Use of primary filters for removal of characteristic tube lines

verfasst von : R. Jenkins, J. L. de Vries

Erschienen in: Worked Examples in X-Ray Analysis

Verlag: Macmillan Education UK

Aktivieren Sie unsere intelligente Suche, um passende Fachinhalte oder Patente zu finden.

search-config
loading …

It is necessary to measure low concentrations of chromium using a spectrometer equipped with a chromium anode X-ray tube. A titanium filter can be fitted over the tube window to minimize the intensity of the Cr Kα radiation arising from the scattered tube radiation. Two thicknesses of titanium are available, at 0.13 mm and 0.011 mm.