Automatic quality assessment and peak identification of auditory brainstem responses with fitted parametric peaks

https://doi.org/10.1016/j.cmpb.2014.02.015Get rights and content

Abstract

The recording of the auditory brainstem response (ABR) is used worldwide for hearing screening purposes. In this process, a precise estimation of the most relevant components is essential for an accurate interpretation of these signals. This evaluation is usually carried out subjectively by an audiologist. However, the use of automatic methods for this purpose is being encouraged nowadays in order to reduce human evaluation biases and ensure uniformity among test conditions, patients, and screening personnel. This article describes a new method that performs automatic quality assessment and identification of the peaks, the fitted parametric peaks (FPP). This method is based on the use of synthesized peaks that are adjusted to the ABR response. The FPP is validated, on one hand, by an analysis of amplitudes and latencies measured manually by an audiologist and automatically by the FPP method in ABR signals recorded at different stimulation rates; and on the other hand, contrasting the performance of the FPP method with the automatic evaluation techniques based on the correlation coefficient, FSP, and cross correlation with a predefined template waveform by comparing the automatic evaluations of the quality of these methods with subjective evaluations provided by five experienced evaluators on a set of ABR signals of different quality. The results of this study suggest (a) that the FPP method can be used to provide an accurate parameterization of the peaks in terms of amplitude, latency, and width, and (b) that the FPP remains as the method that best approaches the averaged subjective quality evaluation, as well as provides the best results in terms of sensitivity and specificity in ABR signals validation. The significance of these findings and the clinical value of the FPP method are highlighted on this paper.

Introduction

The auditory brainstem response (ABR) is the electrical activity of the auditory nerve generated in the brainstem associated with a stimulus [1]. The recording of the ABR has been extensively used in human and animal studies for both clinical and research purposes due to its noninvasive nature. The recording of this signal is commonly used in hospitals and clinics worldwide as a hearing screening tool, to detect the hearing threshold and to detect peripheral and central lesions. Furthermore, the analysis of the ABR may help understand the underlying mechanisms of the process of hearing [2], [3], [4], [5], [6], [7], [8]. The ABR comprises a number of waves that occur during the first 10 ms from stimulus onset [9]. These waves are indicated by sequential Roman numerals as originally proposed by Jewett and Williston [10]. Although up to seven peaks can be identified in the ABR, the most robust are III and V.

The quality of the responses is related to the probability that a response is present, which is usually associated with the amount of noise of the recording [11], [12]. The use of automatic methods for quality assessment and response detection of ABR signals may help improve the process of automatically stopping averaging, avoiding the recording of unnecessary sweeps when there already exists an ABR of sufficient quality and consequently, making a more efficient use of the recording time [13], [14], [15]. Furthermore, the automated identification of the peaks, i.e., amplitudes and latencies, is also a useful tool to provide an automatic interpretation of the ABR [16]. Additionally, automated methods eliminate the need for subjective interpretations of ABR, reduce human biases, and improve uniformity among test conditions, patients, and screening assistants [17], [18], [19], [20], [21], [22]. These advantages promote the use of automated response detection in audiology screening in order to help the operator interpretation and decision making [23].

A number of methods have been proposed in automatic evaluation of ABR [11]. Some of them include the Raleigh test, Watson's U2 test, Kuiper's test, Hodges–Ajne's test, Cochran's Q-test, and Friedman test [24], [25]; automatic computer-assisted recognition of the pattern for ABR latency/intensity functions [26]; MASTER, a Windows-based data acquisition system designed to assess human hearing by recording auditory steady-state responses [27]; zero crossing method [28]; adaptive signal enhancement [29]; multifilters and attributed automaton [30]; single-trial covariance analysis [31]; and automatic analysis methods for peak identification based on a database of ABR signals from a large (>80) number of normal hearing subjects [32], [33]. Despite the large number of automatic evaluation techniques, few of them have been implemented in commercial devices [34]. The most common reported strategies of automated ABR analysis are the correlation coefficient and the F distribution based estimation of the signal to noise ratio (SNR) using a single point of the response (FSP). The correlation coefficient procedure relies on the reproducibility of two consecutive ABR signals obtained in similar conditions to determine the presence or absence of the ABR [35]. FSP provides an estimation of the response SNR evaluated from the distribution of amplitudes of a single point of the response for different sweeps. The power of noise is evaluated by matching the single point distribution of amplitudes with an F distribution, while the power of the signal is estimated from the averaged response [36].

This article describes a new method that performs an automatic evaluation of the quality of ABR signals and identification of the peaks based on the use of templates. We have called this method fitted parametric peaks (FPP). The FPP method can be useful (a) to automatically parameterize the most relevant waves of ABR signals in terms of amplitude, latency, and width, and (b) to provide an automatic estimation of the quality of ABR signals based on the individual assessment of the quality of each wave. Preliminary results of this work were presented in [37].

The rest of the paper is organized as follows. Section 2 describes in detail the fitted parametric peaks (FPP) method. In Section 3, the performance of the described method is assessed by two experiments. Experiment 1 compares the automatic parameterization of the peaks provided by the FPP method with a manual procedure performed by an audiologist in a number of ABR signals obtained at different stimulation rates. Experiment 2 compares the automatic quality assessment of the FPP method with the automatic quality evaluation techniques based on the correlation coefficient, FSP, and cross correlation with a predefined template in terms of the grade of similarity to a subjective evaluation provided by a number of experts on ABR signals of different quality. Additionally, this experiment includes a comparative study of response validation in terms of sensitivity and specificity. Section 4 presents a summary and a discussion of the results. Finally, Section 5 highlights the significance and the main contributions of this article.

Section snippets

Description of the method

The most usual approach for assessing the quality of ABR signals is based in subjective evaluations provided by audiologists. However, it is well known that subjective evaluations may differ from one evaluator to another [33], [38], [39]. This bias represents a problem that could be solved using automatic quality evaluation techniques [17], [18], [19], [20], [21], [22], [23]. This section describes the fitted parametric peaks (FPP) method, a new technique that provides an automatic evaluation

Assessment of the method

The FPP method is validated in this study with two experiments. Experiment 1 evaluates the performance of the FPP method through a comparison of the latencies and amplitudes of waves III and V measured manually by an audiologist and automatically by the FPP method in a number of ABR signals obtained at different stimulation rates. In experiment 2, the performance of the automatic quality evaluation techniques based on the FPP, correlation coefficient (r), FSP, and cross correlation with a

Discussion

This paper describes in detail and evaluates the fitted parametric peaks (FPP) method, a new approach of automatic quality assessment and peak parameterization based on the use of templates. The use of templates for this purpose was first proposed by C. Elberling in [40]. In his work, a cross correlation method between the ABR signal used for test and a template waveform is described. This method has the limitation of requiring a database of predefined templates for each recording condition,

Conclusion and significance

A novel automatic method for quality assessment and peak identification of ABR signals, the fitted parametric peaks (FPP), is described and evaluated in this article. The approach of FPP opens a new paradigm in template-matching algorithms, avoiding the need of a database of templates and including additional information regarding the most relevant components of ABR signals. The computational efficiency of the FPP method could be appropriate for its implementation in real time processing

Conflict of interest

The authors declare no conflicts of interest related to this research work.

Acknowledgments

The authors gratefully acknowledge the participation of the voluntary subjects and evaluators involved in this study. This research is granted by the project “Design, implementation and evaluation of an advanced system for recording Auditory Brainstem Response (ABR) based in encoded signalling” (TEC2009-14245), R&D National Plan (2008-2011), Ministry of Economy and Competivity (Government of Spain) and “European Regional Development fund Programme” (2007-2013); by the Granada Excellence Network

References (61)

  • P.G. Gillespie et al.

    Mechanotransduction by hair cells: models, molecules, and mechanisms

    Cell

    (2009)
  • R.E. Lasky

    A developmental study on the effect of stimulus rate on the auditory evoked brain-stem response

    Electroencephalography and Clinical Neurophysiology

    (1984)
  • G. Lina-Granade et al.

    Maturation and effect of stimulus rate on brainstem auditory evoked potentials

    Brain and Development

    (1993)
  • K.J. Doyle et al.

    Comparison of newborn hearing screening by transient otoacoustic emissions and auditory brainstem response using ALGO-2

    International Journal of Pediatric Otorhinolaryngology

    (1998)
  • K.J. Doyle et al.

    Newborn hearing screening by otoacoustic emissions and automated auditory brainstem response

    International Journal of Pediatric Otorhinolaryngology

    (1997)
  • V. Bajaj et al.

    Automatic classification of sleep stages based on the time-frequency image of EEG signals

    Computer Methods and Programs in Biomedicine

    (2013)
  • D.L. Jewett et al.

    The use of QSD (q-sequence deconvolution) to recover superposed, transient evoked-responses

    Clinical Neurophysiology

    (2004)
  • J.W. Hall

    New Handbook of Auditory Evoked Responses

    (2007)
  • A.R.D. Thornton et al.

    Auditory brainstem responses recorded at fast stimulation rates using maximum length sequences

    British Journal of Audiology

    (1993)
  • H. Zabala-Fernandez et al.

    Identification enhancement of auditory evoked potentials in EEG by epoch concatenation and temporal decorrelation

    Computer Methods and Programs in Biomedicine

    (2012)
  • A. Erenberg et al.

    Newborn and infant hearing loss: detection and intervention

    Pediatrics

    (1999)
  • Z.D. Jiang et al.

    Maximum length sequence brainstem auditory evoked responses in term neonates who have perinatal hypoxia ischemia

    Pediatric Research

    (2000)
  • R.F. Burkard et al.

    The Auditory Brainstem Response, in Auditory Evoked Potentials. Basic Principles and Clinical Application

    (2007)
  • D.L. Jewett et al.

    Auditory-evoked far fields averaged from the scalp of humans

    Brain

    (1971)
  • C. Elberling et al.

    Detecting and Assessing Synchronous Neural Activity in the Temporal Domain (SNR, Response Detection), in Auditory Evoked Potentials. Basic Principles and Clinical Application

    (2007)
  • M. Don et al.

    Use of quantitative measures of auditory brain-stem response peak amplitude and residual background noise in the decision to stop averaging

    Journal of the Acoustical Society of America

    (1996)
  • O. Ozdamar et al.

    Measurement of signal and noise characteristics in ongoing auditory brainstem response averaging

    Annals of Biomedical Engineering

    (1996)
  • R.E. Delgado et al.

    Automated auditory brainstem response interpretation

    IEEE Engineering in Medicine and Biology Magazine

    (1994)
  • G.G. Gentiletti-Faenze et al.

    Evaluation of automatic identification algorithms for auditory brainstem response used in universal hearing loss screening

  • R.E. Eilers et al.

    Optimization of automated hearing test algorithms: simulations using an infant response model

    Ear and Hearing

    (1991)
  • Cited by (25)

    • Machine learning approaches used to analyze auditory evoked responses from the human auditory brainstem: A systematic review

      2022, Computer Methods and Programs in Biomedicine
      Citation Excerpt :

      To improve the objectivity of the assessment, several techniques have been introduced in the literature. Automatic peak detection, latency estimation, and template matching algorithms are a few such techniques [18,21,52]. Several traditional statistical methods have also been used [4,5,14,53].

    • Real-time threshold determination of auditory brainstem responses by cross-correlation analysis

      2021, iScience
      Citation Excerpt :

      In order to accurately detect mild hearing threshold elevation in the diagnosis of, e.g., progressive hearing loss (Barreira-Nielsen et al., 2016) and age-related hearing loss (Gates and Mills, 2005; Sergeyenko et al., 2013), unbiased automatic approaches with high precision and reliability are essential, particularly when screening is involved. Over decades, many attempts were made to automate the procedure based on, e.g., (1) waveform similarity by means of comparing either existing templates (Davey et al., 2007; Elberling, 1979; Valderrama et al., 2014) or matching features learned by artificial neural network from human annotation (Acır et al., 2006; Alpsan and Ozdamar, 1991; McKearney and MacKinnon, 2019; Sanchez et al., 1995; Vannier et al., 2002); (2) waveform stability quantified by cross-correlation function between single-sweeps (Bershad and Rockmore, 1974; Weber and Fletcher, 1980), interleaved responses (Berninger et al., 2014; Ozdamar et al., 1994; Xu et al., 1995), or responses at adjacent stimulus levels (Suthakar and Liberman, 2019); (3) the “signal quality” through scoring procedures like F-ratios (Cebulla et al., 2000; Don and Elberling, 1994; Elberling and Don, 1984; Sininger, 1993); (4) neurophysiological parameters from fitting the responses to different stimulus intensities (Nizami, 2002; Schilling et al., 2019). However, owing to heterogeneity in inter-subject waveform and signal-to-noise-ratio (SNR) introduced by variations in test subject conditions, electrode placement/impedance, as well as acquisition settings, the accurate threshold determination is only possible under a narrow range of experimental settings, which hampers direct comparisons of ABR results across laboratories.

    • Controlling test specificity for auditory evoked response detection using a frequency domain bootstrap

      2021, Journal of Neuroscience Methods
      Citation Excerpt :

      Ultimately, the aim for the detection method is to reduce subjectivity, and to provide an overall more reliable, sensitive, and efficient (i.e. shorter in duration) test. Some of the more well-known methods for assisting examiners with response detection are the Fsp (Elberling and Don, 1984), the magnitude squared coherence (Dobie and Wilson, 1989), the spectral F-test (Valdes et al., 1997), the Hotelling’s T2 test (Golding et al., 2009) and correlation coefficients (e.g. Valderrama et al., 2014). These methods are all parametric, i.e. they use theoretical distributions to evaluate test significance, and thus make various assumptions regarding the EEG data.

    • Automated extraction of auditory brainstem response latencies and amplitudes by means of non-linear curve registration

      2020, Computer Methods and Programs in Biomedicine
      Citation Excerpt :

      Two different approaches have been proposed to address this problem. The computationally simpler, but also less powerful approach is the “wave template” approach [26–29], which fits separate templates to individual ABR waves (see Fig. 1C, for a simplified example). Unlike the response templates of Elberling [22], the wave templates are not derived from the measured responses, but implemented as synthetic functions, representing the waves’ idealized shapes.

    • Peak window correlation method for drug screening using Raman spectroscopy

      2019, Journal of Pharmaceutical and Biomedical Analysis
      Citation Excerpt :

      These problem can be solved and method sensitivity enhanced by some data treatment algorithms, e.g., moving window correlation [3,5], peak search, various types of multivariate calibration, etc. Peak detection and parametrization are most important stages in the signal pre-processing and identification in many methods of instrumental analysis [6–8] used in analytical chemistry, physics, medicine and many other fields [9,10]. Gaussian function combined with other functional dependences is often applied for the peak parametrization [11].

    View all citing articles on Scopus
    View full text