Study of a Nitinol Stent Deployed into Anatomically Accurate Artery Geometry and Subjected to Realistic Service Loading

verfasst von: Nuno Rebelo, Rachel Fu, Michael Lawrenchuk

Erschienen in: Journal of Materials Engineering and Performance | Ausgabe 5-6/2009

Einloggen

Aktivieren Sie unsere intelligente Suche, um passende Fachinhalte oder Patente zu finden.

search-config
loading …

Abstract

Realistic simulation of a stent in service loading is necessary for the design of Nitinol stents. This type of analysis is complicated by several issues: artery geometry and material properties are largely unknown and may vary from person to person; the loading on the stent is due to interactions between pulsatile blood flow, the artery, and stent. This study addresses the first issue by obtaining realistic artery geometry from magnetic resonance imaging (MRI) scans of a patient in the fetal and supine positions, and comparing deformation of the stent deployed into the two different artery configurations. The second issue is addressed by investigating two different methods of applying service loading. The first method applies sinusoidal pressure waves directly to the artery wall. The second method considers the effect of blood flow within the artery by performing a fluid-structure interaction (FSI) simulation. Simulation of expansion, annealing, crimping, and deployment of a Nitinol stent into the two artery configurations was performed using Abaqus. Following the deployment, sinusoidal pressure pulses, emulating systolic and diastolic blood pressures, were applied to the inner surface of the artery. To simulate the effect of blood within the artery, the coupled Eulerian-Lagrangian method in Abaqus was used. The blood was modeled as a partially filled Eulerian domain, whose material boundary contacted the Lagrangian artery. Transient inlet velocity and outlet pressure conditions were applied to the Eulerian domain to simulate blood flow. Since fatigue life calculations are important for the design of Nitinol stents, the maximum principal strains at the diastolic pressure troughs and the difference between the maximum principal strains at diastolic and systolic pressure were investigated for five cycles. Different maximum principal strains resulted for the different artery geometries (fetal or supine), while the difference between maximum principal strains was similar. For the simulations including the effect of blood flow, the maximum principal strains at the diastolic pressure troughs were somewhat higher than the simulations in which pressure was directly applied to the artery. In conclusion, this study established and compared various methods of improving the service loading in the simulation of Nitinol stents, specifically including anatomically correct artery geometry and the effect of blood flow.

Sie haben noch keine Lizenz? Dann Informieren Sie sich jetzt über unsere Produkte:

Springer Professional "Wirtschaft+Technik"

Online-Abonnement

Mit Springer Professional "Wirtschaft+Technik" erhalten Sie Zugriff auf:

  • über 102.000 Bücher
  • über 537 Zeitschriften

aus folgenden Fachgebieten:

  • Automobil + Motoren
  • Bauwesen + Immobilien
  • Business IT + Informatik
  • Elektrotechnik + Elektronik
  • Energie + Nachhaltigkeit
  • Finance + Banking
  • Management + Führung
  • Marketing + Vertrieb
  • Maschinenbau + Werkstoffe
  • Versicherung + Risiko

Jetzt Wissensvorsprung sichern!

Springer Professional "Technik"

Online-Abonnement

Mit Springer Professional "Technik" erhalten Sie Zugriff auf:

  • über 67.000 Bücher
  • über 390 Zeitschriften

aus folgenden Fachgebieten:

  • Automobil + Motoren
  • Bauwesen + Immobilien
  • Business IT + Informatik
  • Elektrotechnik + Elektronik
  • Energie + Nachhaltigkeit
  • Maschinenbau + Werkstoffe




 

Jetzt Wissensvorsprung sichern!

Literatur
1.
Zurück zum Zitat N. Rebelo, M. Perry, Finite element analysis for the design of Nitinol medical devices, Min. Invas. Ther. & Allied Technol., 9, 2000, 75–80CrossRef N. Rebelo, M. Perry, Finite element analysis for the design of Nitinol medical devices, Min. Invas. Ther. & Allied Technol., 9, 2000, 75–80CrossRef
2.
Zurück zum Zitat N. Rebelo, S. Prabhu, C. Feezor, and A. Denison, Deployment of a Self-Expanding Stent in an Artery, Proceedings of the International Conference on Shape Memory and Superelastic Technologies (Kurhaus Baden-Baden, Baden-Baden, Germany), 2004 N. Rebelo, S. Prabhu, C. Feezor, and A. Denison, Deployment of a Self-Expanding Stent in an Artery, Proceedings of the International Conference on Shape Memory and Superelastic Technologies (Kurhaus Baden-Baden, Baden-Baden, Germany), 2004
3.
Zurück zum Zitat RESIStent Program: SRI International, Stanford University and a Consortium of Stent Manufacturers, 2004-2006, private communication RESIStent Program: SRI International, Stanford University and a Consortium of Stent Manufacturers, 2004-2006, private communication
4.
Zurück zum Zitat F. Auricchio, R. Taylor, Shape-memory alloys: modeling and numerical simulations of the finite-strain superelastic behavior, Comput. Methods. Appl. Mech. Engng., 143, 1996, 175–94CrossRef F. Auricchio, R. Taylor, Shape-memory alloys: modeling and numerical simulations of the finite-strain superelastic behavior, Comput. Methods. Appl. Mech. Engng., 143, 1996, 175–94CrossRef
5.
Zurück zum Zitat F. Auricchio, R. Taylor, Shape-memory alloys: macromodeling and numerical simulations of the superelastic behavior, Comput. Methods. Appl. Mech. Engng., 146, 1997, 281–312CrossRef F. Auricchio, R. Taylor, Shape-memory alloys: macromodeling and numerical simulations of the superelastic behavior, Comput. Methods. Appl. Mech. Engng., 146, 1997, 281–312CrossRef
6.
Zurück zum Zitat E.M. Tittelbaugh, R. Fu, and S. Sett, Coupling FEA to CFD to Investigate the Effects of Pulsatile Blood Flow on the Dilatation of Artery Walls, Proceedings of the NAFEMS World Congress (Vancouver, BC, Canada), 2007 E.M. Tittelbaugh, R. Fu, and S. Sett, Coupling FEA to CFD to Investigate the Effects of Pulsatile Blood Flow on the Dilatation of Artery Walls, Proceedings of the NAFEMS World Congress (Vancouver, BC, Canada), 2007
7.
Zurück zum Zitat Dassault Systemes SIMULIA Corp., Abaqus User’s Manual 6.8, Providence, Rhode Island, 2008 Dassault Systemes SIMULIA Corp., Abaqus User’s Manual 6.8, Providence, Rhode Island, 2008
8.
Zurück zum Zitat D. Steinman, B. Vinh, C. Ethier, M. Ojha, R. Cobbold, and K. Johnston, A numerical simulation of flow in a two-dimensional end-to-side anastomosis model, J. Biomed. Eng.115, 1993, 112–8 D. Steinman, B. Vinh, C. Ethier, M. Ojha, R. Cobbold, and K. Johnston, A numerical simulation of flow in a two-dimensional end-to-side anastomosis model, J. Biomed. Eng.115, 1993, 112–8
Metadaten
Titel
Study of a Nitinol Stent Deployed into Anatomically Accurate Artery Geometry and Subjected to Realistic Service Loading
verfasst von
Nuno Rebelo
Rachel Fu
Michael Lawrenchuk
Publikationsdatum
01.08.2009
Verlag
Springer US
Erschienen in
Journal of Materials Engineering and Performance / Ausgabe 5-6/2009
Print ISSN: 1059-9495
Elektronische ISSN: 1544-1024
DOI
https://doi.org/10.1007/s11665-009-9375-0

Weitere Artikel der Ausgabe 5-6/2009

Journal of Materials Engineering and Performance 5-6/2009 Zur Ausgabe

EditorialNotes

Editorial

    Marktübersichten

    Die im Laufe eines Jahres in der „adhäsion“ veröffentlichten Marktübersichten helfen Anwendern verschiedenster Branchen, sich einen gezielten Überblick über Lieferantenangebote zu verschaffen.