A Regional Comparison of Calibration Results for Type K Thermocouple Wire from (100 to 1,100) °C

  • Published:
International Journal of Thermophysics Aims and scope Submit manuscript

Abstract

Under the auspices of the Inter-American Metrology System (SIM), the National Institute of Standards and Technology (NIST) initiated a regional comparison for type K thermocouples from (100 to 1,100) °C with 11 participating countries. The use of type K material above approximately 200 °C is considered destructive. Therefore, each participating laboratory was sent new, unused wire from a lot of material characterized by NIST. The uniformity of the lot was remarkable, especially at temperatures above 500 °C; the standard deviation of the thermocouple emf values of multiple cuts tested at NIST was 2.7 μV or less over the full temperature range. The high uniformity eliminated any need to correct for variations of the transfer standard among the laboratories, greatly simplifying the analysis. The level of agreement among the laboratories’ results was quite good. Even though test procedures and equipment varied significantly among the participants, the standard deviation of all emf values at each test temperature was less than the equivalent of 0.20 °C at 200 °C and below, and less than 0.60 °C from (400 to 1,100) °C. Of the 380 total bilateral combinations of the data at the eight test temperatures, only 13 (i.e., 3.4% of all combinations) are outside the k = 2 limits, and of these 13, only 3 are outside k = 3 limits. All the outliers occur at temperatures of 800 °C and below, which suggests that drift of the type K wire due to high-temperature oxidation did not cause changes in the thermocouple emf comparable to or larger than the claimed uncertainties.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  1. G.W. Burns, M.G. Scroger, The Calibration of Thermocouples and Thermocouple Materials, NIST Special Pub. 250–235 (U.S. Govt. Printing Office, Washington, D.C., 1989)

  2. D. Ripple, G.W. Burns, M.G. Scroger, Assessment of Uncertainties of Thermocouple Calibrations at NIST, NISTIR 5340 (National Institute of Standards and Technology, Gaithersburg, Maryland, 1994)

  3. D.C. Ripple, K.M. Garrity, M. Araya, C.R. Cabrera, L. Cordova Murillo, M.E. de Vanegas, D.J. Gee, E. Guillen, S. Martinez-Martinez, E. Mendez-Lango, L. Mussio, S.G. Petkovic, K.N. Quelhas, G. Rangugni, O. Robatto, E. von Borries Rocha, Metrologia 44 Tech. Suppl. 03007 (2007)

  4. Cox M.G. (2002). Metrologia 39: 589

    Article  ADS  Google Scholar 

  5. J.G. Hust, R.L. Powell, L.L. Sparks, in Temperature: Its Measurement and Control in Science and Industry, Vol. 4, ed. by H.H. Plumb (ISA, Pittsburgh, Pennsylvania, 1972), pp. 1525–1535

  6. Ripple D.C. and Garrity K.M. (2006). Measure 1: 28

    Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to K. M. Garrity.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Garrity, K.M., Ripple, D.C., Araya, M. et al. A Regional Comparison of Calibration Results for Type K Thermocouple Wire from (100 to 1,100) °C. Int J Thermophys 29, 1828–1837 (2008). https://doi.org/10.1007/s10765-007-0354-0

Download citation

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s10765-007-0354-0

Keywords

Navigation