Toward Ultrasensitive Surface Plasmon Resonance Sensors

verfasst von : Vitali Scherbahn, Shavkat Nizamov, Vladimir M. Mirsky

Erschienen in: Label-Free Biosensing

Verlag: Springer International Publishing

Aktivieren Sie unsere intelligente Suche um passende Fachinhalte oder Patente zu finden.

search-config
loading …

Abstract

Despite the history of application of surface plasmon resonance (SPR) for chemo- and biosensing being over 30 years long, the development of this technique is still in progress.
This review is focused on the technological aspects of further improvement of analytical performance of SPR transducers based on Kretschmann configuration. We describe basic measurement configurations, their improvements and optimizations, and their drawbacks and limitations. An importance of referencing in SPR sensors is highlighted. The referencing approaches are classified into the following domains: (1) macroscopic spatially separated referencing, (2) self-referencing based on micro-patterning, (3) in-place referencing, (4) spatiotemporal referencing, and (5) electrochemically assisted referencing. The underlying principles of these approaches, examples of their implementation, and resulting improvements of sensor performance are described. Finally, an analysis of SPR data and an extraction of affinity properties are discussed.

Sie haben noch keine Lizenz? Dann Informieren Sie sich jetzt über unsere Produkte:

Springer Professional "Wirtschaft+Technik"

Online-Abonnement

Mit Springer Professional "Wirtschaft+Technik" erhalten Sie Zugriff auf:

  • über 102.000 Bücher
  • über 537 Zeitschriften

aus folgenden Fachgebieten:

  • Automobil + Motoren
  • Bauwesen + Immobilien
  • Business IT + Informatik
  • Elektrotechnik + Elektronik
  • Energie + Nachhaltigkeit
  • Finance + Banking
  • Management + Führung
  • Marketing + Vertrieb
  • Maschinenbau + Werkstoffe
  • Versicherung + Risiko

Jetzt Wissensvorsprung sichern!

Springer Professional "Technik"

Online-Abonnement

Mit Springer Professional "Technik" erhalten Sie Zugriff auf:

  • über 67.000 Bücher
  • über 390 Zeitschriften

aus folgenden Fachgebieten:

  • Automobil + Motoren
  • Bauwesen + Immobilien
  • Business IT + Informatik
  • Elektrotechnik + Elektronik
  • Energie + Nachhaltigkeit
  • Maschinenbau + Werkstoffe




 

Jetzt Wissensvorsprung sichern!

Literatur
1.
Zurück zum Zitat Bartell FE, Sloan CK (1929) An interferometric investigation of adsorption by pure carbon from non-aqueous binary systems. J Am Chem Soc 584:1637–1643CrossRef Bartell FE, Sloan CK (1929) An interferometric investigation of adsorption by pure carbon from non-aqueous binary systems. J Am Chem Soc 584:1637–1643CrossRef
2.
Zurück zum Zitat Damaskin BB, Petrii OA, Batrakov VV (1972) Adsorption of organic compounds on electrodes. Plenum Press, New York Damaskin BB, Petrii OA, Batrakov VV (1972) Adsorption of organic compounds on electrodes. Plenum Press, New York
3.
Zurück zum Zitat Sauerbrey G (1959) Verwendung von schwingquarzen zur wägung dünner schichten und zur mikrowägung. Z Phys 155:206–222CrossRef Sauerbrey G (1959) Verwendung von schwingquarzen zur wägung dünner schichten und zur mikrowägung. Z Phys 155:206–222CrossRef
4.
Zurück zum Zitat King WH (1964) Piezoelectric sorption detector. Anal Chem 36:1735–1739CrossRef King WH (1964) Piezoelectric sorption detector. Anal Chem 36:1735–1739CrossRef
5.
Zurück zum Zitat Liedberg B, Nylander C, Lunström I (1983) Surface plasmon resonance for gas detection and biosensing. Sensors Actuators 4:299–304CrossRef Liedberg B, Nylander C, Lunström I (1983) Surface plasmon resonance for gas detection and biosensing. Sensors Actuators 4:299–304CrossRef
6.
Zurück zum Zitat Lofas S, Malmqvist M, Ronnberg I, Stenberg E, Liedberg B, Lundstrom I (1991) Bioanalysis with surface plasmon resonance. Sensors Actuators B Chem 5:79–84CrossRef Lofas S, Malmqvist M, Ronnberg I, Stenberg E, Liedberg B, Lundstrom I (1991) Bioanalysis with surface plasmon resonance. Sensors Actuators B Chem 5:79–84CrossRef
7.
Zurück zum Zitat Liedberg B, Nylander C, Lundström I (1995) Biosensing with surface plasmon resonance - how it all started. Biosens Bioelectron 10:1–9CrossRef Liedberg B, Nylander C, Lundström I (1995) Biosensing with surface plasmon resonance - how it all started. Biosens Bioelectron 10:1–9CrossRef
8.
Zurück zum Zitat Homola J (2006) Surface plasmon resonance based sensors. Springer, BerlinCrossRef Homola J (2006) Surface plasmon resonance based sensors. Springer, BerlinCrossRef
9.
Zurück zum Zitat Schasfoort RBM, Tudos AJ (2008) Handbook of surface plasmon resonance. The Royal Society of Chemistry, LondonCrossRef Schasfoort RBM, Tudos AJ (2008) Handbook of surface plasmon resonance. The Royal Society of Chemistry, LondonCrossRef
10.
Zurück zum Zitat Homola J (2003) Present and future of surface plasmon resonance biosensors. Anal Bioanal Chem 377:528–539PubMedCrossRef Homola J (2003) Present and future of surface plasmon resonance biosensors. Anal Bioanal Chem 377:528–539PubMedCrossRef
11.
Zurück zum Zitat Safina G (2012) Application of surface plasmon resonance for the detection of carbohydrates, glycoconjugates, and measurement of the carbohydrate-specific interactions: a comparison with conventional analytical techniques. A critical review. Anal Chim Acta 712:9–29PubMedCrossRef Safina G (2012) Application of surface plasmon resonance for the detection of carbohydrates, glycoconjugates, and measurement of the carbohydrate-specific interactions: a comparison with conventional analytical techniques. A critical review. Anal Chim Acta 712:9–29PubMedCrossRef
12.
Zurück zum Zitat Hoa XD, Kirk AG, Tabrizian M (2007) Towards integrated and sensitive surface plasmon resonance biosensors: a review of recent progress. Biosens Bioelectron 23:151–160PubMedCrossRef Hoa XD, Kirk AG, Tabrizian M (2007) Towards integrated and sensitive surface plasmon resonance biosensors: a review of recent progress. Biosens Bioelectron 23:151–160PubMedCrossRef
13.
Zurück zum Zitat Homola J, Yee SS, Gauglitz G (1999) Surface plasmon resonance sensors: review. Sensors Actuators B Chem 54:3–15CrossRef Homola J, Yee SS, Gauglitz G (1999) Surface plasmon resonance sensors: review. Sensors Actuators B Chem 54:3–15CrossRef
14.
Zurück zum Zitat Puiu M, Bala C (2016) SPR and SPR imaging: recent trends in developing nanodevices for detection and real-time monitoring of biomolecular events. Sensors 16:870CrossRef Puiu M, Bala C (2016) SPR and SPR imaging: recent trends in developing nanodevices for detection and real-time monitoring of biomolecular events. Sensors 16:870CrossRef
15.
Zurück zum Zitat Kretschmann E, Raether H (1968) Radiative decay of non-radiative surface plasmons excited by light. Z Naturforsch 23:2135–2136 Kretschmann E, Raether H (1968) Radiative decay of non-radiative surface plasmons excited by light. Z Naturforsch 23:2135–2136
16.
Zurück zum Zitat Boardman AD (1982) Electromagnetic surface modes. Wiley, Chichester Boardman AD (1982) Electromagnetic surface modes. Wiley, Chichester
17.
Zurück zum Zitat Raether H (1988) Surface plasmons on smooth and rough surfaces and on gratings. Springer, BerlinCrossRef Raether H (1988) Surface plasmons on smooth and rough surfaces and on gratings. Springer, BerlinCrossRef
18.
Zurück zum Zitat Nizamov S, Scherbahn V, Mirsky VM (2015) Self-referencing SPR-sensor based on integral measurements of light intensity reflected by arbitrarily distributed sensing and referencing spots. Sensors Actuators B Chem 207:740–747CrossRef Nizamov S, Scherbahn V, Mirsky VM (2015) Self-referencing SPR-sensor based on integral measurements of light intensity reflected by arbitrarily distributed sensing and referencing spots. Sensors Actuators B Chem 207:740–747CrossRef
19.
Zurück zum Zitat Jung LS, Campbell CT, Chinowsky TM, Mar MN, Yee SS (1998) Quantitative interpretation of the response of surface plasmon resonance sensors to adsorbed films. Langmuir 14:5636–5648CrossRef Jung LS, Campbell CT, Chinowsky TM, Mar MN, Yee SS (1998) Quantitative interpretation of the response of surface plasmon resonance sensors to adsorbed films. Langmuir 14:5636–5648CrossRef
20.
Zurück zum Zitat Mcmeekin TL, Wilensky M, Groves ML (1962) Refractive indices of proteins in relation to amino acid composition and specific volume. Biochem Biophys Res Commun 7:151–156CrossRef Mcmeekin TL, Wilensky M, Groves ML (1962) Refractive indices of proteins in relation to amino acid composition and specific volume. Biochem Biophys Res Commun 7:151–156CrossRef
21.
Zurück zum Zitat Barer R, Tkaczyk S (1954) Refractive index of concentrated protein solutions. Nature 173:821–822PubMedCrossRef Barer R, Tkaczyk S (1954) Refractive index of concentrated protein solutions. Nature 173:821–822PubMedCrossRef
23.
Zurück zum Zitat Klemm F, Johnson R, Mirsky VM (2015) Binding of protein nanoparticles to immobilized receptors. Sensors Actuators B Chem 208:616–621CrossRef Klemm F, Johnson R, Mirsky VM (2015) Binding of protein nanoparticles to immobilized receptors. Sensors Actuators B Chem 208:616–621CrossRef
24.
Zurück zum Zitat Liedberg B, Lundström I, Stenberg E (1993) Principles of biosensing with an extended coupling matrix and surface plasmon resonance. Sensors Actuators B Chem 11:63–72CrossRef Liedberg B, Lundström I, Stenberg E (1993) Principles of biosensing with an extended coupling matrix and surface plasmon resonance. Sensors Actuators B Chem 11:63–72CrossRef
25.
Zurück zum Zitat Zybin A, Boecker D, Mirsky VM, Niemax K (2007) Enhancement of the detection power of surface plasmon resonance measurements by optimization of the reflection angle. Anal Chem 79:4233–4236PubMedCrossRef Zybin A, Boecker D, Mirsky VM, Niemax K (2007) Enhancement of the detection power of surface plasmon resonance measurements by optimization of the reflection angle. Anal Chem 79:4233–4236PubMedCrossRef
26.
Zurück zum Zitat Akimoto T, Sasaki S, Ikebukuro K, Karube I (2000) Effect of incident angle of light on sensitivity and detection limit for layers of antibody with surface plasmon resonance spectroscopy. Biosens Bioelectron 15:355–362PubMedCrossRef Akimoto T, Sasaki S, Ikebukuro K, Karube I (2000) Effect of incident angle of light on sensitivity and detection limit for layers of antibody with surface plasmon resonance spectroscopy. Biosens Bioelectron 15:355–362PubMedCrossRef
27.
Zurück zum Zitat Eum NS, Kim DE, Yeom SH, Kang BH, Kim KJ, Park CS, Kang SW (2009) Variable wavelength surface plasmon resonance (SPR) in biosensing. Biosystems 98:51–55PubMedCrossRef Eum NS, Kim DE, Yeom SH, Kang BH, Kim KJ, Park CS, Kang SW (2009) Variable wavelength surface plasmon resonance (SPR) in biosensing. Biosystems 98:51–55PubMedCrossRef
28.
Zurück zum Zitat Akimoto T, Wada S, Karube I (2008) A surface plasmon resonance probe without optical fibers as a portable sensing device. Anal Chim Acta 610:119–124PubMedCrossRef Akimoto T, Wada S, Karube I (2008) A surface plasmon resonance probe without optical fibers as a portable sensing device. Anal Chim Acta 610:119–124PubMedCrossRef
29.
Zurück zum Zitat Nelson BP, Frutos AG, Brockman JM, Corn RM (1999) Near-infrared surface plasmon resonance measurements of ultrathin films. Anal Chem 71:3928–3934CrossRef Nelson BP, Frutos AG, Brockman JM, Corn RM (1999) Near-infrared surface plasmon resonance measurements of ultrathin films. Anal Chem 71:3928–3934CrossRef
30.
Zurück zum Zitat Frutos AG, Weibel SC, Corn RM (1999) Near-infrared surface plasmon resonance measurements of ultrathin films. 2. Fourier transform SPR spectroscopy. Anal Chem 71:3935–3940CrossRef Frutos AG, Weibel SC, Corn RM (1999) Near-infrared surface plasmon resonance measurements of ultrathin films. 2. Fourier transform SPR spectroscopy. Anal Chem 71:3935–3940CrossRef
31.
Zurück zum Zitat Rothenhausler B, Knoll W (1988) Surface plasmon microscopy. Nature 332:615–617CrossRef Rothenhausler B, Knoll W (1988) Surface plasmon microscopy. Nature 332:615–617CrossRef
32.
Zurück zum Zitat Brockman JM, Nelson BP, Corn RM (2000) Surface plasmon resonance imaging measurements of ultrathin organic films. Annu Rev Phys Chem 51:41–63PubMedCrossRef Brockman JM, Nelson BP, Corn RM (2000) Surface plasmon resonance imaging measurements of ultrathin organic films. Annu Rev Phys Chem 51:41–63PubMedCrossRef
33.
Zurück zum Zitat Campbell CT, Kim G (2007) SPR microscopy and its applications to high-throughput analyses of biomolecular binding events and their kinetics. Biomaterials 28:2380–2392PubMedCrossRef Campbell CT, Kim G (2007) SPR microscopy and its applications to high-throughput analyses of biomolecular binding events and their kinetics. Biomaterials 28:2380–2392PubMedCrossRef
34.
Zurück zum Zitat Boecker D, Zybin A, Niemax K, Grunwald C, Mirsky VM (2008) Noise reduction by multiple referencing in surface plasmon resonance imaging. Rev Sci Instrum 79:2–8CrossRef Boecker D, Zybin A, Niemax K, Grunwald C, Mirsky VM (2008) Noise reduction by multiple referencing in surface plasmon resonance imaging. Rev Sci Instrum 79:2–8CrossRef
35.
Zurück zum Zitat Berger CEH, Beumer TAM, Kooyman RPH, Greve J (1998) Surface plasmon resonance multisensing. Anal Chem 70:703–706CrossRef Berger CEH, Beumer TAM, Kooyman RPH, Greve J (1998) Surface plasmon resonance multisensing. Anal Chem 70:703–706CrossRef
36.
Zurück zum Zitat Guo X (2012) Surface plasmon resonance based biosensor technique: a review. J Biophotonics 5:483–501PubMedCrossRef Guo X (2012) Surface plasmon resonance based biosensor technique: a review. J Biophotonics 5:483–501PubMedCrossRef
37.
Zurück zum Zitat Nylander C, Liedberg B, Lind T (1982) Gas detection by means of surface plasmon resonance. Sensors Actuators 3:79–88CrossRef Nylander C, Liedberg B, Lind T (1982) Gas detection by means of surface plasmon resonance. Sensors Actuators 3:79–88CrossRef
38.
Zurück zum Zitat Nuzzo RG, Allara DL (1983) Adsorption of bifunctional organic disulfides on gold surfaces. J Am Chem Soc 105:4481–4483CrossRef Nuzzo RG, Allara DL (1983) Adsorption of bifunctional organic disulfides on gold surfaces. J Am Chem Soc 105:4481–4483CrossRef
39.
40.
Zurück zum Zitat Riepl M, Mirsky VM, Novotny I, Tvarozek V, Rehacek V, Wolfbeis OS (1999) Optimization of capacitive affinity sensors: drift suppression and signal amplification. Anal Chim Acta 392:77–84CrossRef Riepl M, Mirsky VM, Novotny I, Tvarozek V, Rehacek V, Wolfbeis OS (1999) Optimization of capacitive affinity sensors: drift suppression and signal amplification. Anal Chim Acta 392:77–84CrossRef
41.
Zurück zum Zitat Mirsky VM (2002) New electroanalytical applications of self-assembled monolayers. TrAC Trends Anal Chem 21:439–450CrossRef Mirsky VM (2002) New electroanalytical applications of self-assembled monolayers. TrAC Trends Anal Chem 21:439–450CrossRef
42.
Zurück zum Zitat Love JC, Estroff LA, Kriebel JK, Nuzzo RG, Whitesides GM (2005) Self-assembled monolayers of thiolates on metals as a form of nanotechnology. Chem Rev 105:1103–1169PubMedCrossRef Love JC, Estroff LA, Kriebel JK, Nuzzo RG, Whitesides GM (2005) Self-assembled monolayers of thiolates on metals as a form of nanotechnology. Chem Rev 105:1103–1169PubMedCrossRef
43.
Zurück zum Zitat Hirsch T, Shaporenko A, Mirsky VM, Zharnikov M (2007) Monomolecular films of phthalocyanines: formation, characterization, and expelling by alkanethiols. Langmuir 23:4373–4377PubMedCrossRef Hirsch T, Shaporenko A, Mirsky VM, Zharnikov M (2007) Monomolecular films of phthalocyanines: formation, characterization, and expelling by alkanethiols. Langmuir 23:4373–4377PubMedCrossRef
44.
Zurück zum Zitat Stewart A, Zheng S, McCourt MR, Bell SEJ (2012) Controlling assembly of mixed thiol monolayers on silver nanoparticles to tune their surface properties. ACS Nano 6:3718–3726PubMedPubMedCentralCrossRef Stewart A, Zheng S, McCourt MR, Bell SEJ (2012) Controlling assembly of mixed thiol monolayers on silver nanoparticles to tune their surface properties. ACS Nano 6:3718–3726PubMedPubMedCentralCrossRef
45.
Zurück zum Zitat Crooks RM, Ricco AJ (1998) New organic materials suitable for use in chemical sensor arrays. Acc Chem Res 31:219–227CrossRef Crooks RM, Ricco AJ (1998) New organic materials suitable for use in chemical sensor arrays. Acc Chem Res 31:219–227CrossRef
46.
Zurück zum Zitat Glebe U, Baio JE, Árnadóttir L, Siemeling U, Weidner T (2013) Molecular suction pads: self-assembled monolayers of subphthalocyaninatoboron complexes [BCl{Subpc’(SR)6}] on gold. ChemPhysChem 14:1155–1160PubMedPubMedCentralCrossRef Glebe U, Baio JE, Árnadóttir L, Siemeling U, Weidner T (2013) Molecular suction pads: self-assembled monolayers of subphthalocyaninatoboron complexes [BCl{Subpc’(SR)6}] on gold. ChemPhysChem 14:1155–1160PubMedPubMedCentralCrossRef
47.
Zurück zum Zitat Ferretti S, Paynter S, Russell DA, Sapsford KE, Richardson DJ (2000) Self-assembled monolayers: a versatile tool for the formulation of bio-surfaces. TrAC Trends Anal Chem 19:530–540CrossRef Ferretti S, Paynter S, Russell DA, Sapsford KE, Richardson DJ (2000) Self-assembled monolayers: a versatile tool for the formulation of bio-surfaces. TrAC Trends Anal Chem 19:530–540CrossRef
48.
Zurück zum Zitat Faucheux N, Schweiss R, Lützow K, Werner C, Groth T (2004) Self-assembled monolayers with different terminating groups as model substrates for cell adhesion studies. Biomaterials 25:2721–2730PubMedCrossRef Faucheux N, Schweiss R, Lützow K, Werner C, Groth T (2004) Self-assembled monolayers with different terminating groups as model substrates for cell adhesion studies. Biomaterials 25:2721–2730PubMedCrossRef
49.
Zurück zum Zitat Ostuni E, Yan L, Whitesides GM (1999) The interaction of proteins and cells with self-assembled monolayers of alkanethiolates on gold and silver. Colloids Surf B Biointerfaces 15:3–30CrossRef Ostuni E, Yan L, Whitesides GM (1999) The interaction of proteins and cells with self-assembled monolayers of alkanethiolates on gold and silver. Colloids Surf B Biointerfaces 15:3–30CrossRef
50.
Zurück zum Zitat Schlenoff JB, Li M, Ly H (1995) Stability and self-exchange in alkanethiol monolayers. J Am Chem Soc 117:12528–12536CrossRef Schlenoff JB, Li M, Ly H (1995) Stability and self-exchange in alkanethiol monolayers. J Am Chem Soc 117:12528–12536CrossRef
51.
Zurück zum Zitat Noh J, Kato HS, Kawai M, Hara M (2002) Surface and adsorption structures of dialkyl sulfide self-assembled monolayers on Au(111). J Phys Chem B 106:13268–13272CrossRef Noh J, Kato HS, Kawai M, Hara M (2002) Surface and adsorption structures of dialkyl sulfide self-assembled monolayers on Au(111). J Phys Chem B 106:13268–13272CrossRef
52.
Zurück zum Zitat Phares N, White RJ, Plaxco KW (2009) Improving the stability and sensing of electrochemical biosensors by employing trithiol-anchoring groups in a six-carbon self-assembled monolayer. Anal Chem 81:1095–1100PubMedPubMedCentralCrossRef Phares N, White RJ, Plaxco KW (2009) Improving the stability and sensing of electrochemical biosensors by employing trithiol-anchoring groups in a six-carbon self-assembled monolayer. Anal Chem 81:1095–1100PubMedPubMedCentralCrossRef
53.
54.
Zurück zum Zitat Hermanson GT (2013) Bioconjugate techniques.3rd edn. Academic Press, Boston Hermanson GT (2013) Bioconjugate techniques.3rd edn. Academic Press, Boston
55.
Zurück zum Zitat Chaki NK, Vijayamohanan K (2002) Self-assembled monolayers as a tunable platform for biosensor applications. Biosens Bioelectron 17:1–12PubMedCrossRef Chaki NK, Vijayamohanan K (2002) Self-assembled monolayers as a tunable platform for biosensor applications. Biosens Bioelectron 17:1–12PubMedCrossRef
56.
Zurück zum Zitat Gooding JJJ, Hibbert DBB (1999) The application of alkanethiol self-assembled monolayers to enzyme electrodes. TrAC Trends Anal Chem 18:525–533CrossRef Gooding JJJ, Hibbert DBB (1999) The application of alkanethiol self-assembled monolayers to enzyme electrodes. TrAC Trends Anal Chem 18:525–533CrossRef
57.
Zurück zum Zitat Wrobel N, Schinkinger M, Mirsky VM (2002) A novel ultraviolet assay for testing side reactions of carbodiimides. Anal Biochem 305:135–138PubMedCrossRef Wrobel N, Schinkinger M, Mirsky VM (2002) A novel ultraviolet assay for testing side reactions of carbodiimides. Anal Biochem 305:135–138PubMedCrossRef
58.
Zurück zum Zitat Yamada H, Imoto T, Fujita K, Okazaki K, Motomura M (1981) Selective modification of aspartic acid-101 in lysozyme by carbodiimide reaction. Biochemistry 20:4836–4842PubMedCrossRef Yamada H, Imoto T, Fujita K, Okazaki K, Motomura M (1981) Selective modification of aspartic acid-101 in lysozyme by carbodiimide reaction. Biochemistry 20:4836–4842PubMedCrossRef
59.
Zurück zum Zitat Pei Z, Anderson H, Myrskog A, Dunér G, Ingemarsson B, Aastrup T (2010) Optimizing immobilization on two-dimensional carboxyl surface: pH dependence of antibody orientation and antigen binding capacity. Anal Biochem 398:161–168PubMedCrossRef Pei Z, Anderson H, Myrskog A, Dunér G, Ingemarsson B, Aastrup T (2010) Optimizing immobilization on two-dimensional carboxyl surface: pH dependence of antibody orientation and antigen binding capacity. Anal Biochem 398:161–168PubMedCrossRef
60.
Zurück zum Zitat Wrobel N, Deininger W, Hegemann P, Mirsky VM (2003) Covalent immobilization of oligonucleotides on electrodes. Colloids Surf B Biointerfaces 32:157–162CrossRef Wrobel N, Deininger W, Hegemann P, Mirsky VM (2003) Covalent immobilization of oligonucleotides on electrodes. Colloids Surf B Biointerfaces 32:157–162CrossRef
61.
Zurück zum Zitat Johnsson B, Löfås S, Lindquist G (1991) Immobilization of proteins to a carboxymethyldextran-modified gold surface for biospecific interaction analysis in surface plasmon resonance sensors. Anal Biochem 198:268–277PubMedCrossRef Johnsson B, Löfås S, Lindquist G (1991) Immobilization of proteins to a carboxymethyldextran-modified gold surface for biospecific interaction analysis in surface plasmon resonance sensors. Anal Biochem 198:268–277PubMedCrossRef
62.
Zurück zum Zitat Kolb HC, Finn MG, Sharpless KB (2001) Click chemistry: diverse chemical function from a few good reactions. Angew Chem Int Ed 40:2004–2021CrossRef Kolb HC, Finn MG, Sharpless KB (2001) Click chemistry: diverse chemical function from a few good reactions. Angew Chem Int Ed 40:2004–2021CrossRef
63.
Zurück zum Zitat Mehlich J, Ravoo BJ (2011) Click chemistry by microcontact printing on self-assembled monolayers: a structure–reactivity study by fluorescence microscopy. Org Biomol Chem 9:4108–4415PubMedCrossRef Mehlich J, Ravoo BJ (2011) Click chemistry by microcontact printing on self-assembled monolayers: a structure–reactivity study by fluorescence microscopy. Org Biomol Chem 9:4108–4415PubMedCrossRef
64.
Zurück zum Zitat Fleminger G, Hadas E, Wolf T, Solomon B (1990) Oriented immobilization of periodate-oxidized monoclonal antibodies on amino and hydrazide derivatives of Eupergit C. Appl Biochem Biotechnol 23:123–137PubMedCrossRef Fleminger G, Hadas E, Wolf T, Solomon B (1990) Oriented immobilization of periodate-oxidized monoclonal antibodies on amino and hydrazide derivatives of Eupergit C. Appl Biochem Biotechnol 23:123–137PubMedCrossRef
65.
Zurück zum Zitat Nizamov S, Mirsky VM (2011) Self-referencing SPR-biosensors based on penetration difference of evanescent waves. Biosens Bioelectron 28:263–269PubMedCrossRef Nizamov S, Mirsky VM (2011) Self-referencing SPR-biosensors based on penetration difference of evanescent waves. Biosens Bioelectron 28:263–269PubMedCrossRef
66.
Zurück zum Zitat Fägerstam LG, Frostell-Karlsson Å, Karlsson R, Persson B, Rönnberg I (1992) Biospecific interaction analysis using surface plasmon resonance detection applied to kinetic, binding site and concentration analysis. J Chromatogr A 597:397–410CrossRef Fägerstam LG, Frostell-Karlsson Å, Karlsson R, Persson B, Rönnberg I (1992) Biospecific interaction analysis using surface plasmon resonance detection applied to kinetic, binding site and concentration analysis. J Chromatogr A 597:397–410CrossRef
67.
Zurück zum Zitat Knight MW, King NS, Liu L, Everitt HO, Nordlander P, Halas NJ (2014) Aluminum for plasmonics. ACS Nano 8:834–840PubMedCrossRef Knight MW, King NS, Liu L, Everitt HO, Nordlander P, Halas NJ (2014) Aluminum for plasmonics. ACS Nano 8:834–840PubMedCrossRef
68.
Zurück zum Zitat Zynio SA, Samoylov AV, Surovtseva ER, Mirsky VM, Shirshov YM (2002) Bimetallic layers increase sensitivity of affinity sensors based on surface plasmon resonance. Sensors 2:62–70CrossRef Zynio SA, Samoylov AV, Surovtseva ER, Mirsky VM, Shirshov YM (2002) Bimetallic layers increase sensitivity of affinity sensors based on surface plasmon resonance. Sensors 2:62–70CrossRef
69.
Zurück zum Zitat Zhai P, Guo J, Xiang J, Zhou F (2007) Electrochemical surface plasmon resonance spectroscopy at bilayered silver/gold films. J Phys Chem C 111:981–986CrossRef Zhai P, Guo J, Xiang J, Zhou F (2007) Electrochemical surface plasmon resonance spectroscopy at bilayered silver/gold films. J Phys Chem C 111:981–986CrossRef
70.
Zurück zum Zitat Chen Y, Zheng RS, Zhang DG, Lu YH, Wang P, Ming H, Luo ZF, Kan Q (2011) Bimetallic chips for a surface plasmon resonance instrument. Appl Opt 50:387–391PubMedCrossRef Chen Y, Zheng RS, Zhang DG, Lu YH, Wang P, Ming H, Luo ZF, Kan Q (2011) Bimetallic chips for a surface plasmon resonance instrument. Appl Opt 50:387–391PubMedCrossRef
71.
Zurück zum Zitat Nesterenko DV, Saif-Ur-Rehman, Sekkat Z (2012) Surface plasmon sensing with different metals in single and double layer configurations. Appl Opt 51:6673–6682PubMedCrossRef Nesterenko DV, Saif-Ur-Rehman, Sekkat Z (2012) Surface plasmon sensing with different metals in single and double layer configurations. Appl Opt 51:6673–6682PubMedCrossRef
72.
Zurück zum Zitat Zekriti M, Nesterenko DV, Sekkat Z (2015) Long-range surface plasmons supported by a bilayer metallic structure for sensing applications. Appl Opt 54:2151–2157PubMedCrossRef Zekriti M, Nesterenko DV, Sekkat Z (2015) Long-range surface plasmons supported by a bilayer metallic structure for sensing applications. Appl Opt 54:2151–2157PubMedCrossRef
73.
Zurück zum Zitat Slavík R, Homola J, Vaisocherová H (2006) Advanced biosensing using simultaneous excitation of short and long range surface plasmons. Meas Sci Technol 17:932–938CrossRef Slavík R, Homola J, Vaisocherová H (2006) Advanced biosensing using simultaneous excitation of short and long range surface plasmons. Meas Sci Technol 17:932–938CrossRef
74.
Zurück zum Zitat Hastings JT, Guo J, Keathley PD, Kumaresh PB, Wei Y, Law S, Bachas LG (2007) Optimal self-referenced sensing using long- and short- range surface plasmons. Opt Express 15:17661–17672PubMedCrossRef Hastings JT, Guo J, Keathley PD, Kumaresh PB, Wei Y, Law S, Bachas LG (2007) Optimal self-referenced sensing using long- and short- range surface plasmons. Opt Express 15:17661–17672PubMedCrossRef
75.
Zurück zum Zitat El-Gohary SH, Choi M, Kim YL, Byun KM (2016) Dispersion curve engineering of TiO2/silver hybrid substrates for enhanced surface plasmon resonance detection. Sensors 16:1442–1452CrossRef El-Gohary SH, Choi M, Kim YL, Byun KM (2016) Dispersion curve engineering of TiO2/silver hybrid substrates for enhanced surface plasmon resonance detection. Sensors 16:1442–1452CrossRef
76.
Zurück zum Zitat Chen X, Jiang K (2010) Effect of aging on optical properties of bimetallic sensor chips. Opt Express 18:1105–1112PubMedCrossRef Chen X, Jiang K (2010) Effect of aging on optical properties of bimetallic sensor chips. Opt Express 18:1105–1112PubMedCrossRef
77.
Zurück zum Zitat Genslein C, Hausler P, Kirchner E-M, Bierl R, Baeumner AJ, Hirsch T (2017) Detection of small molecules with surface plasmon resonance by synergistic plasmonic effects of nanostructured surfaces and graphene. Proc SPIE 10080:100800FCrossRef Genslein C, Hausler P, Kirchner E-M, Bierl R, Baeumner AJ, Hirsch T (2017) Detection of small molecules with surface plasmon resonance by synergistic plasmonic effects of nanostructured surfaces and graphene. Proc SPIE 10080:100800FCrossRef
78.
Zurück zum Zitat Genslein C, Hausler P, Kirchner EM, Bierl R, Baeumner AJ, Hirsch T (2016) Graphene-enhanced plasmonic nanohole arrays for environmental sensing in aqueous samples. Beilstein J Nanotechnol 7:1564–1573PubMedPubMedCentralCrossRef Genslein C, Hausler P, Kirchner EM, Bierl R, Baeumner AJ, Hirsch T (2016) Graphene-enhanced plasmonic nanohole arrays for environmental sensing in aqueous samples. Beilstein J Nanotechnol 7:1564–1573PubMedPubMedCentralCrossRef
79.
Zurück zum Zitat Pang K, Dong W, Zhang B, Zhan S, Wang X (2015) Sensitivity-enhanced and noise-reduced surface plasmon resonance sensing with microwell chips. Meas Sci Technol 26:8ppCrossRef Pang K, Dong W, Zhang B, Zhan S, Wang X (2015) Sensitivity-enhanced and noise-reduced surface plasmon resonance sensing with microwell chips. Meas Sci Technol 26:8ppCrossRef
80.
Zurück zum Zitat Shalabney A, Abdulhalim I (2010) Electromagnetic fields distribution in multilayer thin film structures and the origin of sensitivity enhancement in surface plasmon resonance sensors. Sensors Actuators A Phys 159:24–32CrossRef Shalabney A, Abdulhalim I (2010) Electromagnetic fields distribution in multilayer thin film structures and the origin of sensitivity enhancement in surface plasmon resonance sensors. Sensors Actuators A Phys 159:24–32CrossRef
81.
Zurück zum Zitat Shalabney A, Khare C, Rauschenbach B, Abdulhalim I (2011) Sensitivity of surface plasmon resonance sensors based on metallic columnar thin films in the spectral and angular interrogations. Sensors Actuators B Chem 159:201–212CrossRef Shalabney A, Khare C, Rauschenbach B, Abdulhalim I (2011) Sensitivity of surface plasmon resonance sensors based on metallic columnar thin films in the spectral and angular interrogations. Sensors Actuators B Chem 159:201–212CrossRef
82.
Zurück zum Zitat Vaskevich A, Rubinstein I (2008) Localized surface plasmon resonance (LSPR) spectroscopy in biosensing. Part four. Handbook of biosensors and biochips. Wiley, Chichester, p 26 Vaskevich A, Rubinstein I (2008) Localized surface plasmon resonance (LSPR) spectroscopy in biosensing. Part four. Handbook of biosensors and biochips. Wiley, Chichester, p 26
83.
Zurück zum Zitat Kedem O, Tesler AB, Vaskevich A, Rubinstein I (2011) Sensitivity and optimization of localized surface plasmon resonance transducers. ACS Nano 5:748–760PubMedCrossRef Kedem O, Tesler AB, Vaskevich A, Rubinstein I (2011) Sensitivity and optimization of localized surface plasmon resonance transducers. ACS Nano 5:748–760PubMedCrossRef
84.
Zurück zum Zitat Drexler C, Shishkanova TV, Lange C, Danilov SN, Weiss D, Ganichev SD, Mirsky VM (2014) Terahertz split-ring metamaterials as transducers for chemical sensors based on conducting polymers: a feasibility study with sensing of acidic and basic gases using polyaniline chemosensitive layer. Microchim Acta 181:1857–1862CrossRef Drexler C, Shishkanova TV, Lange C, Danilov SN, Weiss D, Ganichev SD, Mirsky VM (2014) Terahertz split-ring metamaterials as transducers for chemical sensors based on conducting polymers: a feasibility study with sensing of acidic and basic gases using polyaniline chemosensitive layer. Microchim Acta 181:1857–1862CrossRef
85.
Zurück zum Zitat Shumaker-Parry J, Campbell C (2004) Quantitative methods for spatially resolved adsorption/desorption measurements in real time by surface plasmon resonance microscopy. Anal Chem 76:918–929PubMedCrossRef Shumaker-Parry J, Campbell C (2004) Quantitative methods for spatially resolved adsorption/desorption measurements in real time by surface plasmon resonance microscopy. Anal Chem 76:918–929PubMedCrossRef
86.
Zurück zum Zitat Nelson BP, Grimsrud TE, Liles MR, Goodman RM, Corn RM (2001) Surface plasmon resonance imaging measurements of DNA and RNA hybridization adsorption onto DNA microarrays. Anal Chem 73:1–7PubMedCrossRef Nelson BP, Grimsrud TE, Liles MR, Goodman RM, Corn RM (2001) Surface plasmon resonance imaging measurements of DNA and RNA hybridization adsorption onto DNA microarrays. Anal Chem 73:1–7PubMedCrossRef
87.
Zurück zum Zitat Zybin A, Kuritsyn YA, Gurevich EL, Temchura VV, Überla K, Niemax K (2010) Real-time detection of single immobilized nanoparticles by surface plasmon resonance imaging. Plasmonics 5:31–35CrossRef Zybin A, Kuritsyn YA, Gurevich EL, Temchura VV, Überla K, Niemax K (2010) Real-time detection of single immobilized nanoparticles by surface plasmon resonance imaging. Plasmonics 5:31–35CrossRef
88.
Zurück zum Zitat Zybin A, Shpacovitch V, Skolnik J, Hergenröder R (2016) Optimal conditions for SPR-imaging of nano-objects. Sensors Actuators B Chem 239:338–342CrossRef Zybin A, Shpacovitch V, Skolnik J, Hergenröder R (2016) Optimal conditions for SPR-imaging of nano-objects. Sensors Actuators B Chem 239:338–342CrossRef
89.
Zurück zum Zitat Nizamov S, Kasian O, Mirsky VM (2016) Individual detection and electrochemically assisted identification of adsorbed nanoparticles by using surface plasmon microscopy. Angew Chem Int Ed 55:1–6CrossRef Nizamov S, Kasian O, Mirsky VM (2016) Individual detection and electrochemically assisted identification of adsorbed nanoparticles by using surface plasmon microscopy. Angew Chem Int Ed 55:1–6CrossRef
90.
Zurück zum Zitat Nizamov S, Scherbahn V, Mirsky VM (2016) Detection and quantification of single engineered nanoparticles in complex samples using template matching in wide-field surface plasmon microscopy. Anal Chem 88:10206–10214PubMedCrossRef Nizamov S, Scherbahn V, Mirsky VM (2016) Detection and quantification of single engineered nanoparticles in complex samples using template matching in wide-field surface plasmon microscopy. Anal Chem 88:10206–10214PubMedCrossRef
91.
Zurück zum Zitat Scherbahn V, Nizamov S, Mirsky VM (2016) Plasmonic detection and visualization of directed adsorption of charged single nanoparticles to patterned surfaces. Microchim Acta 183:2837–2845CrossRef Scherbahn V, Nizamov S, Mirsky VM (2016) Plasmonic detection and visualization of directed adsorption of charged single nanoparticles to patterned surfaces. Microchim Acta 183:2837–2845CrossRef
92.
Zurück zum Zitat Grigorenko AN, Beloglazov AA, Nikitin PI, Kuhne C, Steiner G, Salzer R (2000) Dark-field surface plasmon resonance microscopy. Opt Commun 174:151–155CrossRef Grigorenko AN, Beloglazov AA, Nikitin PI, Kuhne C, Steiner G, Salzer R (2000) Dark-field surface plasmon resonance microscopy. Opt Commun 174:151–155CrossRef
93.
Zurück zum Zitat Kabashin AV, Nikitin PI (1997) Interferometer based on a surface-plasmon resonance for sensor applications. Quantum Electron 27:653–655CrossRef Kabashin AV, Nikitin PI (1997) Interferometer based on a surface-plasmon resonance for sensor applications. Quantum Electron 27:653–655CrossRef
94.
Zurück zum Zitat Nikitin PI, Grigorenko AN, Beloglazov A, Valeiko MV, Savchuk AI, Savchuk OA, Steiner G, Kuhne C, Huebner A, Salzer R (2000) Surface plasmon resonance interferometry for micro-array biosensing. Sensors Actuators A Phys 85:189–193CrossRef Nikitin PI, Grigorenko AN, Beloglazov A, Valeiko MV, Savchuk AI, Savchuk OA, Steiner G, Kuhne C, Huebner A, Salzer R (2000) Surface plasmon resonance interferometry for micro-array biosensing. Sensors Actuators A Phys 85:189–193CrossRef
95.
Zurück zum Zitat Nelson SG, Johnston KS, Yee SS (1996) High sensitivity surface plasmon resonance sensor based on phase detection. Sensors Actuators B Chem 35:187–191CrossRef Nelson SG, Johnston KS, Yee SS (1996) High sensitivity surface plasmon resonance sensor based on phase detection. Sensors Actuators B Chem 35:187–191CrossRef
96.
Zurück zum Zitat Piliarik M, Vaisocherová H, Homola J (2005) A new surface plasmon resonance sensor for high-throughput screening applications. Biosens Bioelectron 20:2104–2110PubMedCrossRef Piliarik M, Vaisocherová H, Homola J (2005) A new surface plasmon resonance sensor for high-throughput screening applications. Biosens Bioelectron 20:2104–2110PubMedCrossRef
97.
Zurück zum Zitat Kashif M, Bakar A, Arsad N, Shaari S (2014) Development of phase detection schemes based on surface plasmon resonance using interferometry. Sensors 14:15914–15938PubMedCrossRef Kashif M, Bakar A, Arsad N, Shaari S (2014) Development of phase detection schemes based on surface plasmon resonance using interferometry. Sensors 14:15914–15938PubMedCrossRef
98.
Zurück zum Zitat Tao NJ, Boussaad S, Huang WL, Arechabaleta RA, D’Agnese J (1999) High resolution surface plasmon resonance spectroscopy. Rev Sci Instrum 70:4656–4660CrossRef Tao NJ, Boussaad S, Huang WL, Arechabaleta RA, D’Agnese J (1999) High resolution surface plasmon resonance spectroscopy. Rev Sci Instrum 70:4656–4660CrossRef
99.
Zurück zum Zitat Berger CEH, Greve J (2000) Differential SPR immunosensing. Sensors Actuators B Chem 63:103–108CrossRef Berger CEH, Greve J (2000) Differential SPR immunosensing. Sensors Actuators B Chem 63:103–108CrossRef
100.
Zurück zum Zitat Jory MJ, Bradberry GW, Cann PS, Sambles JR (1995) A surface-plasmon-based optical sensor using acousto-optics. Meas Sci Technol 6:1193–1200CrossRef Jory MJ, Bradberry GW, Cann PS, Sambles JR (1995) A surface-plasmon-based optical sensor using acousto-optics. Meas Sci Technol 6:1193–1200CrossRef
101.
Zurück zum Zitat Zybin A, Grunwald C, Mirsky VM, Kuhlmann J, Wolfbeis OS, Niemax K (2005) Double-wavelength technique for surface plasmon resonance measurements: basic concept and applications for single sensors and two-dimensional sensor arrays. Anal Chem 77:2393–2399PubMedCrossRef Zybin A, Grunwald C, Mirsky VM, Kuhlmann J, Wolfbeis OS, Niemax K (2005) Double-wavelength technique for surface plasmon resonance measurements: basic concept and applications for single sensors and two-dimensional sensor arrays. Anal Chem 77:2393–2399PubMedCrossRef
102.
Zurück zum Zitat Boecker D, Zybin A, Horvatic V, Grunwald C, Niemax K (2007) Differential surface plasmon resonance imaging for high-throughput bioanalyses. Anal Chem 79:702–709PubMedCrossRef Boecker D, Zybin A, Horvatic V, Grunwald C, Niemax K (2007) Differential surface plasmon resonance imaging for high-throughput bioanalyses. Anal Chem 79:702–709PubMedCrossRef
103.
Zurück zum Zitat Boozer C, Yu Q, Chen S, Lee C-Y, Homola J, Yee SS, Jiang S (2003) Surface functionalization for self-referencing surface plasmon resonance (SPR) biosensors by multi-step self-assembly. Sensors Actuators B Chem 90:22–30CrossRef Boozer C, Yu Q, Chen S, Lee C-Y, Homola J, Yee SS, Jiang S (2003) Surface functionalization for self-referencing surface plasmon resonance (SPR) biosensors by multi-step self-assembly. Sensors Actuators B Chem 90:22–30CrossRef
104.
Zurück zum Zitat Eddings MA, Eckman JW, Arana CA, Papalia GA, Connolly JE, Gale BK, Myszka DG (2009) “Spot and hop”: internal referencing for surface plasmon resonance imaging using a three-dimensional microfluidic flow cell array. Anal Biochem 385:309–313PubMedCrossRef Eddings MA, Eckman JW, Arana CA, Papalia GA, Connolly JE, Gale BK, Myszka DG (2009) “Spot and hop”: internal referencing for surface plasmon resonance imaging using a three-dimensional microfluidic flow cell array. Anal Biochem 385:309–313PubMedCrossRef
105.
Zurück zum Zitat Sigal GB, Mrksich M, Whitesides GM (1997) Using surface plasmon resonance spectroscopy to measure the association of detergents with self-assembled monolayers of hexadecanethiolate on gold. Langmuir 13:2749–2755CrossRef Sigal GB, Mrksich M, Whitesides GM (1997) Using surface plasmon resonance spectroscopy to measure the association of detergents with self-assembled monolayers of hexadecanethiolate on gold. Langmuir 13:2749–2755CrossRef
106.
Zurück zum Zitat O’Brien II MJ, Brueck SRJ, Perez-Luna VH, Tender LM, Lopez GP (1999) SPR biosensors: simultaneously removing thermal and bulk-composition effects. Biosens Bioelectron 14:145–154CrossRef O’Brien II MJ, Brueck SRJ, Perez-Luna VH, Tender LM, Lopez GP (1999) SPR biosensors: simultaneously removing thermal and bulk-composition effects. Biosens Bioelectron 14:145–154CrossRef
107.
Zurück zum Zitat Homola J, Lu HB, Yee SS (1999) Dual-channel surface plasmon resonance sensor with spectral discrimination of sensing channels using dielectric overlayer. Electron Lett 35:1105–1106CrossRef Homola J, Lu HB, Yee SS (1999) Dual-channel surface plasmon resonance sensor with spectral discrimination of sensing channels using dielectric overlayer. Electron Lett 35:1105–1106CrossRef
108.
Zurück zum Zitat Lide DR (1971) Handbook of chemistry and physics. CRC Press, Boca Raton Lide DR (1971) Handbook of chemistry and physics. CRC Press, Boca Raton
109.
Zurück zum Zitat Grassi JH, Georgiadis RM (1999) Temperature-dependent refractive index determination from critical angle measurements: implications for quantitative SPR sensing. Anal Chem 71:4392–4396PubMedCrossRef Grassi JH, Georgiadis RM (1999) Temperature-dependent refractive index determination from critical angle measurements: implications for quantitative SPR sensing. Anal Chem 71:4392–4396PubMedCrossRef
110.
Zurück zum Zitat Kaufmann T, Ravoo BJ (2010) Stamps, inks and substrates: polymers in microcontact printing. Polym Chem 1:371–387CrossRef Kaufmann T, Ravoo BJ (2010) Stamps, inks and substrates: polymers in microcontact printing. Polym Chem 1:371–387CrossRef
111.
Zurück zum Zitat Pimpin A, Srituravanich W (2012) Review on micro- and nanolithography techniques and their applications. Eng J 16:37–55CrossRef Pimpin A, Srituravanich W (2012) Review on micro- and nanolithography techniques and their applications. Eng J 16:37–55CrossRef
112.
Zurück zum Zitat Yu F, Tian S, Yao D, Knoll W (2004) Surface plasmon enhanced diffraction for label-free biosensing. Anal Chem 76:3530–3535PubMedCrossRef Yu F, Tian S, Yao D, Knoll W (2004) Surface plasmon enhanced diffraction for label-free biosensing. Anal Chem 76:3530–3535PubMedCrossRef
113.
Zurück zum Zitat Yu F, Knoll W (2004) Immunosensor with self-referencing based on surface plasmon diffraction. Anal Chem 76:1971–1975PubMedCrossRef Yu F, Knoll W (2004) Immunosensor with self-referencing based on surface plasmon diffraction. Anal Chem 76:1971–1975PubMedCrossRef
114.
Zurück zum Zitat Peterlinz KA, Georgiadis R (1996) In situ kinetics of self-assembly by surface plasmon resonance spectroscopy. Langmuir 12:4731–4740CrossRef Peterlinz KA, Georgiadis R (1996) In situ kinetics of self-assembly by surface plasmon resonance spectroscopy. Langmuir 12:4731–4740CrossRef
115.
Zurück zum Zitat Zacher T, Wischerhoff E (2002) Real-time two-wavelength surface plasmon resonance as a tool for the vertical resolution of binding processes in biosensing hydrogels. Langmuir 18:1748–1759CrossRef Zacher T, Wischerhoff E (2002) Real-time two-wavelength surface plasmon resonance as a tool for the vertical resolution of binding processes in biosensing hydrogels. Langmuir 18:1748–1759CrossRef
116.
Zurück zum Zitat Slavík R, Homola J (2006) Simultaneous excitation of long and short range surface plasmons in an asymmetric structure. Opt Commun 259:507–512CrossRef Slavík R, Homola J (2006) Simultaneous excitation of long and short range surface plasmons in an asymmetric structure. Opt Commun 259:507–512CrossRef
117.
Zurück zum Zitat Medina C, Santos-Martinez MJ, Radomski A, Corrigan OI, Radomski MW (2007) Nanoparticles: pharmacological and toxicological significance. Br J Pharmacol 150:552–558PubMedPubMedCentralCrossRef Medina C, Santos-Martinez MJ, Radomski A, Corrigan OI, Radomski MW (2007) Nanoparticles: pharmacological and toxicological significance. Br J Pharmacol 150:552–558PubMedPubMedCentralCrossRef
118.
Zurück zum Zitat Hoet PH, Brüske-Hohlfeld I, Salata OV (2004) Nanoparticles – known and unknown health risks. J Nanobiotechnol 2:12CrossRef Hoet PH, Brüske-Hohlfeld I, Salata OV (2004) Nanoparticles – known and unknown health risks. J Nanobiotechnol 2:12CrossRef
119.
Zurück zum Zitat Raz SR, Leontaridou M, Bremer MGEG, Peters R, Weigel S (2012) Development of surface plasmon resonance-based sensor for detection of silver nanoparticles in food and the environment. Anal Bioanal Chem 403:2843–2850CrossRef Raz SR, Leontaridou M, Bremer MGEG, Peters R, Weigel S (2012) Development of surface plasmon resonance-based sensor for detection of silver nanoparticles in food and the environment. Anal Bioanal Chem 403:2843–2850CrossRef
120.
Zurück zum Zitat Wang S, Shan X, Patel U, Huang X, Lu J, Li J, Tao N (2010) Label-free imaging, detection, and mass measurement of single viruses by surface plasmon resonance. PNAS 107:16028–16032PubMedCrossRef Wang S, Shan X, Patel U, Huang X, Lu J, Li J, Tao N (2010) Label-free imaging, detection, and mass measurement of single viruses by surface plasmon resonance. PNAS 107:16028–16032PubMedCrossRef
121.
Zurück zum Zitat Halpern AR, Wood JB, Wang Y, Corn RM (2014) Single-nanoparticle near infrared surface plasmon resonance microscopy for real-time measurements of DNA hybridization adsorption. ACS Nano 8:1022–1030PubMedCrossRef Halpern AR, Wood JB, Wang Y, Corn RM (2014) Single-nanoparticle near infrared surface plasmon resonance microscopy for real-time measurements of DNA hybridization adsorption. ACS Nano 8:1022–1030PubMedCrossRef
122.
Zurück zum Zitat Sidorenko I, Nizamov S, Hergenröder R, Zybin A, Kuzmichev A, Kiwull B, Niessner R, Mirsky VM (2016) Computer assisted detection and quantification of single adsorbing nanoparticles by differential surface plasmon microscopy. Microchim Acta 183:101–109CrossRef Sidorenko I, Nizamov S, Hergenröder R, Zybin A, Kuzmichev A, Kiwull B, Niessner R, Mirsky VM (2016) Computer assisted detection and quantification of single adsorbing nanoparticles by differential surface plasmon microscopy. Microchim Acta 183:101–109CrossRef
123.
Zurück zum Zitat Nizamov S, Scherbahn V, Mirsky VM (2017) Advanced wide-field surface plasmon microscopy of single adsorbing nanoparticles. Proc SPIE 10231:102312FCrossRef Nizamov S, Scherbahn V, Mirsky VM (2017) Advanced wide-field surface plasmon microscopy of single adsorbing nanoparticles. Proc SPIE 10231:102312FCrossRef
124.
Zurück zum Zitat Nizamov S, Scherbahn V, Mirsky VM (2017) Wide-field surface plasmon microscopy of nano- and microparticles: features, benchmarking, limitations, and bioanalytical applications. Proc SPIE 10231:1023108CrossRef Nizamov S, Scherbahn V, Mirsky VM (2017) Wide-field surface plasmon microscopy of nano- and microparticles: features, benchmarking, limitations, and bioanalytical applications. Proc SPIE 10231:1023108CrossRef
125.
Zurück zum Zitat Gurevich EL, Temchura VV, Überla K, Zybin A (2011) Analytical features of particle counting sensor based on plasmon assisted microscopy of nano objects. Sensors Actuators B Chem 160:1210–1215CrossRef Gurevich EL, Temchura VV, Überla K, Zybin A (2011) Analytical features of particle counting sensor based on plasmon assisted microscopy of nano objects. Sensors Actuators B Chem 160:1210–1215CrossRef
127.
Zurück zum Zitat Nizamov S, Scherbahn V, Mirsky VM (2017) Ionic referencing in surface plasmon microscopy: visualization of the difference in surface properties of patterned monomolecular layers. Anal Chem 89:3873–3878PubMedCrossRef Nizamov S, Scherbahn V, Mirsky VM (2017) Ionic referencing in surface plasmon microscopy: visualization of the difference in surface properties of patterned monomolecular layers. Anal Chem 89:3873–3878PubMedCrossRef
128.
Zurück zum Zitat Radke RJ, Andra S, Al-Kofahi O, Roysam B (2005) Image change detection algorithms: a systematic survey. IEEE Trans Image Process 14:294–307PubMedCrossRef Radke RJ, Andra S, Al-Kofahi O, Roysam B (2005) Image change detection algorithms: a systematic survey. IEEE Trans Image Process 14:294–307PubMedCrossRef
129.
Zurück zum Zitat Yu F, Yao D, Knoll W (2003) Surface plasmon field-enhanced fluorescence spectroscopy studies of the interaction between an antibody and its surface-coupled antigen. Anal Chem 75:2610–2617PubMedCrossRef Yu F, Yao D, Knoll W (2003) Surface plasmon field-enhanced fluorescence spectroscopy studies of the interaction between an antibody and its surface-coupled antigen. Anal Chem 75:2610–2617PubMedCrossRef
130.
Zurück zum Zitat Mao L, Yuan R, Chai Y, Zhuo Y, Xiang Y (2011) Signal-enhancer molecules encapsulated liposome as a valuable sensing and amplification platform combining the aptasensor for ultrasensitive ECL immunoassay. Biosens Bioelectron 26:4204–4028PubMedCrossRef Mao L, Yuan R, Chai Y, Zhuo Y, Xiang Y (2011) Signal-enhancer molecules encapsulated liposome as a valuable sensing and amplification platform combining the aptasensor for ultrasensitive ECL immunoassay. Biosens Bioelectron 26:4204–4028PubMedCrossRef
131.
Zurück zum Zitat Terrettaz S, Stora T, Duschl C, Vogel H (1993) Protein binding to supported lipid membranes: investigation of the cholera toxin-ganglioside interaction by simultaneous impedance spectroscopy and surface plasmon resonance. Langmuir 9:1361–1369CrossRef Terrettaz S, Stora T, Duschl C, Vogel H (1993) Protein binding to supported lipid membranes: investigation of the cholera toxin-ganglioside interaction by simultaneous impedance spectroscopy and surface plasmon resonance. Langmuir 9:1361–1369CrossRef
132.
Zurück zum Zitat Patskovsky S, Latendresse V, Dallaire A-M, Doré-Mathieu L, Meunier M (2014) Combined surface plasmon resonance and impedance spectroscopy systems for biosensing. Analyst 139:596–602PubMedCrossRef Patskovsky S, Latendresse V, Dallaire A-M, Doré-Mathieu L, Meunier M (2014) Combined surface plasmon resonance and impedance spectroscopy systems for biosensing. Analyst 139:596–602PubMedCrossRef
133.
Zurück zum Zitat Mirsky VM (2011) Quantitative characterization of affinity properties of immobilized receptors. Artificial receptors for chemical sensors. Willey, Weinheim, p 486 Mirsky VM (2011) Quantitative characterization of affinity properties of immobilized receptors. Artificial receptors for chemical sensors. Willey, Weinheim, p 486
134.
Zurück zum Zitat Umpleby RJ, Baxter SC, Chen Y, Shah RN, Shimizu KD (2001) Characterization of molecularly imprinted polymers with the Langmuir - Freundlich isotherm. Anal Chem 73:4584–4591PubMedCrossRef Umpleby RJ, Baxter SC, Chen Y, Shah RN, Shimizu KD (2001) Characterization of molecularly imprinted polymers with the Langmuir - Freundlich isotherm. Anal Chem 73:4584–4591PubMedCrossRef
135.
Zurück zum Zitat Kim H, Kaczmarski K, Guiochon G (2006) Thermodynamic analysis of the heterogenous binding sites of molecularly imprinted polymers. J Chromatogr A 1101:136–152PubMedCrossRef Kim H, Kaczmarski K, Guiochon G (2006) Thermodynamic analysis of the heterogenous binding sites of molecularly imprinted polymers. J Chromatogr A 1101:136–152PubMedCrossRef
136.
Zurück zum Zitat Mirsky VM (2001) Affinity sensors in non-equilibrium conditions: highly selective chemosensing by means of low selective chemosensors. Sensors 1:13–17CrossRef Mirsky VM (2001) Affinity sensors in non-equilibrium conditions: highly selective chemosensing by means of low selective chemosensors. Sensors 1:13–17CrossRef
137.
Zurück zum Zitat Naghibi H, Tamura A, Sturtevant JM (1995) Significant discrepancies between van’t Hoff and calorimetric enthalpies. PNAS 92:5597–5599PubMedCrossRef Naghibi H, Tamura A, Sturtevant JM (1995) Significant discrepancies between van’t Hoff and calorimetric enthalpies. PNAS 92:5597–5599PubMedCrossRef
138.
Zurück zum Zitat Zhurkov A, Karlsson R (2007) Statistical aspects of van’t Hoff analysis: a simulation study. J Mol Recognit 20:379–385CrossRef Zhurkov A, Karlsson R (2007) Statistical aspects of van’t Hoff analysis: a simulation study. J Mol Recognit 20:379–385CrossRef
139.
Zurück zum Zitat Wynne-Jones WFK, Eyring H (1935) The absolute rate of reactions in condensed phases. J Chem Phys 3:492–502CrossRef Wynne-Jones WFK, Eyring H (1935) The absolute rate of reactions in condensed phases. J Chem Phys 3:492–502CrossRef
140.
Zurück zum Zitat Glaser RW (1993) Antigen-antibody binding and mass transport by convection and diffusion to a surface: a two dimensional computer model of binding and dissociation kinetics. Anal Biochem 213:152–161PubMedCrossRef Glaser RW (1993) Antigen-antibody binding and mass transport by convection and diffusion to a surface: a two dimensional computer model of binding and dissociation kinetics. Anal Biochem 213:152–161PubMedCrossRef
141.
Zurück zum Zitat Lok BK, Cheng YL, Robertson CR (1983) Total internal reflection fluorescence: a technique for examining interactions of macromolecules with solid surfaces. J Colloid Interface Sci 91:87–103CrossRef Lok BK, Cheng YL, Robertson CR (1983) Total internal reflection fluorescence: a technique for examining interactions of macromolecules with solid surfaces. J Colloid Interface Sci 91:87–103CrossRef
142.
Zurück zum Zitat Svitel J, Boukari H, Van Ryk D, Willson RC, Schuck P (2007) Probing the functional heterogeneity of surface binding sites by analysis of experimental binding traces and the effect of mass transport limitation. Biophys J 92:1742–1758PubMedCrossRef Svitel J, Boukari H, Van Ryk D, Willson RC, Schuck P (2007) Probing the functional heterogeneity of surface binding sites by analysis of experimental binding traces and the effect of mass transport limitation. Biophys J 92:1742–1758PubMedCrossRef
143.
Zurück zum Zitat Filippov LK, Filippova NL (1997) Adsorption kinetics of polyelectrolytes on planar surfaces under flow conditions. J Colloid Interface Sci 189:1–16CrossRef Filippov LK, Filippova NL (1997) Adsorption kinetics of polyelectrolytes on planar surfaces under flow conditions. J Colloid Interface Sci 189:1–16CrossRef
144.
Zurück zum Zitat Winzor DJ, Jackson CM (2006) Interpretation of the temperature dependence of equilibrium and rate constants. J Mol Recognit 19:389–407PubMedCrossRef Winzor DJ, Jackson CM (2006) Interpretation of the temperature dependence of equilibrium and rate constants. J Mol Recognit 19:389–407PubMedCrossRef
Metadaten
Titel
Toward Ultrasensitive Surface Plasmon Resonance Sensors
verfasst von
Vitali Scherbahn
Shavkat Nizamov
Vladimir M. Mirsky
Copyright-Jahr
2018
DOI
https://doi.org/10.1007/5346_2017_21

    Marktübersichten

    Die im Laufe eines Jahres in der „adhäsion“ veröffentlichten Marktübersichten helfen Anwendern verschiedenster Branchen, sich einen gezielten Überblick über Lieferantenangebote zu verschaffen.