Microscale solute flow probed with rotating microbead trapped in optical vortex

verfasst von: Weronika Lamperska, Jan Masajada, Sławomir Drobczyński

Erschienen in: Experiments in Fluids | Ausgabe 6/2021

Aktivieren Sie unsere intelligente Suche um passende Fachinhalte oder Patente zu finden.

search-config
loading …

Abstract

The dynamics of solute flow in the microscopic chamber can be studied with optical tweezers. A method based on the metallic microbeads trapped in the focused optical vortex beam is proposed. This annular beam of a twisted wavefront exerts torque on a reflective object placed inside the dark core of the vortex. The induced rotational movement of the bead is sensitive to local viscosity changes in the surrounding medium, for example, during the ongoing dissolution process. Two experimental configurations are described, both relying on tracing the angular velocity of the bead in time. In one-bead configuration, the dynamics of local solute concentration can be studied. In two-bead case, the direction and speed of solute flow can be probed with a spatial resolution of single micrometers. We approach the elementary problem of sucrose dissolution and diffusion in water. The surprising impression of the reverse solute flow was observed. Further experimental investigation led to the discovery that this phenomenon originates from the sucrose stream-like diffusion in the mid-depth of the measurement chamber. The rotating microbead method applies for various solid and liquid substances and may become a useful technique for microfluidics research.

Graphic abstract

Begleitmaterial
Supplementary file1 (MP4 1490 KB)
Hinweise

Supplementary Information

The online version contains supplementary material available at https://​doi.​org/​10.​1007/​s00348-021-03223-6.

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

1 Introduction

Optical vortex is a singular beam of a helical wavefront, which means that the phase of the vortex beam circulates around the optical axis (Vasnetsov 1999, Soskin 2001). In the cross section of the vortex phase distribution (Fig. 1a), the equiphase lines converge to a single point, called a vortex point, in which the phase is undetermined. In terms of intensity, the cross section of the optical vortex is a bright ring with a dark disk in the center (Fig. 1b).
Vortex beams can be used for optical trapping in optical tweezers (Torres 2011, Padgett 2011, Jones 2015). Transparent particles of the radius smaller than the radius of the vortex beam (e.g., submicron-sized particles) are typically trapped by the bright ring and move along it (Simpson 2010, Bacia 2015). On the other hand, the transparent particles large enough to cover the bright ring of the optical vortex beam are trapped at the beam center (Simpson 2010) and so are opaque particles of any size. An example of an opaque particle is a reflective micrometer-sized metallic bead (MB) used in this study. The MB placed inside the dark disk is surrounded by the bright ring of high light intensity and gets repelled from the ring toward its center due to radiation pressure. Thus, the MB is trapped in the dark disk in xy plane and partially along the optical axis z.
The twisted curvature of the wavefront results in nonzero angular momentum carried by the optical vortex (Torres 2011). This type of angular momentum is called orbital angular momentum (OAM) in contrast to the spin angular momentum (SAM) arising from circular or elliptical polarization of light. In other words, a characteristic, helical spatial distribution of the optical field is responsible for the presence of OAM in optical vortex. Each photon in the vortex beam carries OAM equal to ± mℏ, where the sign depends on the handedness of the wavefront helix, m is a parameter called topological charge, and ℏ is a reduced Planck constant. The higher the vortex charge m, the larger the diameter of the bright ring. The transfer of orbital angular momentum between light and matter was observed for the first time by He et al. (1995). After directing a vortex beam onto absorptive CuO particles suspended in water, the particles rotated in the direction determined by the sign of the topological charge of the vortex. Quantitative measurements of the angular momentum transfer were also reported (Parkin 2006, Asavei 2013, Lamperska 2020). The transfer of OAM can be also achieved with reflective particles such as MBs. The MB illuminated with an optical vortex experiences a constant net torque. Therefore, trapping the reflective particle with a vortex beam not only prevents bead’s translation but also causes its rotations (Fig. 1c). The rotational speed depends on beam parameters (laser power, vortex charge), size of the MB and viscosity of the surrounding medium. For fixed beam parameters, the rotating MB becomes a sensitive detector of any changes in viscosity of the medium. In particular, the viscosity growth is indicated by the drop in rotational speed of the trapped bead. The change in viscosity may result from the change in concentration of the solution. The concentration can be changed dynamically, for example, during the process of solute diffusion across the sample. We used this approach to study the microscale sucrose diffusion dynamics.
In this paper, we report on the measurement method based on vortex-induced rotations of MB and then apply it to study the diffusion dynamics in microscale. We revisit the elementary problem of sucrose dissolution in water in an unconventional manner. Two-laser optical tweezers (Lamperska 2017) based on the inverted biological microscope are used to generate vortex traps, manipulate the trapped beads and record the magnified image of the sample. A single rotating MB acts as a microsensor for viscosity changes at the specific point of the sample. Three different phases of MB motion were identified, and their background was discussed. The one-MB method proved useful in studying the dynamics of local solute concentration. Furthermore, by using two rotating MBs separated by less than one-tenth of millimeter, one can detect the direction and the propagation speed of sucrose diffusion. The quantity of interest is variations of the MB rotation due to viscosity changes. We report on the nonintuitive observation of distinctive solute flow in the chamber. The origin of this phenomenon was attributed to the sucrose stream-like diffusion in the central depth of the measurement chamber. Two-MB method offers the opportunity to probe the diffusion flow with a spatial resolution of single micrometers. There exist a few methods of studying fluid flows with comparable or greater resolution, such as micro-particle image velocimetry (Lindken 2009, Wereley 2010), surface plasmon resonance (Ahangar 2020, Jeong 2021), micro-laser Doppler velocimetry (Czarske 2015), positron emission particle tracking (Langford 2016), NRM (Lee 2001) and MRI (Elkins 2007). Most of these techniques require advanced research equipment and are limited to specific experimental conditions (in terms of substrate, substances and physical properties). Here, we present a method designed for studying diffusion based on optical tweezers. Nowadays, optical tweezers can be found in many optical laboratories, and thus, we believe the proposed method could be implemented and tested by a number of research groups.

2 Experimental setup

The experiments were performed in two-laser optical tweezers shown in Fig. 2a. In this system, there are two optical paths–holographic and non-holographic one. Each path is equipped with a separate laser source emitting the infrared beam of Gaussian intensity distribution. The holographic path is built of Nd:YAG laser (Quantum Ventus, wavelength: 1064 nm, 4 W) and spatial light modulator (SLM, HoloEye-Pluto). SLM converts the incident Gaussian beam into one or more vortex beams. Vortex charge m is determined by the structure of the computer-generated phase hologram displayed on the SLM matrix (Fig. 2b, c). Moreover, combining vortex-generating hologram with the phase diffraction grating (Fig. 2d) enables to control the vortex trap position in the observation plane. The non-holographic path consists of Nd:YAG laser (Quantum Ventus, wavelength: 1064 nm, 3 W) and a system of two scanning galvanometer mirrors (Thorlabs, GVS002). Electrically addressed mirrors deflect the incident Gaussian beam providing precise beam steering. In both paths, half-wave plates are used to set two orthogonal linear polarization states of the beams. The beams are combined by a polarizing beamsplitter (PBS) and then focused by a microscope objective (100x, oil-immersion, NA = 1.3, Olympus UPLAN FL N) in the sample plane. The sample plane is illuminated with a halogen lamp, imaged onto the CMOS camera and displayed on the computer screen.
In the sample plane, tightly focused beams become optical traps of two kinds–vortex traps from the holographic path and Gaussian traps from the non-holographic path. The holographic beam suffers from substantial power losses at the reflection from the SLM, whereas the non-holographic beam leaves the system of galvanic mirrors almost lossless. Thus, despite the higher output power of the vortex-generating laser, vortex traps are weaker than non-holographic traps. Vortex traps are used for trapping the particles in a fixed position and induce their rotations due to angular momentum transfer from the beam to the particle. The non-holographic path supports the experiments with an auxiliary Gaussian trap used for detaching the MBs from the bottom coverslip.

3 Materials and methods

The sample consisted of a chamber sandwiched between two coverslips (Fig. 3). The chamber was made by cutting a desired shape in a one-millimeter-high foam tape. Whereas the bottom coverslip covered the entire chamber, the top coverslip left a small gap. At the beginning of each experiment, the chamber was filled with the metallic microbead suspension. During the experiment, the desired substance (either solid or liquid) could be added to the sample through the gap. Chambers of two different geometries were used. The first one was a simple circular chamber (Fig. 3a). According to its name, it had a single circular cavity (10 mm in diameter) of the volume of 78 µl. The second one was a bicircular chamber (Fig. 3b, c) consisting of two circular cavities (both 5 mm in diameter) connected with a channel. The total volume of bicircular chamber was about 41 µl. For the sake of clarity, the cavity with a gap is referred to as cavity A and the fully enclosed cavity is named cavity B.
The metallic microbead suspension was made by suspending 0.5 µl of dry silver-coated soda lime glass microspheres (Cospheric SLGMS-AG, particle size: 5–22 µm) in 1 ml of purified water and adding 0.1 µl of powdered BSA (bovine serum albumin). Metallic beads tend to sink and accumulate at the bottom glass of the sample. BSA was used for reducing the unwanted effect of beads attaching to the glass. Otherwise, it would be challenging to detach the beads with just the use of optical forces. The solution was stored in 4 °C and mixed for 5 min before use. During the experiments, the crystals of sucrose (saccharose, table sugar) were added to the chamber. The average mass of a typical crystal used in experiment was estimated at (1.0 ± 0.1) mg.

4 Measurement procedure

The measurement procedure in the bicircular chamber was as follows. In the first step, the sample was prepared–the chamber was cut and filled with metallic beads suspension. Then, in the desired region of the chamber, two beads of a similar size (8–12 µm) were trapped in two separate vortex traps. The distance between traps was about 70–80 µm. This separation (7–10 times as large as the bead) is large enough to prevent one bead affecting the other. The vortex charge m was adjusted to the bead size so as to achieve the stable rotational speed (Fig. 4b). For a narrow vortex ring (low charge), the rotations are very fast and may cause escaping the bead from the trap (Fig. 4a). On the other hand, wide ring (high charge) causes slow, nonuniform rotations (Fig. 4c). Since the measurements are based on tracking the change in rotational speed, its absolute value is irrelevant and may be different for both beads. Apart from inducing rotations, optical vortex also serves to lift the bead a few micrometers away from the bottom glass in order to eliminate bead-surface interaction. For the lifting to occur, the radiation pressure imparted on the bead needs to overcome the mass of the bead. In the oversized vortex (Fig. 4c), there is too little radiation pressure inside wide central dark disk and the bead stays at the bottom of the chamber. The vortex charge used for 8–12 µm beads ranged from m = 24 to m = 30.
In the third step, the recording of beads rotations starts. As a rule of thumb, after 4–5 min of undisturbed motion the rotations were considered stable which was confirmed by the visual inspection. Next, one sucrose crystal is added through the gap at the top of the chamber. As the sucrose dissolves, the viscosity inside the chamber increases. In reaction to the viscous drag growth, the rotations of the beads slow down. Comparing the reaction time for both beads one can find the direction and the speed of sucrose diffusion in the chamber. Fortunately, the beads are not perfectly smooth, but there are some impurities attached to their surface which make the rotations visible (e.g., see Online Resource 1). Therefore, the exact values of the angular velocity can be reconstructed straight from the recording. The movie was processed by hand as the rotational velocities were relatively small (of the order of one rotation per second). In case of faster rotations, one would probably need to process the movie automatically with some dedicated software. Depending on the rotational velocity, a proper frame rate of the recording should be provided in order to follow the bead motion. In our experiments, the sufficient recording speed was 60 fps.
Although the presented procedure refers to two-bead experiments, a similar treatment is used for the one-bead case (in circular chamber). Instead of two optical vortex traps, only one trap is generated. Since the laser power is no longer divided between two traps, there is greater radiation pressure (lifting force) in the single trap. This allows for using larger beads (12–16 µm) in one-bead experiments. Vortex charge ranged from m = 30 to m = 35.

5 Results

5.1 Circular chamber

A single rotating bead was located in the central region of the cavity (Fig. 5a, inset). Subsequent crystals of sucrose were added in 5-min intervals. There is a characteristic pattern in the graph shown in Fig. 5a. After each addition, the rotations of the bead speed up for about 20–40 s, then slow down substantially and speed up again until another addition occurs. These three identified phases of bead motion originate from three different effects. The crystal of sucrose put into the chamber triggers a mechanical disturbance in liquid resembling a stone thrown into the water. However, if instead the insoluble substance is added to the chamber, for example, a grain of sand, the generated disturbance is many times weaker. It led us to conclusion that the induced mechanical wave is additionally driven by the initiated dissolution. This assumption was confirmed for other soluble substances—glycerol and KMnO4.The mechanical wave propagates uniformly from the crystal location. Thereby, the wave reaching the bead position is directed toward the rear wall of the chamber. The bead gets displaced from the center of the trap and is pushed into the bright ring of optical vortex. Due to the high intensity of light in the ring, angular momentum transfer to the bead increases. Consequently, greater torque is exerted on the bead leading to the observed angular acceleration in the first phase. Simultaneously, there is another wave, the “viscosity wave” of the dissolved sugar. Compared to the mechanical wave, the propagation speed of the viscosity wave is many times lower due to the slow process of dissolution (usually about 2–3 min for one crystal to dissolve completely). This explains the delay in bead reaction. The sudden drop in rotational speed indicates the arrival of the high solute concentration at the bead location. In the third phase, a continuous increase in angular velocity is observed. Due to the ongoing diffusion, the sucrose extends to the entire volume of the chamber. At the bead position, local sucrose concentration decreases until the solution becomes homogeneous and angular velocity stabilizes. In Fig. 5a, the stabilization point was not reached because of the insufficient time intervals between subsequent crystal additions.
Changes in angular velocity of the bead can be recalculated into changes in viscosity of the medium. Viscous torque (drag torque) acting upon a rotating bead equals (Landau 1987)
$$\tau_{{{\text{drag}}}} = - 8\pi \mu r^{3} {\Omega }$$
(1)
where µ—viscosity of a medium, r–bead radius, Ω–angular velocity of a bead. In the described experiment (Fig. 5a), velocity changes were slow (maximal angular deceleration of about 0.004 rad/s2) and nearly linear in all three phases of bead motion. Thus, the entire process is quasi-static and the drag force remains constant. According to Eq. (1), for constant drag torque viscosity is inversely proportional to angular velocity—an increase in viscosity causes a proportional decrease in velocity. Relative viscosity of the medium (Fig. 5b) can be obtained from the measured angular velocity Ω(t) (Fig. 5a) in the following way:
$$\frac{\mu \left( t \right)}{{\mu_{0} }} = \frac{{{\Omega }_{0} }}{{{\Omega }\left( {\text{t}} \right)}}$$
(2)
where µ0 and Ω0 are respectively viscosity and angular velocity at the reference point. The reference point was the moment just before the first sucrose crystal addition (188. second of the experiment). Earlier fluctuations of the velocity and thus the calculated viscosity are the artifacts and should be neglected. As mentioned before, time intervals (about 5 min) between each addition were too short for velocity to reach a stabilization point. However, relative viscosity just before the next addition should be close to the target values. As shown in Fig. 5b, relative viscosity just before second, third and fourth blue dot lied within or close to the expected values range (blue strips). On the other hand, evaporation of water occurred during the measurement. Water evaporated from the gap region of the chamber until the substrate in the open area got dry. Due to the capillary forces, the fluid between two coverslips did not flood the dry substrate. As the evaporation continued, concave meniscus appeared and then penetrated the fluid layer. It was estimated that the meniscus started to form after about 20 min of the experiment. Knowing the evaporation time of the water volume in the open area (about 7.3 µl), one can calculate the modified expected viscosity values of the 1-, 2- and 3-crystal solution (red dashed line). For details, see Supplementary Information. The experiment was rerun seven times and all the observations proved repeatable, including the behavior of the mechanical wave and the viscosity wave.
The major technical difficulty connected with the circular chamber is the strong mechanical disturbance after the addition of sucrose crystal. If the bead gets pushed out of the vortex trap (outside the bright ring), the experiment is disrupted and needs to be repeated. This problem can be resolved with the use of the bicircular chamber, for which the mechanical wave generated in cavity A is separated from the cavity B with the relatively narrow channel. This way the stability of the bead is improved and allows for two-bead experiments.

5.2 Bicircular chamber

Initially, the aim of the two-bead experiments in the bicircular chamber was to estimate the propagation speed of the viscosity wave observed in the circular chamber. Meanwhile, the observed effect of distinctive solute flow provoked an experimental investigation on its origin.
In the first experiment, the beads were located in the center of cavity B and arranged along the chamber axis (Fig. 6a, inset). Therefore, one bead (black) was closer to the exit of the channel while the other (red) was closer to the rear wall of the cavity. One crystal of sugar was added to the cavity A. The first reaction to the viscosity wave visible as a significant drop in angular velocity was noted after about 40 s (Fig. 6a). Surprisingly, the earlier reaction came from red bead. In other words, the bead situated further away from the exit of the channel reacted first. According to the dynamics of standard, uniform diffusion, the sugar should spread in the cavity A, go through the channel and then fill the cavity B starting from the exit of the channel. The experiment was repeated several times and the equivalent, nonintuitive observations were made. The relative size of two beads (one slightly bigger than the other or equally sized) had no influence on the results. In order to exclude the occurrence of any direction-related rotational effect, the experiment was repeated for the two beads rotating in opposite directions. Reversing the handedness of the optical vortex is done by changing its topological charge to the opposite value, e.g., from + 30 to − 30. However, the results (Fig. 6b) confirmed previous observations—again, the red bead reacted earlier.
In the second experiment, the beads were located inside the channel and arranged in line perpendicular to the chamber axis (Fig. 7a, inset). This way one bead was closer to the center of the channel, while the other was closer to the side wall of the channel. As shown in Fig. 7a, first reaction to the viscosity wave came from the bead situated at the center of the channel (black bead). This observation corresponds to the expected pattern of the laminar flow through the channel.
In the third experiment, the beads were located near the side wall of cavity B and arranged in line perpendicular to the chamber axis (Fig. 7b, inset). It means that one bead was closer to the center of the cavity while the other was closer to its side wall. The outcome of the experiment is presented in Fig. 7b. Soon after the addition of sucrose, there is a clear increase in angular velocity due to the mechanical wave followed by the velocity drop at the arrival of the viscosity wave. The earlier reaction was noted for the red, near-the-wall bead, which is again a peculiar observation likewise in the experiment in Fig. 6.
All the bicircular chamber measurements described in the manuscript were rerun 15 times. The reported observations were confirmed in 11 trials. The remaining attempts were inconclusive as low angular velocity of one or both beads made it impossible to judge which bead reacted first. In fact, angular velocity can be regarded as sampling frequency and thus the higher the velocity, the better the time resolution in the measurement.

6 Discussion

It has been speculated that the surprising diffusion dynamics in the bicircular chamber is a heat-related effect stemming from the halogen lamp used as an external white light illumination (see Fig. 2). Even though the measurements were performed in very low illumination intensity, halogen lamp could locally heat up the sample thus inducing convection currents. In order to verify this possibility, halogen lamp was replaced with a cold light source (LED illumination) and the bicircular chamber experiment with both beads in the center of cavity B was repeated (Fig. 8). The arrangement of beads was equivalent to the one in Fig. 6, and the only difference lied in the illumination source. The results remained unchanged—the red bead was first to react to the viscosity wave. Therefore, the thermal origin of the phenomenon was disqualified. Interestingly, the lack of heating visibly slowed down the dissolution process. The beads reacted to sucrose after about 60 s, while for halogen illumination, the reaction time ranged from 30 to 40 s.
It should be noted that the observed effect of nonintuitive diffusion is not limited to sucrose. Analogous experiments were conducted for glycerol—instead of adding one crystal of sucrose, 0.2 µl of 50% aqueous glycerol solution was injected into the cavity A of the bicircular chamber (Fig. 9). The two rotating beads were located at the center of the cavity B and their reactions to were compatible with the ones obtained for sucrose.
The attempt was made to observe the trajectory of the diffusion directly using two imaging techniques. The first one was phase contrast microscopy (Burch 1942, Pluta 1988) suitable for specimens which modulate phase of light instead of its intensity. In principle, the higher sucrose concentration, the higher the refractive index and therefore the stronger phase modulation. Thus, the regions of high and low sucrose concentration should appear differently (in different colors) under the phase contrast microscope. The observations were made at 10 × and 20 × magnification. Unfortunately, the resolution of the detectable phase change was too low and the diffusion flow pattern could not be determined. Another tested imaging technique was polarized light microscopy (Pluta 1988) dedicated for optically active materials, i.e., materials which rotate the plane of polarization of transmitted light. Optical activity results from asymmetry in material structure (in sucrose molecule due to asymmetric carbon atoms). The angle of the polarization rotation is proportional to sugar concentration and chamber depth. Since the bicircular chamber was only 1 mm deep and one sucrose crystal raised the total concentration by about 2%, the resolution of polarized light microscope was insufficient to observe the ongoing diffusion process.
Although the direct imaging of phase change failed, an alternative approach gave important insight into the diffusion dynamics. During dissolution of sucrose, small particles of contaminants, most likely dust from the crystal surface, were noticed under the microscope in the optical tweezers setup. These particles were barely visible at the bottom of the chamber where the metallic beads were located. However, they came into focus when the observation plane lied slightly below the middle depth of the chamber. Refocusing from the bottom to middle surface could not be made with 100 × objective used for optical trapping due to its short working distance. Applying 20 × objective revealed the stream of particles spreading well above the beads level. Based on this finding, a plausible explanation of the peculiar impression of the reverse solute flow was concluded and sketched in Fig. 10. The stream of dissolved sucrose propagates about 250–300 µm above the bottom glass along the entire length of the chamber. During its way from cavity A to B, it causes the forward movement of the nearby layers of water and therefore creates the mechanical wave. This is consistent with the finding that the mechanical wave is driven by the diffusion (see Sect. 5.1). Free beads are carried by the mechanical wave, which is directly observed in the experiment (Online Resource 1), while trapped beads are displaced from the trap centers into the bright ring, hence their slight angular acceleration. At the same time, the mid-depth solute stream reaches cavity B, gets reflected from the rear chamber wall and splits into upper and lower streams. The lower stream is the above-mentioned viscosity wave which causes sudden drop in beads rotational speed. For beads located in the center of cavity B, the lower stream arrives at their position “from behind,” i.e., in the opposite direction to the mid-depth flow, hence the misleading observation of the reverse solute flow. The similar happens for the beads located at the side wall of cavity B. In the case of beads placed in the channel, the trend is reversed—the bead further away from the channel wall reacts earlier—due to the slower stream flow near the channel wall (according to the laminar flow).
The temporal resolution of our measurements is somehow dictated by angular velocity of the bead. While measuring the period of each subsequent rotation, the “sampling frequency” and thus the temporal resolution equals the period (in our setup it is about 0.5–1 s). This approach is particularly useful while processing the recording by hand. On the other hand, reading the angular bead position more frequently, e.g., every quarter turn or even at each frame, leads to the increase in temporal resolution, but requires the automation of data processing. Another idea to improve the temporal resolution is to induce faster bead rotations, either by increasing the laser power or by using smaller beads (smaller drag force, see Eq. 1). The better the resolution, the more rapid viscosity changes can be detected.
The spatial resolution of our method is limited by the interaction between two beads located close to each other. It has been estimated that the separation of two-bead diameters between bead centers is enough to neglect their mutual influence. A similar observation was reported by Curran et al. (2012) for beads oscillating along a line (instead of rotating). However, in order to guarantee no interaction, the “safe” distance was assessed at three times the diameter, i.e., about 24–30 µm for 8–10 µm beads. The spatial resolution could be improved by using smaller beads.
In order to support the results with numerical or theoretical analysis, one would need a relevant model of the experimental setup. However, neither Monte Carlo modeling nor numerical solution of the differential equations (via finite element method) reconstructed the experimental observations. One of the problems here is the boundary effects due to capillary forces. Another complex issue is the mechanical wave and its strict relation to sucrose dissolution (as mentioned in Sect. 5.1 the effect diminished for insoluble substances). Thus, theoretical modeling of our system seems to be a rather nontrivial task. Experiments described in this paper are a starting point for the development of more rigorous yet much more demanding approach, which is a main goal for our future research.

7 Conclusions

In this paper, two-laser optical tweezers were used for trapping metallic microbeads in optical vortex traps. The trapped beads underwent regular rotations. Their angular velocity (either absolute or relative between two simultaneously rotating beads) became the source of information about the medium surrounding the beads. The method of tracing the direction and speed of solute diffusion process was proposed. The experimental verification of the method concerned the dynamics of sucrose crystal dissolution in two types of measurement chambers. The circular chamber experiments showed the presence of mechanical wave due to the first phase of dissolution process. This wave is followed by the stream of higher sugar concentration, which results in a rapid drop in MB rotational speed. The results obtained for bicircular chamber were surprising and nonintuitive as they indicated the effect of sucrose diffusion in the form of directional stream in the middle depth of the measurement chamber rather than the uniform solute dispersion. The observed effect was related neither to heat transfer nor to the interaction between two MBs. It also applied to glycerol. Unfortunately, the experiment could not be rerun for salts (attempts were made for NaCl and KI) since the chemical interaction between the salt and thin metal coating on the bead leads to the fast degradation of the layer and precludes the measurements.
The provided explanation of the observed effect cannot be considered definitive. Nevertheless, all the repetitive experiments with trapped MBs as well as the movement of free MBs and contaminant particles observed directly under the microscope led us to these conclusions as the most relevant so far.
The bottom line is that the two trapped MBs separated by tens of micrometers can be used to detect minor viscosity changes occurring dynamically in a very small volume of liquid. It may lead to the discovery of some counterintuitive effects as the one reported in this paper.

Declaration

Conflict of interest

The authors declare that they have no conflict of interest.
Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​.

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Anhänge

Supplementary Information

Below is the link to the electronic supplementary material.
Supplementary file1 (MP4 1490 KB)
Literatur
Zurück zum Zitat Burch CR, Stock JPP (1942) Phase-contrast microscopy. J Sci Instrum 19(5):71–75CrossRef Burch CR, Stock JPP (1942) Phase-contrast microscopy. J Sci Instrum 19(5):71–75CrossRef
Zurück zum Zitat Czarske J, Büttner L (2015) Micro laser Doppler velocimetry (μ-LDV). In: Li D (ed) Encyclopedia of microfluidics and nanofluidics. Springer, New York Czarske J, Büttner L (2015) Micro laser Doppler velocimetry (μ-LDV). In: Li D (ed) Encyclopedia of microfluidics and nanofluidics. Springer, New York
Zurück zum Zitat Jones PH, Maragó M, Onofrio VG (2015) Optical Tweezers principles and applications. Cambridge University Press, CambridgeCrossRef Jones PH, Maragó M, Onofrio VG (2015) Optical Tweezers principles and applications. Cambridge University Press, CambridgeCrossRef
Zurück zum Zitat Landau LD, Lifshitz EM (1987) Fluid mechanics. Course of theoretical physics, 2nd edn. Pergamon Press, Oxford Landau LD, Lifshitz EM (1987) Fluid mechanics. Course of theoretical physics, 2nd edn. Pergamon Press, Oxford
Zurück zum Zitat Pluta M (1988) Advanced Light Microscopy. PWN, Warsaw Pluta M (1988) Advanced Light Microscopy. PWN, Warsaw
Zurück zum Zitat Soskin M, Vasnetsov MV (2001) Singular optics, Prog Opt. North-Holland Publishing Co, Amsterdam Soskin M, Vasnetsov MV (2001) Singular optics, Prog Opt. North-Holland Publishing Co, Amsterdam
Zurück zum Zitat Torres JP, Torner L (2011) Twisted photons: applications of light with orbital angular momentum. Wiley-VCH, WeinheimCrossRef Torres JP, Torner L (2011) Twisted photons: applications of light with orbital angular momentum. Wiley-VCH, WeinheimCrossRef
Zurück zum Zitat Vasnetsov M, Staliunas K (1999) Optical vortices. Nova Science, Huntington Vasnetsov M, Staliunas K (1999) Optical vortices. Nova Science, Huntington
Metadaten
Titel
Microscale solute flow probed with rotating microbead trapped in optical vortex
verfasst von
Weronika Lamperska
Jan Masajada
Sławomir Drobczyński
Publikationsdatum
01.06.2021
Verlag
Springer Berlin Heidelberg
Erschienen in
Experiments in Fluids / Ausgabe 6/2021
Print ISSN: 0723-4864
Elektronische ISSN: 1432-1114
DOI
https://doi.org/10.1007/s00348-021-03223-6

Weitere Artikel der Ausgabe 6/2021

Experiments in Fluids 6/2021 Zur Ausgabe

    Marktübersichten

    Die im Laufe eines Jahres in der „adhäsion“ veröffentlichten Marktübersichten helfen Anwendern verschiedenster Branchen, sich einen gezielten Überblick über Lieferantenangebote zu verschaffen.