Skip to main content
Erschienen in: International Journal of Computer Assisted Radiology and Surgery 8/2018

29.03.2018 | Original Article

Radiation exposure for intraoperative 3D scans in a hybrid operating room: how to reduce radiation exposure for the surgical team

verfasst von: K. Schuetze, M. Kraus, A. Eickhoff, F. Gebhard, P. H. Richter

Erschienen in: International Journal of Computer Assisted Radiology and Surgery | Ausgabe 8/2018

Einloggen

Aktivieren Sie unsere intelligente Suche, um passende Fachinhalte oder Patente zu finden.

search-config
loading …

Abstract

Background

Hybrid operating rooms are used in different fields of surgery. In orthopedic surgery, the possibility of a 3D scan of difficult anatomical regions (spine, pelvis) showed promising results not only in navigated screw placement. The associated exposure to radiation raises questions regarding potential risks for the operating room personal and the patient. The present study focuses on scatter radiation during 3D scans in a hybrid operating room, the adjacent rooms, and methods to reduce radiation exposure.

Material and Methods

\(\hbox {RaySafe}^{\mathrm{TM}}\) i2 dosimeters were used to measure scatter radiation during 3D scans of different anatomical regions in different distances and heights in a hybrid operating room. The 3D scans were performed with a floor-based flat-panel robotic C-arm with 3D scan capability (Artis Zeego, Siemens; Germany). The 3D scans were performed using a human cadaver. The 3D scans were performed using a standard and a dose reduction protocol (DRP).

Results

The highest scatter radiation was measured during 3D scans of the pelvis on the side of the surgical assistant (39.5 Sv in height of 1 m) compared to the side of the main surgeon (23 Sv in height of 1 m). Scatter radiation was less on the position of the scrub nurse (6.8 Sv in height of 1 m) and during 3D scans of the other anatomical regions. The radiation dosage was about 66% less with the DRP. Low values of scatter radiation were measured behind a radiation protection wall and with open doors in the adjacent rooms.

Conclusion

While performing a scan scatter radiation was measured everywhere in the operating room especially during 3D scans of the pelvic girdle. Therefore, settings with lower dosage should be used whenever possible. Personnel should stand behind a radiation safety wall or leave the operating room and close the doors. For this study, six behavioral rules to avoid radiation in a hybrid operating room were established.

Sie haben noch keine Lizenz? Dann Informieren Sie sich jetzt über unsere Produkte:

Springer Professional "Wirtschaft+Technik"

Online-Abonnement

Mit Springer Professional "Wirtschaft+Technik" erhalten Sie Zugriff auf:

  • über 102.000 Bücher
  • über 537 Zeitschriften

aus folgenden Fachgebieten:

  • Automobil + Motoren
  • Bauwesen + Immobilien
  • Business IT + Informatik
  • Elektrotechnik + Elektronik
  • Energie + Nachhaltigkeit
  • Finance + Banking
  • Management + Führung
  • Marketing + Vertrieb
  • Maschinenbau + Werkstoffe
  • Versicherung + Risiko

Jetzt Wissensvorsprung sichern!

Springer Professional "Wirtschaft"

Online-Abonnement

Mit Springer Professional "Wirtschaft" erhalten Sie Zugriff auf:

  • über 67.000 Bücher
  • über 340 Zeitschriften

aus folgenden Fachgebieten:

  • Bauwesen + Immobilien
  • Business IT + Informatik
  • Finance + Banking
  • Management + Führung
  • Marketing + Vertrieb
  • Versicherung + Risiko




Jetzt Wissensvorsprung sichern!

Springer Professional "Technik"

Online-Abonnement

Mit Springer Professional "Technik" erhalten Sie Zugriff auf:

  • über 67.000 Bücher
  • über 390 Zeitschriften

aus folgenden Fachgebieten:

  • Automobil + Motoren
  • Bauwesen + Immobilien
  • Business IT + Informatik
  • Elektrotechnik + Elektronik
  • Energie + Nachhaltigkeit
  • Maschinenbau + Werkstoffe




 

Jetzt Wissensvorsprung sichern!

Literatur
1.
Zurück zum Zitat Tsagakis K, Konorza T, Dohle DS, Kottenberg E, Buck T, Thielmann M, Erbel R, Jakob H (2013) Hybrid operating room concept for combined diagnostics, intervention and surgery in acute type A dissection. Eur J Cardiothorac Surg Off J Eur Assoc Cardiothorac Surg 43(2):397–404CrossRef Tsagakis K, Konorza T, Dohle DS, Kottenberg E, Buck T, Thielmann M, Erbel R, Jakob H (2013) Hybrid operating room concept for combined diagnostics, intervention and surgery in acute type A dissection. Eur J Cardiothorac Surg Off J Eur Assoc Cardiothorac Surg 43(2):397–404CrossRef
2.
Zurück zum Zitat Hlavac M, Konig R, Halatsch M, Wirtz CR (2012) Intraoperative magnetic resonance imaging. Fifteen years’ experience in the neurosurgical hybrid operating suite. Der Unfallchirurg 115(2):121–4CrossRefPubMed Hlavac M, Konig R, Halatsch M, Wirtz CR (2012) Intraoperative magnetic resonance imaging. Fifteen years’ experience in the neurosurgical hybrid operating suite. Der Unfallchirurg 115(2):121–4CrossRefPubMed
3.
Zurück zum Zitat Fandino J, Taussky P, Marbacher S, Muroi C, Diepers M, Fathi AR, Remonda L (2013) The concept of a hybrid operating room: applications in cerebrovascular surgery. Acta Leurochirurgica Suppl 115:113–7 Fandino J, Taussky P, Marbacher S, Muroi C, Diepers M, Fathi AR, Remonda L (2013) The concept of a hybrid operating room: applications in cerebrovascular surgery. Acta Leurochirurgica Suppl 115:113–7
4.
Zurück zum Zitat Terra RM, Andrade JR, Mariani AW, Garcia RG, Succi JE, Soares A, Zimmer PM (2016) Applications for a hybrid operating room in thoracic surgery: from multidisciplinary procedures to—image-guided video-assisted thoracoscopic surgery. J Bras Pneumol 42(5):387–90CrossRefPubMedPubMedCentral Terra RM, Andrade JR, Mariani AW, Garcia RG, Succi JE, Soares A, Zimmer PM (2016) Applications for a hybrid operating room in thoracic surgery: from multidisciplinary procedures to—image-guided video-assisted thoracoscopic surgery. J Bras Pneumol 42(5):387–90CrossRefPubMedPubMedCentral
5.
Zurück zum Zitat Varu VN, Greenberg JI, Lee JT (2013) Improved efficiency and safety for EVAR with utilization of a hybrid room. Eur J Casc Endovasc Surg Off J Eur Soc Vasc Surg 46(6):675–9CrossRef Varu VN, Greenberg JI, Lee JT (2013) Improved efficiency and safety for EVAR with utilization of a hybrid room. Eur J Casc Endovasc Surg Off J Eur Soc Vasc Surg 46(6):675–9CrossRef
6.
Zurück zum Zitat Steinbauer M, Katsargyris A, Greindl M, Topel I, Verhoeven E (2013) Hybrid operation theatre in vascular surgery. Options and perspectives. Der Chirurg; Zeitschrift fur alle Gebiete der operativen Medizen 84(12):1030–5CrossRefPubMed Steinbauer M, Katsargyris A, Greindl M, Topel I, Verhoeven E (2013) Hybrid operation theatre in vascular surgery. Options and perspectives. Der Chirurg; Zeitschrift fur alle Gebiete der operativen Medizen 84(12):1030–5CrossRefPubMed
7.
Zurück zum Zitat Richter PH, Gebhard F, Dehner C, Scola A (2016) Accuracy of computer-assisted iliosacral screw placement using a hybrid operating room. Injury 47(2):402–7CrossRefPubMed Richter PH, Gebhard F, Dehner C, Scola A (2016) Accuracy of computer-assisted iliosacral screw placement using a hybrid operating room. Injury 47(2):402–7CrossRefPubMed
8.
Zurück zum Zitat Kendrick DE, Miller CP, Moorehead PA, Kim AH, Baele HR, Wong VL, Jordan DW, Kashyap VS (2016) Comparative occupational radiation exposure between fixed and mobile imaging systems. J Vasc Surg 63(1):190–7CrossRefPubMed Kendrick DE, Miller CP, Moorehead PA, Kim AH, Baele HR, Wong VL, Jordan DW, Kashyap VS (2016) Comparative occupational radiation exposure between fixed and mobile imaging systems. J Vasc Surg 63(1):190–7CrossRefPubMed
9.
Zurück zum Zitat van den Haak RF, Hamans BC, Zuurmond K, Verhoeven BA, Koning OH (2015) Significant radiation dose reduction in the hybrid operating room using a novel X-ray imaging technology. Eur J Vasc Endovasc Surg Off J Eur Soc Vasc Surg 50(4):480–6CrossRef van den Haak RF, Hamans BC, Zuurmond K, Verhoeven BA, Koning OH (2015) Significant radiation dose reduction in the hybrid operating room using a novel X-ray imaging technology. Eur J Vasc Endovasc Surg Off J Eur Soc Vasc Surg 50(4):480–6CrossRef
10.
Zurück zum Zitat Kesavachandran CN, Haamann F, Nienhaus A (2012) Radiation exposure of eyes, thyroid gland and hands in orthopaedic staff: a systematic review. Eur J Med Res 17:28CrossRefPubMedPubMedCentral Kesavachandran CN, Haamann F, Nienhaus A (2012) Radiation exposure of eyes, thyroid gland and hands in orthopaedic staff: a systematic review. Eur J Med Res 17:28CrossRefPubMedPubMedCentral
11.
Zurück zum Zitat Pokrywka M, Byers K (2013) Traffic in the operating room: a review of factors influencing air flow and surgical wound contamination. Infect Disord Drug Targets 13(3):156–61CrossRefPubMed Pokrywka M, Byers K (2013) Traffic in the operating room: a review of factors influencing air flow and surgical wound contamination. Infect Disord Drug Targets 13(3):156–61CrossRefPubMed
12.
Zurück zum Zitat Darouiche RO, Green DM, Harrington MA, Ehni BL, Kougias P, Bechara CF, O’Connor DP (2017) Association of airborne microorganisms in the operating room with implant infections: a randomized controlled trial. Infect Control Hosp Epidemiol 38(1):3–10CrossRefPubMed Darouiche RO, Green DM, Harrington MA, Ehni BL, Kougias P, Bechara CF, O’Connor DP (2017) Association of airborne microorganisms in the operating room with implant infections: a randomized controlled trial. Infect Control Hosp Epidemiol 38(1):3–10CrossRefPubMed
13.
Zurück zum Zitat Cristina ML, Sartini M, Schinca E, Ottria G, Spagnolo AM (2016) Operating room environment and surgical site infections in arthroplasty procedures. J Prev Med Hyg 57(3):E142–E8PubMedPubMedCentral Cristina ML, Sartini M, Schinca E, Ottria G, Spagnolo AM (2016) Operating room environment and surgical site infections in arthroplasty procedures. J Prev Med Hyg 57(3):E142–E8PubMedPubMedCentral
14.
Zurück zum Zitat Knobben BA, van Horn JR, van der Mei HC, Busscher HJ (2006) Evaluation of measures to decrease intra-operative bacterial contamination in orthopaedic implant surgery. J Hosp Infect 62(2):174–80CrossRefPubMed Knobben BA, van Horn JR, van der Mei HC, Busscher HJ (2006) Evaluation of measures to decrease intra-operative bacterial contamination in orthopaedic implant surgery. J Hosp Infect 62(2):174–80CrossRefPubMed
15.
Zurück zum Zitat Andersson AE, Bergh I, Karlsson J, Eriksson BI, Nilsson K (2012) Traffic flow in the operating room: an explorative and descriptive study on air quality during orthopedic trauma implant surgery. Am J Infect Control 40(8):750–5CrossRefPubMed Andersson AE, Bergh I, Karlsson J, Eriksson BI, Nilsson K (2012) Traffic flow in the operating room: an explorative and descriptive study on air quality during orthopedic trauma implant surgery. Am J Infect Control 40(8):750–5CrossRefPubMed
16.
Zurück zum Zitat Romanova K, Vassileva J, Alyakov M (2015) Radiation exposure to the eye lens of orthopaedic surgeons during various orthopaedic procedures. Radiat Prot Dosim 165(1–4):310–3CrossRef Romanova K, Vassileva J, Alyakov M (2015) Radiation exposure to the eye lens of orthopaedic surgeons during various orthopaedic procedures. Radiat Prot Dosim 165(1–4):310–3CrossRef
17.
Zurück zum Zitat Schulz B, Heidenreich R, Heidenreich M, Eichler K, Thalhammer A, Naeem NN, Vogl TJ, Zangos S (2012) Radiation exposure to operating staff during rotational flat-panel angiography and C-arm cone beam computed tomography (CT) applications. Eur J Radiol 81(12):4138–42CrossRefPubMed Schulz B, Heidenreich R, Heidenreich M, Eichler K, Thalhammer A, Naeem NN, Vogl TJ, Zangos S (2012) Radiation exposure to operating staff during rotational flat-panel angiography and C-arm cone beam computed tomography (CT) applications. Eur J Radiol 81(12):4138–42CrossRefPubMed
18.
Zurück zum Zitat Hendee WR, Edwards FM (1986) ALARA and an integrated approach to radiation protection. Semin Nucl Med 16(2):142–50CrossRefPubMed Hendee WR, Edwards FM (1986) ALARA and an integrated approach to radiation protection. Semin Nucl Med 16(2):142–50CrossRefPubMed
19.
Zurück zum Zitat Bahari S, Morris S, Broe D, Taylor C, Lenehan B, McElwain J (2006) Radiation exposure of the hands and thyroid gland during percutaneous wiring of wrist and hand procedures. Acta Orthop Belg 72(2):194–8PubMed Bahari S, Morris S, Broe D, Taylor C, Lenehan B, McElwain J (2006) Radiation exposure of the hands and thyroid gland during percutaneous wiring of wrist and hand procedures. Acta Orthop Belg 72(2):194–8PubMed
20.
Zurück zum Zitat Srinivasan D, Than KD, Wang AC, La Marca F, Wang PI, Schermerhorn TC, Park P (2014) Radiation safety and spine surgery: systematic review of exposure limits and methods to minimize radiation exposure. World Neurosurg 82(6):1337–43CrossRefPubMed Srinivasan D, Than KD, Wang AC, La Marca F, Wang PI, Schermerhorn TC, Park P (2014) Radiation safety and spine surgery: systematic review of exposure limits and methods to minimize radiation exposure. World Neurosurg 82(6):1337–43CrossRefPubMed
21.
Zurück zum Zitat Kaplan DJ, Patel JN, Liporace FA, Yoon RS (2016) Intraoperative radiation safety in orthopaedics: a review of the ALARA (As low as reasonably achievable) principle. Patient Saf Surg. 10:27CrossRefPubMedPubMedCentral Kaplan DJ, Patel JN, Liporace FA, Yoon RS (2016) Intraoperative radiation safety in orthopaedics: a review of the ALARA (As low as reasonably achievable) principle. Patient Saf Surg. 10:27CrossRefPubMedPubMedCentral
22.
Zurück zum Zitat Kraus M, Fischer E, Gebhard F, Richter PH (2016) Image quality and effective dose of a robotic flat panel 3D C-arm vs computed tomography. Int J Med Robot 12(4):743–50CrossRefPubMed Kraus M, Fischer E, Gebhard F, Richter PH (2016) Image quality and effective dose of a robotic flat panel 3D C-arm vs computed tomography. Int J Med Robot 12(4):743–50CrossRefPubMed
23.
Zurück zum Zitat Racadio J, Nachabe R, Carelsen B, Racadio J, Hilvert N, Johnson N, Kukreja K, Patel M (2014) Effect of real-time radiation dose feedback on pediatric interventional radiology staff radiation exposure. J Vasc Interv Radiol: JVIR 25(1):119–26CrossRefPubMed Racadio J, Nachabe R, Carelsen B, Racadio J, Hilvert N, Johnson N, Kukreja K, Patel M (2014) Effect of real-time radiation dose feedback on pediatric interventional radiology staff radiation exposure. J Vasc Interv Radiol: JVIR 25(1):119–26CrossRefPubMed
24.
Zurück zum Zitat Baumann F, Katzen BT, Carelsen B, Diehm N, Benenati JF, Pena CS (2015) The effect of realtime monitoring on dose exposure to staff within an interventional radiology setting. Cardiovasc Interv Radiol 38(5):1105–11CrossRef Baumann F, Katzen BT, Carelsen B, Diehm N, Benenati JF, Pena CS (2015) The effect of realtime monitoring on dose exposure to staff within an interventional radiology setting. Cardiovasc Interv Radiol 38(5):1105–11CrossRef
25.
Zurück zum Zitat Muller MC, Welle K, Strauss A, Naehle PC, Pennekamp PH, Weber O, Burger C (2014) Real-time dosimetry reduces radiation exposure of orthopaedic surgeons. Orthop Traumatol Surg Res: OTSR. 100(8):947–51CrossRefPubMed Muller MC, Welle K, Strauss A, Naehle PC, Pennekamp PH, Weber O, Burger C (2014) Real-time dosimetry reduces radiation exposure of orthopaedic surgeons. Orthop Traumatol Surg Res: OTSR. 100(8):947–51CrossRefPubMed
Metadaten
Titel
Radiation exposure for intraoperative 3D scans in a hybrid operating room: how to reduce radiation exposure for the surgical team
verfasst von
K. Schuetze
M. Kraus
A. Eickhoff
F. Gebhard
P. H. Richter
Publikationsdatum
29.03.2018
Verlag
Springer International Publishing
Erschienen in
International Journal of Computer Assisted Radiology and Surgery / Ausgabe 8/2018
Print ISSN: 1861-6410
Elektronische ISSN: 1861-6429
DOI
https://doi.org/10.1007/s11548-018-1747-1

Weitere Artikel der Ausgabe 8/2018