11. Cantilevers and Detection Methods in Atomic Force Microscopy

Author : Bert Voigtländer

Published in: Atomic Force Microscopy

Publisher: Springer International Publishing

Activate our intelligent search to find suitable subject content or patents.

search-config
loading …

Abstract

We consider basic requirements for force sensors and introduce a fabrication process for cantilevers. Subsequently, the most common detection method for measuring the cantilever deflection, the beam deflection method, is discussed in detail.

Dont have a licence yet? Then find out more about our products and how to get one now:

Springer Professional "Wirtschaft+Technik"

Online-Abonnement

Mit Springer Professional "Wirtschaft+Technik" erhalten Sie Zugriff auf:

  • über 102.000 Bücher
  • über 537 Zeitschriften

aus folgenden Fachgebieten:

  • Automobil + Motoren
  • Bauwesen + Immobilien
  • Business IT + Informatik
  • Elektrotechnik + Elektronik
  • Energie + Nachhaltigkeit
  • Finance + Banking
  • Management + Führung
  • Marketing + Vertrieb
  • Maschinenbau + Werkstoffe
  • Versicherung + Risiko

Jetzt Wissensvorsprung sichern!

Springer Professional "Technik"

Online-Abonnement

Mit Springer Professional "Technik" erhalten Sie Zugriff auf:

  • über 67.000 Bücher
  • über 390 Zeitschriften

aus folgenden Fachgebieten:

  • Automobil + Motoren
  • Bauwesen + Immobilien
  • Business IT + Informatik
  • Elektrotechnik + Elektronik
  • Energie + Nachhaltigkeit
  • Maschinenbau + Werkstoffe




 

Jetzt Wissensvorsprung sichern!

Footnotes
1
The bending of the cantilever end by the angle \(\theta \) results in an angle of \(2 \theta \) for the reflected beam. Thus, the linear deflection of the reflected laser beam on the photodiode results (for small angles) as \(\Delta x = 2 \theta L\).
 
2
In dynamic AFM the primary bandwidth detecting the oscillatory motion of the cantilever is much larger than 1 kHz used in this quantitative example, however, in this case also the oscillation amplitudes to be detected exceed the pm range by far.
 
3
If the cantilever is tilted with respect to the surface by an angle \(\alpha \), the relation between the force perpendicular to the surface and the deflection perpendicular to the surface is modified [9] to \(F = k \Delta z/ \cos ^2 \alpha \). Since \(\alpha \) is usually small (in the range between \(10^{\circ }\) and \(15^{\circ }\)), this correction is small and will be neglected it in the following.
 
4
Specific effects occurring close to the kink between the two regions are discussed in Sect. 12.​5.
 
5
The parameters \(\omega _0\) and Q can be obtained by measuring a resonance curve of the cantilever in response to an external excitation (frequency sweep over the resonance). Alternatively, the thermal noise spectrum can be measured, as described in the next section.
 
6
The density and viscosity for the most frequently used fluids (air and water) are: \(\rho _{\mathrm {air}} = 1.2\,\)kg/m\(^3\), \(\eta _{\mathrm {air}} = 1.85\times 10^{-5}\) kg/(m s), and \(\rho _{\mathrm {water}} = 1\times 10^3\) kg/m\(^3\), \(\eta _{\mathrm {water}} = 8.9\times 10^{-4}\) kg/(m s), respectively, under ambient conditions and at sea level [12].
 
7
Details of how to extract the noise power spectral density from the time signal without using a spectrum analyzer are given in Appendix B or [15].
 
Literature
3.
go back to reference F. Reif, Fundamentals of Statistical and Thermal Physics, (Waveland Pr. Inc., 1965) ISBN 1577666127 F. Reif, Fundamentals of Statistical and Thermal Physics, (Waveland Pr. Inc., 1965) ISBN 1577666127
11.
go back to reference J.E. Sader, J.A. Sanelli, B.D. Adamson, J.P. Monty, X. Wei, S.A. Crawford, J.R. Friend, I. Marusic, P. Mulvaney, E.J. Bieske, Spring constant calibration of atomic force microscope cantilevers of arbitrary shape. Rev. Sci. Instrum. 83, 103705 (2012). https://doi.org/10.1063/1.4757398 J.E. Sader, J.A. Sanelli, B.D. Adamson, J.P. Monty, X. Wei, S.A. Crawford, J.R. Friend, I. Marusic, P. Mulvaney, E.J. Bieske, Spring constant calibration of atomic force microscope cantilevers of arbitrary shape. Rev. Sci. Instrum. 83, 103705 (2012). https://​doi.​org/​10.​1063/​1.​4757398
16.
go back to reference M.J. Higgins, R. Proksch, J.E. Sader, M. Polcik, S. Mc Endoo, J.P. Cleveland, S.P. Jarvis, Noninvasive determination of optical lever sensitivity in atomic force microscopy. Rev. Sci. Instrum. 77, 013701 (2006). https://doi.org/10.1063/1.2162455 M.J. Higgins, R. Proksch, J.E. Sader, M. Polcik, S. Mc Endoo, J.P. Cleveland, S.P. Jarvis, Noninvasive determination of optical lever sensitivity in atomic force microscopy. Rev. Sci. Instrum. 77, 013701 (2006). https://​doi.​org/​10.​1063/​1.​2162455
Metadata
Title
Cantilevers and Detection Methods in Atomic Force Microscopy
Author
Bert Voigtländer
Copyright Year
2019
DOI
https://doi.org/10.1007/978-3-030-13654-3_11