Skip to main content

1978 | Buch | 2. Auflage

Worked Examples in X-Ray Analysis

verfasst von: R. Jenkins, J. L. de Vries

Verlag: Macmillan Education UK

Buchreihe : Philips Technical Library

insite
SUCHEN

Inhaltsverzeichnis

Frontmatter
Question 1. Characteristic spectra Moseley’s law

Moseley’s law relates the wavelength X of a characteristic line in a given series with atomic number Z. The usual expression of the law is 1<math display='block'> <mrow> <mfrac> <mn>1</mn> <mrow> <msqrt> <mi>&#x03BB;</mi> </msqrt> </mrow> </mfrac> <mtext>&#x2009;</mtext><mo>=</mo><mtext>&#x2009;</mtext><mi>K</mi><mtext>&#x2009;</mtext><mo stretchy='false'>(</mo><mi>Z</mi><mo>&#x2212;</mo><mi>&#x03C3;</mi><mo stretchy='false'>)</mo></mrow> </math>]]</EquationSource><EquationSource Format="TEX"><![CDATA[$$\frac{1}{{\sqrt \lambda }}\; = \;K\;(Z - \sigma )$$ where K is a constant for a given spectral series and σ is a shielding constant.

R. Jenkins, J. L. de Vries
Question 2. Characteristic spectra — prediction of possible transitions

The electronic structure of Cu (Z = 29) is denoted 1s2, 2s22p6, 3s23p63d9, 4s2.

R. Jenkins, J. L. de Vries
Question 3. Continuous spectra — use of Kramers’ formula

Plot the approximate distribution of intensity that would be obtained from a chromium anode X-ray tube operated at 50 kV. Use Kramers’ formula4) to give the relative intensity in the number of photons (I) over the wave­length range 0–5 Å.

R. Jenkins, J. L. de Vries
Question 4. Continuous radiation — Scattered radiation

A The continuous radiation from an X-ray tube is scattered by the sample, diffracted by the analyzing crystal and measured by the detector.

R. Jenkins, J. L. de Vries
Question 5. Detector resolution

The resolution R of a proportional counter is expressed in terms of the peak width at half height multiplied by 100, and divided by the value of the maximum pulse amplitude.

R. Jenkins, J. L. de Vries
Question 6. Dispersion as a function of ‘d’ spacing

LiF is commonly used as an analysing crystal. The (200) reflection planes are commonly used, having a “2d” value of 4.028 Å. It is, however, possible to use reflecting planes with other Miller indices (hkl) for example the (220), the (420) and the (422) planes.

R. Jenkins, J. L. de Vries
Question 7. Choice of collimator/crystal combination

In the measurement of a mixture of elements a problem of the overlap of two lines is encountered. The lines can be separated by use of the fol­lowing combinations: (i)topaz crystal (having good resolution but poor reflecting power) plus a coarse collimator, placed between sample and crystal.(ii)A LiF (200) crystal (having poor resolution but good reflectivity) plus a fine collimator, placed between sample and crystal.

R. Jenkins, J. L. de Vries
Question 8. Pulse height selection — crystal fluorescence

It is necessary to set up an X-ray spectrometer for the measurement of sodium in a series of rock specimens. An argon/methane flow count­er is being employed and a KAP (potassium hydrogen phthalate) analysing crystal. The rock samples contain quite a lot of calcium and this leads to strong crystal fluorescence.

R. Jenkins, J. L. de Vries
Question 9. Pulse height selection — pulse height variation

A mixture of sodium silicates and sodium phosphates is being analysed for phosphorus. The experimental conditions used are: tungsten anode X-ray tube, 50 kV-40 mA; coarse collimator; gypsum analysing crystal; gas flow proportional counter (Ar/10% methane) with pulse height selection.

R. Jenkins, J. L. de Vries
Question 10. Detectors — double plateaux

A series of measurements are made in which counting rate is plotted as a function of counter voltage. The counter in this case is an argon/methane gas flow proportional counter. A single plateau is found for a sample of pure sulphur, using the S Kα line, but pure samples of iron and chromium are both found to give a second plateau, again when measuring their Kα lines.

R. Jenkins, J. L. de Vries
Question 11. Choice of detectors

A choice has to be made between the scintillation counter and the flow proportional counter for the measurement of a certain wavelength.

R. Jenkins, J. L. de Vries
Question 12. Count rate differences with Geiger and proportional counters

A silicon specimen is being examined using nickel filtered Cu Kα radiation. The intensities of two reflections are measured with a Geiger-Müller counter and the following counting rates obtained:

R. Jenkins, J. L. de Vries
Question 13. Dispersion of the diffractometer

The dispersion of the diffractometer (dθ/dλ) can be expressed in the given form obtained by differentiating the Bragg law nλ = 2dsinθ5<math display='block'> <mrow> <mfrac> <mrow> <mi>d</mi><mi>&#x03B8;</mi></mrow> <mrow> <mi>d</mi><mi>&#x03BB;</mi></mrow> </mfrac> <mtext>&#x2009;</mtext><mo>=</mo><mtext>&#x2009;</mtext><mfrac> <mi>n</mi> <mrow> <mn>2</mn><mi>d</mi></mrow> </mfrac> <mtext>&#x2009;</mtext><mo>.</mo><mtext>&#x2009;</mtext><mfrac> <mn>1</mn> <mrow> <mi>cos</mi><mi>&#x03B8;</mi></mrow> </mfrac> </mrow> </math>]]</EquationSource><EquationSource Format="TEX"><![CDATA[$$\frac{{d\theta }}{{d\lambda }}\; = \;\frac{n}{{2d}}\;.\;\frac{1}{{\cos \theta }}$$where the angular separation (dθ) of two wavelengths separated by (dλ) is expressed in terms of the diffraction angle θ for a certain spacing d.

R. Jenkins, J. L. de Vries
Question 14. Removal of sample fluorescence

The diffraction pattern of an iron specimen is being recorded with a copper target X-ray tube operating at 40 kV-50 mA. A Ni-filter is placed between the X-ray tube and specimen. A Xe proportional counter is used to record counts on one of the haematite reflections but even though this is one of the strongest reflections usable, the peak to background is only about 1.5 to 1.

R. Jenkins, J. L. de Vries
Question 15. Spurious peaks in X-ray diffractograms

Adiffractogram of quartz exhibited an additional spurious peak at approximately 6° (2θ).The effect was observed both in specimens containing only low atomic number elements and specimens containing high atomic number elements. The patterns were always taken using Cu radiation, proportional counter (Xe-filled) and pulse height selection. The pulse height selector was set at the 45% acceptance level since the resolution of the counter (peak width at half height) was 18% for Cu Kα radiation.

R. Jenkins, J. L. de Vries
Question 16. Extra lines in X-ray diffractometry

A diffractometer trace is recorded of a compound known to contain a high concentration of Ba. The pattern is characterized by a very strong line with a d -spacing of 3.08 Å, plus many others. A weak, broad line was observed above a rather high background at approximately 6.5 o(2θ). It was suspected from the shape of this line that it did not belong to the regular diffraction pattern. Can there be an alternative explanation for this line.

R. Jenkins, J. L. de Vries
Question 17. Counting statistics — variation of standard deviation with counting time

A certain analysis line gave 800 c/s under certain instrumental conditions. Given that the standard deviation (σ) of a given number of counts (N) is equal to √ N, calculate the standard deviation expected in a counting time of 20 seconds.

R. Jenkins, J. L. de Vries
QUESTION 18. Estimation of Error Introduced by Ignoring the Background

Whenever X-ray intensity measurements are made on a selected peak, occuring at a certain setting of the goniometer, some fraction of the measured counting rate is always due to background. The significance of this background is dependent on its relative magnitude, and it is often difficult to judge whether or not it can be ignored in the calculation of the counting error. The true counting error ɛ % is related to the peak R p and background R b counting rates in the following way: 6<math display='block'> <mrow> <mi>&#x03B5;</mi><mi>&#x0025;</mi><mo>=</mo><mtext>&#x2009;</mtext><mfrac> <mrow> <mn>100</mn></mrow> <mrow> <msqrt> <mi>T</mi> </msqrt> </mrow> </mfrac> <mo>.</mo><mfrac> <mn>1</mn> <mrow> <msqrt> <mrow> <msub> <mi>R</mi> <mi>p</mi> </msub> </mrow> </msqrt> <mtext>&#x2009;</mtext><mo>&#x2212;</mo><msqrt> <mrow> <msub> <mi>R</mi> <mi>b</mi> </msub> </mrow> </msqrt> </mrow> </mfrac> </mrow> </math>]]</EquationSource><EquationSource Format="TEX"><![CDATA[$$\varepsilon \% = \;\frac{{100}}{{\sqrt T }}.\frac{1}{{\sqrt {{R_p}} \; - \sqrt {{R_b}} }}$$ where T is the total analysis time.

R. Jenkins, J. L. de Vries
Question 19. Trace analysis — derivation of detection limit expression

The lower limit of detection is often defined as that concentration equal to two standard deviations of the background count rate. Derive an ex­pression which defines the lower limit of detection in terms of the anal­ysis time T, the background counting rate R b and the response m of an element given in c/s per %. Remember that trace analysis always re­quires two measurements, i.e. peak plus background.

R. Jenkins, J. L. de Vries
Question 20. Choice between fixed time and fixed count

A series of samples of alloyed steels were found to give 300 c/s/% for the Si Kα line. A calibration procedure has to be prepared over the range 0.20 – 0.60% Si and the choice has to be made between fixed count and fixed time operation.

R. Jenkins, J. L. de Vries
Question 21. Choice between ratio and absolute counting methods

The choice has to be made between the ratio and absolute method of counting in a counting programme which already involves collection of counts on a standard sample (i.e. whether the absolute or ratio method is chosen the standard is counted anyway, so no time is to be saved by not counting the standard).

R. Jenkins, J. L. de Vries
Question 22. Variation of limit of detection with analysis time

A series of measurements of fluorine in cement sample yielded a peak counting rate of 4 c/s/% for fluorine (F Kα) over a background of 6.5 c/s.

R. Jenkins, J. L. de Vries
Question 23. Analytical error due to counting statistics

A series of measurements are being made to determine the concentration of BaO in a mixture of BaO and BaO.6Fe2O3.

R. Jenkins, J. L. de Vries
Question 24. Particle statistics in X-ray diffractometry

Only those crystallites having the reflecting planes almost parallel to the specimen surface can contribute to a certain reflection. The intensity of the resulting diffraction is thus dependent on this number of crystallites. Intensities of different diffraction lines or between different specimens can only be compared if the number of particles contributing is the same fraction of the total number of particles. If this total number is too small to warrant a random distribution, an error in the intensity measurement is introduced8). This may occur for instance when the particles are too large.

R. Jenkins, J. L. de Vries
Question 25. Determination of traces of lead in oil (absolute measurement)

A series of oil samples gave the following intensities on the Pb Lα line: Prepare a calibration curve and work out the equation of the line. What is the concentration of lead in sample 6.

R. Jenkins, J. L. de Vries
Question 26. Determination of CaO in cement (ratio measurement)

The following data were obtained using a series of cement samples. Counting ratios were determined on the Ca Kα line using sample 5 as the ratio standard.

R. Jenkins, J. L. de Vries
Question 27. Measurement of, and correction for, absorption in trace analysis of whole rock

A whole rock specimen was being analyzed for trace concentrations of strontium. The method of Norrish and Chappell(10) was used to measure and correct for absorption and the following data were obtained.

R. Jenkins, J. L. de Vries
Question 28. Calculation of matrix absorption and prediction of absorption effects

The total secondary absorption of a matrix is given by 16<math display='block'> <mrow> <msub> <mi>&#x03BC;</mi> <mrow> <mi>m</mi><mi>a</mi><mi>t</mi><mi>r</mi><mi>i</mi><mi>x</mi><mtext>&#x2009;</mtext></mrow> </msub> <mo>=</mo><munder> <mrow> <mo>&#x2211;</mo><mtext>&#x2009;</mtext><mo stretchy='false'>(</mo><msub> <mi>&#x03BC;</mi> <mi>i</mi> </msub> <msub> <mi>w</mi> <mi>i</mi> </msub> <mo stretchy='false'>)</mo></mrow> <mi>i</mi> </munder> </mrow> </math>]]</EquationSource><EquationSource Format="TEX"><![CDATA[$${\mu _{matrix\;}} = \mathop {\sum \;({\mu _i}{w_i})}\limits_i $$ where μ i and w i are the individual mass absorption coefficients and weight fractions for each element i making up the matrix.

R. Jenkins, J. L. de Vries
Question 29. Calculation of absorption coefficients and absorption effects

Calculate the total secondary absorption of a cement sample for Ca Kα radiation. Use the following mass absorption coefficients for Ca Kα.

R. Jenkins, J. L. de Vries
Question 30. Absorption correction involving primary and secondary absorption

The slope of a calibration curve for a certain element is dependent upon the total absorption of the measured wavelength within the sample matrix. Although it is common practice to assume that only secondary absorption is of significance, in fact both primary and secondary absorption should be considered. The following expression relates the so-called “efficiency factor” C(λλ j ) with the mass absorption coefficient µ i of each element i for the measured wavelength λ j and an effective primary wavelength λ which represents the primary spectrum.

R. Jenkins, J. L. de Vries
Question 31. Effect of matric absorption on sensitivity

An ore specimen consisting predominantly of calcium and barium carbonates contained traces of nickel. A specimen containing no barium sulphate with a mass absorption coefficient for Ni K

α

of 124 gave the following data

240 c/s per % for nickel

background 120 c/s

(a)

Estimate the lower limit of detection for nickel in an analysis time of 200 s.

(b)

What would be the effect on this detection limit of adding 20% of barium sulphate to the specimen, assuming that the c/s per % value is inversely proportional to the matrix absorption? Assume also that the background is reduced by 8% by the addition of the barium sulphate. The atomic weights of Ba = 137.4, S = 32.1 and O = 16.

R. Jenkins, J. L. de Vries
Question 32. Calculation of quantity of heavy absorber needed to minimize matrix effects

In the analysis of tin ores it is found that a very curved calibration line is obtained due to the large difference in the mass absorption coefficients of the two matrix components Sn (µ = 13) and SiO2 (µ = 1.80) for the analysis line (Sn Kα).

R. Jenkins, J. L. de Vries
Question 33. Use of α correction factor (intensity correction)

A series of eight samples were analysed for a certain element A. An element B, also present in major concentration, was found to strongly absorb element A. The following intensity data were obtained: The concentrations of element B are unknown, although it is suspected that the eight standard samples are almost pure binary mixtures.

R. Jenkins, J. L. de Vries
Question 34. Use of α correction factors (concentration correction)

A series of samples are being analysed for lead using the Pb Lα line. It is found that zinc which is also present strongly absorbs the Pb Lα wavelength resulting in a very poor calibration curve. The following data were obtained:

R. Jenkins, J. L. de Vries
Question 35. Excitation probability

It is often thought that the wavelength in the primary spectrum most efficient in exciting the fluorescent radiation of an element A is that which is closest to the absorption edge of element A. The probability function of the primary spectrum should follow the absorption curve for A.

R. Jenkins, J. L. de Vries
Question 36. Quantitative diffractometry — determination of Anatase in Rutile

A sample is known to contain two crystalline modifications of TiO2, namely rutile and anatase. Spectrographic analysis failed to reveal the presence of other elements, so the sample can be assumed to consist only of TiO2. A scan is made with the diffractometer and integrated intensities measured for the (110) reflection of rutile (I R ) and for the (101) reflection of anatase (I A ). Background measurements were made from a suitable part of the background. The measured intensities are as follows: The following relationship is known to exist between the concentration of rutile (C R ) and the concentration of anatase (C A ): 21<math display='block'> <mrow> <mfrac> <mrow> <msub> <mi>C</mi> <mi>R</mi> </msub> </mrow> <mrow> <msub> <mi>C</mi> <mi>A</mi> </msub> </mrow> </mfrac> <mtext>&#x2009;</mtext><mo>=</mo><mtext>&#x2009;</mtext><mi>K</mi><mfrac> <mrow> <msub> <mi>I</mi> <mi>R</mi> </msub> </mrow> <mrow> <msub> <mi>I</mi> <mi>A</mi> </msub> </mrow> </mfrac> </mrow> </math>]]</EquationSource><EquationSource Format="TEX"><![CDATA[$$\frac{{{C_R}}}{{{C_A}}}\; = \;K\frac{{{I_R}}}{{{I_A}}}$$ where K is a proportionality constant. Previous experiments 9) indicated a value of 1.33 for K.

R. Jenkins, J. L. de Vries
Question 37. Quantitative X-ray diffractometry use of internal standard

The weight fraction of a quartz (SiO2) has to be determined in a natural specimen. To correct for the unknown absorption in this specimen, 200 mg KCl are added to 1000 mg of the sample, both very finely ground, and well mixed. Determine the weight fraction of α quartz, using the following data:

R. Jenkins, J. L. de Vries
Question 38. Estimation of expected intensities in mixtures

The intensity of the 3.34 Å line of α-quartz (SiO2) has to be estimated in a mixture consisting of 10% α-quartz and 90% CaSiO3. Nickel filtered Cu Kα radiation is used. The intensity of the 3.34 Å line in pure α quartz was 100 c/s, measured under the same conditions. What is the expected intensity from the mixture.

R. Jenkins, J. L. de Vries
Question 39. Quantitative x-ray diffractometry

Very often the quantity of a compound has to be determined in a complex mineral, the composition and absorption coefficients of which are unknown. To correct for these unknown facts, the attenuation of a diffraction line arising from the bottom of the sample holder may be used. The following intensities were collected in measuring the percentage quartz in a mineral with a conventional diffractometer, utilizing a sample holder with a nickel bottom.

R. Jenkins, J. L. de Vries
Question 40. Determination of dead time

It is necessary to determine the dead time of an X-ray spectrometer using the Kβ/Kα counting ratio method. The following data were obtained on the Kα and Kβ lines of tin by varying the X-ray tube current and keeping the kV constant.

R. Jenkins, J. L. de Vries
Question 41. Contamination Lines In X-ray Spectra

A pellet of pure sugar was placed in the spectrometer and a spectrum run between 10° – 130° 2

θ

in order to ascertain the ‘blank’ spectrum of the tube. The following conditions were employed:

Cr anode tube at 60 kV 32 mA (this tube had been run for about 5000 h).

LiF (200) crystal, 2

d

= 4.028Å. Angle between incident and take-off = 90°.

Flow plus scintillation detectors in tandem.

Fine primary collimator, vacuum condition.

R. Jenkins, J. L. de Vries
Question 42. Measurement of coating thickness on irregular shaped objects

The thickness of Cr plating on a piece of iron of rather irregular but flat shape has to be measured.

R. Jenkins, J. L. de Vries
Question 43. Background variation

A scan is being made on a rock sample in the region of the barium K spectra.

R. Jenkins, J. L. de Vries
Question 44. Use of primary filters for removal of characteristic tube lines

It is necessary to measure low concentrations of chromium using a spectrometer equipped with a chromium anode X-ray tube. A titanium filter can be fitted over the tube window to minimize the intensity of the Cr Kα radiation arising from the scattered tube radiation. Two thicknesses of titanium are available, at 0.13 mm and 0.011 mm.

R. Jenkins, J. L. de Vries
Question 45. Calculation of ß filter thickness and transmission

A Ni filter has been constructed in such a way that only 2% of the incident Cu Kβ radiation may pass. Given that the density of Ni is 8.92 g/cm3 and that the mass absorption of Ni for Cu Kα and Cu Kβ is 49.2 cm2/g and 286 cm2/g respectively. calculate: AThe thickness of the filterBThe percentage of incident Cu Kα radiation that can pass the filter.

R. Jenkins, J. L. de Vries
Question 46. Utilization of the ß filter and pulse height selector in diffractometry

A filter is very often used in X-ray diffractometry to reduce the background level and the intensity of β-radiation. This filter can be placed either between X-ray tube and specimen or between the specimen and detector.

R. Jenkins, J. L. de Vries
Question 47. Determination of lattice constants for a cubic lattice

The following 12 lines were obtained from a crystalline powder. Data were obtained using nickel filtered Cu Kα radiation. The powder is known to belong to the cubic system.

R. Jenkins, J. L. de Vries
Question 48. Relative line intensities obtained on alkali halides

Table 1 lists the relative line intensities obtained on samples of sodium, potassium and rubidium chlorides, using identical instrumental conditions.

R. Jenkins, J. L. de Vries
Question 49. Line broadening

Measurements of the nickel (111) reflection are being made on a series of Raney Nickel catalysts. Severe line broadening problems are encountered due to the very small particle size of the powders (about 50 – 100 Å). An estimate is required of the magnitude of the line broadening.

R. Jenkins, J. L. de Vries
Backmatter
Metadaten
Titel
Worked Examples in X-Ray Analysis
verfasst von
R. Jenkins
J. L. de Vries
Copyright-Jahr
1978
Verlag
Macmillan Education UK
Electronic ISBN
978-1-349-03534-2
Print ISBN
978-1-349-03536-6
DOI
https://doi.org/10.1007/978-1-349-03534-2