In vitro quantification of the performance of model-based mono-planar and bi-planar fluoroscopy for 3D joint kinematics estimation

verfasst von: Luca Tersi, Arnaud Barré, Silvia Fantozzi, Rita Stagni

Erschienen in: Medical & Biological Engineering & Computing | Ausgabe 3/2013

Einloggen

Aktivieren Sie unsere intelligente Suche, um passende Fachinhalte oder Patente zu finden.

search-config
loading …

Abstract

Model-based mono-planar and bi-planar 3D fluoroscopy methods can quantify intact joints kinematics with performance/cost trade-off. The aim of this study was to compare the performances of mono- and bi-planar setups to a marker-based gold-standard, during dynamic phantom knee acquisitions. Absolute pose errors for in-plane parameters were lower than 0.6 mm or 0.6° for both mono- and bi-planar setups. Mono-planar setups resulted critical in quantifying the out-of-plane translation (error < 6.5 mm), and bi-planar in quantifying the rotation along bone longitudinal axis (error < 1.3°). These errors propagated to joint angles and translations differently depending on the alignment of the anatomical axes and the fluoroscopic reference frames. Internal-external rotation was the least accurate angle both with mono- (error < 4.4°) and bi-planar (error < 1.7°) setups, due to bone longitudinal symmetries. Results highlighted that accuracy for mono-planar in-plane pose parameters is comparable to bi-planar, but with halved computational costs, halved segmentation time and halved ionizing radiation dose. Bi-planar analysis better compensated for the out-of-plane uncertainty that is differently propagated to relative kinematics depending on the setup. To take its full benefits, the motion task to be investigated should be designed to maintain the joint inside the visible volume introducing constraints with respect to mono-planar analysis.

Sie haben noch keine Lizenz? Dann Informieren Sie sich jetzt über unsere Produkte:

Springer Professional "Wirtschaft+Technik"

Online-Abonnement

Mit Springer Professional "Wirtschaft+Technik" erhalten Sie Zugriff auf:

  • über 102.000 Bücher
  • über 537 Zeitschriften

aus folgenden Fachgebieten:

  • Automobil + Motoren
  • Bauwesen + Immobilien
  • Business IT + Informatik
  • Elektrotechnik + Elektronik
  • Energie + Nachhaltigkeit
  • Finance + Banking
  • Management + Führung
  • Marketing + Vertrieb
  • Maschinenbau + Werkstoffe
  • Versicherung + Risiko

Jetzt Wissensvorsprung sichern!

Springer Professional "Technik"

Online-Abonnement

Mit Springer Professional "Technik" erhalten Sie Zugriff auf:

  • über 67.000 Bücher
  • über 390 Zeitschriften

aus folgenden Fachgebieten:

  • Automobil + Motoren
  • Bauwesen + Immobilien
  • Business IT + Informatik
  • Elektrotechnik + Elektronik
  • Energie + Nachhaltigkeit
  • Maschinenbau + Werkstoffe




 

Jetzt Wissensvorsprung sichern!

Springer Professional "Wirtschaft"

Online-Abonnement

Mit Springer Professional "Wirtschaft" erhalten Sie Zugriff auf:

  • über 67.000 Bücher
  • über 340 Zeitschriften

aus folgenden Fachgebieten:

  • Bauwesen + Immobilien
  • Business IT + Informatik
  • Finance + Banking
  • Management + Führung
  • Marketing + Vertrieb
  • Versicherung + Risiko




Jetzt Wissensvorsprung sichern!

Literatur
1.
Zurück zum Zitat Acker S, Li R, Murray H, John PS, Banks S, Mu S, Wyss U, Deluzio K (2011) Accuracy of single-plane fluoroscopy in determining relative position and orientation of total knee replacement components. J Biomech 44(4):784–787PubMedCrossRef Acker S, Li R, Murray H, John PS, Banks S, Mu S, Wyss U, Deluzio K (2011) Accuracy of single-plane fluoroscopy in determining relative position and orientation of total knee replacement components. J Biomech 44(4):784–787PubMedCrossRef
2.
Zurück zum Zitat Akbarshahi M, Schache AG, Fernandez JW, Baker R, Banks S, Pandy MG (2010) Non-invasive assessment of soft-tissue artifact and its effect on knee joint kinematics during functional activity. J Biomech 43(7):1292–1301PubMedCrossRef Akbarshahi M, Schache AG, Fernandez JW, Baker R, Banks S, Pandy MG (2010) Non-invasive assessment of soft-tissue artifact and its effect on knee joint kinematics during functional activity. J Biomech 43(7):1292–1301PubMedCrossRef
3.
Zurück zum Zitat Amiri S, Wilson DR, Masri BA, Sharma G, Anglin C (2011) A novel multi-planar radiography method for three dimensional pose reconstruction of the patellofemoral and tibiofemoral joints after arthroplasty. J Biomech 44(9):1757–1764PubMedCrossRef Amiri S, Wilson DR, Masri BA, Sharma G, Anglin C (2011) A novel multi-planar radiography method for three dimensional pose reconstruction of the patellofemoral and tibiofemoral joints after arthroplasty. J Biomech 44(9):1757–1764PubMedCrossRef
4.
Zurück zum Zitat Baka N, de Bruijne M, van Walsum T, Kaptein B, Giphart J, Schaap M, Niessen W, Lelieveldt B (2012) Statistical shape model based femur kinematics from biplane fluoroscopy. IEEE Trans Med Imaging 31(8):1573–1583 Baka N, de Bruijne M, van Walsum T, Kaptein B, Giphart J, Schaap M, Niessen W, Lelieveldt B (2012) Statistical shape model based femur kinematics from biplane fluoroscopy. IEEE Trans Med Imaging 31(8):1573–1583
5.
Zurück zum Zitat Banks SA, Hodge WA (1996) Accurate measurement of three-dimensional knee replacement kinematics using single-plane fluoroscopy. IEEE Trans Bio-Med Eng 43(6):638–649CrossRef Banks SA, Hodge WA (1996) Accurate measurement of three-dimensional knee replacement kinematics using single-plane fluoroscopy. IEEE Trans Bio-Med Eng 43(6):638–649CrossRef
6.
Zurück zum Zitat Benedetti M, Catani F, Leardini A, Pignotti E, Giannini S (1998) Data management in gait analysis for clinical applications. Clin Biomech 13(3):204–215CrossRef Benedetti M, Catani F, Leardini A, Pignotti E, Giannini S (1998) Data management in gait analysis for clinical applications. Clin Biomech 13(3):204–215CrossRef
7.
Zurück zum Zitat Bey MJ, Kline SK, Tashman S, Zauel R (2008) Accuracy of biplane x-ray imaging combined with model-based tracking for measuring in vivo patellofemoral joint motion. J Orthop Surg Res 3(1):38PubMedCrossRef Bey MJ, Kline SK, Tashman S, Zauel R (2008) Accuracy of biplane x-ray imaging combined with model-based tracking for measuring in vivo patellofemoral joint motion. J Orthop Surg Res 3(1):38PubMedCrossRef
8.
Zurück zum Zitat Bey MJ, Peltz CD, Ciarelli K, Kline SK, Divine GW, Van Holsbeeck M, Muh S, Kolowich PA, Lock TR, Moutzouros V (2011) In vivo shoulder function after surgical repair of a torn rotator cuff glenohumeral joint mechanics, shoulder strength, clinical outcomes, and their interaction. Am J Sports Med 39(10):2117–2129PubMedCrossRef Bey MJ, Peltz CD, Ciarelli K, Kline SK, Divine GW, Van Holsbeeck M, Muh S, Kolowich PA, Lock TR, Moutzouros V (2011) In vivo shoulder function after surgical repair of a torn rotator cuff glenohumeral joint mechanics, shoulder strength, clinical outcomes, and their interaction. Am J Sports Med 39(10):2117–2129PubMedCrossRef
9.
Zurück zum Zitat Bey MJ, Zauel R, Brock SK, Tashman S (2006) Validation of a new model-based tracking technique for measuring three-dimensional, in vivo glenohumeral joint kinematics. J Biomech Eng 128(4):604PubMedCrossRef Bey MJ, Zauel R, Brock SK, Tashman S (2006) Validation of a new model-based tracking technique for measuring three-dimensional, in vivo glenohumeral joint kinematics. J Biomech Eng 128(4):604PubMedCrossRef
10.
Zurück zum Zitat Bingham J, Li G (2006) An optimized image matching method for determining in vivo TKA kinematics with a dual-orthogonal fluoroscopic imaging system. J Biomech Eng 128(4):588PubMedCrossRef Bingham J, Li G (2006) An optimized image matching method for determining in vivo TKA kinematics with a dual-orthogonal fluoroscopic imaging system. J Biomech Eng 128(4):588PubMedCrossRef
11.
Zurück zum Zitat Bland JM, Altman DG (1999) Measuring agreement in method comparison studies. Stat Methods Med Res 8(2):135–160PubMedCrossRef Bland JM, Altman DG (1999) Measuring agreement in method comparison studies. Stat Methods Med Res 8(2):135–160PubMedCrossRef
12.
Zurück zum Zitat Börlin N, Thien T, Kärrholm J (2002) The precision of radiostereometric measurements. Manual vs. digital measurements. J Biomech 35(1):69–79PubMedCrossRef Börlin N, Thien T, Kärrholm J (2002) The precision of radiostereometric measurements. Manual vs. digital measurements. J Biomech 35(1):69–79PubMedCrossRef
13.
Zurück zum Zitat Cappozzo A, Cappello A, Croce UD, Pensalfini F (1997) Surface-marker cluster design criteria for 3-D bone movement reconstruction. IEEE Trans Biomed Eng 44(12):1165–1174PubMedCrossRef Cappozzo A, Cappello A, Croce UD, Pensalfini F (1997) Surface-marker cluster design criteria for 3-D bone movement reconstruction. IEEE Trans Biomed Eng 44(12):1165–1174PubMedCrossRef
14.
Zurück zum Zitat Conti G, Cristofolini L, Juszczyk M, Leardini A, Viceconti M (2008) Comparison of three standard anatomical reference frames for the tibia–fibula complex. J Biomech 41(16):3384–3389PubMedCrossRef Conti G, Cristofolini L, Juszczyk M, Leardini A, Viceconti M (2008) Comparison of three standard anatomical reference frames for the tibia–fibula complex. J Biomech 41(16):3384–3389PubMedCrossRef
15.
Zurück zum Zitat Dennis DA, Komistek RD, Hoff WA, Gabriel SM (1996) In vivo knee kinematics derived using an inverse perspective technique. Clin Orthop Relat Res 331:107–117PubMedCrossRef Dennis DA, Komistek RD, Hoff WA, Gabriel SM (1996) In vivo knee kinematics derived using an inverse perspective technique. Clin Orthop Relat Res 331:107–117PubMedCrossRef
16.
Zurück zum Zitat Dennis DA, Mahfouz MR, Komistek RD, Hoff W (2005) In vivo determination of normal and anterior cruciate ligament-deficient knee kinematics. J Biomech 38(2):241–253PubMedCrossRef Dennis DA, Mahfouz MR, Komistek RD, Hoff W (2005) In vivo determination of normal and anterior cruciate ligament-deficient knee kinematics. J Biomech 38(2):241–253PubMedCrossRef
17.
Zurück zum Zitat Fregly BJ, Rahman HA, Banks SA (2005) Theoretical accuracy of model-based shape matching for measuring natural knee kinematics with single-plane fluoroscopy. J Biomech Eng 127(4):692–699PubMedCrossRef Fregly BJ, Rahman HA, Banks SA (2005) Theoretical accuracy of model-based shape matching for measuring natural knee kinematics with single-plane fluoroscopy. J Biomech Eng 127(4):692–699PubMedCrossRef
18.
Zurück zum Zitat Garling EH, Kaptein BL, Mertens B, Barendregt W, Veeger HEJ, Nelissen RGHH, Valstar ER (2007) Soft-tissue artefact assessment during step-up using fluoroscopy and skin-mounted markers. J Biomech 40(Suppl 1):S18–S24PubMedCrossRef Garling EH, Kaptein BL, Mertens B, Barendregt W, Veeger HEJ, Nelissen RGHH, Valstar ER (2007) Soft-tissue artefact assessment during step-up using fluoroscopy and skin-mounted markers. J Biomech 40(Suppl 1):S18–S24PubMedCrossRef
19.
Zurück zum Zitat Glaser D, Komistek RD, Cates HE, Mahfouz MR (2008) Clicking and squeaking: in vivo correlation of sound and separation for different bearing surfaces. J Bone Joint Surg 90(Suppl 4):112–120PubMedCrossRef Glaser D, Komistek RD, Cates HE, Mahfouz MR (2008) Clicking and squeaking: in vivo correlation of sound and separation for different bearing surfaces. J Bone Joint Surg 90(Suppl 4):112–120PubMedCrossRef
20.
Zurück zum Zitat Gronenschild E (1997) The accuracy and reproducibility of a global method to correct for geometric image distortion in the x-ray imaging chain. Med Phys 24(12):1875–1888PubMedCrossRef Gronenschild E (1997) The accuracy and reproducibility of a global method to correct for geometric image distortion in the x-ray imaging chain. Med Phys 24(12):1875–1888PubMedCrossRef
21.
Zurück zum Zitat Grood ES, Suntay WJ (1983) A joint coordinate system for the clinical description of three-dimensional motions: application to the knee. J Biomech Eng 105(2):136–144PubMedCrossRef Grood ES, Suntay WJ (1983) A joint coordinate system for the clinical description of three-dimensional motions: application to the knee. J Biomech Eng 105(2):136–144PubMedCrossRef
22.
Zurück zum Zitat Hanson RJ, Norris MJ (1981) Analysis of measurements based on the singular value decomposition. SIAM J Sci Stat Comput 2(3):363CrossRef Hanson RJ, Norris MJ (1981) Analysis of measurements based on the singular value decomposition. SIAM J Sci Stat Comput 2(3):363CrossRef
23.
Zurück zum Zitat Hirokawa S, Abrar Hossain M, Kihara Y, Ariyoshi S (2008) A 3D kinematic estimation of knee prosthesis using X-ray projection images: clinical assessment of the improved algorithm for fluoroscopy images. Med Biol Eng Comput 46(12):1253–1262PubMedCrossRef Hirokawa S, Abrar Hossain M, Kihara Y, Ariyoshi S (2008) A 3D kinematic estimation of knee prosthesis using X-ray projection images: clinical assessment of the improved algorithm for fluoroscopy images. Med Biol Eng Comput 46(12):1253–1262PubMedCrossRef
24.
Zurück zum Zitat Hurschler C, Seehaus F, Emmerich J, Kaptein BL, Windhagen H (2009) Comparison of the model-based and marker-based roentgen stereophotogrammetry methods in a typical clinical setting. J Arthroplast 24(4):594–606CrossRef Hurschler C, Seehaus F, Emmerich J, Kaptein BL, Windhagen H (2009) Comparison of the model-based and marker-based roentgen stereophotogrammetry methods in a typical clinical setting. J Arthroplast 24(4):594–606CrossRef
25.
Zurück zum Zitat Illingworth J, Kittler J (1988) A survey of the Hough transform. Comput Vis Graph Image Process 44(1):87–116CrossRef Illingworth J, Kittler J (1988) A survey of the Hough transform. Comput Vis Graph Image Process 44(1):87–116CrossRef
26.
Zurück zum Zitat Kaptein BL, Valstar ER, Stoel BC, Reiber HC, Nelissen RG (2007) Clinical validation of model-based RSA for a total knee prosthesis. Clin Orthop Relat Res 464:205–209PubMed Kaptein BL, Valstar ER, Stoel BC, Reiber HC, Nelissen RG (2007) Clinical validation of model-based RSA for a total knee prosthesis. Clin Orthop Relat Res 464:205–209PubMed
27.
Zurück zum Zitat Kaptein BL, Valstar ER, Stoel BC, Rozing PM, Reiber JHC (2003) A new model-based RSA method validated using CAD models and models from reversed engineering. J Biomech 36(6):873–882PubMedCrossRef Kaptein BL, Valstar ER, Stoel BC, Rozing PM, Reiber JHC (2003) A new model-based RSA method validated using CAD models and models from reversed engineering. J Biomech 36(6):873–882PubMedCrossRef
28.
Zurück zum Zitat Kedgley AE, Jenkyn TR (2009) RSA calibration accuracy of a fluoroscopy-based system using nonorthogonal images for measuring functional kinematics. Med Phys 36(7):3176PubMedCrossRef Kedgley AE, Jenkyn TR (2009) RSA calibration accuracy of a fluoroscopy-based system using nonorthogonal images for measuring functional kinematics. Med Phys 36(7):3176PubMedCrossRef
29.
Zurück zum Zitat Kuo M-Y, Tsai T-Y, Lin C-C, Lu T-W, Hsu H-C, Shen W-C (2011) Influence of soft tissue artifacts on the calculated kinematics and kinetics of total knee replacements during sit-to-stand. Gait Posture 33(3):379–384PubMedCrossRef Kuo M-Y, Tsai T-Y, Lin C-C, Lu T-W, Hsu H-C, Shen W-C (2011) Influence of soft tissue artifacts on the calculated kinematics and kinetics of total knee replacements during sit-to-stand. Gait Posture 33(3):379–384PubMedCrossRef
30.
Zurück zum Zitat Lavallee S, Szeliski R (1995) Recovering the position and orientation of free-form objects from image contours using 3D distance maps. IEEE Trans Pattern Anal Mach Intell 17(4):378–390CrossRef Lavallee S, Szeliski R (1995) Recovering the position and orientation of free-form objects from image contours using 3D distance maps. IEEE Trans Pattern Anal Mach Intell 17(4):378–390CrossRef
31.
Zurück zum Zitat Li G, Van de Velde SK, Bingham JT (2008) Validation of a non-invasive fluoroscopic imaging technique for the measurement of dynamic knee joint motion. J Biomech 41(7):1616–1622PubMedCrossRef Li G, Van de Velde SK, Bingham JT (2008) Validation of a non-invasive fluoroscopic imaging technique for the measurement of dynamic knee joint motion. J Biomech 41(7):1616–1622PubMedCrossRef
32.
Zurück zum Zitat Ludewig PM, Reynolds JE (2009) The association of scapular kinematics and glenohumeral joint pathologies. J Orthop Sports Phys Therapy 39(2):90–104 Ludewig PM, Reynolds JE (2009) The association of scapular kinematics and glenohumeral joint pathologies. J Orthop Sports Phys Therapy 39(2):90–104
33.
Zurück zum Zitat Mahfouz MR, Hoff WA, Komistek RD, Dennis DA (2003) A robust method for registration of three-dimensional knee implant models to two-dimensional fluoroscopy images. IEEE Trans Med Imaging 22(12):1561–1574PubMedCrossRef Mahfouz MR, Hoff WA, Komistek RD, Dennis DA (2003) A robust method for registration of three-dimensional knee implant models to two-dimensional fluoroscopy images. IEEE Trans Med Imaging 22(12):1561–1574PubMedCrossRef
34.
Zurück zum Zitat Miranda DL, Schwartz JB, Loomis AC, Brainerd EL, Fleming BC, Crisco JJ (2011) Static and dynamic error of a biplanar videoradiography system using marker-based and markerless tracking techniques. J Biomech Eng 133(12):121002–121008PubMedCrossRef Miranda DL, Schwartz JB, Loomis AC, Brainerd EL, Fleming BC, Crisco JJ (2011) Static and dynamic error of a biplanar videoradiography system using marker-based and markerless tracking techniques. J Biomech Eng 133(12):121002–121008PubMedCrossRef
35.
Zurück zum Zitat Moro-oka T, Hamai S, Miura H, Shimoto T, Higaki H, Fregly BJ, Iwamoto Y, Banks SA (2007) Can magnetic resonance imaging-derived bone models be used for accurate motion measurement with single-plane three-dimensional shape registration? J Orthop Res 25(7):867–872PubMedCrossRef Moro-oka T, Hamai S, Miura H, Shimoto T, Higaki H, Fregly BJ, Iwamoto Y, Banks SA (2007) Can magnetic resonance imaging-derived bone models be used for accurate motion measurement with single-plane three-dimensional shape registration? J Orthop Res 25(7):867–872PubMedCrossRef
36.
Zurück zum Zitat Okamoto N, Breslauer L, Hedley AK, Mizuta H, Banks SA (2011) In vivo knee kinematics in patients with bilateral total knee arthroplasty of 2 designs. J Arthroplast 26(6):914–918CrossRef Okamoto N, Breslauer L, Hedley AK, Mizuta H, Banks SA (2011) In vivo knee kinematics in patients with bilateral total knee arthroplasty of 2 designs. J Arthroplast 26(6):914–918CrossRef
37.
Zurück zum Zitat Önsten I, Berzins A, Shott S, Sumner DR (2001) Accuracy and precision of radiostereometric analysis in the measurement of THR femoral component translations: human and canine in vitro models. J Orthop Res 19(6):1162–1167PubMedCrossRef Önsten I, Berzins A, Shott S, Sumner DR (2001) Accuracy and precision of radiostereometric analysis in the measurement of THR femoral component translations: human and canine in vitro models. J Orthop Res 19(6):1162–1167PubMedCrossRef
38.
Zurück zum Zitat Ploegmakers MJM, Ginsel B, Meijerink HJ, de Rooy JW, de Waal Malefijt MC, Verdonschot N, Banks SA (2010) Physical examination and in vivo kinematics in two posterior cruciate ligament retaining total knee arthroplasty designs. The Knee 17(3):204–209PubMedCrossRef Ploegmakers MJM, Ginsel B, Meijerink HJ, de Rooy JW, de Waal Malefijt MC, Verdonschot N, Banks SA (2010) Physical examination and in vivo kinematics in two posterior cruciate ligament retaining total knee arthroplasty designs. The Knee 17(3):204–209PubMedCrossRef
39.
Zurück zum Zitat Seehaus F, Emmerich J, Kaptein BL, Windhagen H, Hurschler C (2009) Experimental analysis of model-based roentgen stereophotogrammetric analysis (MBRSA) on four typical prosthesis components. J Biomech Eng 131(4):041004PubMedCrossRef Seehaus F, Emmerich J, Kaptein BL, Windhagen H, Hurschler C (2009) Experimental analysis of model-based roentgen stereophotogrammetric analysis (MBRSA) on four typical prosthesis components. J Biomech Eng 131(4):041004PubMedCrossRef
40.
Zurück zum Zitat Söderkvist I, Wedin P-Å (1993) Determining the movements of the skeleton using well-configured markers. J Biomech 26(12):1473–1477PubMedCrossRef Söderkvist I, Wedin P-Å (1993) Determining the movements of the skeleton using well-configured markers. J Biomech 26(12):1473–1477PubMedCrossRef
41.
Zurück zum Zitat Stagni R, Fantozzi S, Cappello A, Leardini A (2005) Quantification of soft tissue artefact in motion analysis by combining 3D fluoroscopy and stereophotogrammetry: a study on two subjects. Clin Biomech 20(3):320–329CrossRef Stagni R, Fantozzi S, Cappello A, Leardini A (2005) Quantification of soft tissue artefact in motion analysis by combining 3D fluoroscopy and stereophotogrammetry: a study on two subjects. Clin Biomech 20(3):320–329CrossRef
42.
Zurück zum Zitat El-mihoub TA, Hopgood AA, Nolle L, Battersby A (2006) Hybrid genetic algorithms: a review. Eng Lett 13(2):124–137 El-mihoub TA, Hopgood AA, Nolle L, Battersby A (2006) Hybrid genetic algorithms: a review. Eng Lett 13(2):124–137
43.
Zurück zum Zitat Tashman S, Kolowich P, Collon D, Anderson K, Anderst W (2007) Dynamic function of the ACL-reconstructed knee during running. Clin Orthop Relat Res 454:66–73PubMedCrossRef Tashman S, Kolowich P, Collon D, Anderson K, Anderst W (2007) Dynamic function of the ACL-reconstructed knee during running. Clin Orthop Relat Res 454:66–73PubMedCrossRef
44.
Zurück zum Zitat Tersi L, Fantozzi S, Stagni R (2010) 3D elbow kinematics with mono-planar fluoroscopy: in silico evaluation. EURASIP J Adv Signal Process 2010, art ID 142989. doi:10.1155/2010/142989 Tersi L, Fantozzi S, Stagni R (2010) 3D elbow kinematics with mono-planar fluoroscopy: in silico evaluation. EURASIP J Adv Signal Process 2010, art ID 142989. doi:10.​1155/​2010/​142989
45.
Zurück zum Zitat Tersi L, Fantozzi S, Stagni R, Cappello A (2012) Fluoroscopic analysis for the estimation of in vivo elbow kinematics: influence of 3D model. J Mech Med Biol 12(3), art ID 1250046. doi:10.1142/S0219519411004769 Tersi L, Fantozzi S, Stagni R, Cappello A (2012) Fluoroscopic analysis for the estimation of in vivo elbow kinematics: influence of 3D model. J Mech Med Biol 12(3), art ID 1250046. doi:10.​1142/​S021951941100476​9
46.
Zurück zum Zitat Tersi L, Stagni R, Fantozzi S, Cappello A (2010) Genetic Algorithm as a robust method for the joint kinematics estimation with mono-planar 3D fluoroscopy. In: XVII ESB Conference, Edinburgh, Scotland, UK, July 2010 Tersi L, Stagni R, Fantozzi S, Cappello A (2010) Genetic Algorithm as a robust method for the joint kinematics estimation with mono-planar 3D fluoroscopy. In: XVII ESB Conference, Edinburgh, Scotland, UK, July 2010
47.
Zurück zum Zitat Tsai T-Y, Lu T-W, Chen C-M, Kuo M-Y, Hsu H-C (2010) A volumetric model-based 2D to 3D registration method for measuring kinematics of natural knees with single-plane fluoroscopy. Med Phys 37(3):1273–1284PubMedCrossRef Tsai T-Y, Lu T-W, Chen C-M, Kuo M-Y, Hsu H-C (2010) A volumetric model-based 2D to 3D registration method for measuring kinematics of natural knees with single-plane fluoroscopy. Med Phys 37(3):1273–1284PubMedCrossRef
48.
Zurück zum Zitat Tsai T-Y, Lu T-W, Kuo M-Y, Lin C-C (2011) Effects of soft tissue artifacts on the calculated kinematics and kinetics of the knee during stair-ascent. J Biomech 44(6):1182–1188PubMedCrossRef Tsai T-Y, Lu T-W, Kuo M-Y, Lin C-C (2011) Effects of soft tissue artifacts on the calculated kinematics and kinetics of the knee during stair-ascent. J Biomech 44(6):1182–1188PubMedCrossRef
49.
Zurück zum Zitat Valstar ER, Gill R, Ryd L, Flivik G, Börlin N, Kärrholm J (2005) Guidelines for standardization of radiostereometry (RSA) of implants. Acta Orthop 76(4):563–572PubMedCrossRef Valstar ER, Gill R, Ryd L, Flivik G, Börlin N, Kärrholm J (2005) Guidelines for standardization of radiostereometry (RSA) of implants. Acta Orthop 76(4):563–572PubMedCrossRef
50.
Zurück zum Zitat Valstar ER, Nelissen RGHH, Reiber JHC, Rozing PM (2002) The use of Roentgen stereophotogrammetry to study micromotion of orthopaedic implants. ISPRS J Photogramm Remote Sens 56(5–6):376–389CrossRef Valstar ER, Nelissen RGHH, Reiber JHC, Rozing PM (2002) The use of Roentgen stereophotogrammetry to study micromotion of orthopaedic implants. ISPRS J Photogramm Remote Sens 56(5–6):376–389CrossRef
51.
Zurück zum Zitat Wright AH (1991) Genetic algorithms for real parameter optimization. Found Genet Algorithms 1(1991):205–218 Wright AH (1991) Genetic algorithms for real parameter optimization. Found Genet Algorithms 1(1991):205–218
52.
Zurück zum Zitat Wu G, Siegler S, Allard P, Kirtley C, Leardini A, Rosenbaum D, Whittle M, D’Lima DD, Cristofolini L, Witte H, Schmid O, Stokes I (2002) ISB recommendation on definitions of joint coordinate system of various joints for the reporting of human joint motion—part I: ankle, hip, and spine. J Biomech 35(4):543–548PubMedCrossRef Wu G, Siegler S, Allard P, Kirtley C, Leardini A, Rosenbaum D, Whittle M, D’Lima DD, Cristofolini L, Witte H, Schmid O, Stokes I (2002) ISB recommendation on definitions of joint coordinate system of various joints for the reporting of human joint motion—part I: ankle, hip, and spine. J Biomech 35(4):543–548PubMedCrossRef
53.
Zurück zum Zitat Yamazaki T, Watanabe T, Nakajima Y, Sugamoto K, Tomita T, Yoshikawa H, Tamura S (2004) Improvement of depth position in 2-D/3-D registration of knee implants using single-plane fluoroscopy. IEEE Trans Med Imaging 23(5):602–612PubMedCrossRef Yamazaki T, Watanabe T, Nakajima Y, Sugamoto K, Tomita T, Yoshikawa H, Tamura S (2004) Improvement of depth position in 2-D/3-D registration of knee implants using single-plane fluoroscopy. IEEE Trans Med Imaging 23(5):602–612PubMedCrossRef
54.
Zurück zum Zitat You B-M, Siy P, Anderst W, Tashman S (2001) In vivo measurement of 3-D skeletal kinematics from sequences of biplane radiographs: application to knee kinematics. IEEE Trans Med Imaging 20(6):514–525PubMedCrossRef You B-M, Siy P, Anderst W, Tashman S (2001) In vivo measurement of 3-D skeletal kinematics from sequences of biplane radiographs: application to knee kinematics. IEEE Trans Med Imaging 20(6):514–525PubMedCrossRef
55.
Zurück zum Zitat Zhu Z, Massimini DF, Wang G, Warner JJP, Li G (2012) The accuracy and repeatability of an automatic 2D–3D fluoroscopic image-model registration technique for determining shoulder joint kinematics. Med Eng Phys 34(9):1303–1309PubMedCrossRef Zhu Z, Massimini DF, Wang G, Warner JJP, Li G (2012) The accuracy and repeatability of an automatic 2D–3D fluoroscopic image-model registration technique for determining shoulder joint kinematics. Med Eng Phys 34(9):1303–1309PubMedCrossRef
56.
Zurück zum Zitat Zuffi S, Leardini A, Catani F, Fantozzi S, Cappello A (1999) A model-based method for the reconstruction of total knee replacement kinematics. IEEE Trans Med Imaging 18(10):981–991PubMedCrossRef Zuffi S, Leardini A, Catani F, Fantozzi S, Cappello A (1999) A model-based method for the reconstruction of total knee replacement kinematics. IEEE Trans Med Imaging 18(10):981–991PubMedCrossRef
Metadaten
Titel
In vitro quantification of the performance of model-based mono-planar and bi-planar fluoroscopy for 3D joint kinematics estimation
verfasst von
Luca Tersi
Arnaud Barré
Silvia Fantozzi
Rita Stagni
Publikationsdatum
01.03.2013
Verlag
Springer-Verlag
Erschienen in
Medical & Biological Engineering & Computing / Ausgabe 3/2013
Print ISSN: 0140-0118
Elektronische ISSN: 1741-0444
DOI
https://doi.org/10.1007/s11517-012-0987-4

Weitere Artikel der Ausgabe 3/2013

Medical & Biological Engineering & Computing 3/2013 Zur Ausgabe

Premium Partner