Numerical study of the components positioning influence on the stability of a reverse shoulder prosthesis

verfasst von: T. Ingrassia, A. Mancuso, V. Nigrelli, D. Tumino

Erschienen in: International Journal on Interactive Design and Manufacturing (IJIDeM) | Ausgabe 3/2014

Einloggen

Aktivieren Sie unsere intelligente Suche, um passende Fachinhalte oder Patente zu finden.

search-config
loading …

Abstract

Aim of this paper is to setup a novel procedure able to analyze performances of a reverse shoulder prosthesis when different geometrical configurations are assumed. Nowadays, such a prosthesis is widely used but, because of its novelty, data in literature give poor information about performances and limits to its applicability. The activity has been divided into the following steps. At the beginning the shape of the prosthesis has been digitally acquired via a 3D scanner. Then, CAD models of all prosthetic components have been geometrically optimized in a way to obtain final entities suitable for numerical simulations. After that, CAD assemblies have been created between prosthetic components and bones (humerus and scapula) involved in the shoulder joint. Following step has been the setup of numerical finite element method models to simulate use conditions. To this scope, analyses have been performed in accordance with experimental conditions found in literature. Stability conditions have been verified under the action of horizontal and vertical instability loads with different version angles between humerus and the humeral implant. In particular, the stability ratios of the prosthesis have been calculated for the analysed loading conditions. Obtained results show how the positioning has a great influence on the shoulder stability and allow the definition of guidelines for the application of this prosthesis.

Sie haben noch keine Lizenz? Dann Informieren Sie sich jetzt über unsere Produkte:

Springer Professional "Wirtschaft+Technik"

Online-Abonnement

Mit Springer Professional "Wirtschaft+Technik" erhalten Sie Zugriff auf:

  • über 102.000 Bücher
  • über 537 Zeitschriften

aus folgenden Fachgebieten:

  • Automobil + Motoren
  • Bauwesen + Immobilien
  • Business IT + Informatik
  • Elektrotechnik + Elektronik
  • Energie + Nachhaltigkeit
  • Finance + Banking
  • Management + Führung
  • Marketing + Vertrieb
  • Maschinenbau + Werkstoffe
  • Versicherung + Risiko

Jetzt Wissensvorsprung sichern!

Springer Professional "Technik"

Online-Abonnement

Mit Springer Professional "Technik" erhalten Sie Zugriff auf:

  • über 67.000 Bücher
  • über 390 Zeitschriften

aus folgenden Fachgebieten:

  • Automobil + Motoren
  • Bauwesen + Immobilien
  • Business IT + Informatik
  • Elektrotechnik + Elektronik
  • Energie + Nachhaltigkeit
  • Maschinenbau + Werkstoffe




 

Jetzt Wissensvorsprung sichern!

Springer Professional "Wirtschaft"

Online-Abonnement

Mit Springer Professional "Wirtschaft" erhalten Sie Zugriff auf:

  • über 67.000 Bücher
  • über 340 Zeitschriften

aus folgenden Fachgebieten:

  • Bauwesen + Immobilien
  • Business IT + Informatik
  • Finance + Banking
  • Management + Führung
  • Marketing + Vertrieb
  • Versicherung + Risiko




Jetzt Wissensvorsprung sichern!

Literatur
1.
Zurück zum Zitat Grammont, P.M., Baulot, E.: Delta shoulder prosthesis for rotator cuff rupture. Orthopedics 16, 65–68 (1993) Grammont, P.M., Baulot, E.: Delta shoulder prosthesis for rotator cuff rupture. Orthopedics 16, 65–68 (1993)
2.
Zurück zum Zitat Matsen, F.A., Boileau, P., Walch, G.: The reverse total shoulder arthroplasty. J. Bone Joint Surg. Am. 88–A(3), 660–667 (2007) Matsen, F.A., Boileau, P., Walch, G.: The reverse total shoulder arthroplasty. J. Bone Joint Surg. Am. 88–A(3), 660–667 (2007)
3.
Zurück zum Zitat Terrier, A., Reist, A., Merlini, F., Farron, A.: Simulated joint and muscle forces in reversed and anatomic shoulder prostheses. J. Bone Joint Surg. Br. 90(6), 751–756 (2008)CrossRef Terrier, A., Reist, A., Merlini, F., Farron, A.: Simulated joint and muscle forces in reversed and anatomic shoulder prostheses. J. Bone Joint Surg. Br. 90(6), 751–756 (2008)CrossRef
4.
Zurück zum Zitat Boileau, P., Watkinson, D.J., Hatzidakis, A.M., Balg, F.: Grammont reverse prosthesis: design, rationale, and biomechanics. J. Shoulder Elbow Surg. 14(1, Supp 1), 147–161 (2005)CrossRef Boileau, P., Watkinson, D.J., Hatzidakis, A.M., Balg, F.: Grammont reverse prosthesis: design, rationale, and biomechanics. J. Shoulder Elbow Surg. 14(1, Supp 1), 147–161 (2005)CrossRef
5.
Zurück zum Zitat Cuff, D., Pupello, D., Virani, N., Levy, J.: Reverse shoulder arthroplasty for the treatment of rotator cuff deficiency. J. Bone Joint Surg. Am. 90, 1244–1251 (2008)CrossRef Cuff, D., Pupello, D., Virani, N., Levy, J.: Reverse shoulder arthroplasty for the treatment of rotator cuff deficiency. J. Bone Joint Surg. Am. 90, 1244–1251 (2008)CrossRef
6.
Zurück zum Zitat Sirveaux, F., Favard, L., Oudet, D., Huquet, D., Walch, G., Molé, D.: Grammont inverted total shoulder arthroplasty in the treatment of glenohumeral osteoarthritis with massive rupture of the cuff. J. Bone Joint Surg. 86–B, 388–395 (2004)CrossRef Sirveaux, F., Favard, L., Oudet, D., Huquet, D., Walch, G., Molé, D.: Grammont inverted total shoulder arthroplasty in the treatment of glenohumeral osteoarthritis with massive rupture of the cuff. J. Bone Joint Surg. 86–B, 388–395 (2004)CrossRef
7.
Zurück zum Zitat Frankle, M.A., Cuff, D., Levy, J.C., Gutiérrez, S.: Evaluation of abduction range of motion and avoidance of inferior scapular impingement in a reverse shoulder model. J. Shoulder Elbow Surg. 17(4), 608–615 (2008)CrossRef Frankle, M.A., Cuff, D., Levy, J.C., Gutiérrez, S.: Evaluation of abduction range of motion and avoidance of inferior scapular impingement in a reverse shoulder model. J. Shoulder Elbow Surg. 17(4), 608–615 (2008)CrossRef
8.
Zurück zum Zitat Simovitch, R.W., Zumstein, M.A., Lohri, E., Helmy, N., Gerber, C.: Predictors of scapular notching in patients managed with the Delta III reverse total shoulder replacement. J. Bone Joint Surg. 89, 588–600 (2007)CrossRef Simovitch, R.W., Zumstein, M.A., Lohri, E., Helmy, N., Gerber, C.: Predictors of scapular notching in patients managed with the Delta III reverse total shoulder replacement. J. Bone Joint Surg. 89, 588–600 (2007)CrossRef
9.
Zurück zum Zitat Franklin, J.L., Barrett, W.P., Jackins, S.E., Matsen, F.A.: Glenoid loosening in total shoulder arthroplasty. Association with rotator cuff deficiency. J. Arthroplasty 3, 39–46 (1988)CrossRef Franklin, J.L., Barrett, W.P., Jackins, S.E., Matsen, F.A.: Glenoid loosening in total shoulder arthroplasty. Association with rotator cuff deficiency. J. Arthroplasty 3, 39–46 (1988)CrossRef
10.
Zurück zum Zitat Gutiérrez, S., Levy, J.C., Lee, W.E.: Center of rotation affects abduction range of motion of reverse shoulder arthroplasty. Clin. Orthop. Relat. Res. 458, 78–82 (2007) Gutiérrez, S., Levy, J.C., Lee, W.E.: Center of rotation affects abduction range of motion of reverse shoulder arthroplasty. Clin. Orthop. Relat. Res. 458, 78–82 (2007)
11.
Zurück zum Zitat Frankle, M.A., Luo, Z.P., Gutiérrez, S., Levy, J.C.: Arc of motion and socket depth in reverse shoulder implants. Clin. Biomech. 24(6), 473–479 (2009)CrossRef Frankle, M.A., Luo, Z.P., Gutiérrez, S., Levy, J.C.: Arc of motion and socket depth in reverse shoulder implants. Clin. Biomech. 24(6), 473–479 (2009)CrossRef
12.
Zurück zum Zitat Zumstein, M.A., Pinedo, M., Old, J., Boileau, P.: Problems, complications, reoperations, and revisions in reverse total shoulder arthroplasty: a systematic review. J. Shoulder Elbow Surg. 20(1), 146–157 (2011)CrossRef Zumstein, M.A., Pinedo, M., Old, J., Boileau, P.: Problems, complications, reoperations, and revisions in reverse total shoulder arthroplasty: a systematic review. J. Shoulder Elbow Surg. 20(1), 146–157 (2011)CrossRef
13.
Zurück zum Zitat Gutiérrez, S., Keller, T.S., Levy, J.C., Lee, W.E.: Hierarchy of stability factors in reverse shoulder arthroplasty. Clin. Orthop. Relat. Res. 466(3), 670–676 (2008)CrossRef Gutiérrez, S., Keller, T.S., Levy, J.C., Lee, W.E.: Hierarchy of stability factors in reverse shoulder arthroplasty. Clin. Orthop. Relat. Res. 466(3), 670–676 (2008)CrossRef
14.
Zurück zum Zitat Favre, P., Sussmann, P.S., Gerber, C.: The effect of component positioning on intrinsic stability of the reverse shoulder arthroplasty. J. Shoulder Elbow Surg. 19, 550–556 (2010)CrossRef Favre, P., Sussmann, P.S., Gerber, C.: The effect of component positioning on intrinsic stability of the reverse shoulder arthroplasty. J. Shoulder Elbow Surg. 19, 550–556 (2010)CrossRef
15.
Zurück zum Zitat A. Clouthier, M. Hetzler, G. Fedorak: Effect of loading direction on force to dislocate in reverse shoulder arthroplasty. In: Proceedings of 35th Annual Meeting of the American Society of biomechanics, August 10–13, 2011, California A. Clouthier, M. Hetzler, G. Fedorak: Effect of loading direction on force to dislocate in reverse shoulder arthroplasty. In: Proceedings of 35th Annual Meeting of the American Society of biomechanics, August 10–13, 2011, California
16.
Zurück zum Zitat Ackland, D.C., Pandy, M.G.: Lines of action and stabilizing potential of the shoulder musculature. J. Anat. 215, 184–197 (2009)CrossRef Ackland, D.C., Pandy, M.G.: Lines of action and stabilizing potential of the shoulder musculature. J. Anat. 215, 184–197 (2009)CrossRef
17.
Zurück zum Zitat Barone, S., Paoli, A., Razionale, A.V.: Assessment of chronic wounds by 3D optical imaging based on integrating geometrical, chromatic and thermal data. J. Eng. Med. 225(2), 181–193 (2011)CrossRef Barone, S., Paoli, A., Razionale, A.V.: Assessment of chronic wounds by 3D optical imaging based on integrating geometrical, chromatic and thermal data. J. Eng. Med. 225(2), 181–193 (2011)CrossRef
18.
Zurück zum Zitat Ausiello, P., Franciosa, P., Martorelli, M., Watts, D.C.: Numerical fatigue 3D-FE modelling of indirect composite-restored posterior teeth. Dent. Mater. 27(5), 423–430 (2011)CrossRef Ausiello, P., Franciosa, P., Martorelli, M., Watts, D.C.: Numerical fatigue 3D-FE modelling of indirect composite-restored posterior teeth. Dent. Mater. 27(5), 423–430 (2011)CrossRef
19.
Zurück zum Zitat Ausiello, P., Franciosa, P., Martorelli, M., Watts, D.C.: Mechanical behavior of post-restored upper canine teeth: a 3D FE analysis. Dent. Mater. 27(12), 1285–1294 (2011)CrossRef Ausiello, P., Franciosa, P., Martorelli, M., Watts, D.C.: Mechanical behavior of post-restored upper canine teeth: a 3D FE analysis. Dent. Mater. 27(12), 1285–1294 (2011)CrossRef
20.
Zurück zum Zitat Ingrassia, T., Nalbone, L., Nigrelli, V., Tumino, D., Ricotta, V.: Finite element analysis of two total knee joint prostheses. Int. J. Interact. Design Manuf. 7(2), 91–101 (2012). doi:10.1007/s12008-012-0167-7 Ingrassia, T., Nalbone, L., Nigrelli, V., Tumino, D., Ricotta, V.: Finite element analysis of two total knee joint prostheses. Int. J. Interact. Design Manuf. 7(2), 91–101 (2012). doi:10.​1007/​s12008-012-0167-7
21.
Zurück zum Zitat Cerniglia, D., Ingrassia, T., D’Acquisto, L., Saporito, M., Tumino, D.: Contact between the components of a knee prosthesis: numerical and experimental study. Frattura ed Integrità Strutturale 22, 56–68 (2012). doi:10.3221/IGF-ESIS.22.07 Cerniglia, D., Ingrassia, T., D’Acquisto, L., Saporito, M., Tumino, D.: Contact between the components of a knee prosthesis: numerical and experimental study. Frattura ed Integrità Strutturale 22, 56–68 (2012). doi:10.​3221/​IGF-ESIS.​22.​07
25.
Zurück zum Zitat Lanzotti, A.: Robust design of car packaging in virtual environment. Int. J. Interact. Design Manuf. 2(1), 39–46 (2008)CrossRef Lanzotti, A.: Robust design of car packaging in virtual environment. Int. J. Interact. Design Manuf. 2(1), 39–46 (2008)CrossRef
26.
Zurück zum Zitat Ingrassia, T., Nigrelli, V.: Design optimization and analysis of a new rear underrun protective device for truck. In: Proceedings of the 8th International Symposium on Tools and Methods of Competitive Engineering, TMCE 2010, 2, 2010, pp. 713–725 (2010) Ingrassia, T., Nigrelli, V.: Design optimization and analysis of a new rear underrun protective device for truck. In: Proceedings of the 8th International Symposium on Tools and Methods of Competitive Engineering, TMCE 2010, 2, 2010, pp. 713–725 (2010)
27.
Zurück zum Zitat Ingrassia, T., Nigrelli, V., Buttitta, R.: A comparison of simplex and simulated annealing for optimization of a new rear underrun protective device. Eng. Comput. (2012). doi:10.1007/s00366-012-0270-1 Ingrassia, T., Nigrelli, V., Buttitta, R.: A comparison of simplex and simulated annealing for optimization of a new rear underrun protective device. Eng. Comput. (2012). doi:10.​1007/​s00366-012-0270-1
28.
Zurück zum Zitat Cazeneuve, J.F., Cristofari, D.J.: The reverse shoulder prosthesis in the treatment of fractures of the proximal humerus in the elderly. J. Bone Joint Surg. Br. 92(4), 535–539 (2010)CrossRef Cazeneuve, J.F., Cristofari, D.J.: The reverse shoulder prosthesis in the treatment of fractures of the proximal humerus in the elderly. J. Bone Joint Surg. Br. 92(4), 535–539 (2010)CrossRef
29.
Zurück zum Zitat Kenderian, S., Djordjevic, B.B., Cerniglia, D., Garcia, G.: Dynamic railroad inspection using the laser-air hybrid ultrasonic technique. Insight Non-Destruct. Test. Cond. Monit. 48(6), 336–341 (2006)CrossRef Kenderian, S., Djordjevic, B.B., Cerniglia, D., Garcia, G.: Dynamic railroad inspection using the laser-air hybrid ultrasonic technique. Insight Non-Destruct. Test. Cond. Monit. 48(6), 336–341 (2006)CrossRef
30.
Zurück zum Zitat Pantano, A., Cerniglia, D.: Simulation of laser generated ultrasound with application to defect detection. Appl. Phys. A Mater. Sci. Process. 91(3), 521–528 (2008)CrossRef Pantano, A., Cerniglia, D.: Simulation of laser generated ultrasound with application to defect detection. Appl. Phys. A Mater. Sci. Process. 91(3), 521–528 (2008)CrossRef
31.
Zurück zum Zitat Cerniglia, D., Pantano, A., Montinaro, N.: Simulation of laser-generated ultrasonic wave propagation in solid media and air with application to NDE. Appl. Phys. A Mater. Sci. Process. 98(2), 327–336 (2010)CrossRef Cerniglia, D., Pantano, A., Montinaro, N.: Simulation of laser-generated ultrasonic wave propagation in solid media and air with application to NDE. Appl. Phys. A Mater. Sci. Process. 98(2), 327–336 (2010)CrossRef
32.
Zurück zum Zitat Cerniglia, D., Nigrelli, V., Mancuso, A., Alberti, A.: Real-time non-contact ultrasonic detection of surface defects on objects moving at high speed. Frattura ed Integrità Strutturale 22, 93–101 (2012). doi:10.3221/IGF-ESIS.22.10 Cerniglia, D., Nigrelli, V., Mancuso, A., Alberti, A.: Real-time non-contact ultrasonic detection of surface defects on objects moving at high speed. Frattura ed Integrità Strutturale 22, 93–101 (2012). doi:10.​3221/​IGF-ESIS.​22.​10
33.
Zurück zum Zitat Godest, A.C., Beaugonin, M., Haug, E., Taylor, M., Gregson, P.J.: Simulation of a knee joint replacement during a gait cycle using explicit finite element analysis. J. Biomech. 35, 267–275 (2002)CrossRef Godest, A.C., Beaugonin, M., Haug, E., Taylor, M., Gregson, P.J.: Simulation of a knee joint replacement during a gait cycle using explicit finite element analysis. J. Biomech. 35, 267–275 (2002)CrossRef
Metadaten
Titel
Numerical study of the components positioning influence on the stability of a reverse shoulder prosthesis
verfasst von
T. Ingrassia
A. Mancuso
V. Nigrelli
D. Tumino
Publikationsdatum
01.08.2014
Verlag
Springer Paris
Erschienen in
International Journal on Interactive Design and Manufacturing (IJIDeM) / Ausgabe 3/2014
Print ISSN: 1955-2513
Elektronische ISSN: 1955-2505
DOI
https://doi.org/10.1007/s12008-014-0215-6

Weitere Artikel der Ausgabe 3/2014

International Journal on Interactive Design and Manufacturing (IJIDeM) 3/2014 Zur Ausgabe