A retrospective analysis of oxygen concentrator maintenance needs and costs in a low-resource setting: experience from The Gambia

verfasst von: Beverly D. Bradley, Samantha Chow, Ebrima Nyassi, Yu-Ling Cheng, David Peel, Stephen R. C. Howie

Erschienen in: Health and Technology | Ausgabe 4/2015

Einloggen

Aktivieren Sie unsere intelligente Suche, um passende Fachinhalte oder Patente zu finden.

search-config
loading …

Abstract

Oxygen is an essential medicine for the treatment of pneumonia, the leading cause of death in children under five worldwide. Yet, providing a sufficient and reliable supply of oxygen is a major challenge for many health facilities in the developing world, particularly in paediatric care units. The cost-effectiveness of oxygen concentrators versus compressed gas cylinders as a source of oxygen in low-resource health facilities has been demonstrated, but evidence of their long-term functionality is scarce. The Biomedical Engineering Department at the Medical Research Council Unit in The Gambia manages and maintains 27 oxygen concentrators at several sites across the country, and has kept electronic records of all preventive maintenance checks and repairs on these devices since 2006. Through a retrospective analysis of these maintenance records, the objective of this study was to assess the long-term reliability and maintenance needs of oxygen concentrators in a low-income setting with biomedical engineering technologist support. We found that the majority of concentrator repairs are low-cost and require a low experience level to complete. We estimate that the useful lifespan of oxygen concentrators in low-resource settings could reasonably exceed 7 years provided a system is in place for routine preventive maintenance. We conclude the paper with additional insights on the broader support ecosystem required to manage and maintain oxygen concentrators in low-resource settings.

Sie haben noch keine Lizenz? Dann Informieren Sie sich jetzt über unsere Produkte:

Springer Professional "Wirtschaft+Technik"

Online-Abonnement

Mit Springer Professional "Wirtschaft+Technik" erhalten Sie Zugriff auf:

  • über 102.000 Bücher
  • über 537 Zeitschriften

aus folgenden Fachgebieten:

  • Automobil + Motoren
  • Bauwesen + Immobilien
  • Business IT + Informatik
  • Elektrotechnik + Elektronik
  • Energie + Nachhaltigkeit
  • Finance + Banking
  • Management + Führung
  • Marketing + Vertrieb
  • Maschinenbau + Werkstoffe
  • Versicherung + Risiko

Jetzt Wissensvorsprung sichern!

Springer Professional "Wirtschaft"

Online-Abonnement

Mit Springer Professional "Wirtschaft" erhalten Sie Zugriff auf:

  • über 67.000 Bücher
  • über 340 Zeitschriften

aus folgenden Fachgebieten:

  • Bauwesen + Immobilien
  • Business IT + Informatik
  • Finance + Banking
  • Management + Führung
  • Marketing + Vertrieb
  • Versicherung + Risiko




Jetzt Wissensvorsprung sichern!

Springer Professional "Technik"

Online-Abonnement

Mit Springer Professional "Technik" erhalten Sie Zugriff auf:

  • über 67.000 Bücher
  • über 390 Zeitschriften

aus folgenden Fachgebieten:

  • Automobil + Motoren
  • Bauwesen + Immobilien
  • Business IT + Informatik
  • Elektrotechnik + Elektronik
  • Energie + Nachhaltigkeit
  • Maschinenbau + Werkstoffe




 

Jetzt Wissensvorsprung sichern!

Fußnoten
1
Age is calculated up until the day a concentrator was retired, was identified as missing, or Aug 14, 2013 – whichever is earliest
 
2
This includes all 23 CMs, which were assumed to be same-day repairs (i.e., resolved on the day the PM was completed).
 
3
Note that the average valve cost was US $5.44 but for some repairs, two or more valves were replaced, increasing the average cost per repair.
 
Literatur
1.
Zurück zum Zitat Fischer-Walker CL, Rudan I, Liu L, Nair H, Theodoratou E, Bhutta ZA, et al. Global burden of childhood pneumonia and diarrhoea. Lancet. 2013;381:1405–16.CrossRef Fischer-Walker CL, Rudan I, Liu L, Nair H, Theodoratou E, Bhutta ZA, et al. Global burden of childhood pneumonia and diarrhoea. Lancet. 2013;381:1405–16.CrossRef
2.
Zurück zum Zitat Duke T, Wandi F, Jonathan M, Matai S, Kaupa M, Saavu M, et al. Improved oxygen systems for childhood pneumonia: a multihospital effectiveness study in Papua New Guinea. Lancet. 2008;372:1328–33.CrossRef Duke T, Wandi F, Jonathan M, Matai S, Kaupa M, Saavu M, et al. Improved oxygen systems for childhood pneumonia: a multihospital effectiveness study in Papua New Guinea. Lancet. 2008;372:1328–33.CrossRef
3.
Zurück zum Zitat Hill SE, Njie O, Sanneh M, Jallow M, Peel D, Njie M, et al. Oxygen for treatment of severe pneumonia in The Gambia, West Africa: a situational analysis. Int J Tuber Lung Dis. 2009;13:587–93. Hill SE, Njie O, Sanneh M, Jallow M, Peel D, Njie M, et al. Oxygen for treatment of severe pneumonia in The Gambia, West Africa: a situational analysis. Int J Tuber Lung Dis. 2009;13:587–93.
4.
Zurück zum Zitat Wandi F, Peel D, Duke T. Hypoxaemia among children in rural hospitals in Papua New Guinea: epidemiology and resource availability—a study to support a national oxygen programme. Ann Trop Paediatr. 2006;26:277–84.CrossRef Wandi F, Peel D, Duke T. Hypoxaemia among children in rural hospitals in Papua New Guinea: epidemiology and resource availability—a study to support a national oxygen programme. Ann Trop Paediatr. 2006;26:277–84.CrossRef
5.
Zurück zum Zitat English M, Esamai F, Wasunna A, Were F, Ogutu B, Wamae A, et al. Assessment of inpatient paediatric care in first referral level hospitals in 13 districts in Kenya. Lancet. 2004;363:1948–53.CrossRef English M, Esamai F, Wasunna A, Were F, Ogutu B, Wamae A, et al. Assessment of inpatient paediatric care in first referral level hospitals in 13 districts in Kenya. Lancet. 2004;363:1948–53.CrossRef
6.
Zurück zum Zitat Ginsburg AS, Van Cleve WC, Thompson MIW, English M. Oxygen and pulse oximetry in childhood pneumonia: a survey of healthcare providers in resource-limited settings. J Trop Pediatr. 2012;58:389–93.CrossRef Ginsburg AS, Van Cleve WC, Thompson MIW, English M. Oxygen and pulse oximetry in childhood pneumonia: a survey of healthcare providers in resource-limited settings. J Trop Pediatr. 2012;58:389–93.CrossRef
7.
Zurück zum Zitat Enarson P, La Vincente S, Gie R, Maganga E, Chokani C. Implementation of an oxygen concentrator system in district hospital paediatric wards throughout Malawi. Bull World Health Organ. 2008;86:344–8.CrossRef Enarson P, La Vincente S, Gie R, Maganga E, Chokani C. Implementation of an oxygen concentrator system in district hospital paediatric wards throughout Malawi. Bull World Health Organ. 2008;86:344–8.CrossRef
8.
Zurück zum Zitat Mokuolu OA, Ajayi OA. Use of an oxygen concentrator in a Nigerian neonatal unit: economic implications and reliability. Ann Trop Paediatr. 2002;22:209–12.CrossRef Mokuolu OA, Ajayi OA. Use of an oxygen concentrator in a Nigerian neonatal unit: economic implications and reliability. Ann Trop Paediatr. 2002;22:209–12.CrossRef
9.
Zurück zum Zitat Duke T, Peel D, Graham S, Howie S, Enarson PM, Jacobson R. Oxygen concentrators: a practical guide for clinicians and technicians in developing countries. Ann Trop Paediatr. 2010;30:87–101.CrossRef Duke T, Peel D, Graham S, Howie S, Enarson PM, Jacobson R. Oxygen concentrators: a practical guide for clinicians and technicians in developing countries. Ann Trop Paediatr. 2010;30:87–101.CrossRef
10.
Zurück zum Zitat Peel D, Howie SR. Oxygen concentrators for use in tropical countries: a survey. J Clin Eng. 2009;34:205–9.CrossRef Peel D, Howie SR. Oxygen concentrators for use in tropical countries: a survey. J Clin Eng. 2009;34:205–9.CrossRef
11.
Zurück zum Zitat Peel D, Neighbour R, Eltringham RJ. Evaluation of oxygen concentrators for use in countries with limited resources. Anaesthesia. 2013;68:706–12.CrossRef Peel D, Neighbour R, Eltringham RJ. Evaluation of oxygen concentrators for use in countries with limited resources. Anaesthesia. 2013;68:706–12.CrossRef
12.
Zurück zum Zitat Perrelet A, Zellweger JP, Talla I, Ndiaye Y, Gautier E, Gehri M. The oxygen concentrator: an appropriate technology for treating hypoxaemic children in developing countries. Int J Tuber Lung Dis. 2004;8:1138–41. Perrelet A, Zellweger JP, Talla I, Ndiaye Y, Gautier E, Gehri M. The oxygen concentrator: an appropriate technology for treating hypoxaemic children in developing countries. Int J Tuber Lung Dis. 2004;8:1138–41.
13.
Zurück zum Zitat Dobson MB. Oxygen concentrators offer cost savings for developing countries. Anaesthesia. 1991;46:217–9.CrossRef Dobson MB. Oxygen concentrators offer cost savings for developing countries. Anaesthesia. 1991;46:217–9.CrossRef
14.
Zurück zum Zitat Litch JA, Bishop RA. Oxygen concentrators for the delivery of supplemental oxygen in remote high-altitude areas. Wilderness Environ Med. 2000;11:189–91.CrossRef Litch JA, Bishop RA. Oxygen concentrators for the delivery of supplemental oxygen in remote high-altitude areas. Wilderness Environ Med. 2000;11:189–91.CrossRef
15.
Zurück zum Zitat Howie SR, Hill S, Ebonyi A, Krishnan G, Njie O, Sanneh M, et al. Meeting oxygen needs in Africa: an options analysis from the Gambia. Bull World Health Organ. 2009;87:763–71.CrossRef Howie SR, Hill S, Ebonyi A, Krishnan G, Njie O, Sanneh M, et al. Meeting oxygen needs in Africa: an options analysis from the Gambia. Bull World Health Organ. 2009;87:763–71.CrossRef
16.
Zurück zum Zitat Dobson M, Peel D, Khallaf N. Field trial of oxygen concentrators in upper Egypt. Lancet. 1996;347:1597–9.CrossRef Dobson M, Peel D, Khallaf N. Field trial of oxygen concentrators in upper Egypt. Lancet. 1996;347:1597–9.CrossRef
17.
Zurück zum Zitat Enarson PM, Gie R, Enarson DA, Mwansambo C. Development and implementation of a national programme for the management of severe and very severe pneumonia in children in Malawi. PLoS Med. 2009;6:e1000137.CrossRef Enarson PM, Gie R, Enarson DA, Mwansambo C. Development and implementation of a national programme for the management of severe and very severe pneumonia in children in Malawi. PLoS Med. 2009;6:e1000137.CrossRef
18.
Zurück zum Zitat Matai S, Peel D, Wandi F, Jonathan M, Subhi R, Duke T. Implementing an oxygen programme in hospitals in Papua New Guinea. Ann Trop Paediatr. 2008;28:71–8.CrossRef Matai S, Peel D, Wandi F, Jonathan M, Subhi R, Duke T. Implementing an oxygen programme in hospitals in Papua New Guinea. Ann Trop Paediatr. 2008;28:71–8.CrossRef
19.
Zurück zum Zitat La Vincente SF, Peel D, Carai S, Weber MW, Enarson P, Maganga E, et al. The functioning of oxygen concentrators in resource-limited settings: a situation assessment in two countries. Int J Tuber Lung Dis. 2011;15:693–9.CrossRef La Vincente SF, Peel D, Carai S, Weber MW, Enarson P, Maganga E, et al. The functioning of oxygen concentrators in resource-limited settings: a situation assessment in two countries. Int J Tuber Lung Dis. 2011;15:693–9.CrossRef
20.
Zurück zum Zitat Taylor K, Mullally S, Osmond D, Nyassi E, Gassama L, Manneh E, et al. Extreme biomedical engineering (Western Africa). J Clin Eng. 2009;34:89–93. Taylor K, Mullally S, Osmond D, Nyassi E, Gassama L, Manneh E, et al. Extreme biomedical engineering (Western Africa). J Clin Eng. 2009;34:89–93.
21.
Zurück zum Zitat Gomez AW, Nyassi E, Mullally S. Towards a biomedical engineering quality system: Writing documentation to support clinical & laboratory accreditation at MRC unit in the Gambia. 7th IET Appropriate Healthcare Technology Conference. London; 2012. Gomez AW, Nyassi E, Mullally S. Towards a biomedical engineering quality system: Writing documentation to support clinical & laboratory accreditation at MRC unit in the Gambia. 7th IET Appropriate Healthcare Technology Conference. London; 2012.
22.
Zurück zum Zitat UK Medical and Healthcare products Regulatory Agency (MHRA), editor. Managing Medical Devices: Guidance for healthcare and social services organisations. 2006 ed. 2006. Report No.: DB2006(05). UK Medical and Healthcare products Regulatory Agency (MHRA), editor. Managing Medical Devices: Guidance for healthcare and social services organisations. 2006 ed. 2006. Report No.: DB2006(05).
23.
Zurück zum Zitat Good Clinical Laboratory Practice (GCLP). Geneva: World Health Organization; 2009. Good Clinical Laboratory Practice (GCLP). Geneva: World Health Organization; 2009.
24.
Zurück zum Zitat Nyassi E, Mullally S. A unique model of biomedical engineering—an HTM success story in Africa. Bangkok: WHO First Global Forum on Medical Devices; 2010. Nyassi E, Mullally S. A unique model of biomedical engineering—an HTM success story in Africa. Bangkok: WHO First Global Forum on Medical Devices; 2010.
25.
Zurück zum Zitat Nyassi E. Health Technology Management in Developing Countries-A model for attaining high level of quality services. World Congress on Medical Physics and Biomedical Engineering, IFMBE Proceedings. Beijing, China; 2013. pp. 1664–7. Nyassi E. Health Technology Management in Developing Countries-A model for attaining high level of quality services. World Congress on Medical Physics and Biomedical Engineering, IFMBE Proceedings. Beijing, China; 2013. pp. 1664–7.
26.
Zurück zum Zitat Bradley BD, Cheng YL, Nyassi E, Peel D, Howie SRC. An evidence based approach to developing a training programme for the maintenance of oxygen concentrators in low-resource settings. 8th IET Appropriate Healthcare Technology Conference. London; 2014. Bradley BD, Cheng YL, Nyassi E, Peel D, Howie SRC. An evidence based approach to developing a training programme for the maintenance of oxygen concentrators in low-resource settings. 8th IET Appropriate Healthcare Technology Conference. London; 2014.
27.
Zurück zum Zitat Dobson MB, editor. Oxygen concentrators and cylinders. Int J Tuberc Lung Dis 2001;5:520–3. Dobson MB, editor. Oxygen concentrators and cylinders. Int J Tuberc Lung Dis 2001;5:520–3.
28.
Zurück zum Zitat WHO Test Schedule for Oxygen Concentrators. Programme for the control of acute respiratory infections. Geneva: World Health Organization; 1991. WHO Test Schedule for Oxygen Concentrators. Programme for the control of acute respiratory infections. Geneva: World Health Organization; 1991.
29.
Zurück zum Zitat Malkin R, Keane A. Evidence-based approach to the maintenance of laboratory and medical equipment in resource-poor settings. Med Biol Eng Comput. 2010;48:721–6.CrossRef Malkin R, Keane A. Evidence-based approach to the maintenance of laboratory and medical equipment in resource-poor settings. Med Biol Eng Comput. 2010;48:721–6.CrossRef
30.
Zurück zum Zitat Malkin RA. Design of health care technologies for the developing world. Annu Rev Biomed Eng. 2007;9:567–87.CrossRef Malkin RA. Design of health care technologies for the developing world. Annu Rev Biomed Eng. 2007;9:567–87.CrossRef
Metadaten
Titel
A retrospective analysis of oxygen concentrator maintenance needs and costs in a low-resource setting: experience from The Gambia
verfasst von
Beverly D. Bradley
Samantha Chow
Ebrima Nyassi
Yu-Ling Cheng
David Peel
Stephen R. C. Howie
Publikationsdatum
01.04.2015
Verlag
Springer Berlin Heidelberg
Erschienen in
Health and Technology / Ausgabe 4/2015
Print ISSN: 2190-7188
Elektronische ISSN: 2190-7196
DOI
https://doi.org/10.1007/s12553-015-0094-2

Weitere Artikel der Ausgabe 4/2015

Health and Technology 4/2015 Zur Ausgabe

Premium Partner