Abstract

Purpose

In minimally invasive interventions assisted by C-arm imaging, there is a demand to fuse the intra-interventional 2D C-arm image with pre-interventional 3D patient data to enable surgical guidance. The commonly used intensity-based 2D/3D registration has a limited capture range and is sensitive to initialization. We propose to utilize an opto/X-ray C-arm system which allows to maintain the registration during intervention by automating the re-initialization for the 2D/3D image registration. Consequently, the surgical workflow is not disrupted and the interaction time for manual initialization is eliminated.

Methods

We utilize two distinct vision-based tracking techniques to estimate the relative poses between different C-arm arrangements: (1) global tracking using fused depth information and (2) RGBD SLAM system for surgical scene tracking. A highly accurate multi-view calibration between RGBD and C-arm imaging devices is achieved using a custom-made multimodal calibration target.

Results

Several in vitro studies are conducted on pelvic-femur phantom that is encased in gelatin and covered with drapes to simulate a clinically realistic scenario. The mean target registration errors (mTRE) for re-initialization using depth-only and RGB \(+\) depth are 13.23 mm and 11.81 mm, respectively. 2D/3D registration yielded 75% success rate using this automatic re-initialization, compared to a random initialization which yielded only 23% successful registration.

Conclusion

The pose-aware C-arm contributes to the 2D/3D registration process by globally re-initializing the relationship of C-arm image and pre-interventional CT data. This system performs inside-out tracking, is self-contained, and does not require any external tracking devices.

Sie haben noch keine Lizenz? Dann Informieren Sie sich jetzt über unsere Produkte:

Springer Professional "Wirtschaft+Technik"

Online-Abonnement

Mit Springer Professional "Wirtschaft+Technik" erhalten Sie Zugriff auf:

  • über 102.000 Bücher
  • über 537 Zeitschriften

aus folgenden Fachgebieten:

  • Automobil + Motoren
  • Bauwesen + Immobilien
  • Business IT + Informatik
  • Elektrotechnik + Elektronik
  • Energie + Nachhaltigkeit
  • Finance + Banking
  • Management + Führung
  • Marketing + Vertrieb
  • Maschinenbau + Werkstoffe
  • Versicherung + Risiko

Jetzt Wissensvorsprung sichern!

Springer Professional "Wirtschaft"

Online-Abonnement

Mit Springer Professional "Wirtschaft" erhalten Sie Zugriff auf:

  • über 67.000 Bücher
  • über 340 Zeitschriften

aus folgenden Fachgebieten:

  • Bauwesen + Immobilien
  • Business IT + Informatik
  • Finance + Banking
  • Management + Führung
  • Marketing + Vertrieb
  • Versicherung + Risiko




Jetzt Wissensvorsprung sichern!

Springer Professional "Technik"

Online-Abonnement

Mit Springer Professional "Technik" erhalten Sie Zugriff auf:

  • über 67.000 Bücher
  • über 390 Zeitschriften

aus folgenden Fachgebieten:

  • Automobil + Motoren
  • Bauwesen + Immobilien
  • Business IT + Informatik
  • Elektrotechnik + Elektronik
  • Energie + Nachhaltigkeit
  • Maschinenbau + Werkstoffe




 

Jetzt Wissensvorsprung sichern!

Literatur
1.
Zurück zum Zitat Bay H, Tuytelaars T, Van Gool L (2006) Surf: speeded up robust features. In: European conference on computer vision. Springer, Berlin, pp 404–417 Bay H, Tuytelaars T, Van Gool L (2006) Surf: speeded up robust features. In: European conference on computer vision. Springer, Berlin, pp 404–417
2.
Zurück zum Zitat Van der Bom M, Bartels L, Gounis M, Homan R, Timmer J, Viergever M, Pluim J (2010) Robust initialization of 2D–3D image registration using the projection-slice theorem and phase correlation. Med Phys 37(4):1884–1892CrossRefPubMed Van der Bom M, Bartels L, Gounis M, Homan R, Timmer J, Viergever M, Pluim J (2010) Robust initialization of 2D–3D image registration using the projection-slice theorem and phase correlation. Med Phys 37(4):1884–1892CrossRefPubMed
3.
Zurück zum Zitat Endres F, Hess J, Engelhard N, Sturm J, Cremers D, Burgard W (2012) An evaluation of the RGB-D SLAM system. In: 2012 IEEE international conference on robotics and automation (ICRA). IEEE, pp 1691–1696 (2012) Endres F, Hess J, Engelhard N, Sturm J, Cremers D, Burgard W (2012) An evaluation of the RGB-D SLAM system. In: 2012 IEEE international conference on robotics and automation (ICRA). IEEE, pp 1691–1696 (2012)
4.
Zurück zum Zitat Euler E, Heining S, Riquarts C, Mutschler W (2003) C-arm-based three-dimensional navigation: a preliminary feasibility study. Comput Aided Surg 8(1):35–41CrossRefPubMed Euler E, Heining S, Riquarts C, Mutschler W (2003) C-arm-based three-dimensional navigation: a preliminary feasibility study. Comput Aided Surg 8(1):35–41CrossRefPubMed
5.
Zurück zum Zitat Fischer M, Fuerst B, Lee SC, Fotouhi J, Habert S, Weidert S, Euler E, Osgood G, Navab N (2016) Preclinical usability study of multiple augmented reality concepts for K-wire placement. Int J Comput Assist Radiol Surg 11(6):1007–1014CrossRefPubMed Fischer M, Fuerst B, Lee SC, Fotouhi J, Habert S, Weidert S, Euler E, Osgood G, Navab N (2016) Preclinical usability study of multiple augmented reality concepts for K-wire placement. Int J Comput Assist Radiol Surg 11(6):1007–1014CrossRefPubMed
6.
Zurück zum Zitat Fischler MA, Bolles RC (1981) Random sample consensus: a paradigm for model fitting with applications to image analysis and automated cartography. Commun ACM 24(6):381–395CrossRef Fischler MA, Bolles RC (1981) Random sample consensus: a paradigm for model fitting with applications to image analysis and automated cartography. Commun ACM 24(6):381–395CrossRef
7.
Zurück zum Zitat Fotouhi J, Fuerst B, Lee SC, Keicher M, Fischer M, Weidert S, Euler E, Navab N, Osgood G (2016) Interventional 3D augmented reality for orthopedic and trauma surgery. In: 16th annual meeting of the international society for computer assisted orthopedic surgery (CAOS) Fotouhi J, Fuerst B, Lee SC, Keicher M, Fischer M, Weidert S, Euler E, Navab N, Osgood G (2016) Interventional 3D augmented reality for orthopedic and trauma surgery. In: 16th annual meeting of the international society for computer assisted orthopedic surgery (CAOS)
8.
9.
Zurück zum Zitat Fuerst B, Fotouhi J, Navab N (2015) Vision-based intraoperative cone-beam ct stitching for non-overlapping volumes. In: International conference on medical image computing and computer-assisted intervention. Springer, Berlin, pp 387–395 Fuerst B, Fotouhi J, Navab N (2015) Vision-based intraoperative cone-beam ct stitching for non-overlapping volumes. In: International conference on medical image computing and computer-assisted intervention. Springer, Berlin, pp 387–395
10.
Zurück zum Zitat Geiger A, Moosmann F, Car Ö, Schuster B (2012) Automatic camera and range sensor calibration using a single shot. In: 2012 IEEE international conference on robotics and automation (ICRA). IEEE, pp 3936–3943 Geiger A, Moosmann F, Car Ö, Schuster B (2012) Automatic camera and range sensor calibration using a single shot. In: 2012 IEEE international conference on robotics and automation (ICRA). IEEE, pp 3936–3943
11.
Zurück zum Zitat Gendrin C, Furtado H, Weber C, Bloch C, Figl M, Pawiro SA, Bergmann H, Stock M, Fichtinger G, Georg D (2012) Monitoring tumor motion by real time 2D/3D registration during radiotherapy. Radiother Oncol 102(2):274–280CrossRefPubMed Gendrin C, Furtado H, Weber C, Bloch C, Figl M, Pawiro SA, Bergmann H, Stock M, Fichtinger G, Georg D (2012) Monitoring tumor motion by real time 2D/3D registration during radiotherapy. Radiother Oncol 102(2):274–280CrossRefPubMed
12.
Zurück zum Zitat Glassner AS (1989) An introduction to ray tracing. Elsevier, Amsterdam Glassner AS (1989) An introduction to ray tracing. Elsevier, Amsterdam
13.
Zurück zum Zitat Goitein M, Abrams M, Rowell D, Pollari H, Wiles J (1983) Multi-dimensional treatment planning: Ii. beam’s eye-view, back projection, and projection through ct sections. Int J Radiat Oncol Biol Phys 9(6):789–797CrossRefPubMed Goitein M, Abrams M, Rowell D, Pollari H, Wiles J (1983) Multi-dimensional treatment planning: Ii. beam’s eye-view, back projection, and projection through ct sections. Int J Radiat Oncol Biol Phys 9(6):789–797CrossRefPubMed
14.
Zurück zum Zitat Gong RH, Güler Ö, Kürklüoglu M, Lovejoy J, Yaniv Z (2013) Interactive initialization of 2D/3D rigid registration. Med Phys 40(12):121911CrossRefPubMed Gong RH, Güler Ö, Kürklüoglu M, Lovejoy J, Yaniv Z (2013) Interactive initialization of 2D/3D rigid registration. Med Phys 40(12):121911CrossRefPubMed
15.
Zurück zum Zitat Grisetti G, Kümmerle R, Stachniss C, Frese U, Hertzberg C (2010) Hierarchical optimization on manifolds for online 2D and 3D mapping. In: 2010 IEEE international conference on robotics and automation (ICRA). IEEE, pp 273–278 Grisetti G, Kümmerle R, Stachniss C, Frese U, Hertzberg C (2010) Hierarchical optimization on manifolds for online 2D and 3D mapping. In: 2010 IEEE international conference on robotics and automation (ICRA). IEEE, pp 273–278
16.
Zurück zum Zitat Jain A, Fichtinger G (2006) C-arm tracking and reconstruction without an external tracker. In: International conference on medical image computing and computer-assisted intervention. Springer, Berlin, pp 494–502 (2006) Jain A, Fichtinger G (2006) C-arm tracking and reconstruction without an external tracker. In: International conference on medical image computing and computer-assisted intervention. Springer, Berlin, pp 494–502 (2006)
17.
Zurück zum Zitat Knaan D, Joskowicz L (2003) Effective intensity-based 2D/3D rigid registration between fluoroscopic X-ray and CT. In: International conference on medical image computing and computer-assisted intervention. Springer, Berlin, pp 351–358 Knaan D, Joskowicz L (2003) Effective intensity-based 2D/3D rigid registration between fluoroscopic X-ray and CT. In: International conference on medical image computing and computer-assisted intervention. Springer, Berlin, pp 351–358
18.
Zurück zum Zitat Kümmerle R, Grisetti G, Strasdat H, Konolige K, Burgard W (2011) g2o: a general framework for graph optimization. In: 2011 IEEE international conference on robotics and automation (ICRA). IEEE, pp 3607–3613 Kümmerle R, Grisetti G, Strasdat H, Konolige K, Burgard W (2011) g2o: a general framework for graph optimization. In: 2011 IEEE international conference on robotics and automation (ICRA). IEEE, pp 3607–3613
19.
Zurück zum Zitat Lee SC, Fuerst B, Fotouhi J, Fischer M, Osgood G, Navab N (2016) Calibration of RGBD camera and cone-beam CT for 3D intra-operative mixed reality visualization. Int J Comput Assist Radiol Surg 11(6):967–975CrossRefPubMed Lee SC, Fuerst B, Fotouhi J, Fischer M, Osgood G, Navab N (2016) Calibration of RGBD camera and cone-beam CT for 3D intra-operative mixed reality visualization. Int J Comput Assist Radiol Surg 11(6):967–975CrossRefPubMed
20.
Zurück zum Zitat Lemieux L, Jagoe R, Fish D, Kitchen N, Thomas D (1994) A patient-to-computed-tomography image registration method based on digitally reconstructed radiographs. Med Phys 21(11):1749–1760CrossRefPubMed Lemieux L, Jagoe R, Fish D, Kitchen N, Thomas D (1994) A patient-to-computed-tomography image registration method based on digitally reconstructed radiographs. Med Phys 21(11):1749–1760CrossRefPubMed
21.
Zurück zum Zitat Mitrović U, Pernuš F, Likar B, Špiclin Ž (2015) Simultaneous 3D–2D image registration and c-arm calibration: application to endovascular image-guided interventions. Med Phys 42(11):6433–6447CrossRefPubMed Mitrović U, Pernuš F, Likar B, Špiclin Ž (2015) Simultaneous 3D–2D image registration and c-arm calibration: application to endovascular image-guided interventions. Med Phys 42(11):6433–6447CrossRefPubMed
22.
Zurück zum Zitat Navab N, Bani-Hashemi A, Nadar MS, Wiesent K, Durlak P, Brunner T, Barth K, Graumann R (1998) 3D reconstruction from projection matrices in a c-arm based 3D-angiography system. In: International conference on medical image computing and computer-assisted intervention. Springer, Berlin, pp 119–129 Navab N, Bani-Hashemi A, Nadar MS, Wiesent K, Durlak P, Brunner T, Barth K, Graumann R (1998) 3D reconstruction from projection matrices in a c-arm based 3D-angiography system. In: International conference on medical image computing and computer-assisted intervention. Springer, Berlin, pp 119–129
23.
Zurück zum Zitat Newcombe RA, Izadi S, Hilliges O, Molyneaux D, Kim D, Davison AJ, Kohi P, Shotton J, Hodges S, Fitzgibbon A (2011) Kinectfusion: real-time dense surface mapping and tracking. In: 2011 10th IEEE international symposium on mixed and augmented reality (ISMAR). IEEE, pp 127–136 Newcombe RA, Izadi S, Hilliges O, Molyneaux D, Kim D, Davison AJ, Kohi P, Shotton J, Hodges S, Fitzgibbon A (2011) Kinectfusion: real-time dense surface mapping and tracking. In: 2011 10th IEEE international symposium on mixed and augmented reality (ISMAR). IEEE, pp 127–136
24.
Zurück zum Zitat Otake Y, Armand M, Armiger RS, Kutzer MD, Basafa E, Kazanzides P, Taylor RH (2012) Intraoperative image-based multiview 2D/3D registration for image-guided orthopaedic surgery: incorporation of fiducial-based c-arm tracking and gpu-acceleration. IEEE Trans Med Imaging 31(4):948–962CrossRefPubMed Otake Y, Armand M, Armiger RS, Kutzer MD, Basafa E, Kazanzides P, Taylor RH (2012) Intraoperative image-based multiview 2D/3D registration for image-guided orthopaedic surgery: incorporation of fiducial-based c-arm tracking and gpu-acceleration. IEEE Trans Med Imaging 31(4):948–962CrossRefPubMed
25.
Zurück zum Zitat Otake Y, Armand M, Sadowsky O, Armiger RS, Kutzer MD, Mears SC, Kazanzides P, Taylor, RH (2010) An image-guided femoroplasty system: development and initial cadaver studies. In: SPIE medical imaging. International Society for Optics and Photonics, pp 76250P–76250P (2010) Otake Y, Armand M, Sadowsky O, Armiger RS, Kutzer MD, Mears SC, Kazanzides P, Taylor, RH (2010) An image-guided femoroplasty system: development and initial cadaver studies. In: SPIE medical imaging. International Society for Optics and Photonics, pp 76250P–76250P (2010)
26.
Zurück zum Zitat Otake Y, Schafer S, Stayman J, Zbijewski W, Kleinszig G, Graumann R, Khanna A, Siewerdsen J (2012) Automatic localization of vertebral levels in X-ray fluoroscopy using 3D–2D registration: a tool to reduce wrong-site surgery. Phys Med Biol 57(17):5485CrossRefPubMedPubMedCentral Otake Y, Schafer S, Stayman J, Zbijewski W, Kleinszig G, Graumann R, Khanna A, Siewerdsen J (2012) Automatic localization of vertebral levels in X-ray fluoroscopy using 3D–2D registration: a tool to reduce wrong-site surgery. Phys Med Biol 57(17):5485CrossRefPubMedPubMedCentral
27.
Zurück zum Zitat Powell MJ (2009) The BOBYQA algorithm for bound constrained optimization without derivatives. Cambridge NA Report NA2009/06, University of Cambridge, Cambridge (2009) Powell MJ (2009) The BOBYQA algorithm for bound constrained optimization without derivatives. Cambridge NA Report NA2009/06, University of Cambridge, Cambridge (2009)
28.
Zurück zum Zitat Rusinkiewicz S, Levoy M (2001) Efficient variants of the ICP algorithm. In: Proceedings. Third international conference on 3-D digital imaging and modeling, 2001. IEEE, pp 145–152 Rusinkiewicz S, Levoy M (2001) Efficient variants of the ICP algorithm. In: Proceedings. Third international conference on 3-D digital imaging and modeling, 2001. IEEE, pp 145–152
29.
Zurück zum Zitat Sadowsky O, Chintalapani G, Taylor RH (2007) Deformable 2D-3D registration of the pelvis with a limited field of view, using shape statistics. In: International conference on medical image computing and computer-assisted intervention. Springer, Berlin, pp 519–526 (2007) Sadowsky O, Chintalapani G, Taylor RH (2007) Deformable 2D-3D registration of the pelvis with a limited field of view, using shape statistics. In: International conference on medical image computing and computer-assisted intervention. Springer, Berlin, pp 519–526 (2007)
30.
Zurück zum Zitat Uneri A, Otake Y, Wang A, Kleinszig G, Vogt S, Khanna AJ, Siewerdsen J (2013) 3D–2D registration for surgical guidance: effect of projection view angles on registration accuracy. Phys Med Biol 59(2):271CrossRefPubMedPubMedCentral Uneri A, Otake Y, Wang A, Kleinszig G, Vogt S, Khanna AJ, Siewerdsen J (2013) 3D–2D registration for surgical guidance: effect of projection view angles on registration accuracy. Phys Med Biol 59(2):271CrossRefPubMedPubMedCentral
31.
Zurück zum Zitat Varnavas A, Carrell T, Penney G (2013) Increasing the automation of a 2D–3D registration system. IEEE Trans Med Imaging 32(2):387–399CrossRefPubMed Varnavas A, Carrell T, Penney G (2013) Increasing the automation of a 2D–3D registration system. IEEE Trans Med Imaging 32(2):387–399CrossRefPubMed
32.
Zurück zum Zitat Varnavas A, Carrell T, Penney G (2015) Fully automated 2D–3D registration and verification. Med Image Anal 26(1):108–119CrossRefPubMed Varnavas A, Carrell T, Penney G (2015) Fully automated 2D–3D registration and verification. Med Image Anal 26(1):108–119CrossRefPubMed
33.
Zurück zum Zitat Wein W (2003) Intensity based rigid 2D-3D registration algorithms for radiation therapy. Master’s thesis, Technische Universität München Wein W (2003) Intensity based rigid 2D-3D registration algorithms for radiation therapy. Master’s thesis, Technische Universität München
34.
Zurück zum Zitat Yang L, Wang J, Kobayashi E, Ando T, Yamashita H, Sakuma I, Chiba T (2015) Image mapping of untracked free-hand endoscopic views to an ultrasound image-constructed 3D placenta model. Int J Med Robot Comput Assist Surg 11(2):223–234CrossRef Yang L, Wang J, Kobayashi E, Ando T, Yamashita H, Sakuma I, Chiba T (2015) Image mapping of untracked free-hand endoscopic views to an ultrasound image-constructed 3D placenta model. Int J Med Robot Comput Assist Surg 11(2):223–234CrossRef
35.
Zurück zum Zitat Zhang Z (2000) A flexible new technique for camera calibration. IEEE Trans Pattern Anal Mach Intell 22(11):1330–1334CrossRef Zhang Z (2000) A flexible new technique for camera calibration. IEEE Trans Pattern Anal Mach Intell 22(11):1330–1334CrossRef
Metadaten
Titel
Pose-aware C-arm for automatic re-initialization of interventional 2D/3D image registration
verfasst von
Javad Fotouhi
Bernhard Fuerst
Alex Johnson
Sing Chun Lee
Russell Taylor
Greg Osgood
Nassir Navab
Mehran Armand
Publikationsdatum
19.05.2017
Verlag
Springer International Publishing
Erschienen in
International Journal of Computer Assisted Radiology and Surgery / Ausgabe 7/2017
Print ISSN: 1861-6410
Elektronische ISSN: 1861-6429
DOI
https://doi.org/10.1007/s11548-017-1611-8

Weitere Artikel der Ausgabe 7/2017

International Journal of Computer Assisted Radiology and Surgery 7/2017 Zur Ausgabe