2012 | OriginalPaper | Buchkapitel
Capacitor Distortion in Magnetic Particle Imaging
verfasst von : Bo Zheng, Patrick Goodwill, Wisely Yang, Steven Conolly
Erschienen in: Magnetic Particle Imaging
Verlag: Springer Berlin Heidelberg
Aktivieren Sie unsere intelligente Suche, um passende Fachinhalte oder Patente zu finden.
Wählen Sie Textabschnitte aus um mit Künstlicher Intelligenz passenden Patente zu finden. powered by
Markieren Sie Textabschnitte, um KI-gestützt weitere passende Inhalte zu finden. powered by
The signal-to-noise ratio in magnetic particle imaging can be limited by distortion interference arising in the imaging system. In this work, we investigate the contribution of resonant transmit capacitors to system interference. Feedthrough interference spectra obtained using four capacitors with varying voltage ratings in a custom MPI interference testbed show a 20dB reduction in distortion interference with higherrated capacitors. Finally, we discuss the applicability of the interference testbed to treat other interference mechanisms in magnetic particle imaging.