Review

CMOS Electrochemical Instrumentation for Biosensor Microsystems: A Review

1
Maxim Integrated Products Inc., 160 Rio Robles, San Jose, CA 95134, USA
2
Xcellcure LLC., 1 City Place Drive Suite 285, St. Louis, MO 63131, USA
3
Department of Electrical and Computer Engineering, Michigan State University, East Lansing, MI 48824, USA
4
Apple Inc., 1 Infinite Loop, Cupertino, CA 95014, USA
5
Department of Electrical and Computer Engineering, University of Toronto, Toronto, ON M5S 3G4, Canada
*
Author to whom correspondence should be addressed.
The authors have equal contributions to this paper.
Sensors 2017, 17(1), 74; https://doi.org/10.3390/s17010074
Submission received: 10 October 2016 / Revised: 15 December 2016 / Accepted: 20 December 2016 / Published: 31 December 2016
(This article belongs to the Special Issue Nanobiosensing for Sensors)

Abstract

:
Modern biosensors play a critical role in healthcare and have a quickly growing commercial market. Compared to traditional optical-based sensing, electrochemical biosensors are attractive due to superior performance in response time, cost, complexity and potential for miniaturization. To address the shortcomings of traditional benchtop electrochemical instruments, in recent years, many complementary metal oxide semiconductor (CMOS) instrumentation circuits have been reported for electrochemical biosensors. This paper provides a review and analysis of CMOS electrochemical instrumentation circuits. First, important concepts in electrochemical sensing are presented from an instrumentation point of view. Then, electrochemical instrumentation circuits are organized into functional classes, and reported CMOS circuits are reviewed and analyzed to illuminate design options and performance tradeoffs. Finally, recent trends and challenges toward on-CMOS sensor integration that could enable highly miniaturized electrochemical biosensor microsystems are discussed. The information in the paper can guide next generation electrochemical sensor design.

1. Introduction

Quantification of biological or biochemical processes is of utmost importance for many healthcare, biotechnology and research applications. The global demand for biosensors is forecasted to reach $20 billion by the year 2020 [1]. Among the technologies for biosensing [2,3], electrochemical techniques have been widely employed to capture biological recognition events and directly translate results into the electronic information domain. Electrochemical biosensors have been developed to detect a range of biological targets and events, including hybridized DNA [4,5,6], neuron tissue [7,8], bacteria [9,10,11] and enzyme reactions [12,13,14]. Electrochemical biosensors have several advantages in performance and utility: they allow label-free detection, eliminating the time, cost and complexity of use associated with sample labeling techniques; many do not rely on affinity binding and, therefore, permit continuous real-time measurement; and they are well suited for miniaturization to, e.g., wearable platforms. Compared to label-free optical biosensors, electrochemical sensors are also less susceptible to interference from non-specific binding. Compared to traditional optical biosensors, electrochemical biosensors have several advantages in performance and utility: many provide label-free detection, eliminating the time, cost and complexity of use associated with sample labeling techniques; and they are well suited for miniaturization to, e.g., wearable platforms.
The tremendous growth and advances in microelectronics over the past few decades have elevated complementary metal oxide semiconductor (CMOS) integrated circuits technology to a robust and inexpensive electronics platform that is well suited for the implementation of electrochemical instrumentation. With the advances of CMOS-compatible micro-electrodes, a trend exists to build miniaturized biosensors on silicon chips. Compared to traditional benchtop sensors and instrumentation, CMOS biosensors provide lower cost, lower power and smaller size [15]. They also allow the creation of high-density CMOS biosensor arrays, which enable new sensor platforms, such as DNA testing and drug screening [16]. The past 10–15 years have seen a wide range of CMOS electrochemical circuits introduced that impart a variety of instrumentation features and performance capabilities. To better understand the body of work that has been generated and promote future work that improves upon this foundation, this paper overviews electrochemical sensor concepts from an instrumentation point of view and analyzes performance tradeoffs of the CMOS instrumentation circuits that have been reported for electrochemical biosensors. In addition, this paper outlines the opportunity to adapt CMOS instrumentation into integrated microsystems, discusses challenges faced in the process and reviews recent efforts to address these issues. Section 2 categorizes and describes the electrochemical techniques available for biosensors. Existing CMOS instrumentation structures for electrochemical biosensors are then reviewed in Section 3. Section 4 presents on-CMOS sensor integration challenges and reviews approaches for on-chip electrode fabrication and chip-scale packaging.

2. Electrochemical Biosensor Techniques

An electrochemical biosensor can be generally described as consisting of four functional components: analyte, bio-recognition element, transducer and instrumentation, as illustrated in Figure 1. The analyte is the target of interest, such as a protein, DNA, bacteria, etc. The bio-recognition element, also called the bio-interface, imparts the capability for selective recognition of a target analyte. The transducer translates interactions between the analyte and bio-recognition element into an electrical signal. The typical transducer for an electrochemical biosensor is an electrode that translates the flow of ions generated by binding reactions between the analyte and the bio-recognition element into an electrical current or voltage. The instrumentation component is typically comprised of electronic circuitry that captures, amplifies and records biorecognition signals from the transducer. The liquid environment consists of an electrolyte solution that supports the analyte’s biological activity and transports analytes to the transducer.
A wide range of electrochemical techniques can be employed to characterize biosensors or measure their response during biorecognition events. Historically, these techniques were developed to characterize electrode materials and interface reactions, and electrochemists classify them as potentiometry, amperometry, voltammetry, impedance spectroscopy, etc. However, these “ometry” classes do not clearly reflect the sensor’s signaling modality or their instrumentation needs. Therefore, this paper adopts a classification scheme based on the type of response signal generated [17], and this section defines electrochemical sensors as potentiometric, amperometric or impedimetric based on their responses as DC potentials, DC currents or AC signals, respectively. This section also describes field-effect electrochemical biosensors, and it discusses the impact of sensor electrode structure on electrochemical instrumentation.

2.1. Potentiometric Biosensor

In traditional potentiometric biosensors, a DC electrical potential is measured when there is no current flow between electrodes. A new approach involves applying a controlled DC current [18] and measuring the resulting potential that is proportional to the analyte concentration. Potentiometric biosensors were popular in the past because of their high selectivity, simplicity and low cost. However, low sensitivity limits their use in many modern applications [19]. The current response in potentiometric biosensors is a combination of the sensor response current and the electrode double layer charging current that is modeled by a double layer capacitance in most electrochemical sensor models. The charging current is frequently large and difficult to remove or filter out and, thus, limits the resolution of potentiometric sensors [18]. Some potentiometric biosensors may involve redox reactions and are called redox-potential biosensors. Redox-potential biosensors have been used for glucose and DNA detection. For more information, the readers are referred to [20].

2.2. Amperometric Biosensor

From an instrumentation point of view, biosensors that are stimulated by a DC voltage and generate a DC response current can be categorized as amperometric sensors. The response current is associated with oxidation and/or reduction of an electro-active species involved in the recognition process and relates (ideally directly) to the target analyte concentration. The DC stimulus source can be a constant, step, multi-step or ramp potential applied across the electrode-electrolyte interface and serves as the driving force for the biorecognition process. In the electrochemistry literature, techniques using constant DC stimulus voltages are typically called “amperometry”, while those with DC voltages that vary over time are referred to as “voltammetry”. When the stimulus varies in the pattern of a triangle waveform, it is called “cyclic voltammetry”. Amperometric sensors are commonly used as immunosensors [21], enzyme biosensors [22] and pesticide monitors [23].

2.3. Impedimetric Biosensor

Impedimetric biosensors measure the AC electrical impedance of the electrode-electrolyte interface at equilibrium [24]. Impedimetric biosensors take advantage of a change that occurs during biorecognition events in bio-interface characteristics, namely its capacitance and/or resistance. An impedimetric biosensor is created by imposing a small sinusoidal stimulus voltage (or current) of variable frequency and measuring the resulting current (or voltage), whose phase and/or amplitude change contains information regarding biorecognition events. This technique is generally referred to as the electrochemical impedance spectroscopy (EIS) and has been employed in a variety of applications, such as monitoring bilayer lipid membranes [16,25], detecting small molecules of biological relevance [26,27] and measuring cell concentrations [28,29]. Compared to potentiometric and amperometric biosensors, an important advantage of impedimetric biosensors is that the stimulus sinusoidal voltage is small (usually 5–10 mV in amplitude) and does not damage or disturb most bio-recognition layers.

2.4. Ion-Sensitive Field Effect Transistor Biosensor

Another method to measure ion concentrations in a solution utilizes ion-sensitive field effect transistors (ISFET). As illustrated in Figure 2, an ISFET operates in a manner similar to a metal-oxide-semiconductor field effect transistors (MOSFET) where the current between source and drain terminals is proportional to the gate voltage that, for an ISFET, is set by reactions during biorecognition events between analytes and bio-recognition immobilized on the gate [30]. ISFET biosensors have been developed for enzyme sensing [31], measuring antigen-antibody bonding reactions [32] and DNA detection [33,34,35]. Immobilization of nanomaterials, such as nanoparticles, nanotubes and nanowires, on the gate has been shown to improve sensitivity [36]. Unlike potentiometric, amperometric and impedimetric biosensors that use an electrode transducer, the ISFET’s transducer is the gate oxide layer. The ISFET response allows the output to be either the gate voltage or source current IDS by fixing one and measuring the other. For example, if IDS is fixed at a value, the gate voltage can be measured using a source follower structure [37]. ISFET output current may be measured by a readout circuit similar to that of an amperometric/impedimetric biosensor. However, it needs more controlled voltages (VRE and VDS) and thus requires a different circuit [38]. The remainder of this paper will focus on biosensors with electrode transducers. For more information on the principles, applications and challenges of ISFET biosensors, the reader is referred to [30,39].

2.5. Electrode Transducer

Electrochemical biosensors employ multiple electrodes to translate biorecognition responses into electrical signals. The electrode where electrochemical reactions of interest take place is called the working electrode (WE). The electrode that provides a stable and known potential is called the reference electrodes (RE). The simplest electrode system utilizes only these two electrodes. In such two-electrode systems, a known stimulus voltage, Vst, is applied between RE and WE. The resulting potential across the electrochemical cell’s electrode-electrolyte interface, Vcell, is, ideally, equal to Vst throughout the electrochemical reaction. However, as shown in Figure 3a, in addition to the electrode-electrolyte interface, represented by impedance Zbio, electrochemical cells inherently possess a solution resistance, Rs, representing the ion flow path to RE through the electrolyte solution. The sensing current Isn is supplied by RE. As a result, Vcell is always less than Vst because of the voltage drop, VError, across Rs, which is commonly referred to as the “IR drop” (I = current, R = resistance). Because the IR drop cannot typically be known in advance and can change with environmental conditions, reaction kinetics, etc., it is generally considered to be an undesirable source of noise/error.
Although Rs cannot be entirely removed from an electrochemical cell, the negative effects of IR drop can be eliminated by adding a third electrode, called a counter electrode (CE), to the system. As shown in Figure 3b, in a three-electrode system, Isn is supplied by CE, and in the steady bias state, no current flows through the RE terminal, thus eliminating any IR drop error voltage. As a result, Vcell is equal to Vst, and the reaction potential can be accurately controlled. Because of the performance advantages of eliminating the IR drop effect, three electrode systems are widely used in electrochemical biosensors. Note that potentiometric sensors are generally measured at zero current, and impedimetric sensors generally have small signal AC currents with negligible IR drop. Thus, three-electrode systems are especially useful with amperometric techniques (typically with redox reactions) where larger response currents are generally desired.
Biosensor interfaces with low turnover rates can result in low response currents and low sensitivity. For the redox active bio-recognition element, the redox recycling effect can be utilized to electrochemically magnify the Faradic response current by increasing the mass transport of the redox active species of interest. To utilize redox recycling in biosensors, four-electrode electrochemical systems have been developed [18,40,41]. Four-electrode systems have one RE, one CE and two WEs with the redox recognition element immobilized on both. They are ideally suited for micro-scale interdigitated electrode arrays where the close proximity of WEs enhances redox recycling. Because redox recycling is rarely used and only available with the redox bio-recognition element, four-electrode systems are not very common, and the remainder of this review will focus on popular three-electrode systems.

3. CMOS Instrumentation for Amperometric and Impedimetric Biosensors

After electrodes translate biorecognition events into the electrical domain, the signals can then be measured by electronic instrumentation that also provides the biasing signals to drive electrochemical reactions. In modern microelectronics, CMOS circuits have come to dominate the industry due to their low cost and solid performance, especially in low frequency applications, such as biosensing. Consequentially, most of the electrochemical instrumentation developed recently has utilized CMOS technology, and thus, this review will be limited to CMOS electrochemical instrumentation circuits.
Potentiometric biosensor circuits measure the voltage difference at sensor electrodes. A voltage follower circuit topology is generally used to provide a high-impedance interface. Potentiometric circuit examples can be found in [42,43,44]. Because the circuits for potentiometric sensors are relatively simple and not as commonly used in recent biosensors, this paper will focus on circuits for amperometric and impedimetric biosensors that are currently receiving much attention.
CMOS electrochemical instrumentation for amperometric and impedimetric biosensors can be organized into four functional blocks, as shown in Figure 4: signal generator, potentiostat, readout circuit and signal processing. The signal generator produces the desired stimulus signal shapes, such as constant, triangle, saw and pulse waveforms. The potentiostat controls the biasing potential of the electrodes. The readout circuit amplifies the sensor response signal and can provide front-end signal conditioning. The signal processing block enhances, e.g., filters, data from the readout circuit. Although a low noise signal generator is necessary for high resolution applications, the signal generator and signal processing blocks do not directly connect to the electrode-electrolyte interface and therefore may be implemented by general circuits. Thus, this review will focus on the potentiostat and readout circuit blocks, which are generally most important to sensor performance.

3.1. Potentiostat

In electrochemical sensors, when Vcell reaches the analyte’s redox potential, biochemical reactions occurs, so it is critical to have circuits that accurately set Vcell during sensor response measurement. The circuit block responsible for establishing Vcell on the electrode-electrolyte interface is called a potentiostat.
The first CMOS potentiostat was presented in 1987 [45] and represents the introduction of microelectronics into electrochemical instrumentation, ushering in a new era of miniaturization for high throughput screening, point-of-care diagnostics and implantable device applications. This first circuit was designed for a two-electrode amperometric chemical sensor where a single operational amplifier (OPA) achieves the potentiostat function. Due to the advantages of three-electrode systems compared to two-electrode systems, most of the newer CMOS potentiostats [15,46,47,48,49,50,51,52,53] are designed for three electrodes.
In three-electrode systems, the basic potentiostat configurations are determined by which electrode, WE, RE or CE, is set to the ground reference, often called the analog ground, of the instrumentation circuitry [47]. Configurations that ground RE or CE are electrically equivalent within the potentiostat block leaving the two configurations we will call grounded-WE and grounded-CE. In its simplest form, the grounded-WE potentiostat can be implemented by a single OPA, as shown in Figure 5a. The two inputs of the OPA are connected to stimulus voltage Vst and RE, respectively, and WE is connected to analog ground. Because feedback in the OPA forces the OPA inputs to the same potential, the RE voltage (VRE) is equal to Vst. Thus, the electrochemical cell potential, Vcell, is also equal to the applied Vst, as desired. The output of the OPA connects to CE, and the electrochemical cell provides a path for the sensing current Isn to the grounded WE, where response current is measured by the readout circuitry.
In the grounded-CE potentiostat, sensor response current is collected from CE, which is connected to the ground. Because the potentials at WE and RE can both vary with time, circuitry is required to ensure that Vcell is equal to the applied stimulus, Vst. The typical three-OPA grounded-CE configuration shown in Figure 5b achieves this goal by setting the voltage on WE, VWE, to Vst + VRE. As a result, VRE − VWE = Vcell = Vst, as desired. Grounded-CE is only rarely used and generally with a specially-designed electrochemical transducer [27]. Because the grounded-CE structure is more complex than the grounded-WE, most reported potentiostats use the grounded-WE structure [15,48,49,50,51,52].
The potentiostat structures in Figure 5 are single-ended potentiostats where the trend to reduce CMOS power supply voltages can drastically limit the range of electrochemical potential applied at Vcell. To address this problem in low-voltage CMOS, the fully-differential potentiostat shown in Figure 6 was developed to increase the potentiostat output swing [46]. This circuit centers around a fully-differential amplifier, OP3, that dynamically controls the difference between VWE and VRE to be equal to 2Vst, effectively doubling the Vcell output range compared to a single-ended potentiostat. The fully-differential potentiostat topology is very useful for increasing the available Vcell potential window when an advanced microelectronics process technology with low supply voltages is desired to reduce power consumption or improve speed.

3.2. DC Current Readout Circuitry

In response to a stimulus signal provided by the potentiostat, electrochemical biosensors produce an output signal that is measured by a readout circuit. Because the potentiometric technique is rarely used in biosensors, only readout circuits that measure current responses from amperometric and impedimetric biosensors will be discussed in this paper. From an instrumentation point of view, current readout circuits can be classified as either DC or AC, based on the electrochemical method employed. This section will review DC current readout circuits, while AC current readout circuits are discussed in the next section.
As defined in Section 2, amperometric biosensors are stimulated by a DC voltage, either constant or varying slowly with time, and generate a DC response current. Thus, amperometric biosensors need DC current readout circuits. In general, the amperometric readout circuit should provide a wide current range, high resolution within the bandwidth of interest and allow the measurement of bidirectional currents (sink/source). Numerous CMOS amperometric circuits have been introduced over the past few decades with varying functionality and performance to meet different application demands [15,45,50,51,53,54,55,56,57,58,59,60,61,62,63]. After years of development, the most common structures for CMOS amperometric current readout circuits are resistive, capacitive feedback transimpedance amplifiers (TIA) and the current conveyor. A performance summary of amperometric readout circuits is provided in Table 1.

3.2.1. Resistive Feedback Current Readout Circuits

The simplest way to measure a DC response current is to translate the current to voltage through a resistor. As shown in Figure 7, a basic resistive feedback current readout circuit uses an operational amplifier with a feedback resistor Rf to convert the biosensor sensing current Isn to a voltage. The sensing current Isn can be calculated by:
I sn = V out R f
where Vout is the output voltage of the operational amplifier. The dynamic range of the current readout circuitry can be adjusted by changing the Rf value. In order to decrease the input referenced current noise, the feedback resistor Rf should be large. It has been shown that a readout circuit with resistive feedback can provide a 4-pA noise performance at 10 kHz [61]. However, due to chip area constraints, on-chip Rf cannot be very large, and this limits gain. An active feedback structure has been used to decrease the feedback resistor area while achieving pA level noise and a 1-MHz bandwidth using an off-chip filter [62,63]. It has been shown that resistive feedback achieves better noise performance in the high bandwidth than capacitive feedback [64]. This combination of low noise and high bandwidth enables a powerful new platform for biotechnology applications, such as DNA sequencing and ion channel recording.

3.2.2. Capacitive Feedback Current Readout Circuits

A drawback of resistive feedback is the thermal noise generated by the feedback resistor, which limits readout resolution. To eliminate this noise source, many current readout circuits with capacitive feedback have been introduced [15,56,57,71,73,74,75,76,77]. It has been shown that, in the low bandwidth, capacitive feedback achieves better noise performance than resistive feedback [64]. The basic capacitive feedback structure shown in Figure 8 is an integrating amplifier that converts the input current to a voltage [74]. The biosensor sensing current Isn can be calculated by:
V out = 1 C f 0 T I sn dt
where Cf is the feedback capacitor, T is the integration time and Vout is the output voltage of the operational amplifier.
The resolution of a capacitive feedback current readout circuit is limited by noise generated in the feedback switch. These noise sources include clock feedthrough, charge injection and KT/C noise and presents as input offset errors. To eliminate this offset error, the correlated double sampling (CDS) technique has been applied to the capacitive feedback readout circuit [15]. The CDS technique can also reduce 1/f noise, which is a significant noise source in DC readout circuitry. The CDS concept involves sampling an input twice and recording the difference of these two samples to reject slow-changing noise/interference. An example CDS circuit is shown in Figure 9, where the two sampling actions are implemented by a set of switches controlled by two inverted phase clocks (ph1 and ph2) [54]. Note that the amplifier should be designed to minimize thermal noise rather than flicker noise because CDS technology can double the thermal noise [61]. Noise aliasing due to sampling can also affect the noise level in CDS circuits [78,79,80]. A current resolution of 10 pA has been achieved using the capacitive feedback structure with CDS technology [54,81].
To reduce the effects of charging currents and currents due to interfering analytes, pseudo-differential capacitive current readout circuits have been developed [48,82]. A schematic of the pseudo-differential readout is presented in Figure 10, where the reading time is controlled by switches Sn and Sp. By using two WEs, the voltage produced by currents from WEn and WEp can be subtracted by a differential amplifier stage (not shown), eliminating nearly all parasitic currents while preserving the analytical signal of interest [48].

3.2.3. Current Conveyor Current Readout Circuits

The third type of amperometric readout circuit that has been widely reported [47,57,66,69,83,84,85,86] is the current conveyor structure, which uses a front-end current mode amplifier to eliminate noise from the following stages. Figure 11 depicts a typical current conveyor schematic for which the measured biosensor current Isn can be calculated by:
I sn I sn * = ( W / L ) 1 ( W / L ) 2
where I*sn is the current conveyor output current and W/L is size of current mirror transistors M1 and M2. The current conveyor circuit performs decoupling and linear operations in current mode in the same way that an operational amplifier performs them in voltage mode. Because no capacitor or resistor is needed for feedback, current conveyors can be very compact [86]. The amplifier in Figure 11 does not need to provide a large output current, allowing the total power consumption of a current conveyor to be very low. Because of these benefits, a current conveyor structure was used in a high throughput biosensor array. Notice that the current conveyor in Figure 11 only measures Isn in one direction. To measure a bidirectional Isn, offset current sources or regulated-cascode current mirrors could be employed [66,69].

3.2.4. Digitization Circuits for DC Current Readouts

To capture and process sensor signals with digital circuitry, digitization of analog signals is important for features in many biosensor systems permitting efficient analysis and storage of sensor data. Many customized analog-to-digital converter (ADC) structures have been used with amperometric sensors, including sigma-delta (Σ∆) ADCs [56,67,70,71,87], integration ADCs [50,53,57,66,88], current-to-frequency (I-F) ADCs [47,89], successive approximation (SAR) ADC [72] and hybrid topologies [68]. Among these circuits, the incremental first-order Σ∆ ADC structures tend to have the best sensitivity (minimum linear response) of any class of ADC while remaining compact and consuming only microwatts of power. This ADC structure combines the benefits of integration ADCs with the noise shaping of Σ∆ ADCs and achieves superior performance for instrumentation and measurement applications [72]. The structure of a current-mode incremental first-order Σ∆ ADC is shown in Figure 12, where two current sources are used as reference currents in two directions to support a bidirectional input current. Many other efforts to improve Σ∆ ADC performance for amperometric sensors have been reported, including: replacing the traditional comparator with a hysteretic comparator to reduce substrate noise and achieve 50-fA sensitivity [56], using a digital modulation stage to expand the input current range by 32× without significant compromise to power and area [71,87] and reusing sensor capacitance as the amplifier feedback Cf to reduce the overall area [70]. Other ADC topologies, such as integration ADC, SAR ADC and I-F ADC, have traditional structures found in textbooks, but have seen limited use in CMOS biosensors because their noise performance is not as good as the incremental Σ∆ ADC.

3.3. AC Current Readout Circuitry

As defined in Section 2, impedimetric sensors are stimulated by a sinusoidal (AC) voltage of known frequency and produce an AC current response that must be measured by readout circuitry. Although materials are often characterized in analytical studies using electrochemical impedance spectroscopy (EIS) over a wide range of sinusoidal stimulus frequencies, in sensing applications, it is more common to use a single frequency or small subset of discrete frequencies where the bio-recognition element has demonstrated a reliable response. Impedimetric sensors are typically modeled by a complex impedance with resistive and capacitive components that can vary with biorecognition events. A typical electrochemical impedimetric sensor interface can be represented by a Randles model circuit, as shown in Figure 13a [18], with model elements defined in the caption. Biorecognition events typically manifest as a phase and/or amplitude shift in the output response relative to the stimulus signal. From this phase-amplitude response, real and imaginary components can be algebraically calculated to extract the sensing parameters of interest from the sensor’s impedance model. If real-time operation or system size is not an application constraint, sensor component extraction can be performed offline. However, several reports have demonstrated that complex impedance extraction can be efficiently implemented within the CMOS readout circuit [16,90,91,92,93,94,95], as highlighted in this section.

3.3.1. Impedance Extraction Algorithms

The real and imaginary components of impedimetric sensors can be extracted in either the digital or analog domain. In the digital domain, the common extraction method is the fast Fourier transform (FFT) algorithm [24]. The FFT algorithm utilizes a broadband stimulus, such as a pulse, and computes the result at all frequency points simultaneously, producing an impedance spectrum. The composite signal source and the computationally-intensive FFT require digital EIS instrumentation to use a digital signal processor (DSP) with extensive computational resources [96]. In the analog domain, real and imaginary components are typically extracted using the frequency response analyzer (FRA) method. FRA processes the response of one frequency point at a time and sweeps over the entire frequency range of interest. Compared to FFT, the FRA method can be realized with simple and compact analog circuits, making it suitable for sensor array microsystems [96]. Therefore, most of the CMOS impedimetric current readout circuits use the FRA method [90,91,92,93,94,97,98].
The operating principle of the FRA method is shown in Figure 13b where a known frequency sinusoid stimulus voltage Asin(ωt) is applied to the electrochemical transducer. The transducer’s response current is defined by A’sin(ωt + θ), where A’ and θ are determined by the transducer’s admittance. By multiplying the response current with two reference signals and then applying low pass filters (LPF), two orthogonal components, real (REAL) and imaginary (IMG), can be extracted, and these signal components are proportional to the real and imaginary components of the transducer’s admittance.
Phasor domain analysis for the transducer response current A’θ in Figure 13, gives:
A θ [ A sin ( ω t + θ ) ] = A 0 0 | Y | θ
where A’sin(ωt + θ) is the time domain representation of A’∠θ, A’∠0 is the stimulus signal and |Y|∠θ is the phasor format of the transducer admittance. From Equation (4), it can be seen that |Y| = A’/A.
The multiplication operations of the FRA circuit give:
{ A sin ( ω t + θ ) sin ( ω t ) = A 2 cos ( θ ) DC A 2 cos ( 2 ω t + θ ) AC A sin ( ω t + θ ) cos ( ω t ) = A 2 sin ( θ ) DC + A 2 sin ( 2 ω t + θ ) AC
where DC and AC labels represent DC component and AC component, respectively. The low pass filters (LPF) in Figure 13 remove the AC components in Equation (5) and generate:
{ REAL = A 2 cos ( θ ) IMG = A 2 sin ( θ )
The transducer’s admittance | Y | θ can be calculated by:
{ | Y | = A A = REAL 2 + IMG 2 2 A θ = tan 1 REAL IMG
To simplify signal generator design, the sinusoidal reference signals shown in Figure 13 can be replaced by rectangular waveforms with the same phase of sin(ωt) and cos(ωt) [16]. The higher odd harmonics caused by the rectangular waveform can be eliminated by the LPF.

3.3.2. Impedance Extraction Algorithm Implementation

Several EIS readout circuits have been developed based on the lock-in technique [16,90,91]. A novel structure combining the sigma-delta technique and a lock-in multiplier to support impedance extraction and digitization has been reported [16]. This circuit, shown in Figure 14, shares resources for impedance extraction and digitization to maximize the hardware efficiency, which permits over 100 readout channels within a 3 mm × 3 mm die. The real and imaginary portions are extracted by digital multipliers, which are inherently integrated into the ADC’s integrator, and the high frequency noise is removed by the integrator.
Another impedance extraction circuit applies a transimpedance amplifier to convert the input current into voltage and then extracts impedance using the lock-in technique [92]. This chip is capable of concurrently measuring admittance values as small as 10−8 Ω−1 with a detection dynamic range of more than 90 dB in the frequency range of 10 Hz–50 MHz [92]. An array of on-chip gold electrodes has been implemented on this chip and used to measure the impedance of the associated electrode-electrolyte interface [92,93].
To address the bandwidth limitation of the circuit in [16], the impedance spectroscopy DNA analyzer with dual-slope multiplying ADC shown in Figure 15 has been reported [94]. This EIS circuit can extract the analyte impedance in one interrogation cycle. It also includes an on-chip signal generator to produce stimulus waveforms for EIS measurement, albeit with larger area and power consumption.
A common limitation of the impedance extraction circuits discussed above is that they cannot support high frequency stimulus; the analog-multiplier-based circuit [92,93] only supports frequencies up to 50 MHz, and the lock-in ADC [16] and multiplying dual-slope ADC [94] only support frequencies below 10 kHz. To address this issue, a fully-integrated impedance extractor has been reported that can support stimulus frequencies from 20 kHz to 150 MHz by using a down/up conversion architecture [95]. However, this circuit consumes a large amount power, occupies a large area and does not have digital output. The performance of several impedimetric readout circuits is summarized in Table 2.

4. CMOS Biosensor Microsystem Integration

The integration of biosensor interfaces and CMOS detection circuitry has many benefits, including miniaturization, automated detection, high performance and reduced reagent costs. To capitalize on these benefits, a recent trend in CMOS biosensors has been to integrate electrochemical biosensors monolithically with readout circuits to form on-chip measurement microsystems. The interior CMOS high density electrical metal routings can be used to connect electrode arrays to instrumentation circuits, greatly reducing the impact of environmental noise sources and allowing electrical signal connections of high density arrays without the wiring complexity of hybrid off-chip sensors. However, the materials and fabrication processes of CMOS electronics have been developed and selected independently from those of biosensors, leading to a contrast of requirements and compatibility that challenge CMOS biosensor microsystem integration. For example, operation in aqueous biosensing environments is not a requirement for the development of the CMOS electronics. Based on integration efforts to date [99,100,101,102,103,104,105,106,107,108], the components involved in CMOS biosensor microsystem integration are illustrated in Figure 16, where the CMOS chip acts as both the electrochemical instrumentation and the sensor’s physical substrate. Post-CMOS processed sensor electrodes and bioprobes are directly attached to the surface of the CMOS chip and connected to the underlying electronics through thin film planar metal routings at pre-design electrical contact openings. As liquid environments are inherent to biosensors, this integrated sensor requires the interface with fluid handling ranging from simply creating a reservoir over sensors [15] to integrating microfluidic channels to deliver fluids [99,102,103,109]. In addition, packaging is also required that can protect active electronics from the corrosive biosensing liquids. Specifically, the packaging should protect the integrated sensor from potential hazards of chemical corrosion of reactive aluminum CMOS contact pads and electrical shorts through conductive electrolyte fluids. Finally, the entire assembly requires both electrical and fluidic connection to the outside world. Having covered the CMOS electronics above, this section will review efforts to form on-chip biosensor electrodes and implement chip-scale packaging for liquid environments, which are two of the major challenges in CMOS biosensor microsystem integration.

4.1. On-Chip Electrodes

In a three-electrode electrochemical system, WE, CE and RE can be implemented with different materials to suit their role in the measurement process. The WE is where bio-probes are immobilized and reactions takes place. Gold or carbon paste is usually used to attach bio-probes through covalent bonding of functional chemical groups of bio-linkers. Platinum is often chosen due to its catalytic role in hydrogen peroxide detection. In recent work, nanomaterials have been conjugated with bio-probes either covalently or non-covalently to amplify the bio-signal. The CE provides a path for reaction current and is expected to be inert during the reaction. The CE is generally much larger than the WE, so that the half reaction at CE can provide sufficient current and does not limit the reaction process at WE. CE is usually made of carbon or a noble metal like gold or platinum. For the RE, whose potential should be stable, independent of the composition of other ions in the analyte and reproducible even after a small electric current flow, a Ag/AgCl, Cl electrode is preferable. Generally, a Ag rod is coated by barely soluble AgCl and surrounded by a Cl anion-containing electrolyte reservoir made of liquid-tight coating with a diaphragm or porous membrane. Pseudo-reference electrodes are sometimes used, where only the metal rod is present without the salt or oxide coating and electrolyte reservoir. Use of a pseudo-reference electrode requires specific operation conditions to calculate its potential. In the development of the CMOS-based electrochemical system, to utilize existing mass production techniques driven by semiconductor manufacturing, the electrodes should be implemented in a planar geometry.
Fabricating biosensor electrodes onto the surface of a CMOS chip requires considerations of process compatibility between CMOS functionality and biosensor needs. Furthermore, in low-volume die-level post-CMOS processing, the small size of the CMOS chip (compared to a full wafer) introduces new challenges. For the WE, gold has been electroplated on the aluminum to make an inert electrode, which has affinity for biotargets [110,111]. Alternatively, the top CMOS aluminum layer can be etched away and replaced by another metal electrode material by electroplating, and Pt-black electrodes have been electroplated after aluminum etching for neural recording and stimulation [101]. This post-CMOS process needed no alignment; but the electroplated electrodes have rough surfaces, and the uniform thickness among all of the electrodes was difficult to control. In other work, photolithography-defined planar gold electrodes were deposited on the CMOS surface and connected to circuitry through pre-designed surface contact pads [50,65,112,113]. An example of on-CMOS planar electrodes is shown in Figure 17. These electrodes were flat and uniform, and the geometries of the electrodes could be varied as long as the surface aluminum contacts were properly located on the CMOS. Alternatively, Ti/TiN and tungsten via have been deposited before gold electrode deposition [44]. These gold electrodes are accurately defined, and the surface is smoother than electroplated electrodes.
In contrast to the successes of WE formation on CMOS, the design of a quality planar RE on CMOS for amperometric and potentiometric sensors still remains an open challenge (note that impedimetric sensors typically do not rely on a true reference potential and can accommodate a pseudo-reference electrode). A planar sieve-printed Ag/AgCl quasi-reference electrode on an electrochemical multi-parameter lab-on-chip configuration has been presented on an aluminum oxide ceramic substrate [114]. For a full reference electrode, thin film Ag can be deposited followed by electrochemical chloridization or coverage of sieve-printed AgCl layer. For a long lifetime, a large volume of, e.g., KCl solution should be placed in close proximity in a very limited volume. A proper charge carrier exchange should be established between the solution of the planar reference electrode and the analyte of interest to ensure low resistance of the electrode. Ideally, the loss of KCl and AgCl can be minimized without contamination of the reference element by substances in the solution. To achieve these goals, e.g., hydrogels like agarose have been used to localize KCl and limit the diffusion of the enclosed chloride [115]. The contact between the reference element and the solution could be realized with heterogeneous barriers, which means a hydrophobic layer is facing the outer analyte solution, while a hydrophilic layer is facing inward to reduce contamination [116,117,118]. In other work, a channel-type planar reference electrode has been constructed with a prolonged ion transport path [119,120]. Furthermore, simple integrated glass fibers or filter materials have been used to establish contact to the solution [121]. The surface of the AgCl film has been insulated with a water-repellent layer to slow the loss of chloride [122]. A solid-state reference electrode for an electrochemical system on a microfluidic chip has been reported using electrodeposited nanoporous platinum as the reference electrode [123]. Another solid-state reference electrode was developed utilizing layer-by-layer polymer coating [124].

4.2. Chip-Scale Packaging

After the electrodes are fabricated on the CMOS chip, packaging is required before on-chip electrochemical measurements can be performed within a liquid environment. The CMOS chips are most typically wire bonded in a ceramic package, and this scheme has been adapted to perform on-chip measurement by encapsulating the flexible and soft bonding wires to avoid electrical shorts and prevent corrosion caused by electrolyte solutions. These early efforts at integration demonstrate that CMOS biosensor packaging must simultaneously consider both electrical connection of CMOS I/O signals and proper encapsulation of the chip for in-liquid measurements.
In wire bond-based CMOS biosensors, the main goal of encapsulation is to insulate the bonding wires. Recently, many packaging schemes were reviewed and compared with parameters like area efficiency, lifetime, barrier, number of steps, vertical step height, etc. [103]. The packaging material is important because it affects the fabrication process and quality of packaging. Polymer materials have been considered to seal bond wires. PDMS (polydimethylsiloxane), an elastomeric polymer, has been deposited and cured around the CMOS chip to insulate bonding wires [106]. PDMS is widely used as a building material for microfluidic structures and is bio-compatible and inert to most chemicals. Various types of epoxy have also been used to seal bond wires [101,105,125,126,127]. The epoxy was applied around the CMOS chip leaving the chip’s surface exposed. Since epoxy is a liquid material, application of epoxy on bond wires while keeping it off the chip’s surface can be challenging. Epoxies are available in different viscosities, and a proper viscosity can be selected to encapsulate wires with better control. Once cured, epoxy is a hard material compared to PDMS. However, epoxy encapsulation has a reliability issues due to its poor adhesion to the chip substrate and stress imposed on wire bonds. In certain circumstances, the absorption of water can cause cracks in the epoxy and failure of the encapsulation. Another approach was reported using parylene as the encapsulation material [128]. As a thin film polymer material deposited in a low temperature process, parylene can uniformly coat all surfaces and is resistive to chemicals. To expose the electrode area, a micromachining laser source can be used to ablate the parylene during patterning. However, the energy of the laser source is hard to control so that it does not damage the sensing region underneath. Alternatively, an ultrasonic bath can be used to pattern parylene via lift-off, but this could compromise the seal around wire bonds.
Overall, PDMS, epoxy and parylene have shown some degree of success in encapsulation, and selection between them depends on application-specific requirements. PDMS and epoxy are readily available, low cost polymers that can be rapidly formed for encapsulation, but existing techniques to work with these materials lack precision, which can challenge working with smaller chips. On the other hand, parylene can be precisely patterned using a laser source, but requires relatively expensive deposition and laser tools, and it does not provide the mechanical strength of PDMS or epoxy.
Recently, to support the integration of CMOS biosensors with microfluidics, an innovative approach called “lab-on-CMOS” was reported using planar metal interconnects and a silicon carrier device [102]. As shown in Figure 18, planar metal interconnects are fabricated on the chip-carrier surface to route CMOS signals to the carrier edge, and finally, a multichannel microfluidic device is attached to the surface. Inspired by chip packaging in wafer level processes [129,130,131,132], the CMOS chip was embedded in a supporting carrier to extend the surface of the chip, so that the increased area could allow the integration of much larger microfluidics. Electrical connections to the chip were realized by planar metal traces deposited by physical vapor deposition. Variations of this approach have recently been reported [103,112]. The lab-on-CMOS platform offers enormous opportunities for integrating CMOS ICs with more complicated multichannel microfluidic structures.

5. Conclusions

This paper describes electrochemical biosensors from an instrumentation point of view and reviews the design and performance of CMOS circuits reported for such sensors over the past three decades. This review focuses on two important types of electrochemical biosensors: amperometric and impedimetric, which are classified by their DC and AC responses, respectively. Basic theory and circuit structures are provided; relevant literature is summarized; and performance metrics are reported and discussed. Because the trend in CMOS biosensors is moving toward chip-scale integration, this paper also reviews the fabrication techniques and challenges in the formation of on-chip electrochemical electrodes and on-CMOS biosensor packaging to enable in-liquid measurements. CMOS electrochemical biosensors have a promising future in many applications, and this review paper serves as a guide to future designers who will evolve these sensors even further.

Acknowledgments

This work is supported by National Science Foundation Grant ECCS-1307939, the National Institute for Occupational Safety and Health (NIOSH) Grant R01OH009644 and the National Institute of Health (NIH) Grants R01ES022302 and R01AI113257.

Conflicts of Interest

The authors declare no conflict of interest. The founding sponsors had no role in the design of the study; in the collection, analyses, or interpretation of data; in the writing of the manuscript, and in the decision to publish the results.

References

  1. Global Industry Analysis Inc. The Global Biosensors Market; Global Industry Analysis Inc.: San Jose, CA, USA, 2014. [Google Scholar]
  2. Mohanty, S.P.; Kougianos, E. Biosensors: A tutorial review. IEEE Potentials 2006, 25, 35–40. [Google Scholar] [CrossRef]
  3. Jang, B.; Hassibi, A. Biosensor systems in standard CMOS processes: Fact or fiction? IEEE Trans. Ind. Electr. 2009, 56, 979–985. [Google Scholar] [CrossRef]
  4. Cai, W.; Peck, J.R.; van der Weide, D.W.; Hamers, R.J. Direct electrical detection of hybridization at DNA-modified silicon surfaces. Biosens. Bioelectron. 2004, 19, 1013–1019. [Google Scholar] [CrossRef] [PubMed]
  5. Gua, T.; Hasebeb, Y. Novel amperometric assay for drug–DNA interaction based on an inhibitory effect on an electrocatalytic activity of DNA–Cu(II) complex. Biosens. Bioelectron. 2012, 33, 222–227. [Google Scholar] [CrossRef] [PubMed]
  6. Chiti, G.; Marrazza, G.; Mascini, M. Electrochemical DNA biosensor for environmental monitoring. Anal. Chim. Acta 2001, 427, 155–164. [Google Scholar] [CrossRef]
  7. Ponzio, F.; Jonsson, G. A rapid and simple method for the determination of picogram levels of serotonin in brain tissue using liquid chromatography with electrochemical detection. J. Neurochem. 2006, 32, 129–132. [Google Scholar] [CrossRef]
  8. Peters, J.L.; Miner, L.H.; Michael, A.C.; Sesack, S.R. Ultrastructure of carbon fiber microelectrode implantation sites after acute voltammetric measurements in the striatum of anesthetized rats. J. Neurosci. Methods 2004, 137, 9–23. [Google Scholar] [CrossRef] [PubMed]
  9. Setterington, E.H.; Alocilja, E.C. Rapid electrochemical detection of polyaniline-labeled Escherichia coli O157:H7. Biosens. Bioelectron. 2001, 26, 2208–2212. [Google Scholar] [CrossRef] [PubMed]
  10. Maalouf, R.; Fournier-Wirth, C.; Coste, J.; Chebib, H.; Saïkali, Y.; Vittori, O.; Errachid, A.; Cloarec, J.P.; Martelet, C.; Jaffrezic-Renault, N. Label-Free Detection of Bacteria by Electrochemical Impedance Spectroscopy: Comparison to Surface Plasmon Resonance. Anal. Chem. 2007, 79, 4879–4886. [Google Scholar] [CrossRef] [PubMed]
  11. Ruan, C.; Yang, L.; Li, Y. Immunobiosensor Chips for Detection of Escherichia coli O157:H7 Using Electrochemical Impedance Spectroscopy. Anal. Chem. 2002, 74, 4814–4820. [Google Scholar] [CrossRef] [PubMed]
  12. Kotanen, C.N.; Moussy, F.G.; Carrara, S.; Guiseppi-Elie, A. Implantable enzyme amperometric biosensors. Biosens. Bioelectron. 2012, 35, 14–26. [Google Scholar] [CrossRef] [PubMed]
  13. Hrapovic, S.; Luong, J.H.T. Picoamperometric detection of glucose at ultrasmall platinum-based biosensors: Preparation and characterization. Anal. Chem. 2003, 75, 3308–3315. [Google Scholar] [CrossRef] [PubMed]
  14. Wu, B.; Hou, S.; Yin, F.; Zhao, Z.; Wang, Y.; Wang, X.; Chen, Q. Amperometric glucose biosensor based on multilayer films via layer-by-layer self-assembly of multi-wall carbon nanotubes, gold nanoparticles and glucose oxidase on the Pt electrode. Biosens. Bioelectron. 2007, 22, 2854–2860. [Google Scholar] [CrossRef] [PubMed]
  15. Li, L.; Liu, X.; Qureshi, W.A.; Mason, A.J. CMOS amperometric instrumentation and packaging for biosensor array applications. IEEE Trans. Biomed. Circuits Syst. 2011, 5, 439–448. [Google Scholar] [CrossRef] [PubMed]
  16. Yang, C.; Jadhav, S.R.; Worden, R.M.; Mason, A.J. Compact low-power impedance-to-digital converter for sensor array microsystems. IEEE J. Solid-State Circuits 2009, 44, 2844–2855. [Google Scholar] [CrossRef]
  17. Thevenot, D.R.; Toth, K.; Durst, R.A.; Wilsond, G.S. Electrochemical Biosensors: Recommended Definitions and Classification. Biosens. Bioelectron. 2001, 16, 121–131. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  18. Bard, A.J.; Faulkner, L.R. Electrochemical Methods: Fundamentals and Applications; Wiley: New York, NY, USA, 2000. [Google Scholar]
  19. Korotcenkov, G. Chemical Sensors, Vol 1: General Approaches (Sensor Technology); Momentum Press: New York, NY, USA, 2010. [Google Scholar]
  20. Komori, H.; Niitsu, K.; Tanaka, J.; Ishige, Y.; Kamahori, M.; Nakazato, K. An extended-gate CMOS sensor array with enzyme-immobilized microbeads for redox-potential glucose detection. In Proceedings of the 2014 IEEE Biomedical Circuits and Systems Conference (BioCAS), Lausanne, Switzerland, 22–24 October 2014; pp. 464–467.
  21. Stefan, R.I.; van Staden, J.F.; Aboul-Enein, H.Y. Immunosensors in clinical analysis. Fresenius J. Anal. Chem. 2000, 366, 659–668. [Google Scholar] [CrossRef] [PubMed]
  22. Wang, J. Amperometric Biosensors for Clinical and Therapeutic Drug Monitoring: A Review. J. Pharm. Biomed. Anal. 1999, 19, 47–53. [Google Scholar] [CrossRef]
  23. Palchettia, I.; Cagnini, A.; Del Carlo, M.; Coppi, C.; Mascini, M.; Turner, A.P.F. Determination of anticholinesterase pesticides in real samples using a disposable biosensor. Anal. Chim. Acta 1997, 337, 315–321. [Google Scholar] [CrossRef]
  24. Barsoukov, E.; Macdonald, J. Impedance Spectroscopy: Theory, Experiment, and Applications, 2nd ed.; John Wiley and Sons: Hoboken, NJ, USA, 2005. [Google Scholar]
  25. Stelzle, M.; Weissmuller, G.; Sackmann, E. On the application of supported bilayers as receptive layers for biosensors with electrical detection. J. Phys. Chem. 1993, 97, 2974–2981. [Google Scholar] [CrossRef]
  26. Yin, F. A novel capacitive sensor based on human serum albumin-chelant complex as heavy metal ions chelating proteins. Anal. Lett. 2004, 37, 1269–1284. [Google Scholar] [CrossRef]
  27. Bontidean, I.; Berggren, C.; Johansson, G.; Csoregi, E.; Mattiasson, B.; Lloyd, J.A.; Jakeman, K.J.; Brown, N.L. Detection of heavy metal ions at femtomolar levels using protein-based biosensors. Anal. Chem. 1998, 70, 4162–4169. [Google Scholar] [CrossRef] [PubMed]
  28. Yang, L.J.; Li, Y.B.; Griffis, C.L.; Johnson, M.G. Interdigitated microelectrode (IME) impedance sensor for the detection of viable salmonella typhimurium. Biosens. Bioelectron. 2004, 19, 1139–1147. [Google Scholar] [CrossRef] [PubMed]
  29. Mishra, N.N.; Retterer, S.; Zieziulewicz, T.J.; Isaacson, M.; Szarowski, D.; Mousseau, D.E.; Lawrence, D.A.; Turner, J.N. On-chip micro-biosensor for the detection of human CD4(+) cells based on AC impedance and optical analysis. Biosens. Bioelectron. 2005, 21, 696–704. [Google Scholar] [CrossRef] [PubMed]
  30. Lee, C.-S.; Kim, S.K.; Kim, M. Ion-Sensitive Field-Effect Transistor for Biological Sensing. Sensors 2009, 9, 7111–7131. [Google Scholar] [CrossRef] [PubMed]
  31. Lee, S.-R.; Sawada, K.; Takao, H.; Ishida, M. An enhanced glucose biosensor using charge transfer techniques. Biosens. Bioelectron. 2008, 24, 650–656. [Google Scholar] [CrossRef] [PubMed]
  32. Kharitonov, A.B.; Wasserman, J.; Katz, E.; Willner, I. The Use of Impedance Spectroscopy for the Characterization of Protein-Modified ISFET Devices: Application of the Method for the Analysis of Biorecognition Processes. J. Phys. Chem. B 2001, 105, 4205–4213. [Google Scholar] [CrossRef]
  33. Purushothaman, S.; Toumazou, C.; Ou, C. Protons and single nucleotide polymorphism detection: A simple use for the ion sensitive field effect transistor. Sens. Actuators B Chem. 2006, 114, 964–968. [Google Scholar] [CrossRef]
  34. Kalofonou, M.; Georgiou, P.; Ou, C.P.; Toumazou, C. An ISFET based translinear sensor for DNA methylation detection. Sens. Actuators B Chem. 2012, 161, 156–162. [Google Scholar] [CrossRef]
  35. Rothberg, J.M.; Hinz, W.; Rearick, T.M.; Schultz, J.; Mileski, W.; Davey, M.; Leamon, J.H.; Johnson, K.; Milgrew, M.J.; Edwards, M. An integrated semiconductor device enabling non-optical genome sequencing. Nature 2011, 475, 348–352. [Google Scholar] [CrossRef] [PubMed]
  36. Wang, G.; Wang, L.; Han, Y.; Zhou, S.; Guan, X. Nanopore Stochastic Detection: Diversity, Sensitivity, and Beyond. Acc. Chem. Res. 2013, 46, 2867–2877. [Google Scholar] [CrossRef] [PubMed]
  37. Ravezzi, L.; Conci, P. ISFET sensor coupled with CMOS read-out circuit microsystem. Electron. Lett. 1998, 34, 2234–2235. [Google Scholar] [CrossRef]
  38. Lee, I.; Lee, S.-W.; Lee, K.-Y.; Park, C.; Kim, D.; Lee, J.-S.; Yi, H.; Kim, B. A Reconfigurable and Portable Highly Sensitive Biosensor Platform for ISFET and Enzyme-Based Sensors. IEEE Sens. J. 2016, 16, 4443–4451. [Google Scholar] [CrossRef]
  39. Janata, J. Historical review. Twenty years of ion-selective field-effect transistors. Analyst 1994, 119, 2275–2278. [Google Scholar] [CrossRef]
  40. Zhu, X.; Ahn, C. On-chip Electrochemical Analysis System Using Nanoelectrodes and Bioelectronic CMOS Chip. IEEE J. Sens. 2006, 6, 1280–1286. [Google Scholar] [CrossRef]
  41. Huang, Y.; Mason, A.J. A redox-enzyme-based electrochemical biosensor with a CMOS integrated bipotentiostat. In Proceedings of the Biomedical Circuits and Systems Conference, BioCAS 2009, Beijing, China, 26–28 November 2009; pp. 29–32.
  42. Das, A.; Bhadri, P.; Beyette, F.R.; Jang, A.; Bishop, P.; Timmons, W. A potentiometric sensor system with integrated circuitry for in situ environmental monitoring. In Proceedings of the 2006 Sixth IEEE Conference on Nanotechnology, Cincinnati, OH, USA, 17–20 July 2006; pp. 917–920.
  43. Sun, A.; Venkatesh, A.; Hall, D.A. A Multi-Technique Reconfigurable Electrochemical Biosensor: Enabling Personal Health Monitoring in Mobile Devices. IEEE Trans. Biomed. Circuits Syst. 2016, 10, 945–954. [Google Scholar] [CrossRef]
  44. Thewes, R.; Hofmann, F.; Frey, A.; Holzapfl, B.; Schienle, M.; Paulus, C.; Schindler, P.; Eckstein, G.; Kassel, C.; Stanzel, M.; et al. Sensor arrays for fully electronic DNA detection on CMOS. In Proceedings of the 2002 IEEE International Solid-State Circuits Conference, San Francisco, CA, USA, 7 February 2002.
  45. Turner, R.F.B.; Harrison, D.J.; Baltes, H.P. A CMOS Potentiostat for Amperometric Chemical Sensors. IEEE J. Solid-State Circuits 1987, 22, 473–478. [Google Scholar] [CrossRef]
  46. Martin, S.M.; Gebara, F.H.; Strong, T.D.; Brown, R.B. A fully differential potentiostat. IEEE Sens. J. 2009, 9, 135–142. [Google Scholar] [CrossRef]
  47. Ahmadi, M.M.; Jullien, G.A. Current-mirror-based potentiostats for three-electrode amperometric electrochemical sensors. IEEE Trans. Circuits Syst. I Regul. Pap. 2009, 56, 1339–1348. [Google Scholar] [CrossRef]
  48. Martin, S.M.; Gebara, F.H.; Larivee, B.J.; Brown, R.B. A CMOS-Integrated Microinstrument for Trace Detection of Heavy Metals. IEEE J. Solid-State Circuits 2005, 40, 2777–2786. [Google Scholar] [CrossRef]
  49. Lauwers, E.; Suls, J.; Gumbrecht, W.; Maes, D.; Gielen, G.; Sansen, W. A CMOS Multiparameter Biochemical Microsensor with Temperature Control and Signal Interfacing. IEEE J. Solid-State Circuit 2001, 36, 2030–2038. [Google Scholar] [CrossRef]
  50. Levine, P.M.; Gong, P.; Levicky, R.; Shepard, K.L. Active CMOS sensor array for electrochemical biomolecular detection. IEEE J. Solid-State Circuits 2008, 43, 1859–1871. [Google Scholar] [CrossRef]
  51. Martin, S.M.; Gebara, F.H.; Strong, T.D.; Brown, R.B. A Low-Voltage, Chemical Sensor Interface for Systems-On-Chip: The Fully-Differential Potentiostat. In Proceedings of the 2004 International Symposium on Circuits and Systems, Vancouver, BC, Canada, 23–26 May 2004; Volume 4, pp. 892–895.
  52. Carminati, M.; Ferrari, G.; Guagliardo, F.; Farina, M.; Sampietro, M. Low-Noise Single-Chip Potentiostat for Nano-Bio-Electrochemistry over a 1MHz Bandwidth. In Proceedings of the 16th IEEE International Conference on Electronics, Circuits, and Systems, Hammamet, Tunisia, 13–16 December 2009; pp. 952–955.
  53. Reay, R.J.; Kounaves, S.P.; Kovacs, G.T.A. An integrated CMOS potentiostat for miniaturized electroanalytical instrumentation. In Proceedings of the IEEE International Solid-State Circuits Conference, San Francisco, CA, USA, 16–18 February 1994; pp. 162–163.
  54. Yang, C.; Huang, Y.; Hassler, B.L.; Worden, R.M.; Mason, A.J. Amperometric electrochemical microsystem for a miniaturized protein biosensor array. IEEE Trans. Biomed. Circuits Syst. 2009, 3, 160–168. [Google Scholar] [CrossRef] [PubMed]
  55. Ayers, S.; Gillis, K.D.; Lindau, M.; Minch, B.A. Design of a CMOS potentiostat circuit for electrochemical detector arrays. IEEE Trans. Circuits Syst. I Regul. Pap. 2007, 54, 736–744. [Google Scholar] [CrossRef] [PubMed]
  56. Gore, A.; Chakrabartty, S.; Pal, S.; Alocilja, E.C. A multichannel femtoampere-sensitivity potentiostat array for biosensing applications. IEEE Trans. Circuits Syst. I Regul. Pap. 2006, 53, 2357–2363. [Google Scholar] [CrossRef]
  57. Narula, H.S.; Harris, J.G. A time-based VLSI potentiostat for ion current measurements. IEEE Sens. J. 2006, 6, 239–247. [Google Scholar] [CrossRef]
  58. Breten, M.; Lehmann, T.; Bruun, E. Integrating Data Converters for Picoampere Currents from Electrochemical Transducers. In Proceedings of the IEEE International Symposium on Circuits and Systems, Geneva, Switzerland, 28–31 May 2000.
  59. Damen, C.W.N.; Groot, E.R.; Heij, M.; Boss, D.S.; Schellens, J.H.M.; Rosing, H.; Beijnen, J.H.; Aarden, L.A. Development and validation of an enzyme-linked immunosorbent assay for the quantification of trastuzumab in human serum and plasma. Anal. Biochem. 2009, 391, 114–120. [Google Scholar] [CrossRef] [PubMed]
  60. Weerakoon, P.; Culurciello, E.; Klemic, K.G.; Sigworth, F.J. An integrated patch-clamp potentiostat with electrode compensation. IEEE Trans. Biomed. Circuits Syst. 2009, 3, 117–125. [Google Scholar] [CrossRef] [PubMed]
  61. Kim, D.; Goldstein, B.; Tang, W.; Sigworth, F.J.; Culurciello, E. Noise analysis and performance comparison of low current measurement systems for biomedical applications. IEEE Trans. Biomed. Circuits Syst. 2013, 7, 52–62. [Google Scholar]
  62. Rosenstein, J.K.; Ramakrishnan, S.; Roseman, J.; Shepard, K.L. Single ion channel recordings with CMOS-anchored lipid membranes. Nano Lett. 2013, 13, 2682–2686. [Google Scholar] [CrossRef] [PubMed]
  63. Rosenstein, J.K.; Wanunu, M.; Merchant, C.A.; Drndic, M.; Shepard, K.L. Integrated nanopore sensing platform with sub-microsecond temporal resolution. Nat. Methods 2012, 9, 487–492. [Google Scholar] [CrossRef] [PubMed]
  64. Crescentini, M.; Bennati, M.; Carminati, M.; Tartagni, M. Noise limits of CMOS current interfaces for biosensors: A review. IEEE Trans. Biomed. Circuits Syst. 2014, 8, 278–292. [Google Scholar] [CrossRef] [PubMed]
  65. Schienle, M.; Paulus, C.; Frey, A.; Hofmann, F.; Holzapfl, B.; Schindler-Bauer, P.; Thewes, R. A fully electronic DNA sensor with 128 positions and in-pixel A/D conversion. IEEE J. Solid-State Circuits 2004, 39, 2438–2445. [Google Scholar] [CrossRef]
  66. Genov, R.; Stanacevic, M.; Naware, M.; Cauwenberghs, G.; Thakor, N. 16-channel integrated potentiostat for distributed neurochemical sensing. IEEE Trans. Circuits Syst. I Regul. Pap. 2006, 53, 2371–2376. [Google Scholar] [CrossRef]
  67. Stanacevic, M.; Murari, K.; Rege, A.; Cauwenberghs, G.; Thakor, N.V. VLSI potentiostat array with oversampling gain modulation for wide-range neurotransmitter sensing. IEEE Trans. Biomed. Circuits Syst. 2007, 1, 63–72. [Google Scholar] [CrossRef] [PubMed]
  68. Nazari, M.H.; Mazhab-Jafari, H.; Leng, L.; Guenther, A.; Genov, R. CMOS neurotransmitter microarray: 96-channel integrated potentiostat with on-die microsensors. IEEE Trans. Biomed. Circuits Syst. 2013, 7, 338–348. [Google Scholar] [CrossRef] [PubMed]
  69. Jafari, H.M.; Genov, R. Chopper-stabilized bidirectional current acquisition circuits for electrochemical amperometric biosensors. IEEE Trans. Circuits Syst. I 2013, 60, 1149–1157. [Google Scholar] [CrossRef]
  70. Sutula, S.; Cuxart, J.P.; Gonzalo-Ruiz, J.; Muñoz-Pascual, F.X.; Teres, L.; Serra-Graells, F. A 25-µW All-MOS Potentiostatic Delta-Sigma ADC for Smart Electrochemical Sensors. IEEE Trans. Circuits Syst. I 2014, 61, 671–679. [Google Scholar] [CrossRef]
  71. Li, H.; Boling, C.S.; Mason, A.J. CMOS amperometric ADC with high sensitivity, dynamic range and power efficiency for air quality monitoring. IEEE Trans. Biomed. Circuits Syst. 2016, 10, 817–827. [Google Scholar] [CrossRef] [PubMed]
  72. Guo, J.; Ng, W.; Yuan, J.; Li, S.; Chan, M. A 200-Channel Area-Power-Efficient Chemical and Electrical Dual-Mode Acquisition IC for the Study of Neurodegenerative Diseases. IEEE Trans. Biomed. Circuits Syst. 2016, 10, 567–578. [Google Scholar] [CrossRef] [PubMed]
  73. Murari, K.; Thakor, N.; Stanacevic, M.; Cauwenberghs, G. Wide-range, picoampere-sensitivity multichannel VLSI potentiostat for neurotransmitter sensing. In Proceedings of the 26th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, Chicago, IL, USA, 1–5 September 2004; Volume 6, pp. 4063–4066.
  74. Kakerow, R.G.; Kappert, H.; Spiegel, E.; Manoli, Y. Low power single chip CMOS potentiostat. In Proceedings of the International Conference on Solid-State Sensors and Actuators, Stockholm, Sweden, 25–29 June 1995; pp. 142–145.
  75. Bandyopadhyay, A.; Mulliken, G.; Cauwenberghs, G.; Thakor, N. VLSI potentiostat array for distributed electrochemical neural recording. In Proceedings of the IEEE International Symposium on Circuits and Systems, Scottsdale, AZ, USA, 26–29 May 2002; pp. 740–743.
  76. Ferrari, G.; Gozzini, F.; Molari, A.; Sampietro, M. Transimpedance Amplifier for High Sensitivity Current Measurements on Nanodevices. IEEE J. Solid-State Circuits 2009, 44, 1609–1616. [Google Scholar] [CrossRef]
  77. Parsnejad, S.; Li, H.; Mason, A.J. Compact CMOS amperometric readout for nanopore arrays in high throughput lab-on-CMOS. In Proceedings of the 2016 IEEE International Symposium on Circuits and Systems (ISCAS), Montreal, QC, Canada, 22–25 May 2016; pp. 2851–2854.
  78. Liu, X.; Li, H.; Yin, L.; Chen, W.; Suo, C.; Zhou, Z. Noise analysis and characterization of a full differential CMOS interface circuit for capacitive closed-loop micro-accelerometer. J. Harbin Inst. Technol. 2010, 17, 684–689. [Google Scholar]
  79. Lemkin, M.; Boser, B.E. A three-axis micromachined accelerometer with a CMOS position-sense interface and digital offset-trim electronics. IEEE J. Solid-State Circuits 1999, 34, 456–468. [Google Scholar] [CrossRef]
  80. Liu, X.; Yin, L.; Li, H.; Zhou, Z. Full differential CMOS interface circuit for closed-loop capacitive micro-accelerometers. Opt. Precis. Eng. 2011, 19, 581–585. [Google Scholar]
  81. Goldstein, B.; Kim, D.; Xu, J.; Vanderlick, T.K.; Culurciello, E. CMOS low current measurement system for biomedical applications. IEEE Trans. Biomed. Circuits Syst. 2012, 6, 111–119. [Google Scholar] [CrossRef] [PubMed]
  82. Nazari, M.H.; Genov, R. A Fully Differential CMOS Potentiostat. In Proceedings of the IEEE International Symposium on Circuits and Systems, Taipei, Taiwan, 24–27 May 2009; pp. 2177–2180.
  83. Ferrari, G.; Gozzini, F.; Sampietro, M. Very high sensitivity CMOS circuit to track fast biological current signals. In Proceedings of the IEEE Biomedical Circuits Systems Conference, New Orleans, LA, USA, 29 November–1 December 2006; pp. 53–56.
  84. Ferrari, G.; Gozzini, F.; Sampietro, M. A current-sensitive front-end amplifier for nano-biosensors with a 2 MHz BW. In Proceedings of the IEEE International Solid-State Circuits Conference on Digest of Technical Papers (ISSCC 2007), San Francisco, CA, USA, 11–15 February 2007; pp. 164–165.
  85. Ferrari, G.; Farina, M.; Guagliardo, F.; Carminati, M.; Sampietro, M. Ultra-low-noise CMOS current preamplifier from DC to 1 MHz. Electron. Lett. 2009, 45, 1278–1280. [Google Scholar] [CrossRef]
  86. Li, H.; Parsnejad, S.; Mason, A.J. Single ion channel CMOS electrochemical instrument for high throughput recording arrays. In Proceedings of the 2015 IEEE 58th International Midwest Symposium on Circuits and Systems (MWSCAS), Fort Collins, CO, USA, 2–5 August 2015; pp. 1–4.
  87. Li, H.; Boling, S.; Mason, A.J. Power efficient instrumentation with 100 fA-sensitivity and 164 dB-dynamic range for wearable chronoamperometric gas sensor arrays. In Proceedings of the IEEE International Symposium on Circuits and Systems (ISCAS), Lisbon, Portugal, 24–27 May 2015; pp. 485–488.
  88. Stagni, C.; Guiducci, C.; Benini, L.; Ricco, B.; Carrara, S.; Samori, B.; Paulus, C.; Schienle, M.; Augustyniak, M.; Thewes, R. CMOS DNA Sensor Array With Integrated A/D Conversion Based on Label-Free Capacitance Measurement. IEEE J. Solid-State Circuits 2006, 41, 2956–2964. [Google Scholar] [CrossRef]
  89. Ahmadi, M.M.; Jullien, G.A. A wireless-implantable microsystem for continuous blood glucose monitoring. IEEE Trans. Biomed. Circuits Syst. 2009, 3, 169–180. [Google Scholar] [CrossRef] [PubMed]
  90. Jafari, H.M.; Genov, R. CMOS Impedance Spectrum Analyzer with Dual-Slope Multiplying ADC. In Proceedings of the IEEE Biomedical Circuits Systems Conference, San Diego, CA, USA, 10–12 November 2011.
  91. Gozzini, F.; Ferrari, G.; Sampietro, M. An Instrumenton-chip for Impedance Measurements on Nanobiosensors with AttoFarad Resolution. In Proceedings of the IEEE International Solid-State Circuits Conference, San Francisco, CA, USA, 8–12 February 2009; pp. 346–347.
  92. Manickam, A.; Chevalier, A.; McDermott, M.; Ellington, A.D.; Hassibi, A. A CMOS electrochemical impedance spectroscopy (EIS) biosensor array. IEEE Trans. Biomed. Circuits Syst. 2010, 4, 379–390. [Google Scholar] [CrossRef] [PubMed]
  93. Manickam, A.; Chevalier, A.; McDermott, M.; Ellington, A.D.; Hassibi, A. A CMOS electrochemical impedance spectroscopy biosensor array for label-free biomolecular detection. In Proceedings of the 2010 IEEE International Solid-State Circuits Conference Digest of Technical Papers (ISSCC), San Francisco, CA, USA, 7–11 February 2010; pp. 130–131.
  94. Jafari, H.; Soleymani, L.; Genov, R. 16-channel CMOS impedance spectroscopy DNA analyzer with dual-slope multiplying ADCs. IEEE Trans. Biomed. Circuits Syst. 2012, 6, 468–478. [Google Scholar] [CrossRef] [PubMed]
  95. Ferrari, G.; Bianchi, D.; Rottigni, A.; Sampietro, M. CMOS impedance analyzer for nanosamples investigation operating up to 150 MHz with Sub-aF resolution. In Proceedings of the 2014 IEEE International Solid-State Circuits Conference Digest of Technical Papers (ISSCC), San Francisco, CA, USA, 9–13 February 2014; pp. 292–293.
  96. Rairigh, D.; Yang, C.; Mason, A. Analysis of On-Chip Impedance Spectroscopy Methodologies for Sensor Arrays. Sens. Lett. 2006, 4, 398–402. [Google Scholar] [CrossRef]
  97. Min, M.; Parve, T. Improvement of lock-in electrical bio-impedance analyzer for implantable medical devices. IEEE Trans. Instrum. Meas. 2007, 56, 968–974. [Google Scholar] [CrossRef]
  98. Yúfera, A.; Rueda, A.; Munoz, J.M.; Doldán, R.; Leger, G.; Rodriguez-Villegas, E.O. A tissue impedance measurement chip for myocardial ischemia detection. IEEE Trans. Circuits Syst. I 2005, 52, 2620–2628. [Google Scholar] [CrossRef]
  99. Welch, D.; Christen, J.B. Seamless integration of CMOS and microfluidics using flip chip bonding. J. Micromech. Microeng. 2013, 23, 035009. [Google Scholar] [CrossRef]
  100. Li, W.; Rodger, D.C.; Meng, E.; Weiland, J.D.; Humayun, M.S.; Tai, Y.C. Wafer-level parylene packaging with integrated RF electronics for wireless retina prostheses. J. Microelectromech. Syst. 2010, 19, 735–742. [Google Scholar] [CrossRef]
  101. Frey, U.; Sedivy, J.; Heer, F.; Pedron, R.; Ballini, M.; Mueller, J.; Bakkum, D.; Hafizovic, S.; Faraci, F.D.; Greve, F. Switch-matrix-based high-density microelectrode array in CMOS technology. IEEE J. Solid-State Circuits 2010, 45, 467–482. [Google Scholar] [CrossRef]
  102. Huang, Y.; Mason, A.J. Lab-on-CMOS integratio nof microfluidics and electrochemical sensors. Lab Chip 2013, 13, 3929–3934. [Google Scholar] [CrossRef] [PubMed]
  103. Datta-Chaudhuri, T.; Abshire, P.; Smela, E. Packaging commercial CMOS chips for lab on a chip integration. Lab Chip 2014, 14, 1753–1766. [Google Scholar] [CrossRef] [PubMed]
  104. Li, L.; Mason, A. Development of an integrated CMOS-microfluidic instrumentation array for high throughput membrane protein studies. In Proceedings of the IEEE International Symposium on Circuits and Systems (ISCAS), Melbourne, Australia, 1–5 June 2014.
  105. Ghafar-Zadeh, E.; Sawan, M.; Therriault, D. Novel direct-write CMOS-based laboratory-on-chip: Design, assembly and experimental results. Sens. Actuators A Phys. 2007, 134, 27–36. [Google Scholar] [CrossRef]
  106. Christen, J.B.; Andreou, A.G. Design, fabrication, and testing of a hybrid CMOS/PDMS microsystem for cell culture and incubation. IEEE Trans. Biomed. Circuits Syst. 2007, 1, 3–18. [Google Scholar] [CrossRef] [PubMed]
  107. Heer, F.; Hafizovic, S.; Franks, W.; Blau, A.; Ziegler, C.; Hierlemann, A. CMOS microelectrode array for bidirectional interaction with neuronal networks. IEEE J. Solid-State Circuits 2006, 41, 1620–1629. [Google Scholar] [CrossRef]
  108. Hogervorst, R.; Tero, J.P.; Eschauzier, R.G.; Huijsing, J.H. A compact power-efficient 3 V CMOS rail-to-rail input/output operational amplifier for VLSI cell libraries. IEEE J. Solid-State Circuits 1994, 29, 1505–1513. [Google Scholar] [CrossRef]
  109. Yin, H.; Li, L.; Mason, A.J. Screen-printed planar metallization for lab-on-CMOS with epoxy carrier. In Proceedings of the 2016 IEEE International Symposium on Circuits and Systems (ISCAS), Montreal, QC, Canada, 22–25 May 2016; pp. 2887–2890.
  110. Hwang, S.; Lafratta, C.N.; Agarwal, V.; Yu, X.; Walt, D.R.; Sonkusale, S. CMOS microelectrode array for electrochemical lab-on-a-cihp applications. IEEE J. Sens. 2009, 9, 609–615. [Google Scholar] [CrossRef]
  111. Niitsu, K.; Ota, S.; Gamo, K.; Kondo, H.; Hori, M.; Nakazato, K. Development of microelectrode arrays using electroless plating for CMOS-based direct counting of bacterial and HeLa cells. IEEE Trans. Biomed. Circuits Syst. 2015, 9, 607–619. [Google Scholar] [CrossRef] [PubMed]
  112. Li, L.; Liu, X.; Mason, A.J. Die-level photolithography and etchless parylene packaging processes for on-CMOS electrochemical biosensors. In Proceedings of the International Symposium on Circuits and Systems, Seoul, Korea, 20–23 May 2012; pp. 2401–2404.
  113. Sansen, W.; Wachter, D.D.; Callewaert, L.; Lambrechts, M.; Claes, A. A smart sensor for the voltammetric measurement of oxygen or glucose concentrations. Sens. Actuators B Chem. 1990, 1, 298–302. [Google Scholar] [CrossRef]
  114. Klare, P.; Herrmann, S.; Berthold, F.; Vonau, W. Lab-on-Chip-Messsystem. In Elektrochemische Grundlagenforschung und deren Anwendung in der Elektroanalytik; Die Deutsche Bibliothek: Frankfurt, Germany, 2006; pp. 111–114. [Google Scholar]
  115. Huang, I.; Huang, R.; Lo, L. Improvement of integrated Ag/AgCl thin-film electodes by KCl-gel coating for ISFET applications. Sens. Actuators B Chem. 2003, 94, 53–64. [Google Scholar] [CrossRef]
  116. Lauks, I.; Maczuszenko, A. Heterogeneous Membrane Electrodes. WO Patent App. PCT/CA2003/001,841, 27 November 2003. [Google Scholar]
  117. Rahman, M.A.; Kwon, N.-H.; Won, M.-S.; Choe, E.S.; Shim, Y.-B. Functionalized conducting polymer as an enzyme-immobilizing substrate: An amperometric glutamate microbiosensor for in vivo measurements. Anal. Chem. 2005, 77, 4854–4860. [Google Scholar] [CrossRef] [PubMed]
  118. Kinlen, P.J.; Heider, J.E.; Hubbard, D.E. A solid-state pH sensor based on a Nafion-coated iridium oxide indicator electrode and a polymer-based silver chloride reference electrode. Sens. Actuators B Chem. 1994, 22, 13–25. [Google Scholar] [CrossRef]
  119. Curry, K.M. Method and Apparatus for Using Flex Circuit Technology to Create a Reference Electrode Channel. WIPO Patent 2,007,100,644, 22 February 2007. [Google Scholar]
  120. Shin, J.H.; Lee, S.D.; Nam, H.; Cha, G.S.; Bae, B.W. Miniaturized Solid-State Reference Electrode with Self-Diagnostic Function. U.S. Patent 6,554,982, 29 April 2003. [Google Scholar]
  121. Cha, G.S.; Cui, G.; Yoo, J.; Lee, J.S.; Nam, H. Planar Reference Electrode. U.S. Patent 6,964,734, 15 November 2005. [Google Scholar]
  122. Suzuki, H.; Shiroishi, H.; Sasaki, S.; Karube, I. Microfabricated liquid junction Ag/AgCl reference electrode and its application to a one-chip potentiometric sensor. Anal. Chem. 1999, 71, 5069–5075. [Google Scholar] [CrossRef]
  123. Han, J.; Park, S.; Boo, H.; Kim, H.C.; Nho, J.; Chung, T.D. Solid-State Reference Electrode Based on Electrodeposited Nanoporous Platinum for Microchip. Electroanalysis 2007, 19, 786–792. [Google Scholar] [CrossRef]
  124. Kwon, N.-H.; Lee, K.-S.; Won, M.-S.; Shim, Y.-B. An all-solid-state reference electrode based on the layer-by-layer polymer coating. Analyst 2007, 132, 906–912. [Google Scholar] [CrossRef] [PubMed]
  125. Prodromakis, T.; Georgiou, P.; Constandinou, T.G.; Michelakis, K.; Toumazou, C. Batch encapsulation technique for CMOS based chemical sensors. In Proceedings of the IEEE Biomedical Circuits and Systems Conference, Baltimore, MD, USA, 18–21 November 2008; pp. 321–324.
  126. Oelßner, W.; Zosel, J.; Guth, U.; Pechstein, T.; Babel, W.; Connery, J.G.; Demuth, C.; Grote Gansey, M.; Verburg, J.B. Encapsulation of ISFET sensor chips. Sens. Actuators B Chem. 2005, 105, 104–117. [Google Scholar] [CrossRef]
  127. Dumschat, C.; Müller, H.; Rautschek, H.; Timpe, H.J.; Hoffmann, W.; Pham, M.T.; Hüller, J. Encapsulation of chemically sensitive field-effect transistors with photocurable epoxy resins. Sens. Actuators B Chem. 1990, 2, 271–276. [Google Scholar] [CrossRef]
  128. Prodromakis, T.; Michelakis, K.; Zoumpoulidis, T.; Dekker, R.; Toumazou, C. Biocompatible encapsulation of CMOS based chemical sensors. In Proceedings of the IEEE Internationa Conference on Sensors, Christchurch, New Zealand, 25–28 October 2009; pp. 791–794.
  129. Brunnbauer, M.; Fürgut, E.; Beer, G.; Meyer, T.; Hedler, H.; Belonio, J.; Nomura, E.; Kiuchi, K.; Kobayashi, K. An embedded device technology based on a molded reconfigured wafer. In Proceedings of the 56th Electronic Components and Technology Conference, San Diego, CA, USA, 30 May–2 June 2006.
  130. Keser, B.; Amrine, C.; Duong, T.; Fay, O.; Hayes, S.; Leal, G.; Lytle, W.; Mitchell, D.; Wenzel, R. The Redistributed Chip Package: A Breakthrough for Advanced Packaging. In Proceedings of the 57th Electronic Components and Technology Conference (ECTC’07), Sparks, NV, USA, 29 May–1 June 2007; pp. 286–291.
  131. Kumar, A.; Dingwei, X.; Sekhar, V.N.; Lim, S.; Keng, C.; Sharma, G.; Rao, V.S.; Kripesh, V.; Lau, J.H.; Kwong, D.-L. Wafer level embedding technology for 3D wafer level embedded package. In Proceedings of the 59th Electronic Components and Technology Conference, San Diego, CA, USA, 26–29 May 2009; pp. 1289–1296.
  132. Souriau, J.-C.; Lignier, O.; Charrier, M.; Poupon, G. Wafer level processing of 3D system in package for RF and data application. In Proceedings of the 55th Electronic Components and Technology Conference, Lake Buena Vista, FL, USA, 30 May–2 June 2005; pp. 356–361.
Figure 1. Components of a general integrated electrochemical biosensor.
Figure 1. Components of a general integrated electrochemical biosensor.
Sensors 17 00074 g001
Figure 2. The functional structure of an ion-sensitive field effect transistors (ISFET) biosensor. RE, reference electrode. S is the MOSFET source, D is the drain, and A represents a current meter.
Figure 2. The functional structure of an ion-sensitive field effect transistors (ISFET) biosensor. RE, reference electrode. S is the MOSFET source, D is the drain, and A represents a current meter.
Sensors 17 00074 g002
Figure 3. The equivalent circuit model of (a) a two-electrode system and (b) a three-electrode system. ZCE is impedance between RE and the counter electrode (CE). WE, working electrode.
Figure 3. The equivalent circuit model of (a) a two-electrode system and (b) a three-electrode system. ZCE is impedance between RE and the counter electrode (CE). WE, working electrode.
Sensors 17 00074 g003
Figure 4. The structure of the instrumentation for electrochemical biosensors.
Figure 4. The structure of the instrumentation for electrochemical biosensors.
Sensors 17 00074 g004
Figure 5. Three-electrode potentiostats with the (a) grounded-WE structure and (b) grounded-CE structure.
Figure 5. Three-electrode potentiostats with the (a) grounded-WE structure and (b) grounded-CE structure.
Sensors 17 00074 g005
Figure 6. Fully-differential potentiostat (adapted from [46]). OP1–3 are operational amplifiers.
Figure 6. Fully-differential potentiostat (adapted from [46]). OP1–3 are operational amplifiers.
Sensors 17 00074 g006
Figure 7. Resistive feedback current readout circuit.
Figure 7. Resistive feedback current readout circuit.
Sensors 17 00074 g007
Figure 8. Capacitive feedback current readout circuit.
Figure 8. Capacitive feedback current readout circuit.
Sensors 17 00074 g008
Figure 9. The structure of CDS circuit (adapted from [47]).
Figure 9. The structure of CDS circuit (adapted from [47]).
Sensors 17 00074 g009
Figure 10. The structure of the pseudo-differential amperometry readout circuit (adapted from [41]).
Figure 10. The structure of the pseudo-differential amperometry readout circuit (adapted from [41]).
Sensors 17 00074 g010
Figure 11. The structure of the current mirror amperometric readout circuit with the potentiostat (adapted from [47]).
Figure 11. The structure of the current mirror amperometric readout circuit with the potentiostat (adapted from [47]).
Sensors 17 00074 g011
Figure 12. The structure of a first-order incremental current mode Σ∆ ADC circuit for amperometric sensing.
Figure 12. The structure of a first-order incremental current mode Σ∆ ADC circuit for amperometric sensing.
Sensors 17 00074 g012
Figure 13. (a) Impedimetric sensor circuitry model where Cdl represents double-layer capacitance, Rs denotes solution resistance and Zf represents Faradic components. (b) Frequency response analyzer (FRA) method for extracting transducer impedance/admittance. Both phasor domain and time domain signal expressions are given. ω is the stimulus signal frequency. |Y|∠θ is the phasor form of the transducer admittance. Asin(ωt) is the stimulus voltage for the electrochemical transducer. A'sin(ωt + θ) is the output current of the transducer. sin(ωt) and cos(ωt) are the reference signals to extract the real part of imaginary part of the signal. LPF is the low pass filter. REAL and IMG are two orthogonal outputs, representing real and imaginary components of impedance/admittance.
Figure 13. (a) Impedimetric sensor circuitry model where Cdl represents double-layer capacitance, Rs denotes solution resistance and Zf represents Faradic components. (b) Frequency response analyzer (FRA) method for extracting transducer impedance/admittance. Both phasor domain and time domain signal expressions are given. ω is the stimulus signal frequency. |Y|∠θ is the phasor form of the transducer admittance. Asin(ωt) is the stimulus voltage for the electrochemical transducer. A'sin(ωt + θ) is the output current of the transducer. sin(ωt) and cos(ωt) are the reference signals to extract the real part of imaginary part of the signal. LPF is the low pass filter. REAL and IMG are two orthogonal outputs, representing real and imaginary components of impedance/admittance.
Sensors 17 00074 g013
Figure 14. The structure of the lock-in Σ∆ impedance extraction circuit. The circuit extracts the imaginary portion of admittance when S = 1 and the real portion when S = 0 (adapted from [16]).
Figure 14. The structure of the lock-in Σ∆ impedance extraction circuit. The circuit extracts the imaginary portion of admittance when S = 1 and the real portion when S = 0 (adapted from [16]).
Sensors 17 00074 g014
Figure 15. Dual-slope multiplying ADC architecture (adapted from [94]).
Figure 15. Dual-slope multiplying ADC architecture (adapted from [94]).
Sensors 17 00074 g015
Figure 16. Components in integrating a CMOS biosensor microsystem.
Figure 16. Components in integrating a CMOS biosensor microsystem.
Sensors 17 00074 g016
Figure 17. A 2 × 2 thin film planar gold microelectrode array fabricated on a CMOS chip, which includes WE, CE and RE.
Figure 17. A 2 × 2 thin film planar gold microelectrode array fabricated on a CMOS chip, which includes WE, CE and RE.
Sensors 17 00074 g017
Figure 18. Concept of lab-on-CMOS integration of CMOS biosensors and multichannel microfluidics using a carrier device and planar electrical interconnections.
Figure 18. Concept of lab-on-CMOS integration of CMOS biosensors and multichannel microfluidics using a carrier device and planar electrical interconnections.
Sensors 17 00074 g018
Table 1. Performance summary of amperometric current readout circuitry. CDS, correlated double sampling.
Table 1. Performance summary of amperometric current readout circuitry. CDS, correlated double sampling.
Maximum Input CurrentSensitivityRMS NoiseDynamic RangePower/ChannelProcessCircuit StructureTargetTarget Concentration
JSSC87 [45]3.5 µA100 nA-31 dB<2 mW5 μmCurrent conveyorGlucose0–24.5 mM
ISSCC94 [53]40 µA100 fA-172 dB5 mW2 μmIntegrator + dual-slope ADCCu2+0.5 ppm
JSSC04 [65]100 nA10 pA->100 dB-0.5 µmIntegration ADCDNA-
TCAS06 [56]100 nA50 fA-60 dB11 μW0.5 μmSemi-synchronous Σ∆ ADCBacillus Cereus0–106 CFU/mL
TCAS06 [66]50 µA-46 pA120 dB781 µW1.2 µmIntegration ADCDopamine0-175 µM
Sensors06 [57]200 nA1 pA-116 dB>130 µW0.5 µmIntegration ADCKCl0.01–1 M
TCAS07 [55]400 pA20 pA110 fA26 dB1 µW0.5 μmHalf-amplifier structure; CDS--
TbioCAS07 [67]0.5 µA100 fA-140 dB1.3 mW0.5 µmFeedback modulation Σ∆ ADCDopamine0–5 µM
JSSC08 [50]110 nA~240 pA-~53 dB-0.25 μmDual-slope ADCDNA-
TCAS09 [47]1 µA1 nA-60 dB70 µW0.18 µmCurrent mirror + I-F ADC 1Glucose0–38 mM
TBioCAS09 [54]47 µA1 pA-120 dB-0.5 μmCDSAlcohol dehydrogenase0–25 mM
TbioCAS13 [68]350 nA24 pA-95 dB188 µW0.35 µmI-F + Single slope ADC 1Dopamine0–80 µM
TCAS13 [69]350 nA8.6 pA0.13 pA92 dB4 µW0.13 µmCurrent conveyorK3[Fe(CN)6]0–2 µM
TCAS14 [70]32 µA---25 µW2.5 µmAll-CMOS Σ∆ ADCK4[Fe(CN)6]0.1–1.1 mM
TBioCAS16 [71]16 µA100 fA-164 dB241 µW0.5 µmInput modulated Σ∆ ADCO20%–20%
TBioCAS15 [72]50 nA-480 fA104 dB12.1 µW0.18 µmSAR ADCDopamine0–40 µM
TI-LMP9100x750 µA5 µA-43.5 dB33 µW-Resistive feedback TIAGas/glucose-
1 I-F represents “current-to-frequency”.
Table 2. Performance summary of impedimetric readout circuitry.
Table 2. Performance summary of impedimetric readout circuitry.
Circuit StructureFrequency RangeCurrent RangeProcessPowerTarget
JSSC09 [16]Lock-in ΣΔ ADC10 MHz–10 kHz78 fA–100 nA0.5 μm6 μWGramicidin ion channel
TBioCAS10 [92]analog mixer10 Hz–50 MHz330 pA–42 µA0.35 μm84.8 mWDNA/protein
TBioCAS12 [94]Dual-slope multiplying ADC0.1 Hz–10 kHz320 fA–500 nA0.13 μm42 µWProstate cancer DNA
TCAS05 [98]Mixer, incremental ADC--0.8 μm2.1 mWKidney ischemia
ISSCC14 [95]down/up conversion20 kHz–150 MHz-0.35 μm0.14 W-

Share and Cite

MDPI and ACS Style

Li, H.; Liu, X.; Li, L.; Mu, X.; Genov, R.; Mason, A.J. CMOS Electrochemical Instrumentation for Biosensor Microsystems: A Review. Sensors 2017, 17, 74. https://doi.org/10.3390/s17010074

AMA Style

Li H, Liu X, Li L, Mu X, Genov R, Mason AJ. CMOS Electrochemical Instrumentation for Biosensor Microsystems: A Review. Sensors. 2017; 17(1):74. https://doi.org/10.3390/s17010074

Chicago/Turabian Style

Li, Haitao, Xiaowen Liu, Lin Li, Xiaoyi Mu, Roman Genov, and Andrew J. Mason. 2017. "CMOS Electrochemical Instrumentation for Biosensor Microsystems: A Review" Sensors 17, no. 1: 74. https://doi.org/10.3390/s17010074