A screen-printed dissolved oxygen sensor was fabricated using cerium oxide-supported silver catalyst and polydimethylsiloxane (PDMS) film. A PDMS film of 3 µm thickness showed good permeability for oxygen and impermeability for hydrogen peroxide. The calibration curve has shown a linear relationship with a correlation coefficient of 0.996 for the dissolved oxygen concentration. The sensitivity and detection limit of the present sensor were calculated at -158 µ,A mM-1 and 8.4 µM, respectively.
Similar content being viewed by others
References
J. R. Stetter and J. Li, Chem. Rev., 2008, 108, 352.
L. Hutton, M. E. Newton, P. R. Unwin, and J. V. Macpherson, Anal. Chem., 2009, 81, 1023.
M. L. Hichman, “The Measurement of Dissolved Oxygen”, 1978, John Wiley, New York.
W. Glasspool and J. Atkinson, Sens. Actuators, B, 1998, 48, 308.
C. C. Yang, A. S. Kumar, and J. M. Zen, Electro analysis, 2006, 18, 64.
Y. Lin, X. Cui, and X. Ye, Electrochem. Commun., 2005, 7, 267.
M. Bron, P. Bogdanoff, S. Fiechter, and H. Tributsch, J. Electroanal. Chem., 2005, 578, 339.
F. S. Damos, R. C. S. Luz, A. A. Tanaka, and L. T. Kubota, Anal. Chim. Acta, 2010, 664, 144.
J. Jiang and A. Kucernak, Electrochim. Acta, 2002, 47, 267.
E. S. Ribeiro, S. L. P. Dias, Y. Gushikem, and L. T. Kubota, Electrochim. Acta, 2004, 49, 829.
S. Tsujimura, K. Kano, and T. Ikeda, J. Electroanal. Chem., 2005,576, 113.
M. Itagaki, Y. Fujimura, I. Shitanda, K. Watanabe, and T. Hachiya, Anal. Sci., 2006, 22, 1315.
E. Ramasamy, W. J. Lee, D. Y. Lee, and J. S. Song, J. Power Sources, 2007, 165, 446.
M. Itagaki, Y. Nakano, I. Shitanda, and K. Watanabe, Electrochim. Acta, in press.
I. Shitanda, S. Takamatsu, K. Watanabe, and M. Itagaki, Electrochim. Acta, 2009, 54, 4933.
F. Ricci, A. Amine, G. Palleschi, and D. Moscone, Biosens. Bioelectron., 2003, 18, 165.
I. Shitanda, A. Okumura, M. Itagaki, K. Watanabe, and Y. Asano, Sens. Actuators, B, 2009, 139, 292.
F. Mizutani, T. Sawaguchi, S. Yabuki, and S. Iijima, Electrochemistry, 1999, 67, 1138.
F.Mizutani, S. Yabuki, and S. Iijima, Electro analysis, 2001, 13, 370.
F. Mizutani, S. Yabuki, T. Sawaguchi, Y. Hirata, Y. Sato, and S. Iijima, Sens. Actuators, B, 2001, 76, 489.
E. Maekawa, N. Kitano, T. Yasukawa, and F. Mizutani, Electrochemistry, 2009, 77, 319.
T. Yasukawa, E. Maekawa, and F. Mizutani, Bunseki Kagaku, 2009, 58, 639.
C. C. Wu, T. Yasukawa, H. Shiku, and T. Matsue, Sens. Actuators, B, 2005, 110, 342.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Shitanda, I., Mori, S. & Itagaki, M. Screen-printed Dissolved Oxygen Sensor Based on Cerium Oxide-supported Silver Catalyst and Polydimethylsiloxane Film. ANAL. SCI. 27, 1049–1052 (2011). https://doi.org/10.2116/analsci.27.1049
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.2116/analsci.27.1049