1 Introduction
The vertebral motor unit is, as for every part of the body, exposed to the possibility of degenerative pathologies, a natural process of global aging of the osteo-ligamentous structures of the vertebral column, and/or traumatic and congenital pathologies, such as to cause a structural disorder. In cases of pathologies in which results are not obtained using corrective aids or with physiotherapy, an intervention, known as pedicle arthrodesis, is indispensable. In particular, arthrodesis of the lumbar vertebrae is a surgical technique that allows to join bones in the lumbar spine to stabilize it in order to reduce pain or deformity. Experience shows that the clinical aid that best allows to guide the screws in the spinal fusion surgery with extreme precision is the use of a surgical mask [
1‐
4], a vertebral drilling template designed for delicate pedicle arthrodesis operations. The surgical device allows to implant the screws in the vertebral tracts of interest with extreme safety and precision; the result is a substantial appreciable decrease in operating times (about 40%) and a massive reduction in the harmful ionizing radiation (about 80%) due to the X-ray control so far essential to perform the intervention [
5,
6].
The templates, customized for individual patients, are designed according to the most modern CAD technologies and made thanks to 3D printing in bio-compatible material [
7]. A normal pedicle arthrodesis operation requires that the screws must be inserted by previously making two suitably positioned holes on the vertebra which, as a rule, pass through the pedicles and are fixed in the vertebra body. Vertebra fixation is the most complex phase for this type of operation; it can be compromised by human error, morphology and cleaning of the vertebra, as extremely delicate tissues and structures are present in the intervention area which, if compromised, also cause irreversible damage to the patient. The solution to this problem consists in the design and construction of a “custom made” template for vertebral surgery, customized and optimized for the individual patient, aimed at the directional drilling of the vertebrae with a structure characterized by two hollow cylindrical geometry guides that have the function of guide the tip of the surgical drilling tool so that the hole is drilled in the position chosen in the pre-operative phase by the surgeon. Although the area to be operated is “clean”, as the surgical team is preparing for a skeletal phase of the affected section, blood spills are inevitable throughout the intervention phase. This causes instability due to the positioning of the drilling mask on the vertebral surface due to the presence of superficial and deep muscle tissues and bundles affecting the area to be operated. The solution to this problem could be to design a template whose geometric profile perfectly follows the shape of the vertebral body on which it concerns. The presence of undercuts that can be found in the anchorage areas of the template can cause, however, problems during the insertion and extraction phase of the same. This type of problem is very similar to that encountered in the manufacture of molds in industrial processes. The mold must generally be designed without cavities that can limit the extraction of the piece: in particular, undercuts must be avoided, i.e. angles less than 90°, which, in fact, make the molded element indivisible from the mold. The parts of the model that during the extraction would ruin the shape are said to be undercut. The extraction of the model is possible only in the absence of undercuts, i.e. all those areas of the model that are in the shadow of the direction of extraction of the model itself. Multiple systems of automatic undercut recognition are now consolidated, according to the geometry to be obtained by means of the mold, starting from three-dimensional CAD models [
8‐
13].
None of these methods, however, has ever been used so far for the design of such surgical devices, leading over time to neglect the problem of undercuts in vertebral drilling templates, focusing on design requirements, such as the direction of screw centering. The aim of this work, instead, is to dwell on the problems resulting from the use of the device, during the insertion and extraction phases, ensuring precision and maintenance of the positioning at the same time. This is possible through the implementation of an algorithm capable of eliminating the undercuts on the vertebra-template interaction surface, evaluating the uniqueness of positioning and the stability. The vertebra-template coupling is even more stable as the contact surface is extended. Thinking of creating a profile of the template that fully matches that of the spinous process would, however, lead to evident problems in the phase of fixing the template on the vertebra as well as in the phase of extraction of the same. Furthermore, carefully analysing the morphology of a lumbar vertebra, it can be noticed that the major surface irregularities are those affecting the spinous process. Imagining, therefore, to position the drilling mask on the vertebra, the undercut problems that are encountered are precisely those affecting the spinous process and the cavity obtained inside the central body of the template. In this regard, it is not immediate to define the correct extraction directions for the model, given that the undercuts present could prevent more than one. The design and subsequent construction of a template is, therefore, bound by the extraction direction established during the design phase. The problem is, however, that by bypassing the undercuts, contact surfaces are eliminated, reducing stability. It is important to evaluate the optimal extraction direction taking also into account the stability.
The aim of this work was to implement a corrective semi-automatic model that, after identifying the presence of any undercuts, is able to modify the anchoring profile of the template according to the extraction direction, to determine its final contact surface and its stability. The profile of the template, therefore, will be that of a ‘perfectly matching’ geometry with the spinous process, subtracted, however, from the relevant undercuts. In this way, it will be possible to position and extract the model from the vertebra without difficulty, without affecting any effect on stability, due to a smaller contact surface.
3 Conclusion
In this work, a solution was found to the problem of interference, represented by the so-called shaded areas, present when inserting a template for vertebral drilling along a generic direction; the problems encountered concern the positioning of the template on the vertebral surface. These limits are mainly related to the presence of undercuts, i.e. those areas of the spinous process which, brought back into the cavity of the model body, entail complications at the time of extraction. The problem of undercuts has been limited in the present work by reducing the contact surface of the prototype with that of the spinous process. The procedure here presented allows to modify the geometric profile of the vertebral template according to the extraction direction chosen, and to compare different directions with each other, in order to identify the best solution that can, on the one hand, overcome the impediments of extraction caused by the undercuts and, on the other hand, guarantee the maximum possible adherence, compensating for the problem of stability and for the identification of a unique positioning.
Open Access This chapter is licensed under the terms of the Creative Commons Attribution 4.0 International License (
http://creativecommons.org/licenses/by/4.0/), which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license and indicate if changes were made.
The images or other third party material in this chapter are included in the chapter's Creative Commons license, unless indicated otherwise in a credit line to the material. If material is not included in the chapter's Creative Commons license and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder.