Abstract

Modifications in texture and surface chemistry are commonly used to increase the biologic response to implants. We evaluated the influence of five different acid etching treatments on the chemical and topographical superficial characteristics of cpTi grade IV discs (test groups). One group of samples were only polished (control group). The samples were analyzed by electron microscopy (SEM, EDS), atomic force microscopy, and grazing incidence XRD. The acid etching treatments which produced higher values for the amplitude roughness parameters showed a combination of strong acids (HF and HCl/H2SO4) at high concentrations, with a relatively high temperature (≥ 60 °C) for a considerable time (≥ 60 min). Titanium and oxygen were the only elements detected by EDS on the surface in all groups, whereas titanium hydride was detected when the samples were analyzed by GIXRD. Only the group subjected to thermal treatment showed presence of rutile phase on the surface.

titanium; surface modification; acid-etching; surface characterization


Study of the influence of acid etching treatments on the superficial characteristics of Ti

Bruno Ramos Chrcanovic* * e-mail: brunochrcanovic@hotmail.com ; Maximiliano Delany Martins

Laboratório de Nanoscopia - LabNano, Centro de Desenvolvimento da Tecnologia Nuclear - CDTN, Comissão Nacional de Energia Nuclear - CNEN, Av. Presidente Antônio Carlos, 6627, Campus da UFMG, Pampulha, CEP 31270-901, Belo Horizonte, MG, Brazil

ABSTRACT

Modifications in texture and surface chemistry are commonly used to increase the biologic response to implants. We evaluated the influence of five different acid etching treatments on the chemical and topographical superficial characteristics of cpTi grade IV discs (test groups). One group of samples were only polished (control group). The samples were analyzed by electron microscopy (SEM, EDS), atomic force microscopy, and grazing incidence XRD. The acid etching treatments which produced higher values for the amplitude roughness parameters showed a combination of strong acids (HF and HCl/H2SO4) at high concentrations, with a relatively high temperature (≥ 60 °C) for a considerable time (≥ 60 min). Titanium and oxygen were the only elements detected by EDS on the surface in all groups, whereas titanium hydride was detected when the samples were analyzed by GIXRD. Only the group subjected to thermal treatment showed presence of rutile phase on the surface.

Keywords: titanium, surface modification, acid-etching, surface characterization

1. INTRODUCTION

Alterations in endosseous dental implant design have been revisited extensively to decrease treatment time frames by reducing the healing period for osseointegration establishment. Thomas and Cook1 examined the variables that potentially influence bone apposition to an implant surface. Of twelve parameters studied, only surface characteristics were shown to have a significant effect on bone integration of an implant. Because the implant surface is the first component to interact with the host, several surface modifications have been extensively investigated in the search for improved bone healing that would allow immediate or early loading of dental implants2. Alterations in surface texture and chemistry are modifications used commonly to increase the biological response to implants. The beneficial effect of rough implant surfaces on peri-implant bone formation is considered to be based on the changes in microtopography and subsequent alterations of surface energy that result in increased interaction with the adjacent biological environment by adsorption of proteins and blood components which in turn can enhance cell attachment and implant integration3.

Recently, commercially pure titanium (cpTi) has become widely used as a biomaterial for dental implants, orthopaedic implants, cardiovascular appliances, and implant-supported dental crowns because of outstanding characteristics that include high specific strength, high resistance to corrosion, greater biocompatibility, low modulus of elasticity, and high capacity to be osseointegrated with bone4. Furthermore, the creation of roughened titanium surfaces has long been demonstrated to be an effective way to promote the interfacial biomechanical properties of bone-anchored implants by means of increasing the interlocking capacity of surface5. The literature shows that when micromechanical anchorage is achieved, the bone-repair process switches from slower bone corticalization around the implant surface to more rapid bone trabeculization and bone ingrowth into the rough surface6. Some biomechanical studies in animal models found that the surface texture of titanium implants has a significant influence on their anchorage in bone7,8.

Bone interlocking or micromechanical anchorage at the interface is not a feature common to all surfaces; to achieve it, a certain level of roughness seems to be required. Some authors9 showed that machined surfaces do not achieve micromechanical anchorage at the interface. When torqued, bone was not found attached to the machined surface, whereas anchorage has been documented for other surfaces with physical and chemical surface modifications9.The surface topography obtained by acid attack can be modulated according to prior treatment as, for example, by sandblasting, using acid mixtures, using different temperatures, and using different etching times. Acid etching is a subtractive method, wherein pits are created in the titanium surface10. Acid etching of titanium is of particular interest because it creates a microtextured surface (fine rough surface with micro pits of 1-3 µm and larger pits of approximately 6-10 µm) that appears to enhance early endosseous integration and stability of the implant7.

Some studies support the belief that the subtractive surface modification by acid etching has a positive effect on the strength of endosseous integration. Lazzara et al.11 compared the bone response to Osseotite (an implant system that has a machined and dual acid-etched surface - MDAS) and machined surfaces (MS) placed in the human posterior maxilla. After 6 months of healing, the bone contact at the MDAS was significantly higher when compared to MS, and the osteoconductive effect (the amount of bone apposition) of the MDAS over the MS was particularly pronounced in the softer trabecular bone. Klokkevold et al.7 compared the anchorage of MDAS and MS in the rabbit tibia model, and observed a statistically significant higher mean removal torque for the MDAS after 1, 2, and 3 months. Treating titanium by HF/HNO3 or by HCl, HF, and H3PO4, He et al.12 and Zareidoost et al.13 respectively, showed that acid-etched titanium surfaces had higher roughness, lower cytotoxicity level and better biocompatibility than the control samples not submitted to acid etching.

Thus, considering the importance of this subtractive method of surface modification, the aim of the present study was to evaluate the influence of five different acid etching treatments described in the literature on the chemical and topographical superficial characteristics of cpTi grade IV.

2. MATERIAL AND METHODS

Eighteen machined cpTi grade IV discs (12.7 × 2.0 mm) were used as the substrate material for the experiment. All discs were cut from a rod using a IsoMet® Low Speed Saw (Buehler®, Lake Bluff, USA) with a Diamond Wafering Blade No. 11-4244 (diameter 4"-102mm / thickness 0.012" - 0.3mm) from the same manufacturer. The samples were embedded in polymethyl methacrylate, in order to be polished in a polishing machine until a #2000 silicon carbide paper.

The discs were separated into a control group with 3 discs and 5 test groups with 3 discs each. Each test group was acid-etched by a different acid treatment (defined as groups AT1 to AT5) previously described in the literature14-18. Table 1 presents a detailed description of these 5 acid-etching treatments. The firing in vacuum was performed in a VITA Vacumat 40T vacuum furnace (VITA Zahnfabrik H. Rauter GmbH & Co.KG, Bad Säckingen, Germany).

The surface morphology of treated samples was examined by scanning electron microscopy (SEM - JEOL, model JSM-5310, Tokyo, Japan). The secondary electron (SE) detection mode with an acceleration voltage of 25 kV was selected for SEM analysis and the vacuum pressure was maintained below 1 × 10 - 5 Torr. The load current (LC) was approximately 85 µA. For a direct comparison of the surface morphology, the same magnification of 1000× was selected for all samples.

In order to obtain quantitative analysis of the surface roughness, atomic force microscopy (AFM - NTegra Aura, NT-MDT, Moscow, Russia) of the samples was performed. AFM images were acquired in air using semicontact mode with a sharpened gold-coated silicon tip (nominal spring constant of 2.5-10 N/m and nominal resonance frequency of 110-200 kHz, NSG01 probe series, NT-MDT, Moscow, Russia). The scanning area for the measurements was 50 × 50 µm2. The images obtained by AFM were characterized by 2nd order extraction filter, using the software Image Analysis 2.1.2 (NT - MDT, Moscow, Russia). The seven amplitude surface roughness parameters calculated by the software were evaluated (Sy, Sz, Sa, Sq, Ssk, and Ska). The mean value and standard deviation of these parameters was obtained from 15 satisfactory scans of each group (5 from each of the 3 samples), from random sites on the surface. A statistical analysis of the mean values of the surface roughness parameters was composed and compared with One-Way Analysis of Variance (ANOVA), since a normal distribution of the variables was observed by the Kolmogorov-Smirnov test. The Tukey's HSD test was also performed, in order to find means that are significantly different from each other. The degree of statistical significance was considered P < 0.05. The Statistical Package for the Social Sciences (SPSS) version 20 software (SPSS Inc., Chicago, USA) was used to perform the statistical analysis.

The surface chemical composition was analyzed by energy dispersive x-ray spectroscopy (EDS - JEOL, model JXA-8900RJ, Tokyo, Japan). The most central area of the samples was chosen, and the analysis was made with a magnification of 200×. The elemental chemical composition was calculated by the mean value and standard deviation from the 3 samples of each group.

Moreover, grazing incidence X-ray diffraction (GIXRD) measurements were carried out in a Ultima IV X-ray diffractometer (Rigaku, Tokyo, Japan), using Cu-Kα1 radiation at 30 kV and tube current of 20 mA, without any filter or monochromator, in the angle range of 10° - 90° (2θ) with a grazing incidence of 3º, making the diffraction sensitive to the surface. The step of measurement was set to 0.05° with a scan rate of 0.5° per minute. The divergence slit was set to 1 mm, with a Div H.L. Slit of 2 mm. The results were analyzed in Search-Match software (Crystallographica, Oxford, United Kingdom). GIXRD experiments were carried out in order to distinguish chemical compounds at the sample surface.

3. RESULTS

3.1. SEM analysis

Figures 1a-f revealed characteristic differences at the microscopic level according to the surface modification methods used for the samples as measured by SEM. Differences of the surfaces were obvious between the groups due to differences of the etching processes. The control group samples were mainly characterized by multidirectional grooves as result of the polishing process (Figure 1a). Samples submitted to acid treatment classified as AT1 showed a homogenous distribution of irregularities throughout the surface (Figure 1b). AT2 (Figure 1c) showed the presence of ridges and valleys with bigger dimensions as shown in AT1 samples, in many regions resembling grooves with a width varying from 1 to 2 µm. AT3 samples showed a surface with a great number of micro pits of 1-3 µm and some larger pits (Figure 1d). AT4 showed an almost homogeneous distribution of valleys and peaks with sharp edges (Figure 1e). Acid etching treatment AT5 did not eliminate the multidirectional grooves (Figure 1f), suggesting that the acid treatment was not effective to change surface morphology.


3.2. AFM analysis

The qualitative and quantitative surface topography demonstrated different degrees of roughness. Table 2 shows the mean values of tridimensional roughness parameters for the control group and acid-etched samples (groups AT1 to AT5) as determined by AFM. It can be observed that the surfaces of the samples from control group showed smaller mean values of the amplitude roughness parameters than the test groups. AT1 showed closest mean values of Sy, Sz, and Average when compared to the control group, and the same mean value of Sa and Sq as the control group. The samples from AT2, AT3, and AT4 groups showed higher mean values. The topographic maps obtained by AFM showed, qualitatively, the difference in roughness between the six surfaces: control group (Figure 2a) and test groups AT1 to AT5 (Figures 2b-f, respectively).


The amplitude parameters characterize the surface based on the vertical deviations of the roughness profile from the mean line. The roughness parameter Ssk represents the degree of symmetry of the surface heights about the mean plane. The sign of Ssk indicates the predominance of peaks (i.e. Ssk > 0) or valley structures (Ssk < 0) comprising the surface. Thus, it can be observed that all groups but AT4 have a predominance of peaks. Sku indicates the presence of inordinately high peaks/deep valleys (Sku > 3.00) or lack thereof (Sku < 3.00) making up the texture. In this case, groups AT1 and AT5 were the only ones with a presence of inordinately high peaks/deep valleys.

Comparison of the mean values of the surface roughness parameters with ANOVA showed that the difference of the values between the groups revealed to be statistically significant for all parameters. Groups AT2 and AT3 showed statistically significant difference for the mean value of the roughness parameters when compared to the other groups, but not between themselves. The difference of the mean values of the control, AT1 and AT5 groups were also statistically significant for most of the parameters when compared to the other study groups, but not between themselves.

3.3. EDS analysis

EDS analysis of the surfaces showed titanium and oxygen to be the only elements in all groups, with a similar average % atomic concentration of 60% and 40% (± 5%), respectively. Figure 3 shows a typical EDS spectrum of the group AT2.


3.4. XRD analysis

Figure 4 presents a typical GIXRD diffractogram for a sample from the control group, where diffraction peaks were labeled according to Miller indices, as described elsewhere19. The diffractogram analysis confirmed only the presence of titanium (Ti). The GIXRD results obtained for the samples from groups submitted to acid etching showed the presence of Ti and titanium hydride (TiH2) in groups AT1 to AT4 (Figures 5a-d, respectively), and Ti, TiH2, and rutile (TiO2) in group AT5 (Figure 5e).



4. DISCUSSION

Chemical texturing is a method of surface roughening through acid etching. This method has advantages compared with current methods of roughening such as grit blasting, plasma spraying, and beads. As the production process does not stress the adjacent material, there is no risk of flaking or delamination and the process does not leave particles of grit20,21.

All test groups have been acid attacked; nevertheless, due to the distinct treatment parameters, every implant displayed distinct surface roughness characteristics and a proper surface aspect that could not be mistaken. Seven distinct roughness parameters have been reported for each surface. However, their significance in terms of predicting an enhanced bone response or an increased implant fixation remains questionable10. Some studies have demonstrated that increases in the numerical surface roughness of an implant enhance biomechanical anchorage in bone, as determined by torque removal tests8 and resistance to push-out forces22. These studies showed a tendency toward an increase in bone contact, and resistance to removal, with an increasing numerical implant surface roughness. In the groups AT2, AT3, and AT4 it can be seen that the etching created micropits, when the SEM image is compared with the control group, which also reflected in the higher values of the amplitude roughness parameters in these samples. Groups AT2 and AT3 also showed statistically significant difference for the value of the roughness parameters when compared to the other groups. This was also true for the group AT4 for some parameters, but not all. Specific to dental implants, studies have shown that histologic and biomechanical characteristics were improved due to increase in the as-machined surface texture by varied methods resulting in average implant surface roughness (Sa) ranging from 0.5 to 2 µm23,24. Thus, the values found in the present study for the roughness parameter Sa seems to be appropriate in these groups. Although biological tests were not conducted in the present study, it can be suggested that these surfaces with the highest values of roughness parameters (AT2, AT3 and AT4) could also present a greater osseous anchorage, if implants had been treated with these acid etchings, and if could only consider the surface topographic characteristics, based on results found in the literature. Several studies reported a good correlation between increased Ra (bi-dimensional correlate of the three-dimensional Sa) and stronger anchorage7,9,22, although the Ra is obviously insufficient by its own to characterize a given surface25. Wong et al.22 evaluated pushed-out titanium cylinders placed in the knee trabecular bone of mini-pigs after 3 months, comparing four surfaces. The surface with the highest Ra showed the highest anchorage; a strong linear correlation, r2 = 0.90, was found between the Ra and the push-out load. Gotfredsen et al.9 measured the anchorage of four treated surfaces in a rabbit tibia model after 4, 6, 9, and 12 weeks with the torque test. The highest torque values were observed at each time point for the roughest surface, as determined by the Ra and Rt. At all time points, these authors reported a high correlation of r = 0.83 between the Ra and implant anchorage. Klokkevold et al.7 compared the torque resistance to removal of screw-shaped titanium implants having a MDAS with implants having either a MS, or a titanium plasma spray surface (TPS), in a rabbit model. Their results showed an enhanced bony anchorage to MDAS implants as compared to MS implants (also showing a much higher value of Ra for the MDAS implants), and that dual acid etching of titanium enhances early endosseous integration to a level which is comparable to that achieved by the topographically more complex TPS surfaces. However, although the height-descriptive parameters of the surfaces of the groups AT2, AT3, and AT4 were significantly higher than the AT1 and AT5 surfaces, one cannot speculate on their biological performance until their bone interlocking capacity is demonstrated. It turns out that clinical implications cannot be drawn by relying on roughness-descriptive parameters alone10.

It has been reported that the etching process modifies the Ti surface composition of SLA-treated implants, and XRD and metallographic microscopy analysis indicated the presence of 20 to 40% of titanium hydride (TiHx, x < 2) in addition to Ti[26]. Before attacking the metallic titanium, the acids must first dissolve the protective titanium oxide layer. During the course of the corrosion process of titanium, native hydrogen ions (H+) are released. These small ions diffuse rapidly into the metal because the latter is left without its dense protective oxide layer; the sub-surface is therefore enriched with hydrogen27. When saturation in hydrogen is reached, titanium hydride is formed. Titanium hydride may be biologically important because a hydride layer is much better suited as a template for binding biomolecules chemically onto a titanium surface28. In the present study, as the surface of the samples were also modified by acid etching, it was not surprising that titanium hydride was found in all test groups (from AT1 to AT5) by GIXRD analysis.

EDS analysis of the surfaces showed titanium and oxygen to be the only elements in all groups, most probably due to the natural formation of a titanium oxide passivation layer (mainly TiO2) just after sample surface preparation. The result (about 60%/40%) agrees with the mass percentage of the TiO2, the main component of a typical titanium oxide passivation layer. It was previously observed, by using XPS analysis, that dental implants surfaces treated only by sandblasting and acid-etching consists of oxidized titanium29. Moreover, TiO2 can exist in three crystalline forms called anatase, rutile, and brookite, all with different physical properties. The quasi-amorphous oxides can be crystallized by heating to temperatures of 400-500 °C or higher, which leads to transformation to more crystallized phases30. Thus, the presence of rutile phase in group AT5 can be explained by the thermal treatment by which the samples of this group were subjected (firing in vacuum at 600 °C for 10 min).

The results showed that the acid etching treatments used here produced a surface with different topographical and chemical features compared to the polished one (control), except for AT5. The present results demonstrated that the acid etching can remove the grooves produced by the polishing process. The etching can create a surface with randomly distributed pits in a micron-scale (as seen with AT3 - Figure 1d), or homogeneously create ridges and valleys throughout the surface (as seen with AT1, AT2, and AT4), enlarging the surface area. Although the ultimate aim of acid attack is to create pits to allow for bone ingrowth, etching varies from weak to strong. The AT1 produced very small pits. AT5 did not change the surface morphology when compared with the control group. The surface aspect is probably caused by the use of either a weak acid mixture, a low etching temperature, or a short etching time10. In the case of AT1, the room temperature might have been the cause, whereas in the case of AT5, the cause was probably a weak acid solution. The formation of great number of micropits in the group AT3 probably resulted from a combination of strong acids (HCl/H2SO4) at high concentrations, a relatively high temperature (60 °C), for a considerable time (60 min). The result of an irregular surface with higher values of roughness parameters in groups AT2 and AT4 may be the consequence of a combination of 2 etching steps with strong acids at high concentrations, and with high temperature (80 °C) in at least one step of the etching bath.

This same research team conducted in 2013 a study comparing the effect of the same five acid etching treatment here described (Table 1) on the superficial characteristics of cpTi grade IV, but after sandblasting the samples with aluminum oxide21. Comparing the present roughness parameters results with the ones obtained by Chrcanovic et al.21, it can be observed that the values did not significantly differ in groups AT2, AT3, and AT4. When this is taken into account it can be suggested that for these acid etching groups, at least when considering the amplitude roughness parameters, there is no need to perform the sandblasting procedure before the acid treatment in order to increase the roughness of the surface; only the acid etching treatment could be enough.

Although some of the surfaces here created using specific acid etching methods resemble chemically and topographically surfaces with osseointegration potential described in the literature, further research is necessary to clear up the influence of acid etching methods on the biological response.

5. CONCLUSIONS

Treating titanium surfaces with acid does not create a standard topography. The latter varies according to several parameters, such as prior treatment, acid mixture composition, temperature, and time of acid treatment. The acid etching treatments here classified as AT2, AT3, and AT4 produced higher values for the amplitude roughness parameters in comparison with the other groups (control, AT1, and AT5). These favorable acid treatments showed a combination of strong acids (HF and HCl/H2SO4) at high concentrations, with a relatively high temperature (≥ 60 °C) for a considerable time (≥ 60 min). EDS analysis of the surfaces showed titanium and oxygen to be the only elements in all groups. Only the group subjected to thermal treatment (AT5) showed presence of rutile phase on the surface. All test group samples showed the presence of titanium hydride on the surface after the acid etching treatment. Since the present study has used discs as the substrate material, the observed results cannot be directly extrapolated to complex surfaces like those in cylindrical threaded titanium implants.

ACKNOWLEDGEMENTS

This work was supported by the Brazilian agencies CNPq and FAPEMIG. The authors would like to thank Eduardo Antônio de Carvalho (CDTN/CNEN) and Rolf Ankli (Dental Atelier Rolf Ankli Art & Experience).

Received: April 8, 2013

Revised: February 6, 2014

  • 1. Thomas KA and Cook S. An evaluation of variables influencing implant fixation by direct bone apposition. Journal of Biomedical Materials Research 1985; 19(8):875-901. http://dx.doi.org/10.1002/jbm.820190802
  • 2. Vervaeke S, Collaert B and De Bruyn H. The effect of implant surface modifications on survival and bone loss of immediately loaded implants in the edentulous mandible. International Journal of Oral and Maxillofacial Implants 2013; 28(5):1352-1357. http://dx.doi.org/10.11607/jomi.3200
  • 3. Boukari A, Francius G and Hemmerle J. AFM force spectroscopy of the fibrinogen adsorption process onto dental implants. Journal of Biomedical Material Research 2006; 78(3):466-472. http://dx.doi.org/10.1002/jbm.a.30778
  • 4. Park YJ, Song YH, An JH, Song HJ and Anusavice KJ. Cytocompatibility of pure metals and experimental binary titanium alloys for implant materials. Journal of Dentistry 2013; 41(12):1251-1258. http://dx.doi.org/10.1016/j.jdent.2013.09.003
  • 5. Butz F, Ogawa T and Nishimura I. Interfacial shear strength of endosseous implants. International Journal of Oral and Maxillofacial Implants 2011; 26(4):746-751.
  • 6. Buser D, Schenk RK, Steinemann SG, Fiorellini JP, Fox CH and Stich H. Influence of surface characteristics on bone integration of titanium implants: A histomorphometric study in miniature pigs. Journal of Biomedical Materials Research. 1991; 25(7):889-902. http://dx.doi.org/10.1002/jbm.820250708
  • 7. Klokkevold PR, Johnson P, Dadgostari S, Caputo A, Davies JE and Nishimura RD. Early endosseous integration enhanced by dual acid etching of titanium: A removal torque study in the rabbit. Clinical Oral Implant Research 2001; 12(4):350-357. http://dx.doi.org/10.1034/j.1600-0501.2001.012004350.x
  • 8. Souza FA, Queiroz TP, Guastaldi AC, Garcia-Júnior IR, Magro-Filho O, Nishioka RS et al. Comparative in vivo study of commercially pure Ti implants with surfaces modified by laser with and without silicate deposition: biomechanical and scanning electron microscopy analysis. Journal of Biomedical Materials Research B Applied Biomaterials 2013; 101(1):76-84. http://dx.doi.org/10.1002/jbm.b.32818
  • 9. Gotfredsen K, Berglundh T and Lindhe J. Anchorage of titanium implants with different surface characteristics: An experimental study in rabbits. Clinical Implant Dentistry and Related Research 2000; 2(3):120-128. http://dx.doi.org/10.1111/j.1708-8208.2000.tb00002.x
  • 10. Szmukler-Moncler S, Testori T and Bernard JP. Etched implants: A comparative surface analysis of 4 dental implant systems. Journal of Biomedical Materials Research B Applied Biomaterials 2004; 69(1):46-57. http://dx.doi.org/10.1002/jbm.b.20021
  • 11. Lazzara RJ, Testori T, Trisi P, Porter SS and Weinstein RL. A human histologic analysis of Osseotite and machined surfaces using implants with 2 opposing surfaces. International Journal of Periodontology and Restorative Dentistry 1999; 19(2):117-129.
  • 12. He FM, Yang GL, Zhao SF and Cheng ZP. Mechanical and histomorphometric evaluations of rough titanium implants treated with hydrofluoric acid/nitric acid solution in rabbit tibia. International Journal of Oral and Maxillofacial Implants 2011; 26(1):115-122.
  • 13. Zareidoost A, Yousefpour M, Ghaseme B and Amanzadeh A. The relationship of surface roughness and cell response of chemical surface modification of titanium. Journal of Materials Science: Materials in Medicine 2012 Jun; 23(6):1479-88. http://dx.doi.org/10.1007/s10856-012-4611-9
  • 14. Orsini G, Assenza B, Scarano A, Piattelli M and Piattelli A. Surface analysis of machined versus sandblasted and acid-etched titanium implants. International Journal of Oral Maxillofacial Implants 2000; 15(6):779-784.
  • 15. Cho SA and Park KT. The removal torque of titanium screw inserted in rabbit tibia treated by dual acid etching. Biomaterials 2003; 24(20):3611-3617. http://dx.doi.org/10.1016/S0142-9612(03)00218-7
  • 16. Iwaya Y, Machigashira M, Kanbara K, Miyamoto M, Noguchi K, Izumi Y et al. Surface properties and biocompatibility of acid-etched titanium. Dental Materials Journal 2008; 27(3):415-421. http://dx.doi.org/10.4012/dmj.27.415
  • 17. Yang GL, He FM, Yang XF, Wang XX and Zhao SF. Bone responses to titanium implants surface-roughened by sandblasted and double etched treatments in a rabbit model. Oral Surgery Oral Medicine Oral Pathology Oral Radiology and Endodontics 2008; 106(4):516-524. http://dx.doi.org/10.1016/j.tripleo.2008.03.017
  • 18. Carvalho DR, Carvalho PS, Magro O Fº, De Mello JD, Beloti MM and Rosa AL. Characterization and in vitro cytocompatibility of an acid-etched titanium surface. Brazilian Dental Journal 2010; 21(1):3-11.
  • 19. Chrcanovic BR, Pedrosa AR and Martins MD. Chemical and topographic analysis of treated surfaces of five different commercial dental titanium implants. Materials Research 2012; 15(3):372-382. http://dx.doi.org/10.1590/S1516-14392012005000035
  • 20. Daugaard H, Elmengaard B, Bechtold JE and Soballe K. Bone growth enhancement in vivo on press-fit titanium alloy implants with acid etched microtexture. Journal of Biomedical Materials Research A 2008; 87(2):434-440. http://dx.doi.org/10.1002/jbm.a.31748
  • 21. Chrcanovic BR, Leão NLC and Martins MD. Influence of different acid etchings on the superficial characteristics of Ti sandblasted with Al2O3 Materials Research 2013; 16(5):1006-1014. http://dx.doi.org/10.1590/S1516-14392013005000067
  • 22. Wong M, Eulenberger J, Schenk R and Hunziker E. Effect of surface topology on the osseointegration of implant materials in trabecular bone. Journal of Biomedical Materials Research 1995; 29(12):1567-1575. http://dx.doi.org/10.1002/jbm.820291213
  • 23. Albrektsson T and Wennerberg A. Oral implant surfaces: Part 1-Review focusing on topographic and chemical properties of different surfaces and in vivo responses to them. International Journal of Prosthodontics 2004; 17(5):536-543.
  • 24. Buser D, Broggini N, Wieland M, Schenk RK, Denzer AJ, Cochran DL et al. Enhanced bone apposition to a chemically modified SLA titanium surface. Journal of Dental Research 2004; 83(7):529-533. http://dx.doi.org/10.1177/154405910408300704
  • 25. Hansson S. Surface roughness parameters as predictors of anchorage strength in bone: a critical analysis. Journal of Biomechanics 2000; 33(10):1297-1303. http://dx.doi.org/10.1016/S0021-9290(00)00045-2
  • 26. Perrin D, Szmukler-Moncler S, Echikou C, Pointaire P and Bernard JP. Bone response to alteration of surface topography and surface composition of sandblasted and acid etched (SLA) implants. Clinical Oral Implants Research 2002; 13(5):465-469. http://dx.doi.org/10.1034/j.1600-0501.2002.130504.x
  • 27. Frank MJ, Walter MS, Lyngstadaas SP, Wintermantel E and Haugen HJ. Hydrogen content in titanium and a titanium-zirconium alloy after acid etching. Materials Science and Engineering: C Materials for Biological Applications. 2013; 33(3):1282-1288. http://dx.doi.org/10.1016/j.msec.2012.12.027
  • 28. Videm K, Lamolle S, Monjo M, Ellingsen JE, Lyngstadaas SP and Haugen HJ. Hydride formation on titanium surfaces by cathodic polarization. Applied Surface Science 2008; 255(5):3011-3015. http://dx.doi.org/10.1016/j.apsusc.2008.08.090
  • 29. Castilho GAA, Martins MD and Macedo WAA. Surface characterization of titanium based dental implants. Brazilian Journal of Physics 2006; 36(3b):1004-1008. http://dx.doi.org/10.1590/S0103-97332006000600055
  • 30. Nam SH, Choi JW, Cho SJ, Kimt KS and Boo JH. Growth of TiO2 anti-reflection layer on textured Si (100) wafer substrate by metal-organic chemical vapor deposition method. Journal of Nanoscience and Nanotechnology 2011; 11(8):7315-7318. http://dx.doi.org/10.1166/jnn.2011.4813
  • *
    e-mail:
  • Publication Dates

    • Publication in this collection
      01 Apr 2014
    • Date of issue
      Apr 2014

    History

    • Received
      08 Apr 2013
    • Accepted
      06 Feb 2014
    ABM, ABC, ABPol UFSCar - Dep. de Engenharia de Materiais, Rod. Washington Luiz, km 235, 13565-905 - São Carlos - SP- Brasil. Tel (55 16) 3351-9487 - São Carlos - SP - Brazil
    E-mail: pessan@ufscar.br