Introduction

Nanostructured materials, because of their important characteristics have been the focus of serious interest for the last decade world-wide. In this field most of the novel and remarkable properties appear when the particle size ranges exist below 100 nm [1, 2]. Ferroelectric compounds are a group of materials that possess low dielectric loss, high dielectric constants, high electrical resistivity and suitable electrooptical and electromechanical activates [3]. Ferroelectric materials that have perovskite structure (ABO3), such as lead titanate (PbTiO3) and barium titanate (BaTiO3), are the most investigated ferroelectric oxide compounds due to their versatile characteristics for application in actuators, thin film capacitors, pyroelectric sensors and nonlinear optics. The materials with nano-sized dimensions have established significant research attention because of the unconventional physical and chemical behaviour at the nanoscale dimensions [4, 5]. PbTiO3 is usually used in numerous electronic devices due to a high pyroelectric coefficient and low permittivity [6, 7]. Some methods have been reported for preparation of PbTiO3, such as sol–gel method [8], hydrothermal synthesis [9], dip-pennanolithography [10] and liquid–solid–solution route [11]. It is known that chemical synthesis routes can be provide precise composition, high chemical purity, uniform microstructures and without high formation temperature of the perovskite phase based on molecular scale mixing in the preparation of the precursor. The sol–gel method is a usually applied chemical route for preparation of homogeneous large area thin films and uniform powders in micrometer, sub-micrometer or nanometer scales [12, 13].

Pentoxifylline (PTX) a tri-substituted xanthine derivative is a hemorheologic agent applied for the treatment of peripheral arterial disease and intermittent claudication. The drug improves blood flow by peripheral circulation via reducing blood viscosity, inhibiting platelet aggregation, improving erythrocyte flexibility, and decreasing fibrinogen concentration [14]. Beside the well-known hemorheological properties, it has been recognized to exert a broad range of immunological activities [15]. The compound also reduces leucocyte–endothelium interaction and shows a therapeutic effect in inhibiting ischemia reperfusion injury in microsurgical processes [16]. PTX can be prevented from strokes and used in managing sickle cell disease and increases blood flow to the brain [17]. Several methods for determination of PTX have been used, containing spectrophotometry [18], high performance liquid chromatography [14], gas chromatography [19], thin layer chromatography [20], and electrochemistry [21]. Some of the methods suffer from problems such as complexity, time consuming and need expensive instrumentation along with sample preparation. Among the various methods reported for PTX determination, electrochemistry technique is often favorable method because of their high sensitivity, selectivity, rapid detection and free of sample preparation steps. Electrochemical techniques that use chemically modified electrodes are usually important methods in medicine and biotechnology, environmental monitoring and industrial process control [22].

In the present study, a sol–gel method based on a metal-citrate system was used for synthesis of nano-sized PbTiO3. Then, a sensitive and simple PbTiO3 nanoparticles modified carbon paste electrode (PTN/CPE) was designed for determination of PTX. Primary investigation about the formation of the nanoparticles and modified electrode were perused. Then, analysis of PTX at the surface of the nanostructured electrode was performed by electrochemical methods.

Experimental

Materials

All solutions were freshly prepared by deionized water and chemicals used were of analytical grade. Citric acid, tetra-n-butyl titanate, H202, Pb(N03)2·6H2O, ammonia solution, PTX, carbon graphite powder, paraffin oil and other chemicals were purchased from Merck and used without any further purification. Stock solutions of 0.2 M H3PO4, CH3COOH, H3BO3 and a saturated solution of sodium hydroxide were used for preparation of 0.2 M Britton–Robinson (B–R) buffer.

Preparation of PbTiO3 nanostructures and working electrodes

For synthesis of PbTiO3 nano-sized powders, tetra-n-butyl titanate was added to stirred deionized water and the precipitation was observed. Then, the precipitate was filtered and washed with deionized water. An appropriate of citric acid and H202 both in a 2:1 molar ratio against Ti(n-0C4H9)4, was added and then, the solution became viscous and maroon in color. The pH value of the solution was adjusted to 6.0 by ammonia solution. For preparation of a water-soluble precursor for Ti(IV) continuous stirring and refluxing at 60 °C was performed. Then, stoichiometric amount of Pb(NO3)2 was dissolved in 5.0 M citric acid solution with pH of 6.0 (adjusted by ammonia) and then, the Pb(II) precursor was mixed with the Ti(IV) precursor. The above mixture was refluxed at 60 °C for 2 h with continuous stirring. After drying in a furnace at 60 °C for 24 h a porous citreous resin was formed. Then, the solid resin precursors were calcined at 500 °C and 550 °C for 2 h to obtain the PbTiO3 nanopowders (Scheme 1).

Scheme 1
scheme 1

Schematic diagram for the synthesis of PbTiO3 nanostructures

At the next step, the carbon paste electrode (CPE) was prepared by hand-mixing 0.50 g graphite powder with about 0.2 mL of paraffin oil in a pestle and mortar. For preparation of the modified electrode, appropriate amount of PbTiO3 nanopowders calcined at 550 °C (in proportions of 2, 4, 6, 8, and 10%, w/w of PbTiO3 to graphite powder) was dispersed in 5.0 mL deionized water by an ultrasound bath and then mixed into graphite powder. After vaporizing water, about 0.2 mL of paraffin oil was added to the above mixture.

Apparatus

The X-ray diffraction (XRD) patterns of the nanopowders were recorded by a Model PI TS 3003 of SEIFERT diffractometer using Cu Kα radiation (k = 1.5418 Å) to examine the crystallization and structural properties of PbTiO3 nanopowders. Scanning electron microscopy (SEM) micrographs were obtained using a KYKY-EM3200. Electrochemical experiments were carried out by an Autolab potentiostat–galvanostat PGSTAT 35 (Eco Chemie Utrecht, Netherlands), equipped with NOVA 1.6 software. The electrochemical cell was equipped with a three-electrode system containing a modified carbon paste, a platinum wire (Metrohm, Switzer-land) and an Ag/AgCl/KCl (3.0 M) (Metrohm, Switzerland) as the working, counter and reference electrodes, respectively. A digital pH meter (Metrohm model 691) was applied for preparation of buffer that served as a supporting electrolyte in the electrochemical analysis. An ultrasound bath (Bandelin Sonorex, Germany) at a constant frequency (35 kHz) was applied for preparing the modified nanostructured electrodes. An ultra-pure water system (smart 2 pure, TKA, Germany) was used for preparation of deionized water. All electrochemical experiments were performed at 25 °C.

Results and discussion

Structural and surface analysis

The structural studies of the samples was investigated using XRD pattern of the PbTiO3 nanopowders calcined at 500 °C and 550 °C. As can be observed in Fig. 1, at 500 °C, the crystallization of tetragonal phase took place along with impurity phase [23]. But, at 550 °C all peaks can be attributed to the PbTiO3 crystals with a tetragonal structure without any impurity peak. The crystallite size (t) of the PbTiO3 nanopowders calcined at the two temperatures was achieved by the Scherrer’s equation (Eq. 1):

$$t \, = \, 0.9\lambda \, / \, \beta \, \cos \theta$$
(1)
Fig. 1
figure 1

XRD pattern of PbTiO3 nanostructures calcined at a 500 °C and b 550 °C for 2 h

where λ, β and θ are wavelength of radiation, the peak width at its half maximum intensity and diffraction angle, respectively. Based on the data the crystallite size of PbTiO3 nanostructures at 500 °C and 550 °C were obtained 30.3 and 39.2 nm, respectively.

The morphologies and particle sizes of the PbTiO3 nanopowders calcined at 550 °C have been investigated using SEM micrographs. Fig 2 shows that PbTiO3 nanopowders with the diameter in the range of 37 nm to 60 nm are uniformly distributed.

Fig. 2
figure 2

SEM micrographs of PbTiO3 nanostructures calcined at 550 °C for 2 h

Also for investigation of the CPE modification process SEM was used. Figure 3 compares the morphological features of (A) CPE and (B) PTN/CPE. SEM was used to estimate the surface characteristics of PbTiO3 nanoparticles at the surface of CPE. The SEM micrograph of CPE showed irregularly shaped particles that appeared to remain isolated (Fig. 3a). Whereas, the SEM micrograph of PTN/CPE (Fig. 3b), demonstrates dissimilar morphology than the unmodified electrode. The micrographs indicate that the modification of the CPE surface with PbTiO3 nanoparticles has been well accomplished.

Fig. 3
figure 3

SEM micrographs of a CPE and b PTN/CPE

Electrochemical properties of the PTN/CPE

For investigation of the catalytic activity of the PTN/CPE, differential pulse voltammetry (DPV) technique was used. The differential pulse voltammograms of PTX at the surface of the CPE and PTN/CPE were recorded (Fig. 4). As can be seen, PTN/CPE shows a well-defined oxidation peak with magnification of about 3.0 times greater than the CPE. An enhancement in current response by the nanostructured modified electrode is clear evidence of the catalytic effect of PbTiO3 nanostructures on the oxidation of PTX.

Fig. 4
figure 4

Differential pulse voltammograms of PTX at the surface of a CPE and b PTN/CPE

Optimization studies

DPV data in Fig. 4 approved that the PTN/CPE shows a higher sensitivity for PTX compared to the CPE. But, a set of experiments were performed at the surface of PTN/CPE to achieve the optimum experimental factors for achieving the best sensitivity for PTX.

Effect of pH

The pH value is a key parameter in electrochemical studies in pH-dependent systems. Therefore, the effect of pH was studied by DPV in 0.2 M B-R buffer with different pH values (7.0, 7.5, 8.0, 8.5, and 9.0) containing PTX. As it is shown in Fig. 5a, with changing in pH of the supporting electrolyte both the peak current and peak potential of PTX is shifted. Thus, the oxidation reaction of PTX at the PTN/CPE is a pH-dependent system. Additionally, the highest oxidation peak current for PTX was obtained at pH 7.5 (Fig. 5b). Thus, pH 7.5 was selected for the sensitive determination of PTX.

Fig. 5
figure 5

a DPV of PTX at the surface of PTN/CPE at different buffered pHs, a to e correspond to pH values of 7.0, 7.5, 8.0, 8.5, and 9.0, respectively. b Plot of anodic peak current (Ipa, mA) vs. pH

Effect of PbTiO3 nanoparticles

The amount of modifier is a main factor for construction of a modified electrode. Therefore, different amount of PbTiO3 nanoparticles (2, 4, 6, 8, and 10%, w/w of PbTiO3 nanoparticles to graphite powder) was used for preparation of different working electrodes. The DPV experiments were performed and the corresponding anodic peak current of PTX were recorded at the surface of PTN/CPE (Fig. 6). According the studies, for a modified electrode with 10% of PbTiO3 nanoparticles, maximum current sensitivity was obtained. Thus, the value was used for construction of the modified nanostructured electrode.

Fig. 6
figure 6

Plot of (Ipa, mA) of PTX with different quantities of PbTiO3 nanoparticles

Effect of instrumental parameters

For achieving the best sensitivity for detection of PTX both chemical (pH and PbTiO3 amount) and instrumental parameters should be optimized. Therefore, the effect of scan rate (υ) and modulation amplitude as instrumental parameters on the oxidation peak current of PTX was studied. So, the experiments were performed at different scan rates from 10.0 to 50.0 mV s−1 (Fig. 7) at the surface of PTN/CPE. The results showed by increasing the scan rate, the heights of oxidation peak were increased. Therefore, scan rate was adjusted on 50.0 mV s−1 for determination of PTX.

Fig. 7
figure 7

Plot of (Ipa, mA) of PTX at the surface of PTN/CPE in different scan rates

Additionally, modulation amplitude effect on the electrochemical behaviour of PTX was studied. Figure 8 shows the oxidation peak current of PTX changes with changing in modulation amplitude from 0.1 to 0.5 V. The results shows maximum of peak current obtains when modulation amplitude was adjusted on 0.3 V.

Fig. 8
figure 8

Plot of (Ipa, mA) of PTX at the surface of PTN/CPE in different modulation amplitudes

Determination of PTX at the surface of PTN/CPE

The main purpose of the work is designing a high sensitive method for determination of PTX. For achieving the best response of PTX at the surface of PTN/CPE, all effective factors on the oxidation peak current of the analyte were adjusted on the optimized conditions. Therefore, the voltammograms were recorded in the presence of B–R buffer (pH 7.5), 10% of PbTiO3 nanoparticles, a scan rate of 50.0 mV s−1 and a modulation amplitude of 0.3 V. Fig 9a shows the differential pulse voltammograms obtained for the oxidation of different PTX concentrations at the PTN/CPE. Additionally, Fig. 9b exhibits clearly that the plot of oxidation peak current versus PTX concentration is constituted of three linear segments, corresponding to three different ranges of analyte concentration. From the data, detection limit (3σ) of PTX was obtained 2.1 nM and linear range estimated to be 0.02–220.0 μM. A low detection limit for PTX is indicating the nanostructured modified electrode is an appropriate sensor for determination of PTX with high sensitivity. Therefore, the nanostructured electrode is a potential sensor for trace analysis of PTX in real samples.

Fig. 9
figure 9

a Differential pulse voltammograms of different concentrations of PTX at the surface of PTN/CPE and b plots of peak current as a function of PTX concentration

Interference study

The selectivity of the PTN/CPE was evaluated in the presence of various interfering molecules. The tolerance limit of DPV studies was measured as the maximum concentration of the interference which caused an about ± 5% relative error in the analysis. The differential pulse voltammetric responses of PTX were studied in the presence of several possible interfering substances like Na+, K+, NH4+, Zn2+, Cl, CO32−, SO42−, citric acid, lactic acid, dextrose, glucose, sucrose, starch, oxalic acid, glycine, L-phenyl alanine, uric acid, and dopamine. These substances may present in biological fluids and can interfere in the detection of PTX with conventional methods. The experiments showed that the molecules did not significantly influence on the oxidation peak current of PTX. The results are listed in Table 1.

Table 1 Influence of some foreign substances on the determination of PTX at the surface of PTN/CPE

The repeatability and stability

The electrode ability for the generation of a reproducible surface was studied by DPV obtained from five prepared PTN/CPEs. The calculated relative standard deviation for various factors accepted as the criteria for a satisfactory surface reproducibility (1.2–3.8%). Furthermore, the long-term stability of the nanostructured modified electrode was evaluated over a 2-week period (Fig. 10). Differential pulse voltammograms recorded in the period showed the peak potential for PTX was unchanged and the current signals reduced only less than 2.1%. These results indicate that the repeatability and stability of the nanostructured electrode are acceptable.

Fig. 10
figure 10

Plot of stability of the PTN/CPE in 0.2 M B–R buffer (pH 7.5) for a 2-week period

Application of the method in real sample analysis

For evaluation the applicability of the method, detection of PTX in industrial wastewater and biological samples (human blood plasma) was performed by standard addition method. The recovery percentages of the method were obtained between 94.0 and 104.7% and 96.5–106.0% for industrial wastewater and biological samples, respectively. The results are listed in Table 2. According to the results, very good recoveries were obtained for the detection of PTX, which indicates that the nanostructured sensor can be applied for the analysis of the analyte in environmental and biological samples.

Table 2 Determination of PTX in wastewater and human blood plasma samples

Conclusions

In the present study, PbTiO3 nanostructures were synthesized by a sol–gel method using citric acid as complexing agent. The structural and microscopic studies were performed using XRD and SEM techniques. Then, the synthesized nanopowders were used for designing a novel PbTiO3 nanostructures modified carbon paste electrode and its application for the determination of PTX. Characterization mechanisms showed the PbTiO3 nanopowders and nanostructured modified electrode were actually formed. The PTN/CPEs exhibited highly catalytic effect to the oxidations of the analyte. The modified electrode showed high sensitivity in voltammetric measurements of PTX. Additionally, the nanostructured sensor exhibited good repeatability and stability for determination of PTX. Additionally, the nanostructures electrode was successfully used for the analysis of PTX in real complicated samples.