Abstract
Purpose
Polymethylpentene membrane oxygenators used in venovenous extracorporeal membrane oxygenation (vvECMO) differ in their physical characteristics. The aim of the study was to analyze the gas transfer capability of different ECMO systems in clinical practice, as the choice of the appropriate system may be influenced by the needs of the patient.
Methods
Retrospective study on prospectively collected data of adults with severe respiratory failure requiring vvECMO support (Regensburg ECMO Registry, 2009–2013). Oxygen (O2) transfer and carbon dioxide (CO2) elimination of four different ECMO systems (PLS system, n = 163; Cardiohelp system (CH), n = 59, Maquet Cardiopulmonary, Rastatt, Germany; Hilite 7000 LT system, n = 56, Medos Medizintechnik, Stolberg, Germany; ECC.05 system, n = 39, Sorin Group, Mirandola (MO), Italy) were analyzed.
Results
Gas transfer depended on type of ECMO system, blood flow, and gas flow (p ≤ 0.05, each). CO2 removal is dependent on sweep gas flow and blood flow, with higher blood flow and/or gas flow eliminating more CO2 (p ≤ 0.001). CO2 elimination capacity was highest with the PLS system (p ≤ 0.001). O2 transfer at blood flow rates below 3 l/min depended on blood flow, at higher blood flow rates on blood flow and gas flow. The system with the smallest gas exchange surface (ECC.05 system) was least effective in O2 transfer, but in terms of the gas exchange surface was the most effective.
Conclusion
Our analysis suggests that patients with severe hypoxemia and need for high flow ECMO benefit more from the PLS/CH or Hilite 7000 LT system. The ECC.05 system is advisable for patients with moderate hypoxemia and/or hypercapnia.
Similar content being viewed by others
References
MacLaren G, Combes A, Bartlett RH (2012) Contemporary extracorporeal membrane oxygenation for adult respiratory failure: life support in the new era. Intensive Care Med 38:210–220
Patroniti N, Zangrillo A, Pappalardo F, Peris A, Cianchi G, Braschi A, Iotti GA, Arcadipane A, Panarello G, Ranieri VM, Terragni P, Antonelli M, Gattinoni L, Oleari F, Pesenti A (2011) The Italian ECMO network experience during the 2009 influenza A(H1N1) pandemic: preparation for severe respiratory emergency outbreaks. Intensive Care Med 37:1447–1457
Karagiannidis C, Lubnow M, Philipp A, Riegger GA, Schmid C, Pfeifer M, Mueller T (2010) Autoregulation of ventilation with neurally adjusted ventilatory assist on extracorporeal lung support. Intensive Care Med 36:2038–2044
Khoshbin E, Roberts N, Harvey C, Machin D, Killer H, Peek GJ, Sosnowski AW, Firmin RK (2005) Poly-methyl pentene oxygenators have improved gas exchange capability and reduced transfusion requirements in adult extracorporeal membrane oxygenation. ASAIO J 51:281–287
Peek GJ, Killer HM, Reeves R, Sosnowski AW, Firmin RK (2002) Early experience with a polymethyl pentene oxygenator for adult extracorporeal life support. ASAIO J 48:480–482
Allen S, Holena D, McCunn M, Kohl B, Sarani B (2011) A review of the fundamental principles and evidence base in the use of extracorporeal membrane oxygenation (ECMO) in critically ill adult patients. J Intensive Care Med 26:13–26
Sidebotham D, McGeorge A, McGuinness S, Edwards M, Willcox T, Beca J (2010) Extracorporeal membrane oxygenation for treating severe cardiac and respiratory failure in adults: part 2-technical considerations. J Cardiothorac Vasc Anesth 24:164–172
Snyder TA, Eash HJ, Litwak KN, Frankowski BJ, Hattler BG, Federspiel WJ, Wagner WR (2006) Blood biocompatibility assessment of an intravenous gas exchange device. Artif Organs 30:657–664
Jegger D, Revelly JP, Horisberger J, Mallabiabarrena I, Seigneul I, Jachertz M, Von Segesser LK (2005) Ex vivo evaluation of a new extracorporeal lung assist device: novaLung membrane oxygenator. Int J Artif Organs 28:985–999
Park M, Costa EL, Maciel AT, Silva DP, Friedrich N, Barbosa EV, Hirota AS, Schettino G, Azevedo LC (2013) Determinants of oxygen and carbon dioxide transfer during extracorporeal membrane oxygenation in an experimental model of multiple organ dysfunction syndrome. PLoS ONE 8:e54954
Lehle K, Philipp A, Muller T, Schettler F, Bein T, Schmid C, Lubnow M (2013) Flow dynamics of different adult ECMO systems: a clinical evaluation. Artif Organs 38(5):391–398
Muller T, Bein T, Philipp A, Graf B, Schmid C, Riegger G (2013) Extracorporeal pulmonary support in severe pulmonary failure in adults: a treatment rediscovered. Dtsch Arztebl Int 110:159–166
Peek GJ, Firmin RK (1999) The inflammatory and coagulative response to prolonged extracorporeal membrane oxygenation. ASAIO J 45:250–263
Agati S, Ciccarello G, Fachile N, Scappatura RM, Grasso D, Salvo D, Undar A, Mignosa C (2006) DIDECMO: a new polymethylpentene oxygenator for pediatric extracorporeal membrane oxygenation. ASAIO J 52:509–512
Pieri M, Turla OG, Calabro MG, Ruggeri L, Agracheva N, Zangrillo A, Pappalardo F (2013) A new phosphorylcholine-coated polymethylpentene oxygenator for extracorporeal membrane oxygenation: a preliminary experience. Perfusion 28:132–137
De Somer F, Francois K, van Oeveren W, Poelaert J, De Wolf D, Ebels T, Van Nooten G (2000) Phosphorylcholine coating of extracorporeal circuits provides natural protection against blood activation by the material surface. Eur J Cardiothorac Surg 18:602–606
Acknowledgments
The authors would like to thank the nursing staff of the intensive care units and the perfusionists for excellent patient care and ECMO control.
Conflicts of interest
The study was supported by departmental resources without external funding. CS is an adviser to Maquet Cardiopulmonary AG, Rastatt, Germany. ML, AP, and TM received travel support and lecture honoraria from Maquet Cardiopulmonary AG, Rastatt, Germany. None of the other authors has a financial relationship with a commercial entity that has an interest in the subject of the presented manuscript or other conflicts of interest to disclose.
Author information
Authors and Affiliations
Corresponding author
Additional information
Take-home message: This is the first clinical study comparing the CO2 and O2 transfer capability of different ECMO systems. Better knowledge of the gas exchange performances of MOs is important to choose a suitable system for the patient initially and to assess the functionality of the system while in use.
Electronic supplementary material
Below is the link to the electronic supplementary material.
Rights and permissions
About this article
Cite this article
Lehle, K., Philipp, A., Hiller, KA. et al. Efficiency of gas transfer in venovenous extracorporeal membrane oxygenation: analysis of 317 cases with four different ECMO systems. Intensive Care Med 40, 1870–1877 (2014). https://doi.org/10.1007/s00134-014-3489-z
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00134-014-3489-z