Abstract

Purpose

To design an augmented reality solution that assists surgeons during the distal locking of intramedullary nailing procedures.

Method

Traditionally, the procedure is performed under X-ray guidance and requires a significant amount of time and radiation exposure. To absolve these complications, we propose video guidance that allows surgeons to achieve both the down-the-beam position of the intramedullary nail and its subsequent locking. For the down-the-beam position, the IM nail pose in X-ray is calculated using a 2D/3D registration scheme and later related to the patient leg pose which is calculated using video-tracked AR markers. For the distal locking, surgeons use an augmented radiolucent drill in which its tip position is detected and tracked in real-time under video guidance.

Validation

To evaluate the feasibility of our solution, we performed a preclinical study on dry bone phantom with the participation of four clinicians.

Results

Participants achieved 100 % success rate in the down-the beam positioning and 93 % success rate in distal locking using only two X-ray images in 100 s.

Conclusions

We confirmed that intra-operative navigation using augmented reality provides an alternative way to perform distal locking in a safe and timely manner.

Sie haben noch keine Lizenz? Dann Informieren Sie sich jetzt über unsere Produkte:

Springer Professional "Wirtschaft+Technik"

Online-Abonnement

Mit Springer Professional "Wirtschaft+Technik" erhalten Sie Zugriff auf:

  • über 102.000 Bücher
  • über 537 Zeitschriften

aus folgenden Fachgebieten:

  • Automobil + Motoren
  • Bauwesen + Immobilien
  • Business IT + Informatik
  • Elektrotechnik + Elektronik
  • Energie + Nachhaltigkeit
  • Finance + Banking
  • Management + Führung
  • Marketing + Vertrieb
  • Maschinenbau + Werkstoffe
  • Versicherung + Risiko

Jetzt Wissensvorsprung sichern!

Springer Professional "Wirtschaft"

Online-Abonnement

Mit Springer Professional "Wirtschaft" erhalten Sie Zugriff auf:

  • über 67.000 Bücher
  • über 340 Zeitschriften

aus folgenden Fachgebieten:

  • Bauwesen + Immobilien
  • Business IT + Informatik
  • Finance + Banking
  • Management + Führung
  • Marketing + Vertrieb
  • Versicherung + Risiko




Jetzt Wissensvorsprung sichern!

Springer Professional "Technik"

Online-Abonnement

Mit Springer Professional "Technik" erhalten Sie Zugriff auf:

  • über 67.000 Bücher
  • über 390 Zeitschriften

aus folgenden Fachgebieten:

  • Automobil + Motoren
  • Bauwesen + Immobilien
  • Business IT + Informatik
  • Elektrotechnik + Elektronik
  • Energie + Nachhaltigkeit
  • Maschinenbau + Werkstoffe




 

Jetzt Wissensvorsprung sichern!

Literatur
1.
Zurück zum Zitat Bong MR, Koval KJ, Egol KA (2006) The history of intramedullary nailing. Bull Hosp Jt Dis New York 64(3/4):94 Bong MR, Koval KJ, Egol KA (2006) The history of intramedullary nailing. Bull Hosp Jt Dis New York 64(3/4):94
3.
Zurück zum Zitat Yaniv Z, Joskowicz L (2005) Precise robot-assisted guide positioning for distal locking of intramedullary nails. Med Imaging IEEE Trans 24(5):624–635CrossRef Yaniv Z, Joskowicz L (2005) Precise robot-assisted guide positioning for distal locking of intramedullary nails. Med Imaging IEEE Trans 24(5):624–635CrossRef
4.
Zurück zum Zitat Leloup T, El Kazzi W, Schuind F, Warzée N (2008) Novel technique for distal locking of intramedullary nail based on two non-constrained fluoroscopic images and navigation. IEEE Trans Med Imaging 27(9):1202–1212CrossRefPubMed Leloup T, El Kazzi W, Schuind F, Warzée N (2008) Novel technique for distal locking of intramedullary nail based on two non-constrained fluoroscopic images and navigation. IEEE Trans Med Imaging 27(9):1202–1212CrossRefPubMed
5.
Zurück zum Zitat Zheng G, Zhang X, Haschtmann D, Gédet P, Langlotz F, Nolte LP (2007) Accurate and reliable pose recovery of distal locking holes in computer-assisted intra-medullary nailing of femoral shaft fractures: a preliminary study. Comput Aid Surg 12(3):138–151CrossRef Zheng G, Zhang X, Haschtmann D, Gédet P, Langlotz F, Nolte LP (2007) Accurate and reliable pose recovery of distal locking holes in computer-assisted intra-medullary nailing of femoral shaft fractures: a preliminary study. Comput Aid Surg 12(3):138–151CrossRef
6.
Zurück zum Zitat Windolf M, Schroeder J, Fliri L, Dicht B, Liebergall M, Richards RG (2012) Reinforcing the role of the conventional C-arm: a novel method for simplified distal interlocking. BMC Musculoskelet Disord 13(1):8PubMedCentralCrossRefPubMed Windolf M, Schroeder J, Fliri L, Dicht B, Liebergall M, Richards RG (2012) Reinforcing the role of the conventional C-arm: a novel method for simplified distal interlocking. BMC Musculoskelet Disord 13(1):8PubMedCentralCrossRefPubMed
7.
Zurück zum Zitat Rohilla R, Singh R, Rohilla S, Magu N, Devgan A, Siwach R (2011) Locked intramedullary femoral nailing without fracture table or image intensifier. Strateg Trauma Limb Reconstr 6:127–135CrossRef Rohilla R, Singh R, Rohilla S, Magu N, Devgan A, Siwach R (2011) Locked intramedullary femoral nailing without fracture table or image intensifier. Strateg Trauma Limb Reconstr 6:127–135CrossRef
8.
Zurück zum Zitat Diotte B, Fallavollita P, Wang L, Weidert S, Thaller P-H, Euler E, Navab N (2012) Radiation-free drill guidance in interlocking of intramedullary nails. LNCS, vol 7510. Springer, pp 18–25 Diotte B, Fallavollita P, Wang L, Weidert S, Thaller P-H, Euler E, Navab N (2012) Radiation-free drill guidance in interlocking of intramedullary nails. LNCS, vol 7510. Springer, pp 18–25
9.
Zurück zum Zitat Londei R, Esposito M, Diotte B, Weidert S, Euler E, Thaller P, Navab N, Fallavollita P (2014) The ‘augmented’circles: a video-guided solution for the down-the-beam positioning of IM nail holes. In: Information processing in computer-assisted interventions. Springer International Publishing, pp 100–107 Londei R, Esposito M, Diotte B, Weidert S, Euler E, Thaller P, Navab N, Fallavollita P (2014) The ‘augmented’circles: a video-guided solution for the down-the-beam positioning of IM nail holes. In: Information processing in computer-assisted interventions. Springer International Publishing, pp 100–107
10.
Zurück zum Zitat Diotte B, Fallavollita P, Wang L, Weidert S, Euler E, Thaller P, Navab N (2015) Multi-modal intra-operative navigation during distal locking of intramedullary nails. IEEE Trans Med Imaging 34(2):487–495 Diotte B, Fallavollita P, Wang L, Weidert S, Euler E, Thaller P, Navab N (2015) Multi-modal intra-operative navigation during distal locking of intramedullary nails. IEEE Trans Med Imaging 34(2):487–495
11.
Zurück zum Zitat Wang L, Landes J, Weidert S, Blum T, Von Der Heide A, Euler E, Navab N (2010) First animal cadaver study for interlocking of intramedullary nails under camera augmented mobile C-arm. In: Information processing in computer-assisted interventions. Springer, Berlin Heidelberg, pp 56–66 Wang L, Landes J, Weidert S, Blum T, Von Der Heide A, Euler E, Navab N (2010) First animal cadaver study for interlocking of intramedullary nails under camera augmented mobile C-arm. In: Information processing in computer-assisted interventions. Springer, Berlin Heidelberg, pp 56–66
12.
Zurück zum Zitat Maqungo S, Horn A, Bernstein B, Keel M, Roche S (2014) Distal interlocking screw placement in the femur: free-hand versus electromagnetic assisted technique (sureshot). J Orthop Trauma 28(1):e281–e283CrossRefPubMed Maqungo S, Horn A, Bernstein B, Keel M, Roche S (2014) Distal interlocking screw placement in the femur: free-hand versus electromagnetic assisted technique (sureshot). J Orthop Trauma 28(1):e281–e283CrossRefPubMed
13.
Zurück zum Zitat Moreschini O, Petrucci V, Cannata R (2014) Insertion of distal locking screws of tibial intramedullary nails: a comparison between the free-hand technique and the SURESHOT\(^{{\rm TM}}\) distal targeting system. Injury 45(2):405–407CrossRefPubMed Moreschini O, Petrucci V, Cannata R (2014) Insertion of distal locking screws of tibial intramedullary nails: a comparison between the free-hand technique and the SURESHOT\(^{{\rm TM}}\) distal targeting system. Injury 45(2):405–407CrossRefPubMed
14.
Zurück zum Zitat Navab N, Heining S-M, Traub J (2010) Camera augmented mobile C-arm (CAMC): calibration, accuracy study, and clinical applications. Med Imaging IEEE Trans 29(7):1412–1423CrossRef Navab N, Heining S-M, Traub J (2010) Camera augmented mobile C-arm (CAMC): calibration, accuracy study, and clinical applications. Med Imaging IEEE Trans 29(7):1412–1423CrossRef
15.
Zurück zum Zitat Wang L, Fallavollita P, Zou R, Chen X, Weidert S, Navab Nassir (2012) Closed-form inverse kinematics for interventional C-arm X-ray imaging with six degrees of freedom: modeling and application. Med Imaging IEEE Trans 31(5):1086–1099CrossRef Wang L, Fallavollita P, Zou R, Chen X, Weidert S, Navab Nassir (2012) Closed-form inverse kinematics for interventional C-arm X-ray imaging with six degrees of freedom: modeling and application. Med Imaging IEEE Trans 31(5):1086–1099CrossRef
16.
Zurück zum Zitat Chen X, Wang L, Fallavollita P, Navab N (2012) Error analysis of the X-ray projection geometry of camera-augmented mobile C-arm. In: SPIE medical imaging, pp 83160P–83160P Chen X, Wang L, Fallavollita P, Navab N (2012) Error analysis of the X-ray projection geometry of camera-augmented mobile C-arm. In: SPIE medical imaging, pp 83160P–83160P
17.
Zurück zum Zitat Chen X, Wang L, Fallavollita P, Navab N (2013) Precise X-ray and video overlay for augmented reality fluoroscopy. Int J Comput Assist Radiol Surg 8(1):29–38CrossRefPubMed Chen X, Wang L, Fallavollita P, Navab N (2013) Precise X-ray and video overlay for augmented reality fluoroscopy. Int J Comput Assist Radiol Surg 8(1):29–38CrossRefPubMed
18.
Zurück zum Zitat Pati S, Erat O, Wang L, Weidert S, Euler E, Navab N, Fallavollita P (2013) Accurate pose estimation using single marker single camera calibration system. In: SPIE medical imaging, pp 867126–867126 Pati S, Erat O, Wang L, Weidert S, Euler E, Navab N, Fallavollita P (2013) Accurate pose estimation using single marker single camera calibration system. In: SPIE medical imaging, pp 867126–867126
19.
Zurück zum Zitat Wang L, Fallavollita P, Brand A, Erat O, Weidert S, Thaller P-H, Euler E, Navab N (2012) Intra-op measurement of the mechanical axis deviation: an evaluation study on 19 human cadaver legs. In: Medical image computing and computer-assisted intervention-MICCAI. Springer, Berlin Heidelberg, pp 609–616 Wang L, Fallavollita P, Brand A, Erat O, Weidert S, Thaller P-H, Euler E, Navab N (2012) Intra-op measurement of the mechanical axis deviation: an evaluation study on 19 human cadaver legs. In: Medical image computing and computer-assisted intervention-MICCAI. Springer, Berlin Heidelberg, pp 609–616
20.
Zurück zum Zitat Erat O, Pauly O, Weidert S, Thaller P, Euler E, Mutschler W, Navab N, Fallavollita P (2013) How a surgeon becomes superman by visualization of intelligently fused multi-modalities. In: SPIE medical imaging, pp 86710L–86710L Erat O, Pauly O, Weidert S, Thaller P, Euler E, Mutschler W, Navab N, Fallavollita P (2013) How a surgeon becomes superman by visualization of intelligently fused multi-modalities. In: SPIE medical imaging, pp 86710L–86710L
21.
Zurück zum Zitat Pauly O, Diotte B, Habert S, Weidert S, Euler E, Fallavollita P, Navab N (2014) Relevance-based visualization to improve surgeon perception. In: Information processing in computer-assisted interventions. Springer International Publishing, pp 178–185 Pauly O, Diotte B, Habert S, Weidert S, Euler E, Fallavollita P, Navab N (2014) Relevance-based visualization to improve surgeon perception. In: Information processing in computer-assisted interventions. Springer International Publishing, pp 178–185
22.
Zurück zum Zitat Pauly O, Katouzian A, Eslami A, Fallavollita P, Navab N (2012) Supervised classification for customized intraoperative augmented reality visualization. In: IEEE international symposium on mixed and augmented reality (ISMAR), pp 311–312 Pauly O, Katouzian A, Eslami A, Fallavollita P, Navab N (2012) Supervised classification for customized intraoperative augmented reality visualization. In: IEEE international symposium on mixed and augmented reality (ISMAR), pp 311–312
Metadaten
Titel
Intra-operative augmented reality in distal locking
verfasst von
Roberto Londei
Marco Esposito
Benoit Diotte
Simon Weidert
Ekkehard Euler
Peter Thaller
Nassir Navab
Pascal Fallavollita
Publikationsdatum
01.09.2015
Verlag
Springer Berlin Heidelberg
Erschienen in
International Journal of Computer Assisted Radiology and Surgery / Ausgabe 9/2015
Print ISSN: 1861-6410
Elektronische ISSN: 1861-6429
DOI
https://doi.org/10.1007/s11548-015-1169-2

Weitere Artikel der Ausgabe 9/2015

International Journal of Computer Assisted Radiology and Surgery 9/2015 Zur Ausgabe