A shear rheometer for measuring shear stress and both normal stress differences in polymer melts simultaneously: the MTR 25

Authors: Thomas Schweizer, Jürg Hostettler, Fredy Mettler

Published in: Rheologica Acta | Issue 8/2008

Log in

Activate our intelligent search to find suitable subject content or patents.

search-config
loading …

Abstract

The MTR 25 is a multitask rheometer (for shear and squeeze flow) with 25 kg of normal force and a partitioned plate. Torque and normal force are measured at both, the inner disk and the outer ring of the plate. The first and second normal stress differences can be determined from a single test. The axial stiffness is high (107 N/m) by using rigid springs and strain gauges for the load cell. Monodisperse polystyrene (M w = 206 kg/mol, 180°C) has been sheared in the range from 0.05 to 47 s − 1. The viscosity and first normal stress difference are highly reproducible. The second normal stress difference scatters and mirrors the instability at the rim. A critical comparison is made between the MTR 25 method and the single transducer evaluation method (RMS 800 method, Schweizer, Rheol Acta 41:337–344, 2002): Both yield excellent and coinciding viscosity and first normal stress difference data. The RMS 800 method gives more stable second normal stress difference data, since the normal force from the outer ring, which is influenced by edge fracture, is not used. Data for the RMS 800 method can be acquired on the MTR 25. The high normal force capacity permits larger samples and higher shear rates than on the RMS 800.

Dont have a licence yet? Then find out more about our products and how to get one now:

Springer Professional "Technik"

Online-Abonnement

Mit Springer Professional "Technik" erhalten Sie Zugriff auf:

  • über 67.000 Bücher
  • über 390 Zeitschriften

aus folgenden Fachgebieten:

  • Automobil + Motoren
  • Bauwesen + Immobilien
  • Business IT + Informatik
  • Elektrotechnik + Elektronik
  • Energie + Nachhaltigkeit
  • Maschinenbau + Werkstoffe




 

Jetzt Wissensvorsprung sichern!

Springer Professional "Wirtschaft+Technik"

Online-Abonnement

Mit Springer Professional "Wirtschaft+Technik" erhalten Sie Zugriff auf:

  • über 102.000 Bücher
  • über 537 Zeitschriften

aus folgenden Fachgebieten:

  • Automobil + Motoren
  • Bauwesen + Immobilien
  • Business IT + Informatik
  • Elektrotechnik + Elektronik
  • Energie + Nachhaltigkeit
  • Finance + Banking
  • Management + Führung
  • Marketing + Vertrieb
  • Maschinenbau + Werkstoffe
  • Versicherung + Risiko

Jetzt Wissensvorsprung sichern!

Footnotes
1
Rheometrics Mechanical Spectrometer, manufactured by Rheometric Scientific with a force rebalance transducer for torque and normal force (up to 20 N). Electrical heating fitted (Meissner et al. 1989).
 
2
The design of a low load rheometer is equally challenging. Since normal forces decrease quadratically with rate, the temperature must be well controlled to get a perfectly stable baseline. Such requirements can only be obtained with electrically temperature control, but not with convection ovens. The ARES, e.g., can keep temperature constant within ±0.05°C at a level of 200°C, which is far too noisy to even explore the full range of its force rebalance transducer.
 
3
Maxon EC motor, Type 276575 with tacho generator.
 
4
Manufacturer HBM, Type 1-LC11-3/350, resistance 350 Ω each.
 
5
MGCplus AB22A from Hottinger Baldwin Messtechnik, Darmstadt, Germany.
 
6
Manufacturer HBM, Type 1-XG21-6/350, resistance 350 Ω each.
 
7
ARES LR2, manufactured by TA Instruments, Eschborn, Germany, with a force rebalance transducer and a maximal normal force of 20 N. Temperature control by a convection oven
 
8
Registered Mark, version 7.1
 
9
Manufactured by Anton Paar, Ostfildern, Germany, with directly electrically heated plate and radiation heating for the cone.
 
10
A direct implication for the experiments is that for the signal F o to be strong enough, the sample radius R should be typically R > 1.25·R i. For the RMS 800 method (considering F i only), this restriction does not exist.
 
Literature
go back to reference Adams N, Lodge AS (1964) Rheological properties of concentrated polymer solutions II. A cone-and-plate and parallel plate pressure distribution apparatus for determining normal stress differences in steady shear flow. Philos Trans R Soc Lond, A 256(1068):149–184CrossRef Adams N, Lodge AS (1964) Rheological properties of concentrated polymer solutions II. A cone-and-plate and parallel plate pressure distribution apparatus for determining normal stress differences in steady shear flow. Philos Trans R Soc Lond, A 256(1068):149–184CrossRef
go back to reference Attané P, Pierrard JM, Turrel G (1985) Steady and transient shear flows of polystyrene solutions II: shear-rate dependence of non-dimensional viscometric functions: characteristic relaxation times. J Non-Newton Fluid Mech 18:319–333CrossRef Attané P, Pierrard JM, Turrel G (1985) Steady and transient shear flows of polystyrene solutions II: shear-rate dependence of non-dimensional viscometric functions: characteristic relaxation times. J Non-Newton Fluid Mech 18:319–333CrossRef
go back to reference Dutcher CS, Venerus DC (2007) Compliance effects on the torsional flow of a viscoelastic fluid. J Non-Newton Fluid Mech 150(2–3):154–161 Dutcher CS, Venerus DC (2007) Compliance effects on the torsional flow of a viscoelastic fluid. J Non-Newton Fluid Mech 150(2–3):154–161
go back to reference Fetters LJ, Lohse DJ, Graessley WW (1999) Chain dimensions and entanglement spacings in dense macromolecular systems. J Polymer Sci, Part B, Polym Phys 37:1023–1033CrossRef Fetters LJ, Lohse DJ, Graessley WW (1999) Chain dimensions and entanglement spacings in dense macromolecular systems. J Polymer Sci, Part B, Polym Phys 37:1023–1033CrossRef
go back to reference Hansen MG, Nazem F (1975) Transient normal force transducer response in a modified Weissenberg Rheogoniometer. Trans Soc Rheol 19(1):21–36CrossRef Hansen MG, Nazem F (1975) Transient normal force transducer response in a modified Weissenberg Rheogoniometer. Trans Soc Rheol 19(1):21–36CrossRef
go back to reference Jobling A, Roberts JE (1959) Flow testing of viscoelastic materials. Design and calibration of the Roberts–Weissenberg model R8 Rheogoniometer. J Polym Sci 36:421–431CrossRef Jobling A, Roberts JE (1959) Flow testing of viscoelastic materials. Design and calibration of the Roberts–Weissenberg model R8 Rheogoniometer. J Polym Sci 36:421–431CrossRef
go back to reference Kalogrianitis SG, van Egmond JW (1997) Full tensor optical rheometry of polymer fluids. J Rheol 41(2):343–364CrossRef Kalogrianitis SG, van Egmond JW (1997) Full tensor optical rheometry of polymer fluids. J Rheol 41(2):343–364CrossRef
go back to reference Kaye A, Lodge AS, Vale DG (1968) Determination of normal stress differences in steady shear flow. II. Flow birefringence, viscosity, and normal stress data for a polyisobutylene liquid. Rheol Acta 7(4):368–379CrossRef Kaye A, Lodge AS, Vale DG (1968) Determination of normal stress differences in steady shear flow. II. Flow birefringence, viscosity, and normal stress data for a polyisobutylene liquid. Rheol Acta 7(4):368–379CrossRef
go back to reference Magda JJ, Baek SG (1994) Concentrated entangled and semidilute entangled polystyrene solutions and the second normal stress difference. Polymer 35(6):1187–1194CrossRef Magda JJ, Baek SG (1994) Concentrated entangled and semidilute entangled polystyrene solutions and the second normal stress difference. Polymer 35(6):1187–1194CrossRef
go back to reference Meissner J (1972) Modification of the Weissenberg Rheogoniometer for measurement of transient rheological properties of molten polyethylene under shear. Comparison with tensile data. J Appl Polym Sci 16:2877–2899CrossRef Meissner J (1972) Modification of the Weissenberg Rheogoniometer for measurement of transient rheological properties of molten polyethylene under shear. Comparison with tensile data. J Appl Polym Sci 16:2877–2899CrossRef
go back to reference Meissner J, Garbella RW, Hostettler J (1989) Measuring normal stress differences in polymer melt shear flow. J Rheol 33(6):843–864CrossRef Meissner J, Garbella RW, Hostettler J (1989) Measuring normal stress differences in polymer melt shear flow. J Rheol 33(6):843–864CrossRef
go back to reference Neergaard J, Park K, Venerus DC, Schieber J (2000) Exponential shear flow of linear, entangled polymeric liquids. J Rheol 44(5):1043–1054CrossRef Neergaard J, Park K, Venerus DC, Schieber J (2000) Exponential shear flow of linear, entangled polymeric liquids. J Rheol 44(5):1043–1054CrossRef
go back to reference Ohl N, Gleissle W (1992) The second normal stress difference for pure and highly filled viscoelastic fluids. Rheol Acta 31:294–305CrossRef Ohl N, Gleissle W (1992) The second normal stress difference for pure and highly filled viscoelastic fluids. Rheol Acta 31:294–305CrossRef
go back to reference Pollett WFO (1955) Rheological behaviour of continuously sheared polyethylene. Br J Appl Phys 6:199–206CrossRef Pollett WFO (1955) Rheological behaviour of continuously sheared polyethylene. Br J Appl Phys 6:199–206CrossRef
go back to reference Schweizer T (2002) Measurement of the first and second normal stress differences in a polystyrene melt with a cone and partitioned plate tool. Rheol Acta 41:337–344CrossRef Schweizer T (2002) Measurement of the first and second normal stress differences in a polystyrene melt with a cone and partitioned plate tool. Rheol Acta 41:337–344CrossRef
go back to reference Schweizer T (2004) A quick guide to better viscosity measurements of highly viscous fluids. Appl Rheol 14(4):197–201 Schweizer T (2004) A quick guide to better viscosity measurements of highly viscous fluids. Appl Rheol 14(4):197–201
go back to reference Schweizer T, Bardow A (2006) The role of instrument compliance in normal force measurements of polymer melts. Rheol Acta 45(4):393–402CrossRef Schweizer T, Bardow A (2006) The role of instrument compliance in normal force measurements of polymer melts. Rheol Acta 45(4):393–402CrossRef
go back to reference Schweizer T, Stöckli M (2008) Departure from linear velocity profile during shear of polystyrene melts in cone–plate geometry. J Rheol 52(3):713–727CrossRef Schweizer T, Stöckli M (2008) Departure from linear velocity profile during shear of polystyrene melts in cone–plate geometry. J Rheol 52(3):713–727CrossRef
go back to reference Schweizer T, van Meerveld J, Öttinger HC (2004) Nonlinear shear rheology of polystyrene melt with narrow molecular weight distribution—experiment and theory. J Rheol 48(6):1345–1363CrossRef Schweizer T, van Meerveld J, Öttinger HC (2004) Nonlinear shear rheology of polystyrene melt with narrow molecular weight distribution—experiment and theory. J Rheol 48(6):1345–1363CrossRef
go back to reference Venerus DC (2007) Free surface effects on normal stress measurements in cone and plate flow. Appl Rheol 17:36494-1–36494-6 Venerus DC (2007) Free surface effects on normal stress measurements in cone and plate flow. Appl Rheol 17:36494-1–36494-6
go back to reference Winter HH (1997) Analysis of dynamic mechanical data: inversion into a relaxation time spectrum and consistency check. J Non-Newton Fluid Mech 68:225–239CrossRef Winter HH (1997) Analysis of dynamic mechanical data: inversion into a relaxation time spectrum and consistency check. J Non-Newton Fluid Mech 68:225–239CrossRef
Metadata
Title
A shear rheometer for measuring shear stress and both normal stress differences in polymer melts simultaneously: the MTR 25
Authors
Thomas Schweizer
Jürg Hostettler
Fredy Mettler
Publication date
01-11-2008
Publisher
Springer-Verlag
Published in
Rheologica Acta / Issue 8/2008
Print ISSN: 0035-4511
Electronic ISSN: 1435-1528
DOI
https://doi.org/10.1007/s00397-008-0300-5

Other articles of this Issue 8/2008

Rheologica Acta 8/2008 Go to the issue

Premium Partners