2012 | OriginalPaper | Chapter
Efficient Positioning of the Field-Free Point in Magnetic Particle Imaging
Authors : Tobias Knopp, Timo F. Sattel, Thorsten M. Buzug
Published in: Magnetic Particle Imaging
Publisher: Springer Berlin Heidelberg
Activate our intelligent search to find suitable subject content or patents.
Select sections of text to find matching patents with Artificial Intelligence. powered by
Select sections of text to find additional relevant content using AI-assisted search. powered by
Magnetic particle imaging uses a field with a single field-free point for spatial encoding. Such a gradient field can be generated by a Maxwell coil pair consisting of two opposing coils driven by current flowing in converse directions. In order to sample the volume in-between the coils, the field-free point has to be moved through space. When keeping the gradient strength constant the electrical power loss of the Maxwell coil pair setup drastically increases when moving the field-free point off-center. In this paper a coil configuration is proposed, which consists of four coils and considerably reduces the electrical power loss for off-center field-free point generation.