Verification of patient specific quality assurance system for volumetric modulated arc therapy (VMAT)

Authors: Kwame Anokye Amoabeng, Anne Beate Langeland Marthinsen, Francis Hasford, Samuel Nii Adu Tagoe, Evelyn Anaafi

Published in: Health and Technology | Issue 4/2022

Log in

Activate our intelligent search to find suitable subject content or patents.

search-config
loading …

Abstract

Volumetric Modulated Arc Therapy (VMAT) plans and its treatment are complex radiation therapy modalities and its clinical implementation requires very accurate acceptance testing and a comprehensive quality assurance program. Failure in their periodic applications can result in errors in treatment delivery. It is therefore needed to compare the calculated dose distribution to the measured dose distribution of treatment plans to ensure accuracy and efficiency to minimize errors in planned radiation dose delivery to the patient. VMAT plans for fifteen patients with cervical cancer with pathological pelvic lymph node metastasis were included in this study. Two VMAT plans were made for each patient, one using conventional 6 MV beam with flattening filter and one with flattening filter free beam (FFF). The VMAT plans were made using two arcs, and then recalculated to give the planned dose distribution to the detectors in a Delta4 phantom. All thirty VMAT plans were irradiated on the Delta4 phantom using three Elekta linear accelerators, and planned and calculated doses to the phantom were compared. Measurements were done both before and after a routine linac optimization, and variation in radiation delivery between the linacs were studied. After machine optimization, there was an increase in the mean total gamma pass rate for both the 6 MV plans from 98.7 to 99.9% and the FFF plans also had an increase in the gamma pass rate from 91.7% to 98.4% for SB5 linac. There was significant statistical difference (p < 0.05) between the pass rates of the 6 MV plans and the FFF plans. 6 MV VMAT plans with conventional flattened beams are delivered more accurately and hence more beneficial compared to flattening filter free (FFF) VMAT plans for external radiation of cervical cancer with affected pelvic lymph nodes.

Dont have a licence yet? Then find out more about our products and how to get one now:

Springer Professional "Wirtschaft+Technik"

Online-Abonnement

Mit Springer Professional "Wirtschaft+Technik" erhalten Sie Zugriff auf:

  • über 102.000 Bücher
  • über 537 Zeitschriften

aus folgenden Fachgebieten:

  • Automobil + Motoren
  • Bauwesen + Immobilien
  • Business IT + Informatik
  • Elektrotechnik + Elektronik
  • Energie + Nachhaltigkeit
  • Finance + Banking
  • Management + Führung
  • Marketing + Vertrieb
  • Maschinenbau + Werkstoffe
  • Versicherung + Risiko

Jetzt Wissensvorsprung sichern!

Springer Professional "Wirtschaft"

Online-Abonnement

Mit Springer Professional "Wirtschaft" erhalten Sie Zugriff auf:

  • über 67.000 Bücher
  • über 340 Zeitschriften

aus folgenden Fachgebieten:

  • Bauwesen + Immobilien
  • Business IT + Informatik
  • Finance + Banking
  • Management + Führung
  • Marketing + Vertrieb
  • Versicherung + Risiko




Jetzt Wissensvorsprung sichern!

Springer Professional "Technik"

Online-Abonnement

Mit Springer Professional "Technik" erhalten Sie Zugriff auf:

  • über 67.000 Bücher
  • über 390 Zeitschriften

aus folgenden Fachgebieten:

  • Automobil + Motoren
  • Bauwesen + Immobilien
  • Business IT + Informatik
  • Elektrotechnik + Elektronik
  • Energie + Nachhaltigkeit
  • Maschinenbau + Werkstoffe




 

Jetzt Wissensvorsprung sichern!

Literature
1.
go back to reference Cihoric N, Tsikkinis C, Tapia AC. Dose escalated intensity modulated radiotherapy in the treatment of cervical cancer. Radiat Oncol. 2015;10:240.CrossRef Cihoric N, Tsikkinis C, Tapia AC. Dose escalated intensity modulated radiotherapy in the treatment of cervical cancer. Radiat Oncol. 2015;10:240.CrossRef
2.
go back to reference Palisca MG, Eichler TJ, Hartford AC. American Society for Therapeutic Radiology and Oncology (ASTRO) and American College of Radiology (ACR) practice guidelines for intensity modulated radiation therapy (IMRT). Int J Radiat Oncol Biol Phys Med. 2009;73(1). Palisca MG, Eichler TJ, Hartford AC. American Society for Therapeutic Radiology and Oncology (ASTRO) and American College of Radiology (ACR) practice guidelines for intensity modulated radiation therapy (IMRT). Int J Radiat Oncol Biol Phys Med. 2009;73(1).
3.
go back to reference Otto K, Clark CH, Wood K, Whitaker S, Nisbet A. Volumetric modulated arc therapy: a review of current literature and clinical use in practice. Br J Radiol. 2011;84:967–96.CrossRef Otto K, Clark CH, Wood K, Whitaker S, Nisbet A. Volumetric modulated arc therapy: a review of current literature and clinical use in practice. Br J Radiol. 2011;84:967–96.CrossRef
4.
go back to reference Higby C, Khafaga Y, Al-Shabanah M, Mousa A, Ilyas M, Nazer G, et al. Volumetric-modulated arc therapy (VMAT) versus 3D-conformal radiation therapy in supra-diaphragmatic Hodgkin’s Lymphoma with mediastinal involvement: A dosimetric comparison. J Egypt Natl Canc Inst. 2016;28(3):163–8.CrossRef Higby C, Khafaga Y, Al-Shabanah M, Mousa A, Ilyas M, Nazer G, et al. Volumetric-modulated arc therapy (VMAT) versus 3D-conformal radiation therapy in supra-diaphragmatic Hodgkin’s Lymphoma with mediastinal involvement: A dosimetric comparison. J Egypt Natl Canc Inst. 2016;28(3):163–8.CrossRef
5.
go back to reference Allison LM. Evaluating Pre-Treatment IMRT & VMAT QA Techniques Using Receiver Operating Characteristic (ROC) Analysis. UK: Duke University; 2013. Allison LM. Evaluating Pre-Treatment IMRT & VMAT QA Techniques Using Receiver Operating Characteristic (ROC) Analysis. UK: Duke University; 2013.
6.
go back to reference Yoo S. Radiotherapy treatment plans with RapidArc for prostate cancer involving seminal vescicles and lymph nodes. Int J Radiat Oncol Biol Phys. 2010;76:935–42.CrossRef Yoo S. Radiotherapy treatment plans with RapidArc for prostate cancer involving seminal vescicles and lymph nodes. Int J Radiat Oncol Biol Phys. 2010;76:935–42.CrossRef
7.
go back to reference Clivio A, Fogliata A, Nicolini G, Wyttenbach R, Cozzi L. Volumetric modulated arc therapy for carcinomas of the anal canal. A treatment planning comparison with fixed field IMRT. Radiother Oncol. 2009;92:118–24. Clivio A, Fogliata A, Nicolini G, Wyttenbach R, Cozzi L. Volumetric modulated arc therapy for carcinomas of the anal canal. A treatment planning comparison with fixed field IMRT. Radiother Oncol. 2009;92:118–24.
8.
go back to reference Wolff D, Stieler F, Welzel G, Lorenz F, Mai S. Volumetric modulated arc therapy vs serial tomotherapy, step and shoot IMRT and 3D conformal RT for treatment of prostate cancer. Radiat Oncol. 2009;93:226–33.CrossRef Wolff D, Stieler F, Welzel G, Lorenz F, Mai S. Volumetric modulated arc therapy vs serial tomotherapy, step and shoot IMRT and 3D conformal RT for treatment of prostate cancer. Radiat Oncol. 2009;93:226–33.CrossRef
9.
go back to reference Abolaban F, Zaman S, Cashmore J. Changes in Patterns of Intensity-modulated Radiotherapy Verification and Quality Assurance in the UK. Clin Oncol (R Coll Radiol). 2016;28(8):e28-34.CrossRef Abolaban F, Zaman S, Cashmore J. Changes in Patterns of Intensity-modulated Radiotherapy Verification and Quality Assurance in the UK. Clin Oncol (R Coll Radiol). 2016;28(8):e28-34.CrossRef
10.
go back to reference Dixon P, O'Sullivan B. Radiotherapy Quality assurance: time for everyone to take it seriously. Eur J Cancer. 2002;39 423–429. Dixon P, O'Sullivan B. Radiotherapy Quality assurance: time for everyone to take it seriously. Eur J Cancer. 2002;39 423–429.
11.
go back to reference Ling CC, Zhang P, Archambault Y, Bocanek J, Tang G, Lasasso T, et al. Commissioning and quality assurance of RapidArc radiotherapy delivery system. Int J Radiat Oncol Biol Phys. 2008;72:575–81.CrossRef Ling CC, Zhang P, Archambault Y, Bocanek J, Tang G, Lasasso T, et al. Commissioning and quality assurance of RapidArc radiotherapy delivery system. Int J Radiat Oncol Biol Phys. 2008;72:575–81.CrossRef
12.
go back to reference Hartford AC. Practice Guidelines for Intensity Modulated Radiatio Therapy (IMRT). Am J Clin Oncol. 2012;35(6):612–7.MathSciNetCrossRef Hartford AC. Practice Guidelines for Intensity Modulated Radiatio Therapy (IMRT). Am J Clin Oncol. 2012;35(6):612–7.MathSciNetCrossRef
13.
go back to reference Ma C, Jiang SB, Pawlicki T, Chen Y, Li JS, Deng J. A quality assurance phantom for IMRT dose verification. Phys Med Biol. 2003;48:561–72. Ma C, Jiang SB, Pawlicki T, Chen Y, Li JS, Deng J. A quality assurance phantom for IMRT dose verification. Phys Med Biol. 2003;48:561–72.
14.
go back to reference Low A. Evaluation of the gamma dose distribution comparison method. Med Phys. 2003;30(9):2455–64.CrossRef Low A. Evaluation of the gamma dose distribution comparison method. Med Phys. 2003;30(9):2455–64.CrossRef
15.
go back to reference Yoo S. Radiotherapy treatment plans with RapidArc for prostate cancer involving seminal vescicles and lymph nodes. Int J Radiat Oncol Biol Phys. 2010;76:935–42.CrossRef Yoo S. Radiotherapy treatment plans with RapidArc for prostate cancer involving seminal vescicles and lymph nodes. Int J Radiat Oncol Biol Phys. 2010;76:935–42.CrossRef
16.
go back to reference Low DA, Harms WB, Mutic S, Purdy JA. A technique for the quantitative evaluation of dose distributions Med Phys. 1998;25:656–61. Low DA, Harms WB, Mutic S, Purdy JA. A technique for the quantitative evaluation of dose distributions Med Phys. 1998;25:656–61.
17.
go back to reference Fontaine J, Taecsh F, Djibo-Sidikou A, Bertosi A, Retif P. 11 Evaluation of delivery analysis software (Accuray), comparison to Delta 4 Phantom. Physica Med. 2018;56:7–8.CrossRef Fontaine J, Taecsh F, Djibo-Sidikou A, Bertosi A, Retif P. 11 Evaluation of delivery analysis software (Accuray), comparison to Delta 4 Phantom. Physica Med. 2018;56:7–8.CrossRef
18.
go back to reference Capomolla C, Zagari A, Quarta S, Luca DD, Carlà A, Cazzato M, et al. 131 Performance analysis of new Delta4 Phantom+ using Flattening-Filter and Flattening Filter-Free beams. Physica Medica. 2018;56. Capomolla C, Zagari A, Quarta S, Luca DD, Carlà A, Cazzato M, et al. 131 Performance analysis of new Delta4 Phantom+ using Flattening-Filter and Flattening Filter-Free beams. Physica Medica. 2018;56.
19.
go back to reference Vieillevigne L, Molinier J, Brun T, Ferrand R. Gamma index comparison of three VMAT QA systems and evaluation of their sensitivity to delivery errors. Phys Med Biol. 2015;31:720–5. Vieillevigne L, Molinier J, Brun T, Ferrand R. Gamma index comparison of three VMAT QA systems and evaluation of their sensitivity to delivery errors. Phys Med Biol. 2015;31:720–5.
20.
go back to reference Kumar L, Yadav G, Samuvel KR, Bhushan M, Kumar P, Suhail M, et al. Dosimetric influence of filtered and flattening filter free photon beam on rapid arc (RA) radiotherapy planning in case of cervix carcinoma. Rep Pract Oncol Radiother. 2017;22(1):10–8.CrossRef Kumar L, Yadav G, Samuvel KR, Bhushan M, Kumar P, Suhail M, et al. Dosimetric influence of filtered and flattening filter free photon beam on rapid arc (RA) radiotherapy planning in case of cervix carcinoma. Rep Pract Oncol Radiother. 2017;22(1):10–8.CrossRef
21.
go back to reference Bedford JL, Lee YK, Wai P, South CP, Warrington AP. Evaluation of the Delta4 phantom for IMRT and VMAT verification. Phys Med Biol. 2009;54(9):N167–76.CrossRef Bedford JL, Lee YK, Wai P, South CP, Warrington AP. Evaluation of the Delta4 phantom for IMRT and VMAT verification. Phys Med Biol. 2009;54(9):N167–76.CrossRef
22.
go back to reference Titt U, Vassiliev ON, Pönisch F, Dong L, Liu H, Mohan RA. flattening filter free photon treatment concept evaluation with Monte Carlo. Med Phys. 2006;33. Titt U, Vassiliev ON, Pönisch F, Dong L, Liu H, Mohan RA. flattening filter free photon treatment concept evaluation with Monte Carlo. Med Phys. 2006;33.
23.
go back to reference Balik S, Qi P, Magnelli A, Chao ST, Suh JH, Zhuang T. Comparison of Spine SRS VMAT plans with Flattening Filter Free 6 MV or 10 MV beams. Int J Radiat Oncol Biol Phys. 2018;102(3). Balik S, Qi P, Magnelli A, Chao ST, Suh JH, Zhuang T. Comparison of Spine SRS VMAT plans with Flattening Filter Free 6 MV or 10 MV beams. Int J Radiat Oncol Biol Phys. 2018;102(3).
Metadata
Title
Verification of patient specific quality assurance system for volumetric modulated arc therapy (VMAT)
Authors
Kwame Anokye Amoabeng
Anne Beate Langeland Marthinsen
Francis Hasford
Samuel Nii Adu Tagoe
Evelyn Anaafi
Publication date
30-05-2022
Publisher
Springer Berlin Heidelberg
Published in
Health and Technology / Issue 4/2022
Print ISSN: 2190-7188
Electronic ISSN: 2190-7196
DOI
https://doi.org/10.1007/s12553-022-00675-6

Other articles of this Issue 4/2022

Health and Technology 4/2022 Go to the issue

Premium Partner