Introduction

Quantum fluctuations in an electromagnetic field give rise to forces between conductors at the same potential when their separation is near 100 nm. Known as the Casimir force, this phenomenon was first predicted by H. B. G. Casimir1 and was later expanded to arbitrary materials2,3. Since then, the Casimir force has been measured experimentally many times4,5,6,7,8,9,10,11 and has been proposed as a practical metrology platform using micro- and nanoelectromechanical systems (MEMS/NEMS)4,5,6,12. The Casimir force is attractive for metrology applications due to its extreme sensitivity to the separation between two objects and the ability to be measured in ambient conditions5. One of the most critical challenges in the development of practical platforms is resilience to Casimir pull-in, which results in stiction in MEMS/NEMS devices and typically irreversible damage13,14. Notably, Casimir pull-in is only an obstacle in devices that employ an attractive force4,5,6,8,9,10,11,13, but a specific combination of materials can generate a repulsive force as well7,15. The repulsive force configuration, however, has only been observed in liquid, which is not ideal for many common MEMS/NEMS applications such as resonant sensing. As such, many have proposed avoiding pull-in in attractive force platforms by optimizing the material dielectric properties16, the radius of curvature in a sphere-plate configuration17, and the roughness of interacting surfaces18,19. A device capable of leveraging the attractive Casimir force while resilient to pull-in would realize great utility for quantum metrology.

Quantum metrology has longstanding application in the measurement of very weak magnetic fields. A superconducting quantum interference device (SQUID) measures changes in the magnetic field associated with a flux quantum20. An atomic magnetometer (AM) measures a quantum effect involving the magnetic spin states of atoms in a vapor cell21. The list of applications for such high-resolution magnetic sensors is vast, spanning the biomagnetic detection of cardiac contractions22,23, electromagnetic brainwaves24, and solid cancerous tumors25 to astronomical observations such as Jupiter’s magnetosphere26. Sensitive magnetometry is often challenged by interference from ambient geomagnetic fields and nearby electromagnetic sources. For real-time measurements, the most common methods to reduce the effects of interference are magnetic shielding and gradiometry. Shielding is both expensive and cumbersome, so there is a great effort to conduct unshielded measurements23,24. Today, gradiometry is only applied via differential measurement, measuring the subtraction from two closely spaced sensors. Such a technique increases immunity to interference by reducing sensitivity to temporal fluctuations in magnetic intensity. This is especially effective for measurement near a magnetic source (such as the heart23,24), where magnetic gradients are high, and for use with magnetic sensors which are robust and reliable. In our recent work, we demonstrate magnetic gradiometry without subtraction at single point27, and at a resolution in the range of magnetocardiography (MCG).

Existing gradiometer technology is reviewed in Fig. 1, which is repurposed from ref. 27 for this work. The fields are illustrated in gradients, assuming the sensors of each technology can be separated by 1 cm. All technologies involve subtraction between two sensors to calculate the gradient, except the MEMS single-point gradiometer27. The first-order gradients of the geomagnetic field (E, 500 pT/cm) and geomagnetic noise (GMN, 500 fT/cm) are illustrated by dash-dotted lines. SQUID and AM technology clearly lead the group in resolution. Myriad other techniques are used to sense magnetic fields, where the bottom band (in purple) represents MEMS designs. MEMS magnetic sensors are typically resonant sensors with high-quality factors (Q greater than 1000), where impressive sensitivity has been achieved at higher frequencies in the kHz range27,28,29. To detect low-frequency biomagnetic fields, nontrivial techniques are necessary to employ resonant sensors.

Fig. 1: Magnetic gradiometer technology.
figure 1

All magnetic fields are illustrated as gradients. Existing sensor technology is adjusted by taking the best resolution in literature and assuming two sensors can be separated by 1 cm, which can measure a gradient field. E is Earth’s field gradient and GMN is the gradient geomagnetic noise. The devices shaded in purple at the bottom are microelectromechanical systems (MEMS) technology. Our previous MEMS single-point gradiometer design27 experimentally achieved a resolution of 100 pT/cm, with a calculated best-case resolution of 30 fT/cm at resonance. This work (bottom) proposes a quantum-derived enhancement to this design, which suggests a 10,000-fold improvement in the best-case resolution with a measurement near DC. This figure is repurposed from ref. 27.

The sensitivity of resonant MEMS devices can be tuned by many techniques such as parametric amplification12,30,31 and modal coupling32,33. Conceptually, parametric amplification is typically achieved by modulating a parameter of the equation of motion at two times the frequency of resonance and controlling the phase relation between the driven mode and the modulated parameter. When applied to the drive system of a MEMS cantilever30, thermomechanical noise was reduced in one phase by 4.9 dB. A Lorentz force MEMS magnetometer was parametrically driven to enhance sensitivity at resonance by over 80-fold31. The gain in such electrostatic systems is typically 10−1000. Parametric modulation can also be applied to coupled resonators, where one object, or mode, oscillates at twice the frequency of another. Parametric pumping was shown to dynamically tune the coupling of two modes in a gyroscopic ring resonator, with application in inertial sensing32. In the design of a Casimir-coupled resonator12, a gold sphere was used to parametrically pump the oscillation of a torsional oscillator, proposed to amplify a DC voltage measurement up to ten orders of magnitude. Using an attractive Casimir force design, the challenge of avoiding Casimir pull-in was also discussed in depth. The tunability of such systems is often critical for experimental utility and to account for small fabrication asymmetries, which have a significant impact on the coupling12,32.

In this work, we propose a coupling between two resonators by a quantum-derived Casimir force. Fundamentally, we combine two metrology platforms: a MEMS Casimir platform and a MEMS single-point gradiometer platform. The Casimir platform5 comprises a MEMS accelerometer functionalized by a gold sphere, where the voltage potential can be controlled. A Casimir force measurement is achieved when a gold plate is brought within 100 nm of the sphere. The MEMS gradiometer platform27 comprises a MEMS accelerometer functionalized by a cube micromagnet. Measurement is achieved by detecting an oscillating force on the permanent magnet at resonance, where the force is proportional to a gradient magnetic field. Both resonators can also be electrostatically driven in an analog environment5,27,34. In essence, these two systems, on the same central axis and within 100 nm of separation, form a Casimir-coupled resonator. Similar to a Casimir oscillator using a torsional plate12, this design is highly sensitive to small changes in DC separation due to Casimir coupling. Therefore, DC forces from gradient magnetic fields are intensely amplified by the coupling. Then, instead of using parametric amplification to further amplify the sensitivity, we design a technique to achieve a steady-state Casimir oscillator resilient to pull-in. In this work, we propose and simulate a highly tunable Casimir-driven gradiometer that is sensitive to slowly varying magnetic fields and resilient to pull-in.

MEMS Casimir gradiometer design

The design in this work is a nontrivial combination of an experimental MEMS gradiometer platform27, an experimental MEMS Casimir force metrology platform5, and a coupling method12 that parametrically modulates the interaction between the two platforms. To approach the divergent gain of this physical design, we anticipate that a tunable experimental platform with precise positional control of the two microobjects (magnet and sphere) is required. When the separation is reduced to near 100 nm, the Casimir force contribution becomes significant, as has been shown previously4,5. If the separation decreases much past this point, the system is predicted to experience Casimir pull-in, an event that is caused by the attractive Casimir force overcoming the restoring force of the spring and causing the device to malfunction.

The proposed sensing platform is illustrated in Fig. 2a. Each device is fabricated individually as described previously5,27 using commercially available MEMS accelerometers. This is accomplished using precise control of a vacuum pick-and-place system and feedback from a live sensor (postrelease MEMS). The micromagnet used in the gradiometer design is functionalized with a gold plate on its side that faces the microsphere. The sensor with the microsphere is inverted such that the sphere has a clear pathway to come into close proximity to the plate. Both platforms enable control of static position, oscillation amplitude, frequency, phase, and detection. The drive parameters are controlled via a built-in electrostatic self-test feature on the ADXL 203 platform, which can be used for analog control of the microobjects via pulsed width modulation34. This feature is also used to calibrate and negate the effect of an anticipated electrostatic coupling between the plate and sphere, as has been shown previously5.

Fig. 2: Proposed MEMS Casimir gradiometer design.
figure 2

The free body diagram a shows a sphere positioned in close proximity to a magnet, where the resonant modes are coupled by the quantum-derived Casimir force. Both objects are driven at resonance, where the micromagnet is driven at two times the frequency of the microsphere, to accomplish parametric modulation. The force from a weak gradient magnetic field induces a small deflection of the magnet, decreasing the separation xSM and inducing a frequency and amplitude shift in the sphere oscillation via the Casimir coupling. The proposed experimental design (b) and the accompanying parameters are chosen based on existing experimental platforms5,27. To successfully couple these two systems, a highly tunable design is necessary. By assembling the sphere and magnet on independent MEMS force transducer platforms, their independent actuation is completely controlled. By inverting the sphere platform, a separation between the sphere and plate of near 100 nm can be achieved.

The functional diagram of this design, shown in Fig. 2b, is based on the interaction of two resonators. The first resonator includes an attached gold microsphere, and the second includes the micromagnet/gold mirror assembly. Independently, both resonators act as damped, driven harmonic oscillators. The coupling of the two resonators is based on the Casimir force, which is dependent on their separation. Assuming that out-of-plane forces are minimal, we analyze a uniaxial system of equations (Eqs. 13) along the central axis of the magnet.

$$m_S\ddot x_S + \frac{{m_S\omega _{0S}}}{{Q_S}}\dot x_S + k_{0S}x_S = F_{{\mathrm{Dr}} - S}(t) + F_{{\mathrm{Cas}}}(x_{SM},t)$$
(1)
$$m_M\ddot x_M + \frac{{m_M\omega _{0M}}}{{Q_M}}\dot x_M + k_{0M}x_M = F_{{\mathrm{Dr}} - M}(t) + F_{{\mathrm{Cas}}}\left( {x_{SM},t} \right) + F_M(t)$$
(2)
$$x_{SM}\left( t \right) = s_0 + x_S(t) - x_M(t)$$
(3)

Here, Eq. 1 relates to the sphere, Eq. 2 to the magnet, and Eq. 3 to their separation. In Eq. 1, xS is the displacement of the sphere, mS is the mass, ω0S is the natural frequency, QS is the quality factor, and k0S is the unperturbed spring constant. The sphere is driven electrostatically by FDr-S and experiences an attractive coupling force FCas when the separation is small (on the order of 100 nm). Similarly, in Eq. 2, xM is the displacement of the magnet, mM is the mass, ω0M is the natural frequency, QM is the quality factor, and k0M is the spring constant. The magnet is also driven electrostatically by a force FDr-M and experiences an equal and opposite coupling force to the sphere FCas at small separations. In addition, a slow, time-varying gradient magnetic field imposes a force FM. In Eq. 3, xSM is the real-time separation distance between the sphere and magnet, and s0 is the separation in the absence of the Casimir coupling force. The forcing terms are expanded in Eqs. 47 below.

$$F_{{\mathrm{Dr}} - S}\left( t \right) = k_{0S}A_S\sin \left( {\omega _{0C}(t + \tau _{1f})} \right)$$
(4)
$$F_{{\mathrm{Dr}} - M}\left( t \right) = k_{0S}A_M\sin \left( {2\omega _{0C}(t + \tau _{2f})} \right)$$
(5)
$$F_M = M\left( {\frac{{{\mathrm{d}}B}}{{{\mathrm{d}}x}}} \right)\cos \left( \theta \right)$$
(6)
$$F_{{\mathrm{Cas}}} = - \frac{{\pi ^3\hbar c}}{{360}}\frac{R}{{x_{SM}^3}}$$
(7)

In Eq. 4, AS is the forcing amplitude of the sphere, ω0C is the natural frequency of the coupled system, and τ1f defines the fixed starting time. Similarly, in Eq. 5, AM is the forcing amplitude of the magnet, and τ2f is the time delay of magnet actuation. Most works12,30,31 involving parametric pumping use a phase delay, and we describe our rationale for a time delay system later, in conjunction with Fig. 3. The magnet is forced at 2ω0C for parametric amplification. Equation 6 describes the force on the magnet from a gradient magnetic field predominantly along its polarized axis, where M is the moment, B is the magnetic field intensity, and θ is the angle between them27. The Casimir force for a sphere-plate geometry is displayed in Eq. 7, where is Plank’s constant, c is the speed of light, R is the radius of the sphere, and xSM is the separation between the sphere and plate as defined above. It should be noted that Eq. 7 is ideal (assuming absolute zero temperature and perfectly smooth infinitely conducting surfaces) and should be modified to reflect real experimental conditions4,5,6,7,8,9,10,11 such as temperature, roughness, and dielectric properties. The nominal values for simulation inputs are shown in Table 1, where the primary tuning parameters are indicated. The magnet’s mass (typically 150 µg; see ref. 27) is not a direct input to the simulation, as the magnet’s dynamics are controlled by feedback (discussed later in Fig. 3).

Fig. 3: Quasistatic illustration of Casimir coupling.
figure 3

Inset: diagram of a simple, nonlinear Casimir oscillator with a a fixed magnet and b an oscillating magnet adhered to a conducting plate. a Potential energy of the sphere spring without the Casimir force (blue) and related to the Casimir force (red). The overall potential energy (black) is shown as a function of sphere position xS. The separation s0 denotes the initial separation when the spring is at equilibrium in absence of the Casimir force. Casimir oscillators with a fixed plate have been discussed previously4,12. b Overall potential energy of sphere spring with a dynamic cavity size s, modulated by movement of the magnet (decreases 1 nm per curve from red to blue). In this work, the cavity is highly dynamic as both the sphere and magnet oscillate, and so we use xSM in the text to represent the real-time distance between the sphere and plate.

Table 1 Parameter Inputs for Simulation of Casimir Gradiometer Design.

When the two resonators are coupled as described in Eqs. 13, there is a spring softening effect analogous to the electrostatic spring softening observed in capacitive systems30. The coupling may then be modulated by tuning the parameters of the magnet resonator. This is necessary to access the most sensitive region of parameter space while preventing Casimir pull-in. Following the analytical model outlined earlier12, the equation of motion for the sphere (Eq. 1) in a Casimir coupled system then becomes Eq. 8 below, where the parametric spring constant kp is defined in Eqs. 9 and 10. For simplicity, we maintain the assumption of a linear spring model because the amplitudes of oscillation are small (<100 nm).

$$m_S\ddot x_S + \frac{{m_S\omega _{0C}}}{{Q_C}}\dot x_S + \left[ {k_{0S} + k_p\left( {x_{SM},t} \right)} \right]x_S = F_{{\mathrm{Dr}} - S}\left( t \right) + F_{{\mathrm{Cas}}}\left( {x_{SM},t} \right)$$
(8)
$$k_p\left( {s_0,t} \right) = \frac{{{\mathrm{d}}F_{{\mathrm{Cas}}}\left( {x_{SM},t} \right)}}{{{\mathrm{d}}x_{SM}}}$$
(9)
$$k_p\left( {s_0} \right) = \frac{{\pi ^3\hbar c}}{{120}}\frac{R}{{x_{SM}^4}}$$
(10)

The effect of the modulated spring constant may be best illustrated by the potential energies of a quasi-static system (Fig. 4). The magnet is fixed (Fig. 4a), and the sphere, connected to a spring, is moved to set the gap between it and the conducting plate. The gap is the Casimir cavity size s0, which is defined by the equilibrium position of the spring in the absence of the Casimir force. This has been described earlier4 for a Casimir oscillator where s0 is constant. In this work, the cavity s is influenced by the movement of both the magnet and spring. The overall potential energy curve is a summation of the elastic potential energy of the spring and the potential energy of the Casimir force as a function of the sphere displacement xS from the equilibrium position. At greater displacements, the Casimir attraction overcomes the restoring force, and pull-in occurs as the overall potential decreases rapidly. The overall potential energy curve is shown for varying Casimir cavity sizes (Fig. 4b), where each curve (from red to blue) represents the cavity decreasing by 1 nm (this is quasistatic, so the cavity size is represented by s). Now that the quasistatic system is understood, it is straightforward that the potential energy curve changes dynamically as the sphere and magnet both move in real-time (a dynamic cavity is represented by xSM). Thus, with the appropriate tuning of the magnet oscillation, one can modulate the sphere oscillation via Casimir coupling. Furthermore, one can make a trade-off to tune the stability and sensitivity of the system as a larger amplitude pushes the sphere close to pull-in, which makes the system more unstable but more sensitive to small perturbations on the magnet.

Fig. 4: Simulation schematic.
figure 4

The simulation is designed based on instruments readily available in a standard electronics laboratory. Following Eqs. 17 in the text, a feedback scheme is employed to tunably synchronize the f and 2f drive waveforms. The detailed schematic can be found in Fig. S1.

The system described is a complex combination of several physical phenomena, leading to a highly nonlinear system and a large parameter space. Therefore, simulations using MATLAB’s Simulink and Simscape are chosen to characterize the system. A simplified block diagram is illustrated (Fig. 3), and a detailed block diagram is provided in the Supplementary Material (Fig. S1). In essence, the nominal parameters in Table 1 are input to the system, and the outputs are tracked in real-time. The sphere amplitude xS is reported to gather a clear sense of the operation of the device, and the change in the coupled system resonant frequency ω0C is the ultimately proposed detection method. These values are gathered at some point in the time response after the initial transients have settled (typically 0.35 s). We designed the simulation only with tools we typically use in the laboratory to facilitate the translation to experimental measurement. A feedback approach is employed, where the sphere’s frequency is detected by a zero-crossing (negative slope). A waveform at double the frequency involving the Casimir coupling and a gradient magnetic field is fed back into the actuation of the sphere.

At the start of the simulation (t = 0), we assume that the sphere is resonating at the unloaded resonance frequency ω0S, and the magnet is resonating at 2ω0S. The objects are proposed to incrementally approach each other from a large separation distance (>1 μm, where the Casimir force is minimal) to reach the prescribed separation. This approach is based on the experimental observation5 of the Casimir force using a similar platform. As Casimir coupling begins to interact, our feedback system adjusts the actuation of the sphere (Fig. 3). The resonant mode of the system is pumped such that there is little amplitude decay due to damping, a technique35 that is well characterized in both simulations and experiments. Previous analytical12 and experimental30,31 works have controlled the phase of objects in a parametrically amplified system, but this is challenging to do in dynamic simulation experiments, such as with Simulink.

This is ultimately why we choose a time delay approach, where the translation to experiment is straightforward with a precision digital delay generator (such as the DG645, SRS). It is worth noting, however, that a constant time delay will result in a changing phase delay for a system with changing oscillation frequency (as is our case). Therefore, this design is notably different from a phase delay parametric pumping system.

A gradient magnetic field is introduced to the system in Fig. 3 by a force FM on the magnet. This results in a linear deflection of the magnet by Eq. 6, changing xM, and dynamically altering the separation xSM. This force is imposed statically in Figs. 5 and 7 and is imposed dynamically in Fig. 6. The sphere amplitude is reported to intuitively convey the behavior of the device, but the resonant frequency of the coupled system is measured to infer the measurand, a gradient magnetic field. As such, we define a sensitivity Sfreq (Hz/pT/cm), shown in Eq. 11, as the ratio between the natural frequency of the coupled system f0C and the change in gradient magnetic field Bx. In Fig. 7, this is computed by taking the slope between two consecutive data points.

$$S_{{\mathrm{freq}}} = \frac{{{\Delta}f_{0C}}}{{{\Delta}(\nabla B_x)}}$$
(11)
Fig. 5: Characterized system response.
figure 5

a Bode plots of uncoupled (xSM > 1 μm) and coupled configuration (xSM ~100 nm). The quality factor is near 1000 in both cases and the resonant frequencies are 1000 and 850 Hz, respectively. The decrease in frequency can be explained by the asymmetric interaction with the nearby plate, which imposes the attractive Casimir force. b, c Are in the coupled configuration, with nominal parameters from Table 1, except τ2f. b A phase delay of the parametrically amplified magnet system can yield a highly unstable and sensitive response (pull-in), a stable and controlled response, and a deamplified response. c The sphere amplitude for a finely swept τ2f (at 0.35 s of the time response after transients have settled). The maximum is near a delay of zero, and the profile is strongly nonlinear.

Fig. 6: Response to slowly varying gradient magnetic field.
figure 6

The controlled configuration at τ2f = 750 μs is tracked for 2 s following an imposed initial condition. a Shows the sphere amplitude response to no magnetic field input, where the two resonators are set to interact and reach a steady-state after approximately 200 ms. b Shows the sphere amplitude response when the magnet is modulated by a 1 Hz, 4 pT/cm peak-to-peak gradient magnetic field. c Tracks the shift in the coupled resonance frequency f0C in both conditions (a) and (b). A 1 Hz oscillation is observed in both (b) and (c), where tracking the frequency shift of the high-quality peak reports a clean signal. For small deflections of the magnet (1 nm peak-to-peak here), the response is nearly linear, but larger deflections will result in an asymmetric sine wave.

Fig. 7: Tunable sensitivity to separation.
figure 7

The separation xSM is varied in a 3 nm range, resulting in an inverse exponential profile in a frequency shift of the coupled resonance f0C. The sensitivity Sfreq is calculated using a gradient magnetic field that would cause an equivalent deflection in xSM. The maximum sensitivity reported is 6 Hz/(pT/cm). Standard frequency detection with 10 ppm and a 1 s gate time yields a best-case resolution of 1.6 aT/cm. While we anticipate limitations due to thermomechanical noise, this exceeds a 10,000-fold improvement on the previous design27 of the MEMS single-point gradiometer. While a zero delay configuration may provide maximum amplification, we report a highly sensitive design using a τ2f = 750 μs delay, which is resilient to Casimir pull-in.

Parametric pumping is expected to enhance the coupling of the Casimir force. Analytically, sphere-plate Casimir coupling showed a maximum gain proportional to the inverse of the 5th power of the separation when a phase delay of 0° (or time delay of 0 s) was used and the sphere amplitude was detected12. To our knowledge, there is no experimental observation of parametrically amplified Casimir coupling. Due to the anticipated danger of pull-in at maximum amplification, we identify a region of design space resilient to pull-in and with parametrically amplified Casimir coupling. We leverage a shift in the coupled resonant frequency to propose a frequency detection scheme.

Results: Magnetic sensing with the MEMS Casimir gradiometer

The system is characterized by simulation (Fig. 5), where the results from the dynamics of the sphere are analyzed. The uncoupled and Casimir coupled systems are illustrated using Bode plots (Fig. 5a). As shown in Table 1, the resonant frequency of the unloaded system is designed to be 1 kHz with a quality factor of 1000. Casimir coupling causes a pronounced decrease in the resonant frequency (tens to hundreds of Hz). The high-quality factor of the system facilitates frequency shift detection. Recall that the uncoupled case occurs when the separation between the sphere and magnet is larger than 1 µm. At such separations, the sphere responds to a harmonic drive with symmetric oscillations shepherded by the spring’s linear restoring force. The coupled case occurs when the separation between the magnet and sphere is in the range of 100 nm. In this regime, the spring’s restoring force deviates from parabolic, becoming asymmetric (Fig. 4) and leading to nonlinear dynamics.

In the coupled configuration, the parameters are tuned to characterize the system and investigate useful areas of design space. The first parameter of interest is the time delay of the magnet’s oscillation τ2f. In Fig. 5b, we show the temporal response to various time delays, labeled pull-in (τ2f = 0 µs), Controlled (τ2f = 750 µs), and Deamplified (τ2f = 150 µs). This characterizes three responses for different values of τ2f, and the responses to finely swept values are characterized (Fig. 5c) by the AC amplitude of the sphere after transients have settled (t = 0.35 s). We find a maximum amplification at τ2f = 0 µs (corresponding to a phase delay of 0°), which is consistent with the previous analytical analysis12 of a phase delay system. While the τ2f = 0 µs delay is the most sensitive region of the design space, we find that the two resonators in the Casimir coupled system always result in pull-in if left to interact for an arbitrarily long time. This indicates malfunction and destruction of the sensor. Consequently, we avoid this region of design space and investigate nonzero time delays for a sensitive controlled condition. For nonzero time delays, however, this system is different than a phase delay system and may not be directly compared. We find a maximum deamplification at τ2f = 150 µs followed by two other maxima (Fig. 5c). After the peak at τ2f = 1200 µs, the pattern repeats with slightly decreased amplitudes due to energy lost per cycle (not shown). At τ2f = 750 µs, we report a stable, controlled oscillation of the sphere (Fig. 5b), which is expanded in Figs. 6 and 7.

The time response of the controlled system (τ2f = 750 µs) is dynamically characterized over an elapsed time of 2 s for zero gradient magnetic field input (Fig. 6a) and for a slowly varying sine wave oscillation of a gradient magnetic field (Fig. 6b). The resonant frequency of the coupled system in these two conditions, f0C, is tracked in both configurations (Fig. 6c). The coupled system with no input field reaches a steady state after approximately 200 ms, after which point the signal is stable. A dynamic gradient field input of a 1 Hz sine wave, 4 pT/cm peak-to-peak, is chosen to demonstrate the response of the system to a slowly varying magnetic field. For a gradiometer27 sensitivity of 1 μV/(fT/cm) and a magnet spring constant of 25 mN/m (40 times softer than the experimental platform in ref. 27), this gradient field yields a 1 nm peak-to-peak oscillation of the magnet. The system responds with a 1 Hz oscillation of an approximately 4 Hz peak-to-peak shift in f0C. For very small gradients, the response is nearly linear, and for larger gradient fields, the response is a nonlinear, asymmetric sine wave. However, the change in the coupled resonant frequency directly maps to a change in the gradient, so the true gradient signal can be easily calculated by ratiometric conversion.

The sensitivity Sfreq is tunable with respect to separation (Fig. 7). System parameters of τ2f = 750 µs and A2f = 1 nm are selected. For changing separation xSM, the change in f0C is recorded on the first y-axis (right). Sensitivity, on the second y-axis (left), is calculated using Eq. 11. The gradient field used to calculate sensitivity is the equivalent field that deflects the magnet, altering xSM. An inverse power function profile is observed as a result. Such a profile is expected, as the Casimir force follows an inverse cubic function with respect to separation (Eq. 7). This design is not static, however, and the parametric pumping (Eq. 8) modifies the dynamics. As introduced earlier, an analytical work12 with an equivalent time delay of 0 s proposed detection proportional to the 5th power of the separation. We find the region with zero delays to lead to pull-in (Fig. 5b) and utilize the controlled response found at τ2f = 750 µs. To compare to the previously proposed detection scheme at maximum amplification, we use an inverse exponential fit y = a/(xb)c + d, where a, b, c, and d are constants and c describes the power relationship. Fitting the data in Fig. 7 yields c = 2.6 for our proposed frequency detection scheme. While this is significantly less sensitive than the 5th power relationship (c = 5), the Casimir coupling is amplified and resilient to pull-in.

The best-case resolution of the system is discussed for frequency shift detection of the high-quality peak. For laboratory-based frequency detection systems, such as Agilent’s 53132A frequency counter, a resolution of 10 parts per million using a 1 s gate time is relatively standard. A maximum sensitivity Sfreq of 6 Hz/(pT/cm) is observed (Fig. 7). Therefore, the frequency detection scheme has a best-case resolution of 1.6 aT/cm at 1 Hz. While we anticipate thermomechanical noise to experimentally limit the technique at much larger gradient magnetic fields, this constitutes a four orders of magnitude improved resolution (10,000-fold) for the best-case scenario of the presently designed MEMS single-point gradiometer27.

While this work focuses on employing a Casimir-driven parametric amplifier for MEMS sensing, there are other configurations that may be useful to consider. The electrostatic force acts at larger separations (>100 nm), which may be more resilient to pull-in. Electrostatic forces may also be parametrically amplified30,31 and are also nonlinear (albeit less sensitive than the Casimir force). Furthermore, it has been discussed that sensitivity varies with separation and therefore with gradient magnetic field input. Although complicated for a highly dynamic system, we suggest the investigation of a null-sensing technique that may enable the device to sit at a single sensitivity. Using an additional feedback mechanism may control the center positions of both oscillators (keeping them constant) and may afford control of a constant high sensitivity, such as the 6 Hz/(pT/cm) reported here.

Resonant MEMS devices such as this design are often limited by several types of noise. The characteristic 1/f noise from mechanical and electrical sources will largely not affect the resonator coupling near 1 kHz, but the low-frequency changes in separation that are sought to be measured here will likely be affected. We suggest that techniques such as chopper stabilization and lock-in amplification be employed to reduce this effect. Low-frequency magnetic noise, such as from power lines in an urban area or the Earth’s magnetic field, will interfere with sensitive magnetic measurements. As was experimentally analyzed previously27, this interference presents itself as a torque on the magnet equal to the cross product of the magnetic moment and interference field (in-plane or out-of-plane). Standard gradiometer designs23,24 reduce this noise from distant sources by subtracting the signals from two closely spaced magnetic sensors. This spacing is typically on the order of 1 cm, and so by design, our system improves this reduction with a spatial element 0.25 mm in length along the sensitive axis. Furthermore, our technique presents a subtractionless measurement, offering a reduction in associated error for gradiometric measurements. Although a central goal of our technique is to develop a sensor for unshielded biomagnetic measurements, shielding may still be employed to further reduce the interference of low-frequency magnetic fields.

Our design is intended for an ambient temperature and pressure environment, so we anticipate thermomechanical noise to be a dominant influence on our measurements. Thermal damping on each of the resonators is mitigated in part by sine-wave feedback pumping. Our simulations did not investigate noise directly, but the design and experimental implementation may be directly compared to the previous work27. Squeeze film damping has been shown36 to be common on MEMS devices with gaps smaller than 5 μm, such as those designed in this work. Although we intend for this design to be used in ambient environments, vacuum packaging or cryogenic environments would further reduce the effects of damping. Finally, it was theoretically shown37 that another source of damping for a dynamic Casimir oscillator may arise from the nonuniform relative acceleration of the sphere and plate, which encloses the nonlinear properties of vacuum. It may be interesting to combine our sensitive platform design with cryogenic and magnetic shielding environments to investigate this effect experimentally. We are confident that the design strategies presented in this work, in addition to those that reduce the effects of noise, will profoundly enhance the performance of single-point MEMS gradiometers.

Conclusion

We have investigated the quantum-derived coupling of two resonant microstructures to achieve extremely high sensitivity to changes in a gradient magnetic field. The resonators are coupled by a nonlinear Casimir force, which arises from the electromagnetic interaction between closely spaced dielectrics (near 100 nm) in a sphere-plate geometry. A customized parametric amplification technique is developed, where one resonator is synchronized at double the frequency of the other, and the time delay is tuned to find a steady-state solution. The frequency shift of a high-quality, coupled resonance peak is detected to infer a measured gradient magnetic field. A slowly varying field at 1 Hz is imposed, where the best-case resolution is calculated to be 1.6 aT/cm at a sensitivity of 6 Hz/pT/cm. This is a 10,000-fold improvement on the best-case resolution of the previously designed MEMS single-point gradiometer. Many applications, especially the measurement of biomagnetic fields, already rely on complex quantum metrology. The MEMS quantum-enhanced gradiometer presented in this work paves the way toward unshielded, ambient temperature measurements of extremely weak gradient magnetic fields.