Published in: International Journal of Computer Assisted Radiology and Surgery 7/2023

Open Access 02-06-2023 | Original Article

Sources of performance variability in deep learning-based polyp detection

Authors: T. N. Tran, T. J. Adler, A. Yamlahi, E. Christodoulou, P. Godau, A. Reinke, M. D. Tizabi, P. Sauer, T. Persicke, J. G. Albert, L. Maier-Hein

Published in: International Journal of Computer Assisted Radiology and Surgery | Issue 7/2023

Activate our intelligent search to find suitable subject content or patents.

search-config
loading …

Abstract

Purpose

Validation metrics are a key prerequisite for the reliable tracking of scientific progress and for deciding on the potential clinical translation of methods. While recent initiatives aim to develop comprehensive theoretical frameworks for understanding metric-related pitfalls in image analysis problems, there is a lack of experimental evidence on the concrete effects of common and rare pitfalls on specific applications. We address this gap in the literature in the context of colon cancer screening.

Methods

Our contribution is twofold. Firstly, we present the winning solution of the Endoscopy Computer Vision Challenge on colon cancer detection, conducted in conjunction with the IEEE International Symposium on Biomedical Imaging 2022. Secondly, we demonstrate the sensitivity of commonly used metrics to a range of hyperparameters as well as the consequences of poor metric choices.

Results

Based on comprehensive validation studies performed with patient data from six clinical centers, we found all commonly applied object detection metrics to be subject to high inter-center variability. Furthermore, our results clearly demonstrate that the adaptation of standard hyperparameters used in the computer vision community does not generally lead to the clinically most plausible results. Finally, we present localization criteria that correspond well to clinical relevance.

Conclusion

We conclude from our study that (1) performance results in polyp detection are highly sensitive to various design choices, (2) common metric configurations do not reflect the clinical need and rely on suboptimal hyperparameters and (3) comparison of performance across datasets can be largely misleading. Our work could be a first step towards reconsidering common validation strategies in deep learning-based colonoscopy and beyond.
Notes

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Introduction

Colorectal cancer is one of the most common cancer types, ranking second in females and third in males [1]. By detecting and subsequently resecting neoplastic polyps during screening colonoscopy, the risk of developing the disease can be reduced significantly. Research focuses on developing deep learning (DL) solutions for automated detection of polyps in colonoscopy videos [26]. However, to date, the metrics with which algorithms are validated receive far too little attention. These metrics are not only important for measuring scientific progress, but also for gauging a method’s potential for clinical translation. While previous work has highlighted general metric pitfalls in the broader context of classification, segmentation and detection [7], we are not aware of any prior studies systematically analyzing common metrics in the context of polyp detection. Our underlying hypothesis was that reported performance values in polyp detection methods are largely misleading as they are sensitive to many validation design choices including (1) the choice of test set and (2) the chosen metric configurations (e.g., threshold for the localization criteria). Our contribution is twofold: Firstly, we present the winning solution of the Endoscopy Computer Vision Challenge (EndoCV) on colon cancer detection, conducted in conjunction with the IEEE International Symposium on Biomedical Imaging (ISBI) 2022. Secondly, based on publicly available challenge data, we demonstrate the sensitivity of commonly used metrics to a range of hyperparameters as well as the consequences of poor metric choices.

Methods

Here, we present the winning method of the EndoCV challenge on colon cancer detection, conducted in conjunction with ISBI 2022 (Sect. 2.1), and revisit common detection metrics including their hyperparameters (Sect. 2.2).

Object detection algorithm

We base our study on a state-of-the-art detection method, namely the winning entry [8] of the EndoCV 2022 polyp detection challenge [4].

Method overview

The method is illustrated in Fig. 1. A heterogeneous ensemble of YOLOv5-based models was trained on the EndoCV2022 challenge training dataset [4], a collection of 46 endoscopic video sequences of polyps. The median length of each video was 51 frames with 3290 frames in total, of which 2631 contained polyps. For training, we split the data into disjoint subsets. Originally, we created four subsets for stratified four-fold cross-validation, but the final ensemble consisted of models which were trained on only two of the four folds to decrease training and inference time. Furthermore, on each fold we trained three models for the ensemble with complementary training setups, as follows: the model was trained with (1) only light data augmentation (horizontal flips, scale, color jitter, and mosaic), (2) heavy data augmentation (light augmentations, vertical flips, translation, rotation, mix-up, shear, and copy-paste augmentations), and (3) light data augmentation, but additional training data from external polyp datasets (CVC-Datasets, Etis-Larib) [911] were added (see [8] for details). Overall, this led to six ensemble members (two folds, three models each). As test sets, we used appropriate subsets of the PolypGen dataset [12] to highlight the variability of the metrics. The dataset consists of a database of single-frame endoscopic images of polyps from six data centers, each with a median of 242 frames, totaling 1512 single testing frames. Details about the exact number of frames and polyp prevalence for each center can be found in Sect. 3.1, Table 2.

Implementation details

The models were trained for 20 epochs using a stochastic gradient descent optimizer, a learning rate of 0.1, and a complete intersection over union [13] loss. The non-maximum suppression algorithm was applied to each ensemble member individually with an intersection over union (IoU) threshold of 0.5. The individual member predictions were merged using the weighted boxes fusion [14] (WBF) algorithm. For the WBF hyperparameters, we chose an IoU threshold of 0.5, a skip box threshold of 0.02, and all models were weighted equally. As we observed a tendency towards oversegmentation, we shrank all bounding boxes with a confidence score higher than 0.4 by 2% of their size during post-processing.

Object detection metrics

Three metric-related design decisions are important when assessing performance of object detection algorithms [15]:
(1)
Localization criterion The localization criterion determines whether a predicted object spatially corresponds to one of the reference objects and vice versa by measuring the spatial similarity between prediction (represented by a bounding box, pixel mask, center point or similar) and reference object. It defines whether the prediction hit/detected (true positive) or missed (false positive) the reference. Any reference object not detected by the algorithm is defined as false negative. The localization criteria that were applied in this work comprise two groups, namely the point-based criteria and the overlap-based criteria (Appendix A).
 
(2)
Assignment strategy As applying the localization criterion might lead to ambiguous matchings, such as two predictions being assigned to the same reference object, an assignment strategy needs to be chosen that determines how potential ambiguities are resolved. As multiple polyps in the same image are rather rare, an assignment strategy is not as relevant as in other applications. With respect to the metric configuration, we therefore focus on the localization criterion and the classification metrics.
 
(3)
Classification metric Based on the choice of localization criterion and assignment strategy, standard classification metrics can be computed at object level [7]. The most popular multi-threshold metric in object detection is Average Precision (AP) (Fig. 9).
 
As a foundation of this work, we determined common metrics in object detection challenges, along with their respective localization criterion and classification metric (Table 1).
Table 1
Common design choices for validation of polyp detection methods according to international competitions
 
Localization criterion
Classification metric
EndoVis 2015 [16]
Point inside mask
PPV
  
Sensitivity
  
Overall/Average F1-score
GIANA 2017 [6]
Point inside mask
PPV
GIANA 2018 [6]
 
Sensitivity
  
F1/F2-score
  
Custom metrics
GIANA 2021 [5]
Box IoU (only 2021)
AP@0.5:0.95 (only 2021)
EndoCV 2021 [3]
Box IoU
AP@0.5, AP@0.75, AP@0.5:0.95
EndoCV 2022 [4]
 
AP across 3 scales of polyp size
  
Mean of 4 APs
PPV positive predictive value, IoU intersection over union, AP average precision

Experiments and results

In this section, we investigate the sensitivity of popular classification metrics to the test set composition (Sect. 3.1) and the localization criterion (Sect. 3.2). We further assess the clinical value of commonly used metric configurations (Sect. 3.3). Details on the metrics and localization criteria are provided in Appendix A.

Effect of test set

In the following, we quantitatively assess the performance variability resulting from the chosen test set, specifically from the target domain (i.e., the clinical validation center) and the distribution of polyp size.

Sensitivity to center

To show the variability of performance resulting from different test sets, we used data from six validation centers [12]. Figure 2 shows the performance of our object detection method (Sect. 2.1) according to commonly used metrics, using Box IoU as criterion and cutoff threshold 0.5. These exhibit high variability between centers. For example, the AP@0.5:0.95 ranges from [0.38, 0.63], which is notable, given that the AP of the top three submissions for EndoCV 2022 ranged from [0.12, 0.33].

Sensitivity to polyp size

We further calculated the AP scores from all six validation centers, stratified by polyp size (Table 2). The polyp sizes are defined in pixels for images with \(1920 \times 1080\) pixel resolution. Polyps smaller than \(100^2\) pixels are considered small, polyps larger than \(200^2\) pixels are counted as large and the rest are counted as medium sized, as suggested in the EndoCV2021 challenge [3]. A high variability can be observed, indicating that algorithm performance is highly affected by the distribution of polyp sizes.
Table 2
AP stratified by polyp size. The results are shown for a fixed IoU threshold of 0.5 (left) as well as for a range of thresholds following the COCO benchmark evaluation standard [17, 18] (center)
Metric
AP@0.5
AP@0.5:0.95
n
\(\phi (\%) \)
Polyp size
Small
Medium
Large
Small
Medium
Large
Center 1
0.24
0.48
0.73
0.14
0.33
0.55
256
98
Center 2
0.00
0.46
0.81
0.00
0.31
0.67
276
89
Center 3
0.16
0.64
0.91
0.10
0.47
0.72
457
99
Center 4
0.14
0.33
0.59
0.06
0.27
0.43
227
64
Center 5
0.19
0.39
0.59
0.12
0.26
0.42
208
99
Center 6
0.00
0.52
0.89
0.00
0.39
0.67
88
94
All centers (SD)
0.14 (0.1)
0.47 (0.11)
0.75 (0.14)
0.07 (0.06)
0.34 (0.08)
0.58 (0.13)
1512
91
We provide additional information on the number of frames (n) and polyp prevalence (\(\phi \)) per center (right). AP average precision, IoU intersection over union, SD standard deviation
To further evaluate how the IoU values relate to polyp size and polyp type and simultaneously account for the hierarchical structure of the data set, we fit a linear mixed effects model (R version 4.1.3, package lme4). In this model, polyp size (small, medium, or large) and polyp type (flat or protruded) were fixed effects, while data center, patient identifier (ID), and image ID were random effects. The results suggest that there are strong effects of polyp type and polyp size on the IoU values. In particular when the polyp is of a protruded as opposed to a flat type, the values of IoU are on average higher by a difference of 0.08 (conditional that the other predictors remain constant). When the polyp is of a medium or small size compared to a large size, the IoU values are lower by a difference of 0.05 and 0.28, respectively (conditional that the remaining predictors remain constant).

Effect of metric configuration

In the case of polyp detection, the goal of high sensitivity (not missing a polyp) is an indispensable priority. We therefore assess the effect of design choices related to the localization criterion on the decision whether a prediction is determined to be a true or false positive. Figures 3, 4 showcase the effect of the reference shape in point-based and overlap-based localization criteria, respectively, while Fig. 5 demonstrates the sensitivity of overlap-based criteria to different localization thresholds. In the following, we provide experimental evidence for the showcased phenomena.

Sensitivity of the AP to the specific choice of overlap-based localization criterion

In this experiment, we investigated the AP scores using Box IoU, Mask IoU and Hull IoU criteria over a range of IoU thresholds [0.05:0.95]. The resulting curves are shown in Fig. 6a). We observe that the Mask IoU and Hull IoU-based AP scores are very similar; conversely, using Box IoU yielded overall higher AP, even at lower thresholds.

Sensitivity of the AP to the IoU range

We investigated the AP scores, using Box IoU as a criterion, over different IoU threshold ranges including the commonly used range of [0.5:0.95]. As shown in Fig. 6b), the AP scores on the commonly-used IoU range substantially differ from those on lower IoU ranges.

IoU versus point in mask

Considering the clinical goal of prioritizing the localization of polyps more than their boundaries, we compared the values of the aggregated metrics Sensitivity, Positive Predictive Value (PPV), F1-Score, F2-Score and AP using point-based localization criteria to the values obtained using Box IoU. Point in Mask returns a true positive (TP) if the center point of the predicted bounding box lies within the respective reference mask. The reference can be the segmentation mask, convex hull or bounding box. The result is shown in Table 3. Point inside Reference criteria yield higher scores across all metrics compared to Box IoU over most IoU thresholds. This especially holds true for detection Sensitivity. Note that the AP score for Center Point in Reference is 0.73 and thus comparable to the AP score for Box IoU criterion with a cutoff range of [0.05:0.5], namely 0.72.
Table 3
Point-based versus overlap-based localization criteria applied to the set of all six centers. Point-based criteria give rise to similar results while the IoU criterion consistently yields lower values. IoU intersection over union
Localization criterion
Box IoU = 0.5
Point inside box
Point inside mask
Point inside hull
Sensitivity
0.68
0.74
0.74
0.74
PPV
0.78
0.86
0.85
0.86
F1-Score
0.73
0.8
0.79
0.8
F2-Score
0.70
0.74
0.74
0.74
AP
0.65
0.73
0.73
0.73

Alignment with clinical interest

In the presence of many sources of variability depending on the metric configuration, we conducted an experiment to determine which configuration aligns most with the clinical goal. We presented colonoscopy images of over 300 patients with their predicted bounding boxes to three gastroenterologists, one with over five years and two with over ten years of experience, who rated the predicted boxes as (clinically) “useful” or “not useful”. Each clinician was responsible for one third of the images and each image was only rated once.
In order to assess the agreement of certain metric configurations with the clinician score, we plotted the number of predictions that met the criterion as a fraction of the predictions rated as “useful”, as well as the number of predictions not meeting the criterion as a fraction of predictions rated as “not useful”. We applied overlap-based and point-based criteria and highlighted the localization granularity that they focus on (rough outline or only position). The result can be seen as a bar plot in Fig. 7. All predictions clinically rated as “not useful” were rejected by all localization criteria. Criteria that focus only on position yielded a higher agreement with the “useful” score than those that localize based on overlap using rough outline. In other words, currently used localization criteria, which put a focus on object outlines through relatively strict IoU localization criteria may not be well-suited for reflecting the clinical interest (Figs. 8, 9 ).

Generalizability of results

As the present work is about performance variability rather than absolute performance we assumed that a high number of images (n = 1512), objects (n = 1386) and centers (n = 6) involved in the analyses is the key to ensure drawing solid conclusions. Specifically, we hypothesized that a state-of-the-art method applied to such a large number of cases would yield a representative distribution of output boxes. To confirm this hypothesis, we have repeated our analyses with two complementary algorithms based on different neural network backbones. (1) EndoCV 2021 winner: We used the open-source implementation (https://github.com/GorkemP/EndoCV2021-EfficientDet-Pytorch) of the EndoCV 2021 polyp detection challenge winner [19], which is based on an ensemble of EfficientDet networks and (2) YOLOv7 ensemble: To reflect the new state of the art in object detection (note that the EndoCV 2022 challenge took place in early 2022) we implemented an ensemble of YOLOv7 [20] models, which are currently considered the state of the art in object detection in terms of speed and accuracy. Both new methods were trained, optimized and tested on the same data splits as the EndoCV 2022 winner. As illustrated in Appendix B, Figs. 10, 11 and Tables 4, 5, in terms of the trend (relative values) the results are in very strong agreement with the results of the EndoCV 2022 winner (only absolute values differ).

Discussion

To our knowledge, we were the first to systematically investigate the variability of polyp detection performance resulting from various validation design choices. The following key insights can be derived from our experiments:
(1)
Performance results are highly sensitive to various design choices Our experiments clearly demonstrate that various validation design choices have a substantial effect on the performance computed for object detection algorithms according to popular metrics. These range from the choice of test set to the specific metric configuration used. While the effect of using different classification metrics may be increasingly well-understood [15], we believe that common metrics, such as AP, are often regarded as black boxes and the effect of the various hyperparameters remains poorly understood. Our findings clearly suggest that hyperparameters—specifically the localization criterion and the corresponding threshold—should not indiscriminately be adopted from other work, but carefully be chosen to match the domain need.
 
(2)
Common metric configurations do not reflect the clinical need According to a usefulness assessment of polyp predictions from over 300 patients by three clinicians from different hospitals, commonly used localization criteria that are popular in the computer vision community do not reflect the clinical domain interest when deciding whether a prediction should be assigned a true positive or false positive. This holds specifically true for the international competitions that have been conducted in the context of Polyp detection. The community should therefore revisit the question of whether a good object detection method must necessarily yield a good outline of a polyp. Restricting the need to just localizing a polyp via its position (reflected by the requirement of \(IoU > 0\), for example) might better approximate the clinical need and at the same time overcome problems resulting from suboptimal IoU thresholds.
 
(3)
Common hyperparameters may be too restrictive Our visual examples (Fig. 5) demonstrate that even fairly well-localized polyps feature an IoU below the commonly used threshold of 0.5, resulting in them being considered a miss even though a clinician might find the prediction useful. The community may therefore want to reconsider commonly used threshold ranges and use a broader range (see Fig. 6b).
 
(4)
Comparison of performance across datasets can be largely misleading Our work finds that detection performance depends crucially on the polyp sizes. Hence, even if the prevalences of polyps across centers are similar, comparison of algorithm results can be largely misleading in case of different polyp size distributions.
 
The closest work to ours was recently presented by Ismail et al. [21] outside the field of deep learning. They provide anecdotal evidence on the non-comparability of confusion matrices between different methods, but do not analyze common multi-threshold metrics such as AP or popular localization criteria that serve as the basis for popular classification metrics. Other related work focused on providing benchmarking data sets [2] or showing limitations of metrics for clinical use cases outside the field of polyp detection [7, 22, 23].
We purposefully investigated only the task of polyp detection (and no other object detection tasks) because performance variability is application-specific and metric hyperparameters should be adjusted to the clinical interest [15]. Future work could be directed to challenging current metric configurations also for other medical detection tasks.
A limitation of our study can be seen in the fact that we reported our findings only on a single data set [12]. However, this data set comprises images from six centers and can therefore be seen as sufficiently representative for the scope of our research question. Our results clearly show that validation results are not comparable unless a method has been tested on the same data with the exact same metric hyperparameters. Future work could investigate the stability of challenge rankings as a function of metric configurations. Finally, there are several other factors related to performance assessment that we did not prioritize in this work. These include the assignment strategy, the prevalence as well the confidence threshold in the case of counting metrics. Future work could hence explore the impact of these factors.
Table 4
IoU versus Point in Mask results of complementary algorithms
Localization criterion
Box IoU = 0.5
Point inside box
Point inside mask
Point inside hull
EfficientDet ensemble (Complementary method #1)
 Sensitivity
0.49
0.61
0.59
0.59
 PPV
0.36
0.46
0.44
0.44
 F1-Score
0.41
0.51
0.49
0.49
 F2-Score
0.45
0.56
0.54
0.54
 AP
0.37
0.49
0.47
0.47
YOLOv7 ensemble (Complementary method #2)
 Sensitivity
0.64
0.76
0.75
0.76
 PPV
0.71
0.78
0.77
0.77
 F1-Score
0.67
0.77
0.76
0.76
 F2-Score
0.65
0.77
0.76
0.76
 AP
0.59
0.74
0.73
0.73
Consistent with the observation shown in Table 3, point-based criteria give rise to similar metric values, while the Box IoU criterion consistently yields lower values. IoU intersection over union, PPV positive predictive value, AP average precision
Table 5
Sensitivity to polyp sizes of complementary algorithms
Metric
AP@0.5
AP@0.5:0.95
Polyp size
Small
Medium
Large
Small
Medium
Large
EfficientDet ensemble (Complementary method #1)
Center 1
0.12
0.2
0.5
0.07
0.08
0.22
Center 2
0.00
0.39
0.65
0.00
0.25
0.38
Center 3
0.02
0.24
0.59
0.01
0.12
0.29
Center 4
0.03
0.0
0.24
0.02
0.0
0.13
Center 5
0.08
0.08
0.17
0.03
0.04
0.08
Center 6
0.00
0.03
0.56
0.00
0.23
0.32
All centers (SD)
0.04 (0.05)
0.2 (0.14)
0.45 (0.2)
00.2 (0.03)
0.12 (0.1)
0.24 (0.12)
YOLOv7 ensemble (Complementary method #2)
Center 1
0.12
0.41
0.59
0.07
0.29
0.45
Center 2
0.00
0.53
0.81
0.00
0.33
0.63
Center 3
0.12
0.57
0.83
0.06
0.36
0.64
Center 4
0.07
0.49
0.58
0.01
0.26
0.44
Center 5
0.22
0.41
0.58
0.13
0.27
0.39
Center 6
0.00
0.46
0.69
0.00
0.3
0.55
All centers (SD)
0.09 (0.08)
0.48 (0.06)
0.68 (0.12)
0.05 (0.05)
0.3 (0.04)
0.52 (0.11)
Consistent with the observation shown in Table 2, the results suggest that there are strong effects of polyp size on the AP values. For both methods and across all centers, smaller polyps have a lower AP@0.5 and AP@0.5:0.95 score than medium sized polyps, while larger polyps generally score higher. AP average precision, SD standard deviation
In conclusion, our study is the first to systematically demonstrate the sensitivity of commonly used performance metrics in deep learning-based colon cancer screening to a range of validation design choices. In showing clear evidence for the disparity between commonly used metric configurations and clinical needs, we hope to raise awareness for the importance of adapting validation in machine learning to clinical relevance in general, and spark the careful reconsideration of common validation strategies in automatic cancer screening applications in particular.

Declarations

Conflict of interest

The authors have no relevant financial or non-financial interests to disclose.

Ethical approval

This work was conducted using public datasets of human subject data made available by [3].
Not applicable.
Not applicable.
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​.

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Appendix

A Appendix

In this work, we distinguish point-based and overlap-based localization criteria (see Fig. 8). Point-based criteria include the (Center) Point in Mask criterion, which defines a TP once the center point of the predicted object or bounding box lies inside of the reference mask (alternatively, reference box or reference hull) and the center distance, which calculates the distance between the reference and predicted center points. This criterion defines a TP if the distance is below a user-defined threshold. Overlap-based criteria compute the overlap between the reference and the prediction, in our case done by the Mask or Box IoU. This criterion returns a TP if the Mask/Box IoU is greater than a user-defined threshold. A specific case is the Mask IoU > 0 criterion, which returns a TP once only a small amount of pixels (e.g., a single pixel) of the prediction overlaps with the reference.
We further utilize the AP as a validation metric of the detection performance. In practice, AP is typically calculated at a specific IoU threshold (e.g., AP@0.5 for a threshold of 0.5) or a range of thresholds. The metric is calculated by scanning over different decision thresholds and calculating the Sensitivity (Recall) and Precision (PPV) at the respective thresholds. The points are connected to form a diagram, as depicted in Fig. 9.

B Complementary experiments

This section presents the results corresponding to our generalizability experiment described in Sect. 3.4
Literature
3.
go back to reference Ali S, Ghatwary N, Jha D, Isik-Polat E, Polat G, Yang C, Li W, Galdran A, Ballester M-ÁG, Thambawita V, Hicks S, Poudel S, Lee S-W, Jin Z, Gan T, Yu C, Yan J, Yeo D, Lee H, Tomar NK, Haithmi M, Ahmed A, Riegler MA, Daul C, Halvorsen P, Rittscher J, Salem OE, Lamarque D, Cannizzaro R, Realdon S, de Lange T, East JE (2022) Assessing generalisability of deep learning-based polyp detection and segmentation methods through a computer vision challenge. https://doi.org/10.48550/arXiv.2202.12031 Ali S, Ghatwary N, Jha D, Isik-Polat E, Polat G, Yang C, Li W, Galdran A, Ballester M-ÁG, Thambawita V, Hicks S, Poudel S, Lee S-W, Jin Z, Gan T, Yu C, Yan J, Yeo D, Lee H, Tomar NK, Haithmi M, Ahmed A, Riegler MA, Daul C, Halvorsen P, Rittscher J, Salem OE, Lamarque D, Cannizzaro R, Realdon S, de Lange T, East JE (2022) Assessing generalisability of deep learning-based polyp detection and segmentation methods through a computer vision challenge. https://​doi.​org/​10.​48550/​arXiv.​2202.​12031
8.
go back to reference Yamlahi A, Godau P, Tran TN, Müller L-R, Adler T, Tizabi MD, Baumgartner M, Jäger P, Maier-Hein L (2022) Heterogeneous model ensemble for polyp detection and tracking in colonoscopy. EndoCV@ISBI Yamlahi A, Godau P, Tran TN, Müller L-R, Adler T, Tizabi MD, Baumgartner M, Jäger P, Maier-Hein L (2022) Heterogeneous model ensemble for polyp detection and tracking in colonoscopy. EndoCV@ISBI
11.
go back to reference Angermann Q, Bernal J, Sánchez-Montes C, Hammami M, Fernández-Esparrach G, Dray X, Romain O, Sánchez FJ, Histace A (2017) Towards real-time polyp detection in colonoscopy videos: adapting still frame-based methodologies for video sequences analysis. Comput Assist Robot Endosc Clin Image-based Proced. https://doi.org/10.1007/978-3-319-67543-5_3CrossRef Angermann Q, Bernal J, Sánchez-Montes C, Hammami M, Fernández-Esparrach G, Dray X, Romain O, Sánchez FJ, Histace A (2017) Towards real-time polyp detection in colonoscopy videos: adapting still frame-based methodologies for video sequences analysis. Comput Assist Robot Endosc Clin Image-based Proced. https://​doi.​org/​10.​1007/​978-3-319-67543-5_​3CrossRef
16.
19.
go back to reference Polat G, Işık Polat E, Kayabay K, Temizel A (2021) Polyp detection in colonoscopy images using deep learning and bootstrap aggregation. EndoCV@ISBI Polat G, Işık Polat E, Kayabay K, Temizel A (2021) Polyp detection in colonoscopy images using deep learning and bootstrap aggregation. EndoCV@ISBI
22.
go back to reference Kofler F, Ezhov I, Isensee F, Balsiger F, Berger C, Koerner M, Paetzold J, Li H, Shit S, McKinley R, et al (2021) Are we using appropriate segmentation metrics? Identifying correlates of human expert perception for cnn training beyond rolling the dice coefficient. https://doi.org/10.48550/arXiv.2103.06205 Kofler F, Ezhov I, Isensee F, Balsiger F, Berger C, Koerner M, Paetzold J, Li H, Shit S, McKinley R, et al (2021) Are we using appropriate segmentation metrics? Identifying correlates of human expert perception for cnn training beyond rolling the dice coefficient. https://​doi.​org/​10.​48550/​arXiv.​2103.​06205
23.
go back to reference Gooding MJ, Smith AJ, Tariq M, Aljabar P, Peressutti D, van der Stoep J, Reymen B, Emans D, Hattu D, van Loon J et al (2018) Comparative evaluation of autocontouring in clinical practice: a practical method using the turing test. Med Phys. https://doi.org/10.1002/mp.13200 Gooding MJ, Smith AJ, Tariq M, Aljabar P, Peressutti D, van der Stoep J, Reymen B, Emans D, Hattu D, van Loon J et al (2018) Comparative evaluation of autocontouring in clinical practice: a practical method using the turing test. Med Phys. https://​doi.​org/​10.​1002/​mp.​13200
Metadata
Title
Sources of performance variability in deep learning-based polyp detection
Authors
T. N. Tran
T. J. Adler
A. Yamlahi
E. Christodoulou
P. Godau
A. Reinke
M. D. Tizabi
P. Sauer
T. Persicke
J. G. Albert
L. Maier-Hein
Publication date
02-06-2023
Publisher
Springer International Publishing
Published in
International Journal of Computer Assisted Radiology and Surgery / Issue 7/2023
Print ISSN: 1861-6410
Electronic ISSN: 1861-6429
DOI
https://doi.org/10.1007/s11548-023-02936-9