Real-Time Gating of IVUS Sequences Based on Motion Blur Analysis: Method and Quantitative Validation
verfasst von : Carlo Gatta, Simone Balocco, Francesco Ciompi, Rayyan Hemetsberger, Oriol Rodriguez Leor, Petia Radeva
Erschienen in: Medical Image Computing and Computer-Assisted Intervention – MICCAI 2010
Verlag: Springer Berlin Heidelberg
Aktivieren Sie unsere intelligente Suche, um passende Fachinhalte oder Patente zu finden.
Wählen Sie Textabschnitte aus um mit Künstlicher Intelligenz passenden Patente zu finden. powered by
Markieren Sie Textabschnitte, um KI-gestützt weitere passende Inhalte zu finden. powered by
Intravascular Ultrasound (IVUS) is an image-guiding technique for cardiovascular diagnostic, providing cross-sectional images of vessels. During the acquisition, the catheter is pulled back (pullback) at a constant speed in order to acquire spatially subsequent images of the artery. However, during this procedure, the heart twist produces a swinging fluctuation of the probe position along the vessel axis. In this paper we propose a real-time gating algorithm based on the analysis of
motion blur
variations during the IVUS sequence. Quantitative tests performed on an in-vitro ground truth data base shown that our method is superior to state of the art algorithms both in computational speed and accuracy.