Adaptive removal of gradients-induced artefacts on ECG in MRI: a performance analysis of RLS filtering

verfasst von: Mario Sansone, Luciano Mirarchi, Marcello Bracale

Erschienen in: Medical & Biological Engineering & Computing | Ausgabe 5/2010

Einloggen

Aktivieren Sie unsere intelligente Suche, um passende Fachinhalte oder Patente zu finden.

search-config
loading …

Abstract

One of the main vital signs used in patient monitoring during Magnetic Resonance Imaging (MRI) is Electro-Cardio-Gram (ECG). Unfortunately, magnetic fields gradients induce artefacts which severely affect ECG quality. Adaptive Noise Cancelling (ANC) is one of the preferred techniques for artefact removal. ANC involves the adaptive estimation of the impulse response of the system constituted by the MRI equipment, the patient and the ECG recording device. Least Mean Square (LMS) adaptive filtering has been traditionally employed because of its simplicity: anyway, it requires the choice of a step-size parameter, whose proper value for the specific application must be estimated case by case: an improper choice could yield slow convergence and unsatisfactory behaviour. Recursive Least Square (RLS) algorithm has, potentially, faster convergence while not requiring any parameter. As far as the authors’ knowledge, there is no systematic analysis of performances of RLS in this scenario. In this study we evaluated the performance of RLS for adaptive removal of artefacts induced by magnetic field gradients on ECG in MRI, in terms of efficacy of suppression. Tests have been made on real signals, acquired via an expressly developed system. A comparison with LMS was made on the basis of opportune performance indices. Results indicate that RLS is superior to LMS in several respects.

Sie haben noch keine Lizenz? Dann Informieren Sie sich jetzt über unsere Produkte:

Springer Professional "Wirtschaft+Technik"

Online-Abonnement

Mit Springer Professional "Wirtschaft+Technik" erhalten Sie Zugriff auf:

  • über 102.000 Bücher
  • über 537 Zeitschriften

aus folgenden Fachgebieten:

  • Automobil + Motoren
  • Bauwesen + Immobilien
  • Business IT + Informatik
  • Elektrotechnik + Elektronik
  • Energie + Nachhaltigkeit
  • Finance + Banking
  • Management + Führung
  • Marketing + Vertrieb
  • Maschinenbau + Werkstoffe
  • Versicherung + Risiko

Jetzt Wissensvorsprung sichern!

Springer Professional "Technik"

Online-Abonnement

Mit Springer Professional "Technik" erhalten Sie Zugriff auf:

  • über 67.000 Bücher
  • über 390 Zeitschriften

aus folgenden Fachgebieten:

  • Automobil + Motoren
  • Bauwesen + Immobilien
  • Business IT + Informatik
  • Elektrotechnik + Elektronik
  • Energie + Nachhaltigkeit
  • Maschinenbau + Werkstoffe




 

Jetzt Wissensvorsprung sichern!

Springer Professional "Wirtschaft"

Online-Abonnement

Mit Springer Professional "Wirtschaft" erhalten Sie Zugriff auf:

  • über 67.000 Bücher
  • über 340 Zeitschriften

aus folgenden Fachgebieten:

  • Bauwesen + Immobilien
  • Business IT + Informatik
  • Finance + Banking
  • Management + Führung
  • Marketing + Vertrieb
  • Versicherung + Risiko




Jetzt Wissensvorsprung sichern!

Literatur
1.
Zurück zum Zitat Abächerli R, Pasquier C, Odille F, Kraemer M, Schmid JJ, Felblinger J (2005) Suppression of MR gradient artefacts on electrophysiological signals based on an adaptive real-time filter with LMS coefficient updates. Magn Reson Mater Phys Biol Med 18(1):41–50 Abächerli R, Pasquier C, Odille F, Kraemer M, Schmid JJ, Felblinger J (2005) Suppression of MR gradient artefacts on electrophysiological signals based on an adaptive real-time filter with LMS coefficient updates. Magn Reson Mater Phys Biol Med 18(1):41–50
2.
3.
Zurück zum Zitat Allen PJ, Josephs O, Turner R (2000) A method for removing imaging artifact from continuous EEG recorded during functional MRI. Neuroimage 12:230–239CrossRef Allen PJ, Josephs O, Turner R (2000) A method for removing imaging artifact from continuous EEG recorded during functional MRI. Neuroimage 12:230–239CrossRef
4.
Zurück zum Zitat Bouchard M, Quednau S (2000) Multichannel recursive-least-squares algorithms and fast transversal filter algorithms for active noise control and sound reproduction systems. IEEE Trans Acoust Speech Signal Process 5(8):606–618 Bouchard M, Quednau S (2000) Multichannel recursive-least-squares algorithms and fast transversal filter algorithms for active noise control and sound reproduction systems. IEEE Trans Acoust Speech Signal Process 5(8):606–618
5.
Zurück zum Zitat Chia JM, Fischer SE, Wickline SA, Lorenz CH (2000) Performance of QRS detection for cardiac magnetic resonance imaging with a novel vectorcardiographic triggering method. J Magn Reson Imaging 12(5):678–688CrossRef Chia JM, Fischer SE, Wickline SA, Lorenz CH (2000) Performance of QRS detection for cardiac magnetic resonance imaging with a novel vectorcardiographic triggering method. J Magn Reson Imaging 12(5):678–688CrossRef
6.
Zurück zum Zitat Felblinger J, Slotboom J, Kreis R, Jung B, Boesch C (1999) Restoration of electrophysiological signals distorted by inductive effects of magnetic field gradients during MR sequences. Magn Reson Med 41(4):715–721CrossRef Felblinger J, Slotboom J, Kreis R, Jung B, Boesch C (1999) Restoration of electrophysiological signals distorted by inductive effects of magnetic field gradients during MR sequences. Magn Reson Med 41(4):715–721CrossRef
7.
Zurück zum Zitat Finn JP, Nael K, Deshpande V, Ratib O, Laub G (2006) Cardiac MR imaging: state of the technology. Radiology 241(2):338–354CrossRef Finn JP, Nael K, Deshpande V, Ratib O, Laub G (2006) Cardiac MR imaging: state of the technology. Radiology 241(2):338–354CrossRef
8.
Zurück zum Zitat Gimbel JR, Johnson D, Levine PA, Wilkoff BL (1996) Safe performance of magnetic resonance imaging on five patients with permanent cardiac pacemakers. Pacing Clin Electrophysiol 19(6):913–919CrossRef Gimbel JR, Johnson D, Levine PA, Wilkoff BL (1996) Safe performance of magnetic resonance imaging on five patients with permanent cardiac pacemakers. Pacing Clin Electrophysiol 19(6):913–919CrossRef
9.
Zurück zum Zitat Grouiller F, Vercueil L, Krainik A, Segebarth C, Kahana P, David O (2007) A comparative study of different artefact removal algorithms for EEG signals acquired during functional MRI. Neuroimage 38:124–137CrossRef Grouiller F, Vercueil L, Krainik A, Segebarth C, Kahana P, David O (2007) A comparative study of different artefact removal algorithms for EEG signals acquired during functional MRI. Neuroimage 38:124–137CrossRef
10.
Zurück zum Zitat Hayes MH (1996) Statistical digital signal processing and modelling. Wiley, New York Hayes MH (1996) Statistical digital signal processing and modelling. Wiley, New York
11.
Zurück zum Zitat Haykin S (1996) Adaptive filter theory, 3rd edn. Prentice Hall International Edition Haykin S (1996) Adaptive filter theory, 3rd edn. Prentice Hall International Edition
12.
Zurück zum Zitat Huber S, Muthupillai R, Cheong B, Wible JH Jr, Shah D, Woodard P, Grothues F, Mahrholdt H, Rochitte CE, Masoli O, Kim RJ, Schwaiger CM, Fuisz A, Kramer C, van Rossum AC, Biederman R, Lombardi M, Martin E, Kevorkian R, Flamm SD (2008) Safety of gadoversetamide in patients with acute and chronic myocardial infarction. J Magn Reson Imaging 28(6):1368–1378CrossRef Huber S, Muthupillai R, Cheong B, Wible JH Jr, Shah D, Woodard P, Grothues F, Mahrholdt H, Rochitte CE, Masoli O, Kim RJ, Schwaiger CM, Fuisz A, Kramer C, van Rossum AC, Biederman R, Lombardi M, Martin E, Kevorkian R, Flamm SD (2008) Safety of gadoversetamide in patients with acute and chronic myocardial infarction. J Magn Reson Imaging 28(6):1368–1378CrossRef
13.
Zurück zum Zitat Kanal E, Shellock FG (1992) Patient monitoring during clinical MR imaging. Radiology 185:623–629 Kanal E, Shellock FG (1992) Patient monitoring during clinical MR imaging. Radiology 185:623–629
14.
Zurück zum Zitat Kettenbach J, Kacher DF, Koskinen SK, Silverman SG, Nabavi A, Gering D, Tempany CM, Schwartz RB, Kikinis R, Black PM, Jolesz FA (2000) Interventional and intraoperative magnetic resonance imaging. Annu Rev Biomed Eng 2:661–690CrossRef Kettenbach J, Kacher DF, Koskinen SK, Silverman SG, Nabavi A, Gering D, Tempany CM, Schwartz RB, Kikinis R, Black PM, Jolesz FA (2000) Interventional and intraoperative magnetic resonance imaging. Annu Rev Biomed Eng 2:661–690CrossRef
15.
Zurück zum Zitat Kligfield P, Gettes LS, Bailey JJ, Childers R, Deal BJ, Hancock EW, van Herpen G, Kors JA, Macfarlane P, Mirvis DM, Pahlm O, Rautaharju P, Wagner GS, Josephson M, Mason JW, Okin P, Surawicz B, Wellens H (2007) Recommendations for the standardization and interpretation of the electrocardiogram: part I: the electrocardiogram and its technology: a scientific statement from the American Heart Association Electrocardiography and Arrhythmias Committee, Council on Clinical Cardiology; the American College of Cardiology Foundation; and the Heart Rhythm Society: endorsed by the International Society for Computerized Electrocardiology. Circulation 115(10):1306–1324 Kligfield P, Gettes LS, Bailey JJ, Childers R, Deal BJ, Hancock EW, van Herpen G, Kors JA, Macfarlane P, Mirvis DM, Pahlm O, Rautaharju P, Wagner GS, Josephson M, Mason JW, Okin P, Surawicz B, Wellens H (2007) Recommendations for the standardization and interpretation of the electrocardiogram: part I: the electrocardiogram and its technology: a scientific statement from the American Heart Association Electrocardiography and Arrhythmias Committee, Council on Clinical Cardiology; the American College of Cardiology Foundation; and the Heart Rhythm Society: endorsed by the International Society for Computerized Electrocardiology. Circulation 115(10):1306–1324
16.
Zurück zum Zitat Kugel H, Bremer C, Püschel M, Fischbach R, Lenzen H, Tombach B, Van Aken H, Heindel W (2003) Hazardous situation in the MR bore: induction in ECG leads causes fire. Eur J Radiol 13(4):690–694 Kugel H, Bremer C, Püschel M, Fischbach R, Lenzen H, Tombach B, Van Aken H, Heindel W (2003) Hazardous situation in the MR bore: induction in ECG leads causes fire. Eur J Radiol 13(4):690–694
17.
Zurück zum Zitat Laudon MK, Webster JG, Frayne R, Grist TM (1998) Minimizing interference from magnetic resonance imagers during electrocardiography. IEEE Trans Biomed Eng 45(2):160–164CrossRef Laudon MK, Webster JG, Frayne R, Grist TM (1998) Minimizing interference from magnetic resonance imagers during electrocardiography. IEEE Trans Biomed Eng 45(2):160–164CrossRef
18.
Zurück zum Zitat Macchi O, Eweda E (1983) Second-order convergence analysis of stochastic adaptive linear filtering. IEEE Trans Automat Contr 1:76–85CrossRefMathSciNet Macchi O, Eweda E (1983) Second-order convergence analysis of stochastic adaptive linear filtering. IEEE Trans Automat Contr 1:76–85CrossRefMathSciNet
19.
Zurück zum Zitat Masterton RAJ, Abbott DF, Fleming SW, Jackson GD (2007) Measurement and reduction of motion and ballistocardiogram artefacts from simultaneous EEG and fMRI recordings. Neuroimage 37:202–211CrossRef Masterton RAJ, Abbott DF, Fleming SW, Jackson GD (2007) Measurement and reduction of motion and ballistocardiogram artefacts from simultaneous EEG and fMRI recordings. Neuroimage 37:202–211CrossRef
20.
Zurück zum Zitat Niazy RK, Beckmann CF, Iannetti GF, Brady JM, Smith SM (2005) Removal of fMRI environment artifacts form EEG data using optimal basis sets. Neuroimage 28:720–737CrossRef Niazy RK, Beckmann CF, Iannetti GF, Brady JM, Smith SM (2005) Removal of fMRI environment artifacts form EEG data using optimal basis sets. Neuroimage 28:720–737CrossRef
21.
Zurück zum Zitat Nijm GM, Swiryn S, Larson AC, Sahakian AV (2008) Extraction of the magnetohydrodynamic blood flow potential from the surface electrocardiogram in magnetic resonance imaging. Med Biol Eng Comput 46(7):729–733CrossRef Nijm GM, Swiryn S, Larson AC, Sahakian AV (2008) Extraction of the magnetohydrodynamic blood flow potential from the surface electrocardiogram in magnetic resonance imaging. Med Biol Eng Comput 46(7):729–733CrossRef
22.
Zurück zum Zitat Odille F, Pasquier C, Abächerli R, Vuissoz PA, Zientara GP, Felblinger J (2007) Noise cancellation signal processing method and computer system for improved real-time electrocardiogram artifact correction during MRI data acquisition. IEEE Trans Biomed Eng 54(4):630–640CrossRef Odille F, Pasquier C, Abächerli R, Vuissoz PA, Zientara GP, Felblinger J (2007) Noise cancellation signal processing method and computer system for improved real-time electrocardiogram artifact correction during MRI data acquisition. IEEE Trans Biomed Eng 54(4):630–640CrossRef
23.
Zurück zum Zitat Park HD, Cho SP, Lee KJ, Park YC (2007) Minimisation of gradient artefacts in cardiac-MRI-gating using adaptive interference cancellation filter with synthesised reference. Electron Lett 43(20):1070–1071CrossRef Park HD, Cho SP, Lee KJ, Park YC (2007) Minimisation of gradient artefacts in cardiac-MRI-gating using adaptive interference cancellation filter with synthesised reference. Electron Lett 43(20):1070–1071CrossRef
24.
Zurück zum Zitat Puthusserypady S (2007) Extraction of fetal electrocardiogram using H(infinity) adaptive algorithms. Med Biol Eng Comput 45(10):927–937CrossRef Puthusserypady S (2007) Extraction of fetal electrocardiogram using H(infinity) adaptive algorithms. Med Biol Eng Comput 45(10):927–937CrossRef
25.
26.
Zurück zum Zitat Schaefer DJ, Bourland JD, Nyenhuis JA (2000) Review of patient safety in time-varying gradient fields. J Magn Reson Imaging 12:20–29CrossRef Schaefer DJ, Bourland JD, Nyenhuis JA (2000) Review of patient safety in time-varying gradient fields. J Magn Reson Imaging 12:20–29CrossRef
27.
Zurück zum Zitat Sijbers J, Michiels I, Verhoye M, Van Audekerke J, Van Der Linden A, Van Dyck D (1999) Restoration of MR-induced artefacts in simultaneously recorded MR/EEG data. Magn Reson Imaging 17(9):1383–1391CrossRef Sijbers J, Michiels I, Verhoye M, Van Audekerke J, Van Der Linden A, Van Dyck D (1999) Restoration of MR-induced artefacts in simultaneously recorded MR/EEG data. Magn Reson Imaging 17(9):1383–1391CrossRef
28.
Zurück zum Zitat Tate CN, Goodyear CC (1983) Note on the convergence of linear predictive filters, adapted using the LMS algorithm. Electron Circ Syst IEE Proc G 130(2):61–64 Tate CN, Goodyear CC (1983) Note on the convergence of linear predictive filters, adapted using the LMS algorithm. Electron Circ Syst IEE Proc G 130(2):61–64
29.
Zurück zum Zitat Tenforde TS (2005) Magnetically induced electric fields and currents in the circulatory system. Prog Biophys Mol Biol 87(2–3):279–288CrossRef Tenforde TS (2005) Magnetically induced electric fields and currents in the circulatory system. Prog Biophys Mol Biol 87(2–3):279–288CrossRef
30.
Zurück zum Zitat Vaidyanathan PP (1985) Optimal design of linear phase FIR digital filters with very flat passband and equiripple stopbands. IEEE Trans Circuits Syst CAS-32:904–907CrossRef Vaidyanathan PP (1985) Optimal design of linear phase FIR digital filters with very flat passband and equiripple stopbands. IEEE Trans Circuits Syst CAS-32:904–907CrossRef
31.
Zurück zum Zitat Westbrook C (1999) Handbook of MRI techniques. Paperback Westbrook C (1999) Handbook of MRI techniques. Paperback
32.
Zurück zum Zitat Widrow B, Glover JR, McCool JM (1975) Adaptive noise cancelling: principles and applications. Proc IEEE 63(12):1692–1716CrossRef Widrow B, Glover JR, McCool JM (1975) Adaptive noise cancelling: principles and applications. Proc IEEE 63(12):1692–1716CrossRef
33.
Zurück zum Zitat Yang B (1994) A note on the error propagation analysis of recursive least squares algorithm. IEEE Trans Signal Process 42:3523–3525CrossRef Yang B (1994) A note on the error propagation analysis of recursive least squares algorithm. IEEE Trans Signal Process 42:3523–3525CrossRef
Metadaten
Titel
Adaptive removal of gradients-induced artefacts on ECG in MRI: a performance analysis of RLS filtering
verfasst von
Mario Sansone
Luciano Mirarchi
Marcello Bracale
Publikationsdatum
01.05.2010
Verlag
Springer-Verlag
Erschienen in
Medical & Biological Engineering & Computing / Ausgabe 5/2010
Print ISSN: 0140-0118
Elektronische ISSN: 1741-0444
DOI
https://doi.org/10.1007/s11517-010-0596-z

Weitere Artikel der Ausgabe 5/2010

Medical & Biological Engineering & Computing 5/2010 Zur Ausgabe