Abstract.

On the basis of kinetic measurements of biomolecular interactions, a reflectometric interference spectroscopy (RIfS) setup is compared with two commercial instruments. These instruments are based on evanescent wave techniques, surface plasmon resonance (SPR) (represented by BIAcore 2000) and resonant mirror (RM) technique (using IAsys plus). All methods allow a label-free and time-resolved optical detection of biomolecular interaction.

These methods are mainly used in biomolecular interaction analysis (BIA). They provide practical techniques for quantifying equilibrium constants and rate constants over several orders of magnitude.

The general parameters of the three detectors, namely baseline noise and drift as well as overall sensitivity and limits of detection were compared. The fluid handling and the related implications on the measurements have also been considered.

The interaction between thrombine and thrombine inhibitor (TI) was investigated as a test system with the three different methods and the kinetic rate constants were determined and compared. For this TI was immobilized on the surface and binding of thrombine was monitored time-resolved. Determination of the kinetic rate constants could prove that the RIfS set-up is comparable with SPR using BIAcore 2000 and RM technique represented by IAsys plus.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

Author information

Authors and Affiliations

Authors

Additional information

Electronic Publication

Rights and permissions

Reprints and permissions

About this article

Cite this article

Hänel, C., Gauglitz, G. Comparison of reflectometric interference spectroscopy with other instruments for label-free optical detection. Anal Bioanal Chem 372, 91–100 (2002). https://doi.org/10.1007/s00216-001-1197-3

Download citation

  • Received:

  • Revised:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00216-001-1197-3

Navigation