Nuklearmedizin 2009; 48(04): 159-165
DOI: 10.3413/nukmed-0215
Original Article
Schattauer GmbH

Resolution improvement and noise reduction in human pinhole SPECT using a multi-ray approach and the SHINE method

Verbesserte Auflösung und geringeres Bildrauschen bei der Einzelphotonen-Emissions-Tomographie (SPECT) mit Pinhole-Kollimator am Menschen unter Verwendung des Mehrstrahl-Ansatzes und der SHINE-Methode
A. Seret
1   Imagerie médicale expérimentale, Université de Liège, Liège, Belgium
,
C. Vanhove
2   Nuclear Medicine Department, University Hospital Brussels, Brussels, Belgium
,
M. Defrise
3   Nuclear Medicine Department, Vrije Universiteit Brussel, Brussels, Belgium
› Author Affiliations
Further Information

Publication History

received: 19 September 2009

accepted in revised form: 27 March 2009

Publication Date:
22 January 2018 (online)

Summary

Purpose: This work aimed at quantifying the gains in spatial resolution and noise that could be achieved when using resolution modelling based on a multi-ray approach and additionally the Statistical and Heuristic Noise Extraction (SHINE) method in human pinhole single photon emission tomography (PH-SPECT). Methods: PH-SPECT of two line phantoms and one homogeneous cylinder were recorded using parameters suited for studies of the human neck area. They were reconstructed using pinhole dedicated ordered subsets expectation maximisation algorithm including a resolution recovery technique based on 7 or 21 rays. Optionally, the SPECT data were SHINE pre-processed. Transverse and axial full widths at half-maximum (FWHM) were obtained from the line phantoms. The noise was quantified using the coefficient of variation (COV) derived from the uniform phantom. Two human PHSPECT studies of the thyroid (a hot nodule and a very low uptake) were processed with the same algorithms. Results: Depending on the number of iterations, FWHM decreased by 30 to 50% when using the multi-ray approach in the reconstruction process. The SHINE method did not affect the resolution but decreased the COV by at least 20% and by 45% when combined with the multi-ray method. The two human studies illustrated the gain in spatial resolution and the decrease in noise afforded both by the multi-ray reconstruction and the SHINE method. Conclusion: Iterative reconstruction with resolution modelling allows to obtain high resolution human PH-SPECT studies with reduced noise content. The SHINE method affords an additional noise reduction without compromising the resolution.

Zusammenfassung

Ziel: Mit dieser Arbeit sollen die Verbesserungen der räumlichen Auflösung und des Bildrauschens quantifiziert werden, die mithilfe eines Bildauflösungsmodells auf der Grundlage des Mehrstrahl-Ansatzes zusammen mit der SHINEMethode (Statistical and Heuristic Image Noise Extraction = statistische und heuristische Rauschunterdrückung bei bildgebenden Verfahren) bei der Einzelphotonen-Emissions-Tomographie mit Pinhole-Kollimator (PH-SPECT) am Menschen erreicht werden. Methodik: Es wurden PH-SPECT von zwei Linien-Phantomen und einem homogenen Zylinder-Phantom unter Verwendung von Parametern, die sich zur Untersuchung der menschlichen Halsregion eignen, aufgenommen. Sie wurden mit Hilfe des OS-EM (Ordered Subsets Expectation Maximisation)- Algorithmus mit Pinhole-Kollimator einschließlich einer auf 7 oder 21 Projektionsstrahlen basierten Wiederherstellungsmethode für die Auflösung rekonstruiert. Wahlweise wurden die SPECT-Daten zuvor mit der SHINEMethode bearbeitet. Diagonale und axiale Halbwertsbreiten (Full Widths at Half-Maximum – FWHM) erhielt man durch die Linien- Phantome. Das Bildrauschen wurde mit dem Varianzkoeffizienten (Coefficient of Variation – COV) quantifiziert, der vom homogenen Phantom abgeleitet wurde. Mit denselben Algorithmen wurden zwei PH-SPECT-Untersuchungen der Schilddrüse am Menschen (ein heißer Knoten und eine sehr schlechte Aufnahme) bearbeitet. Ergebnisse: Je nach Anzahl der Iterationen verringerte sich die FWHM um 30 bis 50 %, wenn beim Rekonstruktionsprozess der Mehrstrahl- Ansatz verwendet wurde. Die SHINEMethode hatte auf die Auflösung keine Auswirkungen, reduzierte jedoch den COV um mindestens 20 % und in Kombination mit der Mehrstrahl-Methode sogar um 45 %. Die zwei Untersuchungen am Menschen zeigen auf, welche Verbesserung bei der räumlichen Auflösung erreicht wurde und wie stark das Bildrauschen sowohl durch die Mehrstrahl-Rekonstruktion wie auch durch die SHINE-Methode reduziert werden kann. Schlussfolgerung: Eine iterative Bildrekonstruktion mit Bildauflösungsmodell ermöglicht eine hohe Auflösung bei PH-SPECT-Untersuchungen am Menschen bei gleichzeitig reduziertem Bildrauschen. Die SHINE-Methode erzeugt eine zusätzliche Rauschunterdrückung, ohne dabei die Auflösung zu beeinträchtigen.

 
  • References

  • 1 Bai C, Zeng GL, Gullberg GT. et al. Slab-by-slab blurring model for geometric point response correction and attenuation correction using iterative reconstruction algorithms. IEEE Trans Nucl Sci 1998; 45: 2168-2172.
  • 2 Beekman F, van der Have F. The pinhole: gateway to ultra-high-resolution three-dimensional radionuclide imaging. Eur J Nucl Med Mol Imaging 2007; 34: 151-161.
  • 3 Béqué D, Nuyts J, Bormans G. et al. Characterization of pinhole SPECT acquisition geometry. IEEE Trans Med Imaging 2003; 22: 599-612.
  • 4 Béqué D, Nuyts J, Suetens P, Bormans G. Optimization of geometrical calibration in pinhole SPECT. IEEE Trans Med Imaging 2005; 24: 180-190.
  • 5 Carlier T, Oudoux A, MiralliZ E. et al. 99mTc-MIBI pinhole SPECT in primary hyperparathyroidism: comparison with conventional SPECT, planar scintigraphy and ultrasonography. Eur J Nucl Med Mol Imaging 2008; 35: 637-643.
  • 6 Desvignes P, Laurette I, Koulibaly PM. et al. Evaluation clinique d'une méthode algébrique de reconstruction 3D pour la détection des nodules thyroïdiens froids par tomoscintigraphie sténopéïque. Médecine Nucléaire 2000; 24: 217-223.
  • 7 Hannequin P, Mas J. Statistical and Heuristic Noise Extraction (SHINE): a new method for processing Poisson noise in scintigraphic images. Phys Med Biol 2002; 47: 4329-4344.
  • 8 Krausz Y, Bettman L, Guralnik L. et al. Technetium-99m-MIBI SPECT/CT in primary hyperparathyroidism. World J Surg 2006; 30: 76-83.
  • 9 Grosev D, Loncaric D, Huic D, Dodig D. Geometric models in dosimetry of thyroid remnant mass. Nuklearmedizin 2008; 47: 120-126.
  • 10 Panin VY, Kehren F, Michel C, Casey M. Fully 3-D PET reconstruction with system matrix derived from point source measurements. IEEE Trans Med Imaging 2006; 25: 907-921.
  • 11 Qi J, Leahy RM, Cherry SR. et al. High resolution 3D Bayesian image reconstruction using the micro- PET small animal scanner. Phys Med Biol 1998; 43: 1001-1013.
  • 12 Rolleman EJ, Bernard BF, Breeman WAP. et al. Molecular imaging of reduced renal uptake of radiolabelled [D0TA0,Tyr3]octreotate by the combination of lysine and Gelofusine in rats. Nuklearmedizin 2008; 47: 110-115.
  • 13 Ruf J, Seehofer D, Denecke T. et al. Impact of image fusion and attenuation correction by SPECT-CT on the scintigraphic detection of parathyroid adenomas. Nuklearmedizin 2007; 46: 15-21.
  • 14 Seret A. Will high-resolution/high-sensitivity SPECT ensure that PET is not the only survivor in nuclear medicine during the next decade?. Eur J Nucl Med Mol Imaging 2009; 36: 533-535.
  • 15 Seret A, Defrise M, Blocklet D. 180°Pinhole SPET with a tilted detector and OS-EM reconstruction: phantom studies and potential clinical applications. Eur J Nucl Med 2001; 28: 1836-1841.
  • 16 Seret A, Fleres D, Defrise M. Body contour 180°pin- hole SPET with or without tilted detector: a phantom study. Eur J Nucl Med Mol Imaging 2003; 30: 1205-1210.
  • 17 Seret A, Hustinx R. Clinical applications of pinhole single photon emission tomography. Current Medical Imaging Reviews 2006; 2: 347-352.
  • 18 Sohlberg A, Ruotsalainen U, Watabe H. et al. Accelerated median root prior reconstruction for pinhole single-photon emission tomography (SPET). Phys Med Biol 2003; 48: 1957-1969.
  • 19 Sohlberg A, Lensu S, Jolkkonen J. et al. Improving the quality of small animal brain pinhole SPECT imaging by Bayesian reconstruction. Eur J Nucl Med Mol Imaging 2004; 31: 986-994.
  • 20 Sohlberg A, Watabe H, Zeniya T, Iida H. Comparison of multi-ray and point-spread function based resolution recovery methods in pinhole SPECT reconstruction. Nucl Med Commun 2006; 27: 823-827.
  • 21 Spanu A, Falchi A, Manca A. et al. The usefulness of neck pinhole SPECT as a complementary tool for planar scintigraphy in primary and secondary hyperparathyroidism. J Nucl Med 2004; 45: 40-48.
  • 22 Vanhove C, Defrise M, Franken PR. et al. Interest of the ordered subsets expectation maximization (OSEM) algorithm in pinhole single-photon emission tomography reconstruction: a phantom study. Eur J Nucl Med 2000; 27: 140-146.
  • 23 Vanhove C, Andreyev A, Defrise M, et al. Resolution recovery in pinhole SPECT based on multi-ray projections: a phantom study. Eur J Nucl Med Mol Imaging 2007; 34: 170-180.
  • 24 Wanet P, Sand A, Abramovici J. Physical and clinical evaluation of high-resolution thyroid pinhole tomography. J Nucl Med 1996; 37: 2017-2020.
  • 25 Weber DA, Ivanovic M. Ultra-high-resolution imaging of small animals: Implications for preclinical and research studies. J. Nucl Cardiol 1999; 6: 332-344.
  • 26 Zeng GL, Gullberg GT, Tsui BMW, Terry JA. Three dimensional iterative reconstruction algorithms with attenuation and geometric point response correction. IEEE Trans Nucl Sci 1991; 38: 693-702.
  • 27 Zeniya T, Watabe H, Aoi T. et al. A new reconstruction strategy for image improvement in pinhole SPECT. Eur J Nucl Med Mol Imaging 2004; 31: 1166-1172.