Topology optimised metallic bone plates produced by electron beam melting: a mechanical and biological study

verfasst von: Abdulsalam Abdulaziz Al-Tamimi, Boyang Huang, Cian Vyas, Miguel Hernandez, Chris Peach, Paulo Bartolo

Erschienen in: The International Journal of Advanced Manufacturing Technology | Ausgabe 1-4/2019

Aktivieren Sie unsere intelligente Suche, um passende Fachinhalte oder Patente zu finden.

search-config
loading …

Abstract

Metallic bone plates are commonly used as a medical implant to treat bone fractures. The gold standard materials for these implants are biocompatible 316L stainless steel, cobalt chromium, titanium and its alloys (e.g. CoCrMo and Ti6Al4V). However, the main disadvantage of these implants is the material stiffness mismatch between the implant and bone. This mismatch may negatively affect the biological processes in bone healing. This paper investigates topology optimization to produce plates with reduced equivalent stiffness and the fabrication of optimised plate designs using an electron beam melting (EBM) system. Non-post-processed EBM plates were assessed against commercially available bone fixation plates in terms of mechanical and biological characteristics. Results show that some redesigned produced plates present mechanical properties similar to the cortical bone and that there is no need to post-process the produced plates in order to establish a good biological bonding with the surrounding tissue.
Hinweise

Publisher’s note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

1 Introduction

Bone is a highly vascular and specialised tissue responsible for maintaining the shape of the skeleton, protecting soft tissues, transmitting the force of muscular contraction during movement, serving as a reservoir for ions and contributes to the regulation of the extracellular matrix, blood cell production and blood pH regulation [1].
Bone is prone to traumatic injuries as well as diseases. In the case of minor fractures, bone heals and remodels with little detectable scarring. However, in the case of pathological and traumatic fractures, traumatic bone loss or primary tumour resection, bone is not able to heal itself [24]. Bone fractures that are displaced or unstable are frequently surgically treated with bone fixation implants such as internal fixation devices, external fixators and intramedullary pins. In the case of internal fixation devices, the locking compression plate (LCP) is widely used to treat various types of fractures due to its hybrid screw system containing two means of fixation techniques. This allows the application of absolute stability or relative stability leading to both primary healing and secondary healing [5]. Similar to other internal fixation plates (e.g. dynamic compression plate and periarticular locking plate system), LCP is made from an inert and biocompatible metal. The most commonly used metallic materials are titanium and its alloys, stainless steel, or cobalt chrome and its alloys [6, 7].
The role of the internal fixation plates is to stabilise the fractured bone during the union/healing process. After the fracture is healed, the implant is no longer required to stay in the body. Currently, around 30% of the metallic implants have to be surgically removed after the healing process [8]. However, this procedure presents significant risks of bone refracture and nerve damage. Bone refracture is due to the lack of density of the bone in the close vicinity of the implant due to the significant mechanical differences between bone and fixation plates. For example, the Young’s modulus of Ti6Al4V is around 120 GPa whilst the Young’s modulus of trabecular bone ranges between 50 and 100 MPa and between 15 and 25 GPa in the case of cortical bone [7, 9, 10]. This Young’s modulus mismatch results in a phenomenon named stress shielding which is usually associated with implant loosening and bone loss. Alternatively, the implant can be left in the body, but this option also presents problems such as risks of periprosthetic fracture due the stress risers at the cortical bone and implant junctions, corrosion, release of metal ions and the risk of allergic and potentially carcinogenic reactions [11, 12].
Several researchers also explored the use of novel additive manufacturing technologies to produce fully personalised metallic implants. A notable example was the fabrication of a metallic complex jaw implant coated with bioceramic and implanted in an 83-year-old lady. Partners in Belgium and the Netherlands (universities of Leuven and Hasselt, AM bureau LayerWise) designed the implant to guarantee good attachment of muscles and space for nerves and used a powder-bed fusion system to produce it [13]. Other research groups also explored powder-bed fusion techniques (selective laser melting and electron beam melting) to produce personalised metallic bone plates without any concerns regarding the stress shielding effect [14]. Very few studies addressed this problem by changing the plate design (e.g. plate thickness) to improve its performance through a trial and error approach [15]. A systematic approach is still required. This paper presents an integrated approach linking computer modelling, simulation and optimisation together with additive manufacturing to address this problem. The following research questions are addressed:
  • Is it possible to use topology optimization to produce plates with reduced equivalent stiffness?
  • Is it possible to automatically produce optimised plate designs using an electron beam system?
  • Are the mechanical properties of the produced plates similar to the designed ones and similar to the cortical bone?
  • Is it necessary to post-process the produced plates in order to establish a good biological bonding with the surrounding tissue?

2 Initial bone plate design

There are no standard dimensions (i.e. length, width and thickness) or number of screw holes for commercially available bone fixation plates. Different commercial companies are also commercialising plates of the same type (e.g. locking compression plates) with different dimensions. Commercial bone plates from the same company present the same width and thickness but different lengths. Therefore, the number of distinct topologies (geometry and number of holes) is high.
In this research, the DePuy Synthes narrow locking compression plate (DePuy Synthes, Synthes GmbH, Switzerland) was considered as the reference design. Plates containing four, six and eight holes, with 180 mm in length, 14 mm in width and 5 mm in thickness, were defined (see Table 1) and considered for redesigning.
Table 1
The design domain of the bone plates
https://static-content.springer.com/image/art%3A10.1007%2Fs00170-019-03866-0/MediaObjects/170_2019_3866_Tab1_HTML.png

3 Computer modelling, simulation and optimisation

Plates were initially designed using Solidworks (Dassault Systèmes, France). Optimisation was conducted using the Abaqus Topology Optimization Module (ATOM) software (Dassault Systèmes, France), and the Solid Isotropic Microstructure with Penalization (SIMP) method proposed by Bendsøe [16]. Mathematically topology optimization is described as follows [17]:
$$ \genfrac{}{}{0pt}{}{\min }{\rho_{\mathrm{e}}}C\left({\rho}_{\mathrm{e}}\right)={f}^{\mathrm{T}}\bullet u $$
(1)
subject to
$$ \left\{\begin{array}{c}{\sum}_{\mathrm{e}=1}^N{\rho}_{\mathrm{e}}{v}_{\mathrm{e}}\le {V}^{\ast },\kern34.25em \left(1\mathrm{a}\right)\\ {}{\sum}_{\mathrm{e}=1}^{\mathrm{N}}{\rho}_{\mathrm{e}}^p{K}_{\mathrm{e}}u=f,\kern34em \left(1\mathrm{b}\right)\\ {}0<{\rho}_{\mathrm{o}}\le {\rho}_{\mathrm{e}}\le 1,\kern34.25em \left(1\mathrm{c}\right).\end{array}\right. $$
where C is the compliance, f is the force vector, Ke is the element stiffness matrix, u is the displacement vector, V* is the user-defined fraction volume, p is the penalisation factor, ρe is the element density, ρo is the initial density and ve is the volume of each element.
The iterative process to achieve the optimal design starts with the definition of a design domain, Ω, discretised into a set of elements with specific density (ρ) and Young’s modulus (E) values. Loading and boundary conditions are imposed, generating stresses and strains at each element, determining the corresponding sensitivity, i.e. the element that will have their density updated. Filtering techniques are implemented to restrict the optimisation from common disadvantages such as checkerboarding. The density is distributed along the design domain and updated after each iteration toward optimality according to the derivative of Eq. (1), considering for each element the value of ρ = 1 for full element stiffness and ρ = 0 for a void. The density update process is repeated until reaching the optimal solution.
Even though topology optimization is a load magnitude independent process, a post-operative case of a 70-kg patient was considered. Post-operation, patients suffering from a fractured tibia walk with crutches and their gait induces an overall load on the legs equivalent to 10% of the bodyweight [18]. Therefore, we assume a static compression force of 70 N applied on both far-end sides of the bone plate. Screw holes were considered a frozen design region. Ti6Al4V (Young’s modulus of 120 GPa and 0.3 Poisson’s ratio) was considered as the bone plates’ material. Different volume reductions (25, 45 and 75%) were considered and a mesh of approximately 50,000 8-node linear hexahedral elements was used for all considered geometries.
The redesigned plates were computationally assessed considering a static mechanical tensile test to determine their equivalent stiffness. In this case, a longitudinal force (Fxx) in the x-axis direction is applied to one side of the plate constraining the movement of the opposite side. Following a previously reported procedure [19], the equivalent stiffness of the designed plates can be determined as follows:
$$ K=\frac{RF_{xx}}{U_{xx}} $$
(2)
where K is the structural stiffness, RFxx is the reaction force on the constraint face and Uxx is the displacement.

4 Fabrication process

Bone fixation plates were prepared (e.g. part orientation, design of support structures and position in the building platform) using the software Magics (Materialise, Belgium) and exported as STL files to the Build Assembler software where the models were sliced and uploaded to the machine. Support structures were designed considering part stability and to reduce both thermal stresses and warping. The EBM Arcam A2 model (Arcam, Sweden) was used for producing the plates. Its schematic configuration is shown in Fig. 1a. The machine consists on an electron beam emitted from a hot tungsten filament, operates at 60 kV under vacuum pressure of 2.0 × 10–3 mBa, scanning speed of 4530 mm/s, beam focus offset of 3 mA, line offset of 0.1 mm and layer thickness of 50 μm. The substrate plate temperature was kept at ~ 600 °C and the build temperature was defined to be ~ 750 °C. Parts were produced using standard Ti6Al4V powder supplied by the machine manufacturer. The powder material consists of spherical gas atomised particles with a size raging between 45 and 100 μm (see Fig. 1b), composed of 6.04% of aluminium, 4.05% of vanadium, 0.013% of carbon, 0.0107% of iron, 0.13% of oxygen and balanced titanium.
The EBM fabrication process initiates with the entire titanium powder being scanned and preheated to 750 °C (80% of its melting temperature) with a low beam current of approximately 0.2 mA and high scanning speed up to 8000 m/s to minimise residual stresses. After fabrication, the powder recovery system was used to remove all trapped powder through blasting a stream of high-pressure air and the support structures were removed using pliers. Parts were not submitted to any post-processing operation or thermal treatment.

5 Characterisation of produced plates

5.1 Mechanical testing

Bone plates were mechanically assessed against the mechanical properties of cortical bone and compared to four commercially available titanium Depuy Synthes LCP 4.5/5.0 narrow bone plates with different screw holes (ten, nine, eight and six holes) purchased from DePuy Synthes (DePuy Synthes, Synthes GmbH, Switzerland). The test is a displacement controlled quasi-static tension using the Instron® 8862 system (Instron, MA, USA) with a maximum force of 100 kN. All tests were performed at room temperature until plate failure. Failure is determined by either plate crack, breakage or extreme plastic deformation, whichever occurred first. The test considered a 0.5-mm/min displacement rate. Strains and displacements were measured using a standard straight profile knife edged extensometer (2630-106, Instron, USA). Tests were performed in duplicate. Force and displacement histories were documented at a sampling rate of 10 Hz, and the mean slope of the curve (i.e. stiffness) measured. The equivalent Young’s modulus was obtained through the following procedure:
i)
From the tensile test, and considering the linear elastic region, the following correlation between applied force (F) and associated displacement (δ) determined by the extensometer can be established as follows:
 
$$ F=\kappa \delta $$
(3)
where κ is the stiffness.
ii)
Knowing the cross-section area and the initial length of each plate Eq. (3) can be transformed into
 
$$ \sigma = E\varepsilon $$
(4)
where σ is the stress, ε is the strain and E is the equivalent Young’s modulus determined from the slope of the stress-strain curve. The yield stress was calculated by measuring the slope of the stress-strain curve from a 0.2% strain offset.

5.2 Surface roughness

Surface roughness was determined using a coaxial laser confocal microscope (Keyence, VK-X200, Japan) by scanning a zone of 500 × 750 μm2. The arithmetic average of the surface roughness, Ra, and the average arithmetic height of the surface, Sa, of both produced EBM plates and commercial plates were calculated considering 10 measurements with an area of 100 × 100 μm2 over a length of 500 μm using the VK Analyzer 3.3 software (Keyence, Japan).

5.3 Hardness

Hardness was measured using the Vickers hardness (HV) tester-Armstrong Pedestal considering a normal load of 10 kgf applied for 12 s as shown in Fig. 2a. Previously, the surface of the EBM bone plate was prepared using silicon carbide paper with a grid of 400, 800, 1200 and 2400 grits and polished with 1-μm diamond particle paste. A total of 5 indentations were taken as shown in Fig. 2b: measurements 1, 4 and 5 along the y-axis and measurements 1, 2 and 3 along the x-axis. The Vickers hardness was calculated replacing the average length of the diagonal, d, measured by optical microscopy from all the indentations in millimetres according to the following equation [20]:
$$ Hv=\frac{18.55}{d_2} $$
(5)
where d2 is the arithmetic mean of the diagonals.

5.4 Morphological characterisation

The morphology of the EBM plates was investigated using scanning electron microscopy (SEM) with a Hitachi S300N microscope (Hitachi, Japan), using an accelerating voltage of 15.0 kV.

5.5 In vitro biological testing

Biological tests were conducted to understand the impact of the surface characteristics of the EBM plates and to investigate the potential need for any post-process in order to establish a good biological bonding with the surrounding tissue. A commercial plate was used as a reference.

5.5.1 Sample preparation

EBM and commercial bone plates were cut into samples to fit a 12-well plate using an electron discharge machine. Samples were sterilised by sonication whilst the samples were immersed within 80% ethanol for 30 min, rinsed with distilled water and then autoclaved for 30 min at 121 °C.

5.5.2 Cell culture and seeding

In vitro tests were performed using human osteosarcoma cells Saos-2 (ATCC® HTB-85™) (ATCC, Manassas, VA, USA). Cells were cultured in T75 tissue culture flasks (Sigma-Aldrich, Dorset, UK) with McCoy’s 5A Medium (ATCC® 30-2007™) (ATCC, Manassas, Virginia, USA) based media containing 15% foetal bovine serum, supplemented with penicillin and streptomycin (1%) until 80% confluence and harvested using 0.05% trypsin-EDTA solution (Thermo Fischer Scientific, Waltham, MA, USA). Cells were seeded on the samples at a density of 5 × 10^4 cells in 200 μL of media per sample, and the cell-seeded samples incubated at standard conditions (37 °C, 5% CO2 and 95% humidity) for 3 h to allow cell attachment, before the addition of 2 mL fresh media. Seeded samples were transferred into a new well plate after 24 h. Cell culture media was changed every 3 days.

5.5.3 Cell viability and morphology

EBM and commercial samples were cultured up to 14 days to assess cell morphology and attachment through SEM. Samples were fixed with a 3% glutaraldehyde solution (Sigma-Aldrich, UK) for 30 min at room temperature, rinsed twice with phosphate-buffered saline (PBS) solution, dehydrated with a graded ethanol series (50%, 60% 70%, 80%, 90% and 100% (twice)), in 50:50 ethanol/hexamethyldisilazane (HMDS, Sigma-Aldrich, Dorset, UK) and then in 100% HMDS, for 15 min at each step, and allowed to evaporate overnight to remove HMDS.
Laser confocal microscopy was employed to examine cell viability through a Live/Dead stain kit (Thermo Fischer Scientific, Waltham, MA, USA) for days 4 and 14. The Live/Dead staining solution was prepared according to manufacturer’s instruction. Briefly, a 4 μm M EtHD-1 and of 2 μm M calcein AM working solution was prepared in PBS. The samples were washed with PBS prior to the addition of the staining solution and incubated for 45 min at room temperature. Cell viability images were obtained using an inverted Leica TCS SP5 confocal microscope (Leica Microsystems, Germany).

5.6 Statistical analysis

The statistical analysis was performed using Minitab 18 software (PA, USA) considering one-way analysis of variance (ANOVA) with Tukey test. Differences were considered statistically significant at p < 0.05.

6 Results

6.1 Topology optimization and produced bone plates

For the three different plate designs (four, six and eight holes), nine designs were obtained as shown in Fig. 3. The EBM produced bone plate after support structure removal are shown in Fig. 4. Parts were not submitted to any post-processing operation or thermal treatment.

6.2 Surface roughness

Surface roughness of the EBM plates are presented in Fig. 5. Results showed arithmetic average of the surface roughness, Ra of 19.15 ± 4.94 μm and average arithmetic height of the surface, Sa of 12.42 ± 1.11 μm for the EBM plates, whereas the commercial plates presented an average Ra of 0.37 ± 0.03 μm and Sa of 0.42 ± 0.03 μm. Significant difference was observed (p < 0.05) between the samples for both Ra and Sa.

6.3 Hardness

EBM-produced plates presented a high hardness compared to commercial plates. Results showed that the EBM plates mean hardness is 326.9 HV 10 ± 3.35 and the mean hardness of the commercial plates is 275.98 HV 10 ± 7.75. The hardness between the samples showed significant difference (p < 0.05).

6.4 Tensile tests

A comparison between experimental and numerically predicted tensile results are presented in Figs. 6, 7 and 8. A good approximation was obtained with differences not higher than 7%.
The highest equivalent stiffness values were observed for the initial designs. Results also show high values of stiffness in plates containing a smaller number of holes (293,810 N/mm for four-hole plate, 242,634 N/mm for six-hole plate and 226,132 N/mm for eight-hole plate). Moreover, stiffness decreases by increasing volume reduction (228,050 N/mm for four-hole plate, 228,418 N/mm for six-hole plate and 216,226 N/mm for eight-hole plate and 25% of volume reduction; 152,648 N/mm for four-hole plate, 118,970 N/mm for six-hole plate, 216,147 N/mm for eight-hole plate and 45% of volume reduction; 103,674 N/mm for four-hole plate, 52,011 N/mm for six-hole plate, 55,824 N/mm for eight-hole plate and 75% of volume reduction).
A comparison of the equivalent Young’s modulus values of all considered plates is provided in Fig. 9. Results show a decrease in the equivalent Young’s modulus in the optimised plates. However, optimised plates considering 25% of volume reduction are three times stiffer than the cortical bone. The commercial plates resulted in an equivalent Young’s modulus of around 40 GPa. Optimised six- and eight-hole plates considering 75% of volume reduction present values within the cortical bone region.
Table 2 presents the yield stress values from the tensile tests for all considered plates. In the case of commercially available plates, it was considered the maximum value between the four commercial plates investigated in this study. The yield stress values of the initial plates, four-hole 25% and 45% volume reduction and eight-hole 25% volume reduction are 1.5 to 3 times higher than the commercial plates, whilst the four-hole 75% volume reduction and six-hole 25% volume reduction plate showed approximately similar values. The two optimised bone plates that resulted with similar cortical bone modulus, the 75% volume reduction for the six- and eight-hole bone plates, resulted in around half of the yield of the commercial plates and the maximum force reached before yield was around 4191 N and 5481 N, respectively. The equivalent stiffness of the four-hole plates (75%, 45% and 25% volume reductions) were significantly different (p < 0.05) than the four-hole initial plates. Six-hole plates (75% and 45% volume reductions) showed significant difference between the six-hole initial plate with no significance when comparing with the 25% volume reduction. Similar, significant difference was observed in the eight-hole plates. In terms of the equivalent Young’s modulus, the four-hole 45% and 75% volume reduction, six-hole 75% volume reduction and eight-hole 45% and 75% volume reduction plates were significantly different (p < 0.05) comparing with the commercial plates. Similarly, the EBM (six-hole bone plates and 45% and 75% of volume reduction and eight-hole plates with 75% of volume reduction) presented significant lower yield stress than commercial plates (p < 0.05).
Table 2
Yield stresses of all bone plates
Plate
Volume reduction (%)
Yield stress (MPa)
Four holes
Initial
300
25
281
45
250
75
160
Six holes
Initial
346
25
200
45
91
75
86
Eight holes
Initial
368
25
359
45
279
75
104
DePuy Synthes plate
170

6.5 Cell morphology and viability

SEM images of the surface of both EBM and commercial plates are shown in Fig. 10a and b, respectively. The cell-seeded EBM plates after days 7 and 14 are shown in Fig. 10c and e, respectively, whilst cells seeded on commercial plates after days 7 and 14 are presented in Fig. 10d and f, respectively. Cell viability results are presented in Fig. 11 for both EBM and commercial plates after days 4 and 14 of cell seeding. Results show that cells are proliferating and are viable on the EBM plates whilst in the case of commercial plates, fewer cells attached with a higher number of dead cells.

7 Discussion

The topology optimization method is based on the material redistribution, eventually leading to a material reduction whilst optimising stiffness. As also previously reported, topology optimization can be used to create lightweight structures capable of keeping satisfactory mechanical performance [21, 22]. By imposing high volume reduction, the effect of material removal becomes the significant contributing factor to an overall reduction in the stiffness of the plates. This was previously reported by the authors with 2D designed plates [23]. Different volume reductions were considered (25%, 45% and 75%) but the obtained volume reduction was not always exactly the same as the imposed value. In the case of eight-hole bone plates and imposed volume reductions of 25% and 45%, no significant differences were observed on the optimised shapes in terms of equivalent stiffness as the optimisation procedure stopped at relatively similar volume reductions. This is due to the fact that the optimisation procedure stops once it reaches a local optimal value, which is not necessarily the global one. It can also be explained by Eq. 1a, which shows that the optimisation process approximates to the user-defined volume, and to the fact that the optimisation process is mesh-dependent. However, the effect of the mesh on the topology optimization is not considered here. An automatic mesh generation procedure was considered.
The function of a bone plate is to resist the stresses occurring in the physiological environment whilst stabilising the fractured bone. Therefore, avoiding mechanical failure as well as plastic deformation is critical. In this research, both the elastic and plastic deformations were considered. The elastic linear region was analysed to measure both the equivalent stiffness and equivalent Young’s modulus. Equivalent stiffness values were used to characterise each family of bone plates (four, six and eight holes). Results show, for all cases, a reduction on the equivalent stiffness by increasing the volume reduction. A good approximation between numerical and experimental results was also obtained, showing that the computational tools considered in this research represent a viable approach to design and simulate their behaviour before expensive fabrication and experimental steps. Results also show that topology optimization is a viable tool to redesign bone plates, reducing their equivalent stiffness and consequently the stress shielding effect. The equivalent Young’s modulus was used to position the mechanical properties of produced plates and commercial ones against cortical bone. Initial designed plates present higher equivalent Young’s modulus values than commercial plates. These differences are reduced through topology optimization. Although as mentioned earlier, taking into consideration that the dimensions of the bone plates defer from one another, the EBM plates containing six holes (45% of volume reduction) and four holes (75% of volume reduction) show equivalent Young’s modulus similar to commercial plates. However, produced plates containing six and eight holes (75% of volume reduction) show a significant reduction in the equivalent Young’s modulus, presenting values in the cortical bone region, which is not possible to achieve with the commercially available plates. This confirms that topology optimization can be used to create lightweight plates (less material volume) with reduced equivalent stiffness and sufficiently strong without impairing the plate’s mechanical performance, thus minimising the stress-shielding phenomena.
This study is limited in terms of considering the mechanical stability of the topology optimised bone plates, assuming that a significant decrease in the material volume could result in stress concentration and plate failure, leading to a delay of bone healing. However, the authors are addressing this limitation by analysing the plastic region of the mechanical tests, considering the commercial plate (i.e. DePuy Synthes LCP, the gold standard for bone fracture healing) as a reference to base the stability of the bone plates in terms of the yield stress results. Most of the bone plates presented a higher yield stress with two plates (four-hole 75% volume reduction and six-hole 25% volume reduction plates) showing similarity with the commercial ones in terms of yield behaviour. The resulting yield stresses for the two bone plates that have similar equivalent Young’s modulus to the cortical bone showed approximately half of the commercial bone plate yield. However, if we consider that the maximum force required to yield the two bone plates are 4191 N (six holes and 75% of volume reduction) and 5481 N (eight hole and 75% of volume reduction), this is still clinically acceptable as it corresponds to 6 to 8 times the bodyweight of a 70-kg patient.
EBM-produced plates were not submitted to any post-processing treatment (e.g. elimination of residual stresses, surface finishing and polishing). This was decided to avoid the use of any additional time, consumables and costly steps and to avoid any additional effects of the post-processing on the topography, microstructure and properties of the plates, partially answering the question: is it necessary to perform any post-processing to produce plates with adequate physical and biological characteristics? Due to the nature of the EBM process (powder-bed fusion technique), rough parts were produced. As observed by Karlsson et al. [24] and Tong et al. [25], the surface characteristics of EBM parts are influenced by the raw powder characteristics (e.g. particle size) and processing conditions (e.g. scan speed) that result in the presence of both partially melted particles and fully melted ones (flattest zones aligned with the build direction). Despite presenting high surface roughness compared to commercial plates, produced plates were not polished to understand if the rough values were within an acceptable range to allow cell attachment and cell-cell communication. This was investigated by accessing the in vitro response of osteosarcoma cells seeded on both EBM and the commercial plates.
Contradictory results were previously reported regarding the biological influence of smooth and rough surfaces. According to De Wild et al. [26], EBM implants should be surface treated as their study showed improved biological response on acid-etched and sand-blasted implants. Pattanayak et al. [27] showed that heat treated smoother implants improve the osteo-integration of the implants. However, Thomson et al. [28] and Bertollo et al. [29] did not observe significant differences between no post-processed implants and conventionally wrought machined and plasma-sprayed implants. Ponader et al. [30] observed high cell attachment and proliferation in EBM parts with Ra values lower than 24.9 μm and reduced cell proliferation values in rough parts (Ra higher than 56.9 μm), surprisingly showing that the highest proliferation rate and cell viability were obtained with highly smooth surfaces (Ra of 0.07 μm). In contrast, the results presented in this paper showed that surface roughness of the produced plates (19.15 ± 1.56 μm) allowed a high level of cell attachment and proliferation suggesting that no polishing post-processing step is required after the plate printing phase. Results also show high cell attachment on the EBM rough plates than in the commercial plates with a smooth surface (0.37 ± 0.01 μm). Cell viability studies also show a significant number of dead cells in the commercial plates after 14 days of cell seeding compared to the EBM plates. This can be partially explained by the surface topography of the commercial bone plates. However, further investigation is required to elucidate the role that ion release and wettability have on cell interactions.
Surface hardness also has an important role on cell attachment and proliferation and differentiation. Several studies reported that harder substrates promote osteogenesis and increase cell attachment [31, 32]. The results show that EBM plates present higher surface hardness (326.9 ± 3.35) than commercial plates (275.98 ± 7.75), both exhibiting significantly higher values compared to cortical bone hardness (i.e. ~ 50 HV) [33].

8 Conclusion

This paper presents an integrated strategy combining topology optimization and additive manufacturing to produce bone fixation plates minimising the stress-shielding phenomena. From the results, it is possible to conclude the following:
  • Topology optimization can be used to design plates minimising the stress-shielding effect. For high values of volume reduction, the effect of material removal strongly contributes to an overall reduction of the equivalent stiffness of the designed bone fixation plates.
  • Designed plates were successfully produced using EBM.
  • No significant mechanical differences were observed between produced and designed, showing a good accuracy between numerical and experimental results. It was also possible to design plates with equivalent Young’s modulus values in the region of cortical bone.
  • No post-processing seems to be required to establish a good biological bonding with the surrounding tissue. High cell attachment and proliferation was observed on rough EBM plates, which also presents higher surface roughness than commercial plates. Contrary to other studies, results also show that EBM parts with no post-processing present higher biocompatibility than commercially available plates.

Acknowledgements

Authors are grateful for the support provided by the Saudi Arabian government, Manchester University Foundation Trust and Royal College of Surgeons of England. The first author is also grateful for the support of the Advanced Manufacturing Institute (AMI), King Saud University, Riyadh, Saudi Arabia. Mechanical tests were conducted with the support of the National Composites Certification and Evaluation Facility (NCCEF), University of Manchester, Manchester, UK.
Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

Publisher’s note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Literatur
1.
Zurück zum Zitat Dorati R, Colonna C, Conti B (2011) Polymer scaffolds for bone tissue regeneration. In: Gefen A (ed) Studies in Mechanobiology, tissue Eng and biomater, 8th edn. Springer, pp 259–288 Dorati R, Colonna C, Conti B (2011) Polymer scaffolds for bone tissue regeneration. In: Gefen A (ed) Studies in Mechanobiology, tissue Eng and biomater, 8th edn. Springer, pp 259–288
9.
Zurück zum Zitat McNamara L (2017) Bone as a material. In: Ducheyne P (ed) Comprehnsive biomaterials II, 2nd edn. Elsevier, Berlin, pp 202–227CrossRef McNamara L (2017) Bone as a material. In: Ducheyne P (ed) Comprehnsive biomaterials II, 2nd edn. Elsevier, Berlin, pp 202–227CrossRef
17.
Zurück zum Zitat Bendsøe M, Sigmund O (2004) Topology optimization. Springer-Verlag, Berlin HeidelbergCrossRefMATH Bendsøe M, Sigmund O (2004) Topology optimization. Springer-Verlag, Berlin HeidelbergCrossRefMATH
26.
Zurück zum Zitat de Wild M, Schumacher R, Mayer K, Schkommodau E, Thoma D, Bredell M, Kruse Gujer A, Grätz KW, Weber FE (2013) Bone regeneration by the osteoconductivity of porous titanium implants manufactured by selective laser melting: a histological and micro computed tomography study in the rabbit. Tissue Eng A 19:2645–2654. https://doi.org/10.1089/ten.tea.2012.0753 CrossRef de Wild M, Schumacher R, Mayer K, Schkommodau E, Thoma D, Bredell M, Kruse Gujer A, Grätz KW, Weber FE (2013) Bone regeneration by the osteoconductivity of porous titanium implants manufactured by selective laser melting: a histological and micro computed tomography study in the rabbit. Tissue Eng A 19:2645–2654. https://​doi.​org/​10.​1089/​ten.​tea.​2012.​0753 CrossRef
28.
30.
Zurück zum Zitat Ponader S, Vairaktaris E, Heinl P, Wilmowsky C, Rottmair A, Körner C, Singer RF, Holst S, Schlegel KA, Neukam FW, Nkenke E (2008) Effects of topographical surface modifications of electron beam melted Ti-6Al-4V titanium on human fetal osteoblasts. J Biomed Mater Res A 84A:1111–1119. https://doi.org/10.1002/jbm.a.31540 CrossRef Ponader S, Vairaktaris E, Heinl P, Wilmowsky C, Rottmair A, Körner C, Singer RF, Holst S, Schlegel KA, Neukam FW, Nkenke E (2008) Effects of topographical surface modifications of electron beam melted Ti-6Al-4V titanium on human fetal osteoblasts. J Biomed Mater Res A 84A:1111–1119. https://​doi.​org/​10.​1002/​jbm.​a.​31540 CrossRef
Metadaten
Titel
Topology optimised metallic bone plates produced by electron beam melting: a mechanical and biological study
verfasst von
Abdulsalam Abdulaziz Al-Tamimi
Boyang Huang
Cian Vyas
Miguel Hernandez
Chris Peach
Paulo Bartolo
Publikationsdatum
25.05.2019
Verlag
Springer London
Erschienen in
The International Journal of Advanced Manufacturing Technology / Ausgabe 1-4/2019
Print ISSN: 0268-3768
Elektronische ISSN: 1433-3015
DOI
https://doi.org/10.1007/s00170-019-03866-0

Weitere Artikel der Ausgabe 1-4/2019

The International Journal of Advanced Manufacturing Technology 1-4/2019 Zur Ausgabe

    Marktübersichten

    Die im Laufe eines Jahres in der „adhäsion“ veröffentlichten Marktübersichten helfen Anwendern verschiedenster Branchen, sich einen gezielten Überblick über Lieferantenangebote zu verschaffen.