Robot-assisted automatic ultrasound calibration

verfasst von: Fereshteh Aalamifar, Alexis Cheng, Younsu Kim, Xiao Hu, Haichong K. Zhang, Xiaoyu Guo, Emad M. Boctor

Erschienen in: International Journal of Computer Assisted Radiology and Surgery | Ausgabe 10/2016

Einloggen

Aktivieren Sie unsere intelligente Suche, um passende Fachinhalte oder Patente zu finden.

search-config
loading …

Abstract

Purpose

Ultrasound (US) calibration is the process of determining the unknown transformation from a coordinate frame such as the robot’s tooltip to the US image frame and is a necessary task for any robotic or tracked US system. US calibration requires submillimeter-range accuracy for most applications, but it is a time-consuming and repetitive task. We provide a new framework for automatic US calibration with robot assistance and without the need for temporal calibration.

Method

US calibration based on active echo (AE) phantom was previously proposed, and its superiority over conventional cross-wire phantom-based calibration was shown. In this work, we use AE to guide the robotic arm motion through the process of data collection; we combine the capability of the AE point to localize itself in the frame of the US image with the automatic motion of the robotic arm to provide a framework for calibrating the arm to the US image automatically.

Results

We demonstrated the efficacy of the automated method compared to the manual method through experiments. To highlight the necessity of frequent ultrasound calibration, it is demonstrated that the calibration precision changed from 1.67 to 3.20 mm if the data collection is not repeated after a dismounting/mounting of the probe holder. In a large data set experiment, similar reconstruction precision of automatic and manual data collection was observed, while the time was reduced by 58 %. In addition, we compared ten automatic calibrations with ten manual ones, each performed in 15 min, and showed that all the automatic ones could converge in the case of setting the initial matrix as identity, while this was not achieved by manual data sets. Given the same initial matrix, the repeatability of the automatic was [0.46, 0.34, 0.80, 0.47] versus [0.42, 0.51, 0.98, 1.15] mm in the manual case for the US image four corners.

Conclusions

The submillimeter accuracy requirement of US calibration makes frequent data collections unavoidable. We proposed an automated calibration setup and showed feasibility by implementing it for a robot tooltip to US image calibration. The automated method showed a similar reconstruction precision as well as repeatability compared to the manual method, while the time consumed for data collection was reduced. The automatic method also reduces the burden of data collection for the user. Thus, the automated method can be a viable solution for applications that require frequent calibrations.

Sie haben noch keine Lizenz? Dann Informieren Sie sich jetzt über unsere Produkte:

Springer Professional "Wirtschaft+Technik"

Online-Abonnement

Mit Springer Professional "Wirtschaft+Technik" erhalten Sie Zugriff auf:

  • über 102.000 Bücher
  • über 537 Zeitschriften

aus folgenden Fachgebieten:

  • Automobil + Motoren
  • Bauwesen + Immobilien
  • Business IT + Informatik
  • Elektrotechnik + Elektronik
  • Energie + Nachhaltigkeit
  • Finance + Banking
  • Management + Führung
  • Marketing + Vertrieb
  • Maschinenbau + Werkstoffe
  • Versicherung + Risiko

Jetzt Wissensvorsprung sichern!

Springer Professional "Wirtschaft"

Online-Abonnement

Mit Springer Professional "Wirtschaft" erhalten Sie Zugriff auf:

  • über 67.000 Bücher
  • über 340 Zeitschriften

aus folgenden Fachgebieten:

  • Bauwesen + Immobilien
  • Business IT + Informatik
  • Finance + Banking
  • Management + Führung
  • Marketing + Vertrieb
  • Versicherung + Risiko




Jetzt Wissensvorsprung sichern!

Springer Professional "Technik"

Online-Abonnement

Mit Springer Professional "Technik" erhalten Sie Zugriff auf:

  • über 67.000 Bücher
  • über 390 Zeitschriften

aus folgenden Fachgebieten:

  • Automobil + Motoren
  • Bauwesen + Immobilien
  • Business IT + Informatik
  • Elektrotechnik + Elektronik
  • Energie + Nachhaltigkeit
  • Maschinenbau + Werkstoffe




 

Jetzt Wissensvorsprung sichern!

Literatur
1.
Zurück zum Zitat Ungi T, Lasso A, Fichtinger G (2015) Tracked US in navigated spine interventions. In: Li S, Yao J (eds) Spinal imaging and image analysis. Springer International Publishing, pp 469–494 Ungi T, Lasso A, Fichtinger G (2015) Tracked US in navigated spine interventions. In: Li S, Yao J (eds) Spinal imaging and image analysis. Springer International Publishing, pp 469–494
2.
Zurück zum Zitat Westwood JD (2014) An US-based navigation system for minimally invasive neck surgery. Med Meets Virtual Real 21 NextMed/MMVR21 196:36 Westwood JD (2014) An US-based navigation system for minimally invasive neck surgery. Med Meets Virtual Real 21 NextMed/MMVR21 196:36
3.
Zurück zum Zitat Monfaredi R, Wilson E, Azizi koutenaei B, Labrecque B, Leroy K, Goldie J, Louis E, Swerdlow D, Cleary K (2014) Robot-assisted US imaging: overview and development of a parallel telerobotic system. Minim Invasive Ther Allied Technol 1–9 Monfaredi R, Wilson E, Azizi koutenaei B, Labrecque B, Leroy K, Goldie J, Louis E, Swerdlow D, Cleary K (2014) Robot-assisted US imaging: overview and development of a parallel telerobotic system. Minim Invasive Ther Allied Technol 1–9
4.
Zurück zum Zitat Sauer F, Khamene A, Bascle B, Schinunang L, Wenzel F, Vogt S (2001) Augmented reality visualization of ultrasound images: system description, calibration, and features. In Augmented reality, 2001. Proceedings IEEE and ACM international symposium on, pp 30–39. IEEE Sauer F, Khamene A, Bascle B, Schinunang L, Wenzel F, Vogt S (2001) Augmented reality visualization of ultrasound images: system description, calibration, and features. In Augmented reality, 2001. Proceedings IEEE and ACM international symposium on, pp 30–39. IEEE
5.
Zurück zum Zitat Mebarki R, Krupa A, Chaumette F (2010) 2-d US probe complete guidance by visual servoing using image moments. Robot IEEE Trans 26(2):296–306CrossRef Mebarki R, Krupa A, Chaumette F (2010) 2-d US probe complete guidance by visual servoing using image moments. Robot IEEE Trans 26(2):296–306CrossRef
6.
Zurück zum Zitat Chen TK, Leung C, Azar RZ, Chan K-K, Zhuang B, Dickie K, Dixon J, Pendziwol L, Pelissier L (2012) Laurent importance of transducer position tracking for automated breast US: initial assessments. In: Ultrasonics symposium (IUS), 2012 IEEE International. IEEE Chen TK, Leung C, Azar RZ, Chan K-K, Zhuang B, Dickie K, Dixon J, Pendziwol L, Pelissier L (2012) Laurent importance of transducer position tracking for automated breast US: initial assessments. In: Ultrasonics symposium (IUS), 2012 IEEE International. IEEE
7.
Zurück zum Zitat Aalamifar F, Jiang D, Zhang HK, Cheng A, Guo X, Khurana R, Iordachita I, Boctor EM (2015) Co-robotic US tomography: dual arm setup and error analysis. In: Bosch JG, Duric N (eds) SPIE medical imaging. International Society for Optics and Photonics, Bellingham, p 94190N Aalamifar F, Jiang D, Zhang HK, Cheng A, Guo X, Khurana R, Iordachita I, Boctor EM (2015) Co-robotic US tomography: dual arm setup and error analysis. In: Bosch JG, Duric N (eds) SPIE medical imaging. International Society for Optics and Photonics, Bellingham, p 94190N
8.
Zurück zum Zitat Mercier L, Langø T, Lindseth F, Collins LD (2005) A review of calibration techniques for freehand 3-D US systems. US Med biol 31(2):143–165CrossRef Mercier L, Langø T, Lindseth F, Collins LD (2005) A review of calibration techniques for freehand 3-D US systems. US Med biol 31(2):143–165CrossRef
9.
Zurück zum Zitat Boctor EM, Iordachita I, Choti MA, Hager G, Fichtinger G (2006) Bootstrapped ultrasound calibration. Stud Health Technol Inform 119:61–66PubMed Boctor EM, Iordachita I, Choti MA, Hager G, Fichtinger G (2006) Bootstrapped ultrasound calibration. Stud Health Technol Inform 119:61–66PubMed
10.
Zurück zum Zitat Chen TK, Thurston AD, Ellis RE, Abolmaesumi P (2009) A real-time freehand ultrasound calibration system with automatic accuracy feedback and control. Ultrasound Med Biol 35(1):79–93CrossRefPubMed Chen TK, Thurston AD, Ellis RE, Abolmaesumi P (2009) A real-time freehand ultrasound calibration system with automatic accuracy feedback and control. Ultrasound Med Biol 35(1):79–93CrossRefPubMed
11.
Zurück zum Zitat Chen TK, Thurston AD, Moghari MH, Ellis RE, Abolmaesumi P (2008) A real-time ultrasound calibration system with automatic accuracy control and incorporation of ultrasound section thickness. In: Miga MI, Cleary KR (eds) Medical imaging. International Society for Optics and Photonics, Bellingham, p 69182A Chen TK, Thurston AD, Moghari MH, Ellis RE, Abolmaesumi P (2008) A real-time ultrasound calibration system with automatic accuracy control and incorporation of ultrasound section thickness. In: Miga MI, Cleary KR (eds) Medical imaging. International Society for Optics and Photonics, Bellingham, p 69182A
12.
Zurück zum Zitat Boctor E, Viswanathan A, Choti M, Taylor RH, Fichtinger G, Hager G (2004) A novel closed form solution for ultrasound calibration. In: Biomedical imaging: nano to macro, 2004. IEEE international symposium on, pp. 527–530. IEEE Boctor E, Viswanathan A, Choti M, Taylor RH, Fichtinger G, Hager G (2004) A novel closed form solution for ultrasound calibration. In: Biomedical imaging: nano to macro, 2004. IEEE international symposium on, pp. 527–530. IEEE
13.
Zurück zum Zitat Viswanathan A, Boctor EM, Taylor RH, Hager G, Fichtinger G (2004) Immediate ultrasound calibration with three poses and minimal image processing. In: Barillot C, Haynor DR, Hellier P (eds) Medical image computing and computer-assisted intervention-MICCAI 2004. Springer, Berlin, pp 446–454 Viswanathan A, Boctor EM, Taylor RH, Hager G, Fichtinger G (2004) Immediate ultrasound calibration with three poses and minimal image processing. In: Barillot C, Haynor DR, Hellier P (eds) Medical image computing and computer-assisted intervention-MICCAI 2004. Springer, Berlin, pp 446–454
14.
Zurück zum Zitat Dandekar S, Li Y, Molloy J, Hossack J (2005) A phantom with reduced complexity for spatial 3-D ultrasound calibration. Ultrasound Med Biol 31(8):1083–1093CrossRefPubMed Dandekar S, Li Y, Molloy J, Hossack J (2005) A phantom with reduced complexity for spatial 3-D ultrasound calibration. Ultrasound Med Biol 31(8):1083–1093CrossRefPubMed
15.
Zurück zum Zitat Cheng A, Ackerman MK, Chirikjian GS, Boctor EM (2014) Design and development of an US calibration phantom and system. In: Yaniv ZR, Holmes DR (eds) SPIE medical imaging. International Society for Optics and Photonics, Bellingham, p 903624A Cheng A, Ackerman MK, Chirikjian GS, Boctor EM (2014) Design and development of an US calibration phantom and system. In: Yaniv ZR, Holmes DR (eds) SPIE medical imaging. International Society for Optics and Photonics, Bellingham, p 903624A
16.
Zurück zum Zitat Hsu P-W, Prager RW, Gee AH, Treece GM (2008) Real-time freehand 3D US calibration. US Med Biol 34(2):239–251CrossRef Hsu P-W, Prager RW, Gee AH, Treece GM (2008) Real-time freehand 3D US calibration. US Med Biol 34(2):239–251CrossRef
17.
Zurück zum Zitat Carbajal G, Lasso A, Gómez Á, Fichtinger G (2013) Improving N-wire phantom-based freehand US calibration. Int J Comput Assist Radiol Surg 8(6):1063–1072CrossRefPubMed Carbajal G, Lasso A, Gómez Á, Fichtinger G (2013) Improving N-wire phantom-based freehand US calibration. Int J Comput Assist Radiol Surg 8(6):1063–1072CrossRefPubMed
18.
Zurück zum Zitat Hsu P-W, Treece GM, Prager RW, Houghton NE, Gee Andrew H (2008) Comparison of freehand 3-D US calibration techniques using a stylus. US Med Biol 34(10):1610–1621 Hsu P-W, Treece GM, Prager RW, Houghton NE, Gee Andrew H (2008) Comparison of freehand 3-D US calibration techniques using a stylus. US Med Biol 34(10):1610–1621
19.
Zurück zum Zitat Melvær EL, Mørken K, Samset E (2012) A motion constrained cross-wire phantom for tracked 2D ultrasound calibration. Int J Comput Assist Radiol Surg 7(4):611–620CrossRefPubMed Melvær EL, Mørken K, Samset E (2012) A motion constrained cross-wire phantom for tracked 2D ultrasound calibration. Int J Comput Assist Radiol Surg 7(4):611–620CrossRefPubMed
20.
Zurück zum Zitat Zhang H, Banovac F, White A, Cleary K (2006) Freehand 3D ultrasound calibration using an electromagnetically tracked needle. In: Cleary KR, Galloway RL Jr (eds) Medical imaging. International Society for Optics and Photonics, Bellingham, p 61412M Zhang H, Banovac F, White A, Cleary K (2006) Freehand 3D ultrasound calibration using an electromagnetically tracked needle. In: Cleary KR, Galloway RL Jr (eds) Medical imaging. International Society for Optics and Photonics, Bellingham, p 61412M
21.
Zurück zum Zitat Prager RW, Rohling RN, Gee AH, Berman L (1998) Rapid calibration for 3-D freehand US. US Med Biol 24(6):855–869CrossRef Prager RW, Rohling RN, Gee AH, Berman L (1998) Rapid calibration for 3-D freehand US. US Med Biol 24(6):855–869CrossRef
22.
Zurück zum Zitat Guo X, Cheng A, Zhang HK, Kang H-J, Etienne-Cummings R,Boctor EM (2014) Active echo: a new paradigm for US calibration. In: Golland P, Hata N, Barillot C, Hornegger J, Howe R (eds) Medical image computing and computer assisted intervention-MICCAI 2014. Springer International Publishing, pp 397–404 Guo X, Cheng A, Zhang HK, Kang H-J, Etienne-Cummings R,Boctor EM (2014) Active echo: a new paradigm for US calibration. In: Golland P, Hata N, Barillot C, Hornegger J, Howe R (eds) Medical image computing and computer assisted intervention-MICCAI 2014. Springer International Publishing, pp 397–404
23.
Zurück zum Zitat Guo X, Kang H-J, Etienne-Cummings R, Boctor EM (2014) Active US Pattern Injection System (AUSPIS) for interventional tool guidance. PLoS One 9(10):e104262CrossRefPubMedPubMedCentral Guo X, Kang H-J, Etienne-Cummings R, Boctor EM (2014) Active US Pattern Injection System (AUSPIS) for interventional tool guidance. PLoS One 9(10):e104262CrossRefPubMedPubMedCentral
24.
Zurück zum Zitat Ackerman MK, Cheng A, Boctor E, Chirikjian G (2014) Online US sensor calibration using gradient descent on the Euclidean group. In: Robotics and automation (ICRA), 2014 IEEE international conference on. IEEE Ackerman MK, Cheng A, Boctor E, Chirikjian G (2014) Online US sensor calibration using gradient descent on the Euclidean group. In: Robotics and automation (ICRA), 2014 IEEE international conference on. IEEE
Metadaten
Titel
Robot-assisted automatic ultrasound calibration
verfasst von
Fereshteh Aalamifar
Alexis Cheng
Younsu Kim
Xiao Hu
Haichong K. Zhang
Xiaoyu Guo
Emad M. Boctor
Publikationsdatum
01.10.2016
Verlag
Springer Berlin Heidelberg
Erschienen in
International Journal of Computer Assisted Radiology and Surgery / Ausgabe 10/2016
Print ISSN: 1861-6410
Elektronische ISSN: 1861-6429
DOI
https://doi.org/10.1007/s11548-015-1341-8

Weitere Artikel der Ausgabe 10/2016

International Journal of Computer Assisted Radiology and Surgery 10/2016 Zur Ausgabe

Premium Partner