Three-dimensional particle image velocimetry: experimental error analysis of a digital angular stereoscopic system

and

Published under licence by IOP Publishing Ltd
, , Citation N J Lawson and J Wu 1997 Meas. Sci. Technol. 8 1455 DOI 10.1088/0957-0233/8/12/009

0957-0233/8/12/1455

Abstract

Experimental error analysis of a digital angular stereoscopic PIV system is presented. The paper firstly describes an experimental rig which includes the design of a novel PIV test block for in situ calibration. This allowed the user to set up a static seeded flow volume which was translated in and out of plane to record PIV images using two megapixel CCD cameras positioned for angular stereoscopic viewing. PIV data were collected for a range of camera angles up to and for a range of flow displacements and processed by cross correlation into a set of two-dimensional calibration and flow displacement vectors. These 2D data were then processed into three-dimensional data by the use of geometric and bicubic spline interpolation algorithms and an error analysis performed on the predicted displacements. Results from this analysis have shown optimum system performance will be obtained by using camera angles of between 20 and and f numbers of f16 and higher. The results have also shown a theoretical prediction of system performance derived in previous work, which considers the ratio of out of plane to in plane errors, matches to within 8 and 18% of the experimental system performance.

Export citation and abstract BibTeX RIS

Please wait… references are loading.
10.1088/0957-0233/8/12/009