Furness G, Reilly M, Kuchi S (2002) An evaluation of ultrasound imaging for identificationof lumbar intervertebral level. Anaesthesia 57(3):277–280CrossRefPubMed Furness G, Reilly M, Kuchi S (2002) An evaluation of ultrasound imaging for identificationof lumbar intervertebral level. Anaesthesia 57(3):277–280CrossRefPubMed
12.
Zurück zum Zitat Karmakar M, Li X, Ho AH, Kwok W, Chui P (2009) Real-time ultrasound-guided paramedian epidural access: evaluation of a novel in-plane technique. Br J Anaesth 102(6):845–854CrossRefPubMed Karmakar M, Li X, Ho AH, Kwok W, Chui P (2009) Real-time ultrasound-guided paramedian epidural access: evaluation of a novel in-plane technique. Br J Anaesth 102(6):845–854CrossRefPubMed
13.
Zurück zum Zitat Kimura A, Seichi A, Inoue H, Endo T, Sato M, Higashi T, Hoshino Y (2012) Ultrasonographic quantification of spinal cord and dural pulsations during cervical laminoplasty in patients with compressive myelopathy. Eur Spine J 21(12):2450–2455CrossRefPubMedCentralPubMed Kimura A, Seichi A, Inoue H, Endo T, Sato M, Higashi T, Hoshino Y (2012) Ultrasonographic quantification of spinal cord and dural pulsations during cervical laminoplasty in patients with compressive myelopathy. Eur Spine J 21(12):2450–2455CrossRefPubMedCentralPubMed
14.
Zurück zum Zitat Liu C, Torralba A, Freeman WT, Durand F, Adelson EH (2005) Motion magnification. In: ACM transactions on graphics (TOG), vol 24. ACM, pp 519–526 Liu C, Torralba A, Freeman WT, Durand F, Adelson EH (2005) Motion magnification. In: ACM transactions on graphics (TOG), vol 24. ACM, pp 519–526
15.
Zurück zum Zitat McLeod AJ, Baxter JS, de Ribaupierre S, Peters TM (2014) Motion magnification for endoscopic surgery. In: Yaniv Z, Holmes D (eds) SPIE medical imaging. International Society for Optics and Photonics, pp 90,360C–90,360C–8 McLeod AJ, Baxter JS, de Ribaupierre S, Peters TM (2014) Motion magnification for endoscopic surgery. In: Yaniv Z, Holmes D (eds) SPIE medical imaging. International Society for Optics and Photonics, pp 90,360C–90,360C–8
16.
Zurück zum Zitat McNames J, Aboy M (2008) Statistical modeling of cardiovascular signals and parameter estimation based on the extended Kalman filter. Biomed Eng IEEE Trans 55(1):119–129CrossRef McNames J, Aboy M (2008) Statistical modeling of cardiovascular signals and parameter estimation based on the extended Kalman filter. Biomed Eng IEEE Trans 55(1):119–129CrossRef
17.
Zurück zum Zitat Pogalin E, Smeulders AW, Thean AH (2008) Visual quasi-periodicity. In: Computer vision and pattern recognition, 2008. CVPR 2008. IEEE Conference on, pp 1–8 Pogalin E, Smeulders AW, Thean AH (2008) Visual quasi-periodicity. In: Computer vision and pattern recognition, 2008. CVPR 2008. IEEE Conference on, pp 1–8
18.
Zurück zum Zitat Polana R, Nelson RC (1997) Detection and recognition of periodic, nonrigid motion. Int J Comput Vis 23(3):261–282CrossRef Polana R, Nelson RC (1997) Detection and recognition of periodic, nonrigid motion. Int J Comput Vis 23(3):261–282CrossRef
19.
Zurück zum Zitat Pöpping DM, Elia N, Van Aken HK, Marret E, Schug SA, Kranke P, Wenk M, Tramèr MR (2014) Impact of epidural analgesia on mortality and morbidity after surgery: systematic review and meta-analysis of randomized controlled trials. Ann Surg 259(6):1056–1067CrossRefPubMed Pöpping DM, Elia N, Van Aken HK, Marret E, Schug SA, Kranke P, Wenk M, Tramèr MR (2014) Impact of epidural analgesia on mortality and morbidity after surgery: systematic review and meta-analysis of randomized controlled trials. Ann Surg 259(6):1056–1067CrossRefPubMed
20.
Zurück zum Zitat Quinn BG, Hannan EJ (2001) The estimation and tracking of frequency. Cambridge University Press, CambridgeCrossRef Quinn BG, Hannan EJ (2001) The estimation and tracking of frequency. Cambridge University Press, CambridgeCrossRef
21.
Zurück zum Zitat Rajakulendran Y, Rahman S, Venkat N (1999) Long-term neurological complication following traumatic damage to the spinal cord with a 25 gauge whitacre spinal needle. Int J Obstet Anesth 8(1):62–66CrossRefPubMed Rajakulendran Y, Rahman S, Venkat N (1999) Long-term neurological complication following traumatic damage to the spinal cord with a 25 gauge whitacre spinal needle. Int J Obstet Anesth 8(1):62–66CrossRefPubMed
22.
Zurück zum Zitat Ran Y, Weiss I, Zheng Q, Davis LS (2007) Pedestrian detection via periodic motion analysis. Int J Comput Vis 71(2):143–160CrossRef Ran Y, Weiss I, Zheng Q, Davis LS (2007) Pedestrian detection via periodic motion analysis. Int J Comput Vis 71(2):143–160CrossRef
23.
Zurück zum Zitat Rasoulian A, Abolmaesumi P, Mousavi P (2012) Feature-based multibody rigid registration of CT and ultrasound images of lumbar spine. Med Phys 39(6):3154–3166CrossRefPubMed Rasoulian A, Abolmaesumi P, Mousavi P (2012) Feature-based multibody rigid registration of CT and ultrasound images of lumbar spine. Med Phys 39(6):3154–3166CrossRefPubMed
24.
Zurück zum Zitat Seitz SM, Dyer CR (1997) View-invariant analysis of cyclic motion. Int J Comput Vis 25(3):231–251CrossRef Seitz SM, Dyer CR (1997) View-invariant analysis of cyclic motion. Int J Comput Vis 25(3):231–251CrossRef
25.
Zurück zum Zitat Shaikh F, Brzezinski J, Alexander S, Arzola C, Carvalho JC, Beyene J, Sung L (2013) Ultrasound imaging for lumbar punctures and epidural catheterisations: systematic review and meta-analysis. BMJ 346:f1720CrossRefPubMed Shaikh F, Brzezinski J, Alexander S, Arzola C, Carvalho JC, Beyene J, Sung L (2013) Ultrasound imaging for lumbar punctures and epidural catheterisations: systematic review and meta-analysis. BMJ 346:f1720CrossRefPubMed
26.
Zurück zum Zitat Simon D, Chia TL (2002) Kalman filtering with state equality constraints. Aerosp Electron Syst IEEE Trans 38(1):128–136CrossRef Simon D, Chia TL (2002) Kalman filtering with state equality constraints. Aerosp Electron Syst IEEE Trans 38(1):128–136CrossRef
27.
Zurück zum Zitat Wadhwa N, Rubinstein M, Durand F, Freeman WT (2013) Phase-based video motion processing. ACM Trans Graph (TOG) 32(4):80CrossRef Wadhwa N, Rubinstein M, Durand F, Freeman WT (2013) Phase-based video motion processing. ACM Trans Graph (TOG) 32(4):80CrossRef
28.
Zurück zum Zitat Wadhwa N, Rubinstein M, Durand F, Freeman WT (2014) Riesz pyramids for fast phase-based video magnification. In: Computational Photography, IEEE International Conference on. doi:10.1109/ICCPHOT.2014.6831820 Wadhwa N, Rubinstein M, Durand F, Freeman WT (2014) Riesz pyramids for fast phase-based video magnification. In: Computational Photography, IEEE International Conference on. doi:10.​1109/​ICCPHOT.​2014.​6831820
29.
Zurück zum Zitat Wu HY, Rubinstein M, Shih E, Guttag JV, Durand F, Freeman WT (2012) Eulerian video magnification for revealing subtle changes in the world. ACM Trans Graph 31(4):65CrossRef Wu HY, Rubinstein M, Shih E, Guttag JV, Durand F, Freeman WT (2012) Eulerian video magnification for revealing subtle changes in the world. ACM Trans Graph 31(4):65CrossRef
30.
Zurück zum Zitat Yuen SG, Novotny PM, Howe RD (2008) Quasiperiodic predictive filtering for robot-assisted beating heart surgery. In: Robotics and automation, 2008. ICRA 2008. IEEE International Conference on, pp 3875–3880. IEEE Yuen SG, Novotny PM, Howe RD (2008) Quasiperiodic predictive filtering for robot-assisted beating heart surgery. In: Robotics and automation, 2008. ICRA 2008. IEEE International Conference on, pp 3875–3880. IEEE
Metadaten
Titel
Detection and visualization of dural pulsation for spine needle interventions
verfasst von
A. Jonathan McLeod
John S. H. Baxter
Golafsoun Ameri
Sugantha Ganapathy
Terry M. Peters
Elvis C. S. Chen
Publikationsdatum
01.06.2015
Verlag
Springer Berlin Heidelberg
Erschienen in
International Journal of Computer Assisted Radiology and Surgery / Ausgabe 6/2015
Print ISSN: 1861-6410
Elektronische ISSN: 1861-6429
DOI
https://doi.org/10.1007/s11548-015-1192-3

Weitere Artikel der Ausgabe 6/2015

International Journal of Computer Assisted Radiology and Surgery 6/2015 Zur Ausgabe