2011 | OriginalPaper | Chapter
Combining Growcut and Temporal Correlation for IVUS Lumen Segmentation
Authors : Simone Balocco, Carlo Gatta, Francesco Ciompi, Oriol Pujol, Xavier Carrillo, Josepa Mauri, Petia Radeva
Published in: Pattern Recognition and Image Analysis
Publisher: Springer Berlin Heidelberg
Activate our intelligent search to find suitable subject content or patents.
Select sections of text to find matching patents with Artificial Intelligence. powered by
Select sections of text to find additional relevant content using AI-assisted search. powered by
The assessment of arterial luminal area, performed by IVUS analysis, is a clinical index used to evaluate the degree of coronary artery disease. In this paper we propose a novel approach to automatically segment the vessel lumen, which combines model-based temporal information extracted from successive frames of the sequence, with spatial classification using the Growcut algorithm. The performance of the method is evaluated by an
in vivo
experiment on 300 IVUS frames. The automatic and manual segmentation performances in general vessel and stent frames are comparable. The average segmentation error in vessel, stent and bifurcation frames are 0.17±0.08 mm, 0.18±0.07 mm and 0.31±0.12 mm respectively.