Abstract
Electrowetting (EW)-based techniques have been widely used in manipulating discrete liquid. However, few articles discussed the controlling of continuous fluid flow by using EW-based techniques. In this paper, an EW-based valve combined with plasma-modified Teflon surface, which serves as a microfluidic guidance, in controlling continuous fluid flow has been demonstrated. The plasma-modified Teflon surface is firstly demonstrated for confining continuous fluid flow. The EW-based microfluidic device possesses the functions of a valve and a microchannel without complex moving parts and grooved microchannels. The quantitative characteristics of the EW-based valve are also studied. Propylene carbonate (PC) is firstly demonstrated as the working liquid in the EW-based device because of its applications in parallel oligonucleotide synthesis. It is found that lower valve actuation voltage reduces the deterioration of the valve and improves the valve stability.
Similar content being viewed by others
References
C.C. Cho, R.M. Wallace, and L.A. Files-Sesler, J. Electron. Mater. 23, 827 (1994).
S.K. Cho, H. Moon, and C.J. Kim, Journal of Microelectromechanical Systems 12, 70 (2003).
D.T. Clark and A.J. Dilks, Polym. Chem. Ed. 17, 957 (1979).
T.R. Hughes, M. Mao, A.R. Jones, J. Burchard, M.J. Marton, K.W. Shannon, S.M. Lefkowitz, M. Ziman, J.M. Schelter, M.R. Meyer, S. Kobayashi, C. Davis, H. Dai, Y.D. He, S.B. Stephaniants, G. Cavet, W.L. Walker, A. West, E. Coffey, D.D. Shoemaker, R. Stoughton, A.P. Blanchard, S.H. Friend, and P.S. Linsley, Nat. Biotechnol. 19, 342 (2001).
P. Herrera-Fierro, W.R. Jones, Jr., and S.V. Pepper, J. Vac. Sci. Technol. A 11, 354 (1993).
D. Huh, A.H. Tkaczyk, J.H. Bahng, Y. Chang, H.H. Wei, J.B. Grotberg, C.J. Kim, K. Kurabayashi, and S. Takayama, J. Am. Chem. Soc. 125, 14678 (2003).
J. Janting, J. Branebjerg, and P. Rombach, in Proceedings of the 14th European Conference on Solid-State Transducers, edited by R.D. Reus et al. (MIC-Mikroelektronik Centret, Denmark, 2000), p. 275.
S. Kaplan and A.J. Dilk, Appl. Polym. Sci., Appl. Polym. Symp. 38, 105 (1984).
D.E. Kataoka and S.M. Troian, Nature 402, 794 (1999).
P. Lam, K.J. Wynne, and G.E. Wnek, Langmuir 18, 948 (2002).
J. Lee, H. Moon, J. Fowler, T. Schoellhammer, and C.J. Kim, Sens. Actuators. A Phys. 95, 259 (2002).
E. LeProust, J.P. Pellois, P. Yu, H. Zhang, and X. Gao, J. Comb. Chem. 2, 349 (2000).
Y. Li, T. Pfohl, J.H. Kim, M. Yasa, Z. Wen, M.W. Kim, and C.R. Safinya, Biomedical Microdevices 3, 239 (2001).
S.A. Makohliso, L. Giovangrandi, D. Léonard, H.J. Mathieu, M. Ilegems, and P. Aebischer, Biosens. Bioelectron. 13, 1227 (1998).
E. Occhiello, M. Morra, and F. Garbassi, Angew. Markromol. Chem. 173, 183 (1989).
P. Paik, V.K. Pamula, and R.B. Fair, Lab Chip 3, 253 (2003).
R.R. Rye, J. Polym. Sci. Part B: Polym. Phys. 26, 2133 (1988).
L.M. Siperko and R.R. Thomas, J. Adhes. Sci. Technol. 3, 157 (1989).
K.L. Tan, L.L. Woon, H.K. Wong, E.T. Kang, and K.G. Neoh, Macro-molecules 26, 2832 (1993).
O.D. Velev, B.G. Prevo, and K.H. Bhatt, Nature 426, 515 (2003).
R. Winter, D. Korzec, N. Sprang, D. Theirich, and J. Engemann, Surf. Coat. Technol. 74/75, 618 (1995).
B. Zhao, J.S. Moore, and D.J. Beebe, Langmuir 19, 1873 (2003).
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Cheng, JY., Hsiung, LC. Electrowetting (EW)-Based Valve Combined with Hydrophilic Teflon Microfluidic Guidance in Controlling Continuous Fluid Flow. Biomedical Microdevices 6, 341–347 (2004). https://doi.org/10.1023/B:BMMD.0000048565.59159.c1
Issue Date:
DOI: https://doi.org/10.1023/B:BMMD.0000048565.59159.c1