Transmission Based Adaptive Automatic Tube Voltage Selection for Computed Tomography

verfasst von : Gábor Lékó, Péter Balázs

Erschienen in: Combinatorial Image Analysis

Verlag: Springer International Publishing

Aktivieren Sie unsere intelligente Suche, um passende Fachinhalte oder Patente zu finden.

search-config
loading …

Abstract

Computed Tomography (CT) is a widely used x-ray based imaging modality in radiodiagnostics and non-destructive testing. For medical considerations as well as for economical and environmental reasons, it is desirable to reduce the dose of radiation and to optimize energy used for acquiring x-ray projection images. Especially, in case of elongated objects, using a constant energy spectrum radiation may not provide realistic information about the interior of the examined object. In this paper we provide an adaptive tube voltage selection method, which determines the proper amount of radiation energy on-the-fly, during the acquisition, based on projection information. By experiments on software phantom images we show that this adaptive approach can produce CT images of better quality, and in the same time, by consuming less energy.

Sie haben noch keine Lizenz? Dann Informieren Sie sich jetzt über unsere Produkte:

Springer Professional "Wirtschaft+Technik"

Online-Abonnement

Mit Springer Professional "Wirtschaft+Technik" erhalten Sie Zugriff auf:

  • über 102.000 Bücher
  • über 537 Zeitschriften

aus folgenden Fachgebieten:

  • Automobil + Motoren
  • Bauwesen + Immobilien
  • Business IT + Informatik
  • Elektrotechnik + Elektronik
  • Energie + Nachhaltigkeit
  • Finance + Banking
  • Management + Führung
  • Marketing + Vertrieb
  • Maschinenbau + Werkstoffe
  • Versicherung + Risiko

Jetzt Wissensvorsprung sichern!

Springer Professional "Technik"

Online-Abonnement

Mit Springer Professional "Technik" erhalten Sie Zugriff auf:

  • über 67.000 Bücher
  • über 390 Zeitschriften

aus folgenden Fachgebieten:

  • Automobil + Motoren
  • Bauwesen + Immobilien
  • Business IT + Informatik
  • Elektrotechnik + Elektronik
  • Energie + Nachhaltigkeit
  • Maschinenbau + Werkstoffe




 

Jetzt Wissensvorsprung sichern!

Springer Professional "Wirtschaft"

Online-Abonnement

Mit Springer Professional "Wirtschaft" erhalten Sie Zugriff auf:

  • über 67.000 Bücher
  • über 340 Zeitschriften

aus folgenden Fachgebieten:

  • Bauwesen + Immobilien
  • Business IT + Informatik
  • Finance + Banking
  • Management + Führung
  • Marketing + Vertrieb
  • Versicherung + Risiko




Jetzt Wissensvorsprung sichern!

Literatur
1.
Zurück zum Zitat Chen, J.H., Jin, E.H., He, W., Zhao, L.Q.: Combining automatic tube current modulation with adaptive statistical iterative reconstruction for low-dose chest CT screening. PloS one 9, e92414 (2014) Chen, J.H., Jin, E.H., He, W., Zhao, L.Q.: Combining automatic tube current modulation with adaptive statistical iterative reconstruction for low-dose chest CT screening. PloS one 9, e92414 (2014)
2.
Zurück zum Zitat Gervaise, A., Naulet, P., Beuret, F., Henry, C., Pernin, M., Portron, Y., Lapierre-Combes, M.: Low-dose CT with automatic tube current modulation, adaptive statistical iterative reconstruction, and low tube voltage for the diagnosis of renal colic: Impact of body mass index. Am. J. Roentgenol. (AJR) 202, 553–560 (2014) Gervaise, A., Naulet, P., Beuret, F., Henry, C., Pernin, M., Portron, Y., Lapierre-Combes, M.: Low-dose CT with automatic tube current modulation, adaptive statistical iterative reconstruction, and low tube voltage for the diagnosis of renal colic: Impact of body mass index. Am. J. Roentgenol. (AJR) 202, 553–560 (2014)
3.
Zurück zum Zitat Herman, G.T.: Fundamentals of Computerized Tomography: Image Reconstruction from Projections, 2nd edn. Springer Publishing Company, Incorporated, London (2009)CrossRef Herman, G.T.: Fundamentals of Computerized Tomography: Image Reconstruction from Projections, 2nd edn. Springer Publishing Company, Incorporated, London (2009)CrossRef
4.
Zurück zum Zitat Jan, S., et al.: GATE: A simulation toolkit for PET and SPECT. Phys. Med. Biol. 49(19), 4543–4561 (2004) Jan, S., et al.: GATE: A simulation toolkit for PET and SPECT. Phys. Med. Biol. 49(19), 4543–4561 (2004)
5.
Zurück zum Zitat Jan, S., et al.: GATE v6: A major enhancement of the GATE simulation platform enabling modelling of CT and radiotherapy. Phys. Med. Biol. 56(4), 881–901 (2011) Jan, S., et al.: GATE v6: A major enhancement of the GATE simulation platform enabling modelling of CT and radiotherapy. Phys. Med. Biol. 56(4), 881–901 (2011)
6.
Zurück zum Zitat Krämer, A., Kovacheva, E., Lanza, G.: Projection based evaluation of CT image quality in dimensional metrology (2015) Krämer, A., Kovacheva, E., Lanza, G.: Projection based evaluation of CT image quality in dimensional metrology (2015)
7.
Zurück zum Zitat Lee, K.H., et al.: Attenuation-based automatic tube voltage selection and tube current modulation for dose reduction at contrast-enhanced liver CT. Radiology 265(2), 437–447 (2012)CrossRef Lee, K.H., et al.: Attenuation-based automatic tube voltage selection and tube current modulation for dose reduction at contrast-enhanced liver CT. Radiology 265(2), 437–447 (2012)CrossRef
8.
Zurück zum Zitat Lv, P., Liu, J., Zhang, R., Jia, Y., Gao, J.: Combined use of automatic tube voltage selection and current modulation with iterative reconstruction for CT evaluation of small hypervascular hepatocellular carcinomas: Effect on lesion conspicuity and image quality. Korean J. Radiol. 16(3), 531–540 (2015) Lv, P., Liu, J., Zhang, R., Jia, Y., Gao, J.: Combined use of automatic tube voltage selection and current modulation with iterative reconstruction for CT evaluation of small hypervascular hepatocellular carcinomas: Effect on lesion conspicuity and image quality. Korean J. Radiol. 16(3), 531–540 (2015)
9.
Zurück zum Zitat Nagel, H.D.: CT Parameters that Influence the Radiation Dose, pp. 51–79. Springer, Berlin, Heidelberg (2007) Nagel, H.D.: CT Parameters that Influence the Radiation Dose, pp. 51–79. Springer, Berlin, Heidelberg (2007)
10.
Zurück zum Zitat Ngaile, J., Msaki, P.: Estimation of patient organ doses from CT examinations in Tanzania. J. Appl. Clin. Med. physics/Am. Coll. Med. Phys. 7, 80–94 (2006) Ngaile, J., Msaki, P.: Estimation of patient organ doses from CT examinations in Tanzania. J. Appl. Clin. Med. physics/Am. Coll. Med. Phys. 7, 80–94 (2006)
11.
Zurück zum Zitat Poludniowski, G., Landry, G., DeBlois, F., Evans, P.M., Verhaegen, F.: SpekCalc: A program to calculate photon spectra from tungsten anode x-ray tubes. Phys. Med. Biol. 54(19), N433–N438 (2009)CrossRef Poludniowski, G., Landry, G., DeBlois, F., Evans, P.M., Verhaegen, F.: SpekCalc: A program to calculate photon spectra from tungsten anode x-ray tubes. Phys. Med. Biol. 54(19), N433–N438 (2009)CrossRef
12.
Zurück zum Zitat Poludniowski, G.: Calculation of x-ray spectra emerging from an x-ray tube. part II. X-ray production and filtration in x-ray targets. Med. Phys. 34, 2175–2186 (2007) Poludniowski, G.: Calculation of x-ray spectra emerging from an x-ray tube. part II. X-ray production and filtration in x-ray targets. Med. Phys. 34, 2175–2186 (2007)
13.
Zurück zum Zitat Poludniowski, G., Evans, P.M.: Calculation of x-ray spectra emerging from an x-ray tube. Part I. Electron penetration characteristics in x-ray targets. Med. Phys. 34, 2164–2174 (2007) Poludniowski, G., Evans, P.M.: Calculation of x-ray spectra emerging from an x-ray tube. Part I. Electron penetration characteristics in x-ray targets. Med. Phys. 34, 2164–2174 (2007)
14.
Zurück zum Zitat Reiter, M., Harrer, B., Heinzl, C., Salaberger, D., Gusenbauer, C., Kuhn, C., Kastner, J.: Simulation Aided Study for Optimising Industrial X-ray CT Scan Parameters for Non-Destructive Testing and Materials Characterisation, vol. 6, pp. 20–22 (2011) Reiter, M., Harrer, B., Heinzl, C., Salaberger, D., Gusenbauer, C., Kuhn, C., Kastner, J.: Simulation Aided Study for Optimising Industrial X-ray CT Scan Parameters for Non-Destructive Testing and Materials Characterisation, vol. 6, pp. 20–22 (2011)
15.
Zurück zum Zitat Reiter, M., Heinzl, C., Salaberger, D., Weiss, D., Kastner, J.: Study on parameter variation of an industrial computed tomography simulation tool concerning dimensional measurement deviations (2010) Reiter, M., Heinzl, C., Salaberger, D., Weiss, D., Kastner, J.: Study on parameter variation of an industrial computed tomography simulation tool concerning dimensional measurement deviations (2010)
16.
Zurück zum Zitat Rhee, D., Kim, S.w., Moon, Y., Kim, J., Jeong, D.: Effects of the difference in tube voltage of the CT scanner on dose calculation. J. Korean Phys. Soc. 67, 123–128 (2015) Rhee, D., Kim, S.w., Moon, Y., Kim, J., Jeong, D.: Effects of the difference in tube voltage of the CT scanner on dose calculation. J. Korean Phys. Soc. 67, 123–128 (2015)
17.
Zurück zum Zitat Schindera, S., Winklehner, A., Alkadhi, H., Goetti, R., Fischer, M., Gnannt, R., Szucs-Farkas, Z.: Effect of automatic tube voltage selection on image quality and radiation dose in abdominal CT angiography of various body sizes: A phantom study. Clin. Radiol. 68, e79–e86 (2012) Schindera, S., Winklehner, A., Alkadhi, H., Goetti, R., Fischer, M., Gnannt, R., Szucs-Farkas, Z.: Effect of automatic tube voltage selection on image quality and radiation dose in abdominal CT angiography of various body sizes: A phantom study. Clin. Radiol. 68, e79–e86 (2012)
18.
Zurück zum Zitat Seyal, A., Arslanoglu, A., Abboud, S., Sahin, A., Horowitz, J., Yaghmai, V.: CT of the Abdomen with Reduced Tube Voltage in Adults: A Practical Approach. Radiographics 35, 1922–1939 (2015). A review publication of the Radiological Society of North America, Inc Seyal, A., Arslanoglu, A., Abboud, S., Sahin, A., Horowitz, J., Yaghmai, V.: CT of the Abdomen with Reduced Tube Voltage in Adults: A Practical Approach. Radiographics 35, 1922–1939 (2015). A review publication of the Radiological Society of North America, Inc
19.
Zurück zum Zitat Shin, H.J., et al.: Radiation dose reduction via sinogram affirmed iterative reconstruction and automatic tube voltage modulation (CARE kV) in abdominal CT. Korean J. Radiol. Off. J. Korean Radiol. Soc. 14, 886–893 (2013) Shin, H.J., et al.: Radiation dose reduction via sinogram affirmed iterative reconstruction and automatic tube voltage modulation (CARE kV) in abdominal CT. Korean J. Radiol. Off. J. Korean Radiol. Soc. 14, 886–893 (2013)
20.
Zurück zum Zitat Tang, K., Wang, L., Li, R., Lin, J., Zheng, X., Cao, G.: Effect of low tube voltage on image quality, radiation dose, and low-contrast detectability at abdominal multidetector CT: phantom study. J. Biomed. Biotechnol. 2012(130169), 1–6 (2012)CrossRef Tang, K., Wang, L., Li, R., Lin, J., Zheng, X., Cao, G.: Effect of low tube voltage on image quality, radiation dose, and low-contrast detectability at abdominal multidetector CT: phantom study. J. Biomed. Biotechnol. 2012(130169), 1–6 (2012)CrossRef
21.
Zurück zum Zitat Yu, L., Li, H., Fletcher, J., McCollough, C.: Automatic selection of tube potential for radiation dose reduction in CT: A general strategy. Med. Phys. 37, 234–243 (2010) Yu, L., Li, H., Fletcher, J., McCollough, C.: Automatic selection of tube potential for radiation dose reduction in CT: A general strategy. Med. Phys. 37, 234–243 (2010)
Metadaten
Titel
Transmission Based Adaptive Automatic Tube Voltage Selection for Computed Tomography
verfasst von
Gábor Lékó
Péter Balázs
Copyright-Jahr
2020
DOI
https://doi.org/10.1007/978-3-030-51002-2_14